Search Filters

Search Results

Found 9 results

510(k) Data Aggregation

    K Number
    K253980

    Validate with FDA (Live)

    Date Cleared
    2026-03-27

    (105 days)

    Product Code
    Regulation Number
    890.3860
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Electrically Powered Wheelchair is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.

    Device Description

    The device consists of front wheel, drive wheel, frame, controller, motor, armrest, backrest, seat cushion, pedal, battery box and charger.

    Electrically Powered Wheelchair is powered by Li-ion Battery pack (25.55V 10.4Ah*2) with 11.8 km range, which can be recharged by an off-board battery charger that can be plugged into an AC socket outlet (100-240V, 50/60Hz) when the device is not in use. The materials of frame is Carbon fiber.

    The patient can activate the controller handle (joystick) to control the speed and direction of the wheelchair movement. In addition, when the patient releases the joystick, the joystick will return back to the central position and the wheelchair will be automatically stopped soon due to automatic electromagnetic brake system. Once the joystick is activated again move to other position, the wheelchair will be re-energized.

    The max loading of the device is 136KG. Only for one person to sit.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K252494

    Validate with FDA (Live)

    Date Cleared
    2025-12-05

    (119 days)

    Product Code
    Regulation Number
    890.3860
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Electrically powered wheelchair is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.

    Device Description

    The Electrically powered wheelchair mainly consists: a frame (a front wheel bracket, a motor frame, a seat frame, a backrest frame), armrests, a controller, a driver, a seat cushion, a back cushion, footrests, 4 wheels, 2 back wheel motors, 2 anti-tilt wheels, a safety belt and 2 lithium batteries.

    The device is composed of two sets of four-bar linkage hinges, which can be unfolded and folded. After unfolding, the two sets of four-bar linkage are locked by a locking structure.

    The device is installed with an electromagnetic brake that will engage automatically when the device is not in use and the brake cannot be used manually. The device only can be operated on the flat and hard road.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K244003

    Validate with FDA (Live)

    Date Cleared
    2025-03-26

    (90 days)

    Product Code
    Regulation Number
    890.3860
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    It is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.

    Device Description

    This product consists of frame, wheels, seat, armrest, lithium battery, motor and controller with a lightweight and compact design.The entire wheelchair can be disassembled, making it easy to carry. The wheelchair comes equipped with an electronically controlled rising seat that allows the user 280mm/11" of seat height adjustment using the joystick. The armrest can be flipped upside down, which is convenient for the elderly to move. Users can drive the wheelchair by themselves through the control device. The wheelchair uses lead-acid Battery as its power source. The controls the drive leftright motor to realize the wheelchair forward, backward and turn functions. The frame of the device is carbon steel. The front wheels suitable for rotation, acceleration, retrograde and other actions of the wheels movement will be achieved by thrust generated from the rear wheels are driving wheels to control the speed and direction. The wheels are Solid PU tires. When in use, the operator drives the motor of the rear wheel by operating the controller joystick to achieve the rear wheels movement. The DC brushless motor and brake system are fixed on the rear wheels. The max loading of the device is 136KG. Only for one person sit.

    AI/ML Overview

    The provided text is a 510(k) summary for an electrically powered wheelchair. It focuses on demonstrating substantial equivalence to a predicate device rather than presenting a study with specific acceptance criteria and detailed device performance metrics in the format requested.

    Therefore, many of the requested elements of information, such as sample sizes, data provenance, expert qualifications, and MRMC study details, are not applicable or not available in this type of document, as it describes the testing and comparison of a physical medical device (a wheelchair) rather than a diagnostic AI algorithm.

    However, I can extract the acceptance criteria as indicated by compliance with various ISO standards and the 'reported device performance' based on the comparison table and the general statements about meeting requirements.

    Here's the closest representation of the requested information based on the provided document:


