Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K240396

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-03-08

    (29 days)

    Product Code
    Regulation Number
    878.4810
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    755mm:

    The Elite iQ PRO Laser System is indicated for stable long-term or permanent hair reduction is defined as long-term stable reduction in the number of hairs regrowing when measured at 6, 9 or 12 months after the completion of a treatment regime. It is used for skin types (Fitzpatrick I-VI) including tanned skin. It is also indicated for the treatment of vascular lesions, benign pigmented lesions, and wrinkles.

    1064nm:

    The Elite iQ PRO Laser System is intended for the coagulation and hemostasis of benign vascular lesions such as, but not limited to, port wine stains, hemangiomas, warts, telangiectasia, rosacea, venous lakes, leg veins and porkiloderma of Civatte; and treatment of beingn cutaneous lesions such as warts, scars, striae and psoriasis. The laser is also intended for the treatment of benign pigmented lesions such as, but not limited to, lentigines (age spots), solar lentigines (sunspots), cafe au lait macules, sebortheic keratoses, nevi, chloasma, verrucae, skin tags, keratosis and plaques.

    The laser is also indicated for the treatment of wrinkles such as, but not limited to, periocular and perioral wrinkles.

    Additionally, the laser is indicated for the treatment of pseudo follicultis barbae (PFB) and for stable long-term or permanent hair reduction. Permanent hair reduction is defined as long-term stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime.

    The Skintel Reader is intended as an objective measurement tool for examining skin melanin content for determining and setting a test spot starting fluence.

    Device Description

    The Elite iQ™ PRO workstation is a dual wavelength system that delivers laser energy in both the Nd:YAG (1064-nm) and Alexandrite (755-nm) wavelengths. Through various spot sizes, fluences and repetition rates, the system offers hair removal treatment and aesthetic treatments across all skin types. An Alexandrite standalone workstation is also available for purchase. The Elite iQ PRO delivers the laser energy through a lens-coupled optical fiber with a wide range of interchangeable, quick-release laser handpieces with electronic spot recognition. The Elite iQ PRO also includes the Skintel® Melanin Reader for objective measurement of the melanin content of skin.

    The locking casters allow this stand-alone laser system to be secured in place, as well as to be conveniently moved or transported. When not in use, handpiece components and other system components can be stowed in the storage area located in the side drawer. All system connections, such as the foot switch and remote interlock connections, are located on the rear of the laser. This includes all applicable device labels. User-selectable controls and functions are located on the front of the laser.

    Elite iQ PRO software features include a Windows® operating system, LCD touch screen and a state-of-the-art graphic user interface. The software of the device was changed to update the fluence levels for the handpieces in order to support the new 30mm handpiece. The new 30 mm handpiece is also part of the DEKA AGAIN PRO device (K233090). The new fluence levels are within range of the previously approved ranges in the Elite iQ device (K193426). The software was also updated to reflect the Elite iQ PRO branding.

    There are no changes to the principle of use of the device compared to the predicates Elite iQ Laser System (K193426) and DEKA AGAIN PRO (K233090).

    The labeling of the device has been updated to reflect the name of the device Elite IQ PRO Laser System.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the Elite iQ PRO (M122K1) device.

    Key Observation: The provided 510(k) summary primarily focuses on establishing substantial equivalence to predicate devices (Elite iQ Laser System (K193426) and DEKA AGAIN PRO (K233090)) through comparison of technical characteristics and intended uses. It does not contain details about a clinical study with acceptance criteria for device performance in terms of efficacy or diagnostic accuracy (e.g., sensitivity, specificity, accuracy) for its intended indications (hair reduction, vascular lesions, etc.).

    Instead, the "acceptance criteria" and "study" described are in the context of verification and validation activities against recognized electrical and laser safety standards, and comparing technical specifications to predicate devices.


    1. Table of acceptance criteria and the reported device performance

    Based on the provided text, the "acceptance criteria" appear to be compliance with established safety and performance standards for medical electrical and laser equipment, as well as maintaining similar technical characteristics to predicate devices. The "reported device performance" is implicitly that the device meets these standards and maintains comparable technical specifications.

    Acceptance Criteria CategorySpecific Standard/CharacteristicReported Device Performance (Implicit)
    Electrical SafetyAAMI/ANSI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012 – Medical Electrical Equipment – Part 1: General Requirements for Basic Safety and Essential Performance.Complies with this standard for basic safety and essential performance.
    EMCIEC 60601-1-2 Ed. 4.1 :2020-09 Medical electrical equipment - Part 1-2: General Requirements for Basic Safety and Essential Performance— Collateral standard: Electromagnetic Disturbances – Requirements and tests.Complies with this standard for electromagnetic disturbances.
    Laser SafetyIEC 60601-2-22 Edition 3.1 2012-10 Medical electrical equipment - Part 2-22: Particular requirements of basic safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment.Complies with this standard for surgical, cosmetic, therapeutic and diagnostic laser equipment.
    Laser Product SafetyIEC 60825-1:2014 Safety of laser products Part 1: Equipment . classification and requirementsComplies with this standard for laser product classification and requirements.
    Skintel Melanin ReaderAccuracy of +/- 5 Melanin Index (MI)Maintains this accuracy, consistent with predicate.
    Skintel Melanin ReaderMeasurement Time Less than one (1) second per measurementMaintains this measurement speed, consistent with predicate.

    Note: The document explicitly states: "The following verification and validation activities have been performed on the modified device." This indicates that the studies performed were primarily focused on demonstrating compliance with these technical and safety standards, rather than clinical efficacy.


    Regarding Device Performance (from comparison tables):

    The device's performance is described in terms of its technical specifications, which are compared to predicate devices. The key updates for the Elite iQ PRO are:

    • Repetition Rate: Up to 12 Hz (compared to 10 Hz for Elite iQ and 12 Hz for DEKA AGAIN PRO).
    • Handpiece (Spot) Size: Addition of a 30mm handpiece, similar to DEKA AGAIN PRO (Elite iQ had up to 24mm).
    • Software update to reflect new fluence levels for the 30mm handpiece and Elite iQ PRO branding.

    These are considered performance improvements or equivalences rather than demonstrating achievement of specific clinical acceptance criteria through a trial.


    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not mention a clinical test set sample size, data provenance, or whether it was retrospective or prospective. The studies appear to be technical verification and validation, not clinical trials on patients.


    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not provided in the document. As no clinical efficacy or diagnostic accuracy study is described, there's no mention of experts establishing ground truth for a patient test set.


    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not provided, as the document does not describe a clinical test set requiring adjudication.


    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    There is no mention of an MRMC comparative effectiveness study, nor any AI assistance/human-in-the-loop performance data. The device is a laser system and a melanin reader; it does not appear to incorporate AI for interpretation or diagnosis that would necessitate such a study.


    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is not applicable as the device is a laser system and a melanin reader, not an AI algorithm performing diagnostic tasks.


    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    Since the studies described are for technical verification and validation against standards, the "ground truth" would be the specifications and requirements of those engineering and safety standards. For the Skintel Melanin Reader accuracy, the ground truth would be established by reference measurements or calibrated instruments. There is no mention of clinical ground truth (e.g., pathology, outcomes data) for the efficacy of hair removal or lesion treatment.


    8. The sample size for the training set

    This is not applicable as the device is not an AI/ML device that requires a training set in the conventional sense.


    9. How the ground truth for the training set was established

    This is not applicable for the same reason as above.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1