    1. Table of Acceptance Criteria and Reported Device Performance

    Feature/TestAcceptance Criteria (Standard/Requirement)Reported Device Performance
    Material Safety (Biocompatibility)ISO10993-5: Biological evaluations of medical devices -- Part 5: Tests for In Vitro cytotoxicity ISO 10993-10: Biological evaluation of medical devices - Part 10: Tests for skin sensitization ISO 10993-23: Biological evaluation of medical devices - Part 23: Tests for irritationAll user directly contacting materials are compliant with these ISO requirements, and tested to be safe.
    Safety - Electromagnetic Compatibility (EMC)IEC 60601-1-2: Medical Electrical Equipment - Part 1-2: General requirements for basic safety and essential performance – Collateral Standard: Electromagnetic disturbances – Requirements and tests ISO7176-21: Wheelchairs - Part 21: Requirements and test methods for electromagnetic compatibility of electrically powered wheelchairs and scootersMeets IEC 60601-1-2 and ISO7176-21 requirements.
    Performance (General)ISO 7176 series (including specific parts for stability, brakes, energy consumption, dimensions, speed, strength, climatic tests, obstacle climbing, power/control systems, batteries/chargers, etc.) ISO 16840-10: Wheelchair seating - Part 10: Resistance to ignition of postural support devices - Requirements and test method FDA Guidance: Submission for Power Wheelchair Risk Analysis developed in accordance with ISO 14971: 2019 Software validationThe device complies with all referenced ISO standards and FDA guidance. All tests indicate a positive conclusion and meet requirements.
    Braking Distance≤1.5 m (as per predicate)≤1.5 m
    Maximum Safe Operational Incline Degree6° (as per predicate)
    Max Loading Weight120kg (predicate), validated with maximum rated weight dummy.136kg (similar performance to predicate, validated with maximum rated weight dummy)
    Max Speed ForwardUp to 6 km/h (1.6 m/s) adjustable (predicate)Up to 5.4km/h (1.5m/s) (similar)
    Max Speed BackwardLess than 3 km/h (0.8 m/s) (predicate)Less than 2.5km/h (0.7m/s) (similar)
    Maximum Distance of Travel on Fully Charged Battery16km (predicate)12km (similar, not causing new safety/effectiveness concerns)
    Electronic Controller PerformanceMeets the requirements of ISO 7176-14Meets the requirements of ISO 7176-14
    Brake System EffectivenessMeets the requirements of ISO 7176-3Meets the requirements of ISO 7176-3
    Stability (Static, Dynamic)Meets the requirements of ISO 7176-1 and ISO 7176-2Meets the requirements of ISO 7176-1 and ISO 7176-2
    Strength (Static, Impact, Fatigue)Meets the requirements of ISO 7176-8Meets the requirements of ISO 7176-8
    Obstacle Climbing AbilityMeets the requirements of ISO 7176-10Meets the requirements of ISO 7176-10 (e.g., 25mm climbing vs predicate's 40mm, deemed not to cause new concerns)
    Batteries and ChargersMeets the requirements of ISO 7176-25Meets the requirements of ISO 7176-25
    Information Disclosure, Documentation, LabelingMeets the requirements of ISO 7176-15 and FDA RegulatoryConforms to ISO 7176-15 and FDA Regulatory

    2. Sample size used for the test set and the data provenance:

    • Sample Size: The document does not specify a distinct "test set" in terms of patient data or case numbers. Instead, it refers to performance and safety testing of the physical device according to various international standards (ISO). For tests involving a dummy, it mentions validation "with the maximum rated weight dummy." The number of devices tested is not explicitly stated.
    • Data Provenance: The testing appears to have been conducted as part of the device development and verification process by the manufacturer in China. The specific location or date of individual tests are not detailed beyond the document preparation date (2025/2/26). The studies are inherently "prospective" in the sense that they are conducted to verify the new device's compliance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This section is not applicable. The device is an electrically powered wheelchair, not an AI diagnostic algorithm or system requiring expert-established ground truth from medical images or patient data. The "ground truth" for the device's performance is established by its compliance with engineering and safety standards, which are defined by standard bodies (like ISO) and regulatory bodies (like FDA), and verified through physical testing.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • This section is not applicable. Adjudication methods like 2+1 or 3+1 are typically used in clinical studies to resolve discrepancies among multiple expert readers for diagnostic tasks. For a physical device like a wheelchair, performance is measured against objective engineering and safety standards, not subjective expert assessment of a diagnostic outcome.

    5. If a multi-reader, multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • This section is not applicable. An MRMC study is relevant for evaluating the impact of AI on human reader performance in diagnostic tasks. This document is for a physical medical device (a wheelchair) and does not involve AI for diagnostic purposes or human readers.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • This section is not applicable. There is no AI algorithm being evaluated for standalone performance in this document. The device is a physical wheelchair.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • The "ground truth" in this context is defined by the technical specifications, requirements, and acceptable tolerances outlined in the referenced international standards (e.g., ISO 7176 series, ISO 10993 series, IEC 60601-1-2) which are based on engineering principles, safety considerations, and recognized testing methodologies. It's essentially engineering and safety standard compliance verified through objective physical testing.

    8. The sample size for the training set:

    • This information is not applicable as this document does not describe the development or testing of an AI model with a training set.

    9. How the ground truth for the training set was established:

    • This information is not applicable as this document does not describe the development or testing of an AI model with a training set.
    Ask a Question

    Ask a specific question about this device

    K Number
    K243961

    Validate with FDA (Live)

    Date Cleared
    2025-02-12

    (51 days)

    Product Code
    Regulation Number
    890.3860
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    It is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.

    Device Description

    This product consists of frame, wheels, seat, armrest, lithium battery, motor and controller with a lightweight and compact design. The whole wheelchair can be folded and it can be easily carried or rolled after folding. The seat cushion is detachable. The armrest can be flipped upside down, which is convenient for the elderly to move. Users can drive the wheelchair by themselves through the control device. The wheelchair uses lithium batteries as its power source. The controls the drive leftright motor to realize the wheelchair forward, backward and turn functions. The frame of the device is aluminum alloy. The front wheels suitable for rotation, acceleration, retrograde and other actions of the wheelchair. The front wheels movement will be achieved by thrust generated from the rear wheels are driving wheels to control the speed and direction. The wheels are Solid PU tires. When in use, the operator drives the motor of the rear wheel by operating the controller joystick to achieve the rear wheels movement. The DC brushless motor and brake system are fixed on the rear wheels. The max loading of the device is 120KG. Only for one person sit.

    AI/ML Overview

    The provided document is a 510(k) Premarket Notification for an Electrically Powered Wheelchair (HP206). It primarily details the device's characteristics and compares it to a predicate device (HP358EA) to demonstrate substantial equivalence.

    However, the document does not contain information about a study proving device performance against specific acceptance criteria, especially not in the context of an AI/human reader study as typically associated with the comprehensive questions asked. The device in question is an Electrically Powered Wheelchair, not a medical imaging or diagnostic AI device that would involve expert readers, ground truth establishment for a test set, multi-reader multi-case studies, or AI algorithm performance metrics like sensitivity, specificity, etc.

    The "acceptance criteria" and "study that proves the device meets the acceptance criteria" in this context refer to the device's compliance with established performance standards and safety requirements for electrically powered wheelchairs.

    Therefore, I cannot populate the requested table or answer most of the specific questions about AI performance, expert adjudication, or MRMC studies for this device, as these concepts are not applicable to the information provided in the 510(k) submission for a powered wheelchair.

    I can, however, extract the performance data and compliance standards mentioned:

    1. Table of acceptance criteria and the reported device performance

    The document frames "acceptance criteria" as compliance with various ISO standards and FDA guidance. "Reported device performance" is demonstrated through testing against these standards. There isn't a specific quantitative "acceptance criteria" table with corresponding numerical "reported performance" easily extractable for each parameter the way one would for an AI diagnostic device. Instead, the document states that the device "met all design specifications" and provided "support of the substantial equivalence determination" by undergoing the listed tests.

    Here's an attempt to structure it based on the available information, understanding that "acceptance criteria" is implied by compliance with the standards, and "reported performance" is that it passed the tests:

    Acceptance Criterion (Standard)Reported Device Performance (Compliance)
    ISO 10993-5 (In Vitro cytotoxicity)Met requirements; material is safe
    ISO 10993-10 (Skin sensitization)Met requirements; material is safe
    ISO 10993-23 (Irritation)Met requirements; material is safe
    ISO 14971:2019 (Risk Analysis)Risk Analysis developed in accordance.
    Software validationSoftware validation carried out and meets requirements.
    ISO 7176-1:2014 (Static stability)Compliance stated (implied by "met all design specifications" and "positive conclusion")
    ISO 7176-2:2017 (Dynamic stability)Compliance stated
    ISO 7176-3:2012 (Brake effectiveness)Compliance stated; meets ISO 7176-3
    ISO 7176-4:2008 (Energy consumption)Compliance stated
    ISO 7176-5:2008 (Dimensions, mass, maneuvering)Compliance stated
    ISO 7176-6:2018 (Max speed, acceleration)Compliance stated
    ISO 7176-7:1998 (Seating & wheel dimensions)Compliance stated
    ISO 7176-8:2014 (Static, impact, fatigue strength)Compliance stated
    ISO 7176-9:2009 (Climatic tests)Compliance stated
    ISO 7176-10:2008 (Obstacle-climbing)Compliance stated; meets ISO 7176-10
    ISO 7176-11:2012 (Test dummies)Compliance stated
    ISO 7176-13:1989 (Coefficient of friction)Compliance stated
    ISO 7176-14:2022 (Power & control systems)Compliance stated; meets ISO 7176-14
    ISO 7176-15:1996 (Information & labeling)Compliance stated
    ISO 16840-10:2021 (Resistance to ignition)Compliance stated; test carried out according to standard
    ISO 7176-21:2009 (EMC)Compliance stated; IEC 60601-1-2 & ISO7176-21
    ISO 7176-22:2014 (Set-up procedures)Compliance stated
    ISO 7176-25:2013 (Batteries & chargers)Compliance stated
    IEC 60601-1-2:2020 (EMC)Compliance stated
    FDA Regulatory Guidance (Label and labeling)Conforms to FDA Regulatory

    2. Sample size used for the test set and the data provenance:

    • Sample Size: Not explicitly stated as a "sample size" in the conventional sense for a clinical trial or algorithm test set with individual cases. The testing refers to the device prototype(s) themselves. For mechanical and electrical performance tests, a limited number of test units (often 1-3) are typically used per test type, depending on the standard. For biocompatibility, material samples are tested.
    • Data Provenance: The tests are performed to demonstrate compliance with international standards (ISO, IEC) generally by test labs. The document does not specify geographical provenance for the testing data in terms of "country of origin of the data" being retrospective or prospective from patient cohorts. It's device performance data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This is not applicable to an electrically powered wheelchair device. "Ground truth" in this context would refer to the physical and electrical safety and performance parameters measured against established engineering and medical device standards. No "experts" are establishing "ground truth" in the diagnostic sense. Compliance is verified by testing according to published, objective standards.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • Not applicable. This refers to consensus reading for diagnostic purposes, which is not relevant here.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This is not an AI diagnostic device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This is not an AI algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • The "ground truth" for this device is based on established international performance and safety standards (ISO and IEC, as listed in the document) and FDA guidance for electrically powered wheelchairs. Compliance is demonstrated through physical, electrical, and materials testing.

    8. The sample size for the training set:

    • Not applicable. This is not an AI device trained on data.

    9. How the ground truth for the training set was established:

    • Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K243962

    Validate with FDA (Live)

    Date Cleared
    2025-02-12

    (51 days)

    Product Code
    Regulation Number
    890.3860
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    It is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.

    Device Description

    This product consists of frame, wheels, seat, armrest, lithium battery, motor and controller with a lightweight and compact design. The whole wheelchair can be folded and it can be easily carried or rolled after folding. The seat cushion is detachable. The armrest can be flipped upside down, which is convenient for the elderly to move. Users can drive the wheelchair by themselves through the control device. The wheelchair uses lithium batteries as its power source. The controls the drive left/right motor to realize the wheelchair forward, backward and turn functions. The frame of the device is Magnesium alloy. The front wheels are driven wheels suitable for rotation, acceleration, retrograde and other actions of the wheelchair. The front wheels movement will be achieved by thrust generated from the rear wheels are driving wheels to control the speed and direction. The wheels are Solid PU tires.When in use, the operator drives the motor of the rear wheel by operating the controller joystick to achieve the rear wheels movement. The DC brushless motor and brake system are fixed on the rear wheels. The max loading of the device is 120KG. Only for one person sit.

    AI/ML Overview

    The provided text is a 510(k) summary for an Electrically Powered Wheelchair (device name: HP558) and does not contain information about acceptance criteria or a study proving device performance in the context of AI/ML or diagnostic applications.

    The document discusses the substantial equivalence of the proposed device to a predicate device (HP458E) by comparing their design, functionality, materials, and compliance with various ISO and IEC standards relevant to wheelchairs. It also mentions performance data to verify that the subject device meets design specifications, but this is in the context of standard engineering and safety testing for a medical device (electrically powered wheelchair), not a study assessing AI/ML model performance.

    Therefore, I cannot provide the requested information, such as:

    1. A table of acceptance criteria and reported device performance: This document lists various technical specifications and safety standards for the wheelchair itself, not performance metrics of an AI algorithm.
    2. Sample size, data provenance, number of experts, adjudication method for the test set: These are relevant to studies evaluating AI/ML models on a dataset, which is not what this document describes.
    3. MRMC comparative effectiveness study: No such study is mentioned.
    4. Standalone performance (algorithm only): The device is a physical wheelchair, not an algorithm.
    5. Type of ground truth used: This concept is not applicable to the evaluation described.
    6. Sample size for the training set: Not applicable.
    7. How ground truth for the training set was established: Not applicable.

    The document is a regulatory submission for a physical medical device (an electrically powered wheelchair) and focuses on ensuring its physical safety, effectiveness, and substantial equivalence to an existing device through engineering and biocompatibility testing against established standards. It does not involve AI or diagnostic imaging.

    Ask a Question

    Ask a specific question about this device

    K Number
    K241337

    Validate with FDA (Live)

    Date Cleared
    2024-10-17

    (157 days)

    Product Code
    Regulation Number
    890.3860
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    It is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.

    Device Description

    The Electrically powered wheelchair is a motor driven, indoor transportation vehicle, which a device for assisting action handicapped people and disabled people to move. It is suitable for disabled people with mobility difficulties and elderly people.

    It is mainly composed of front wheel, drive wheel, frame, controller (joystick), motor and drive devices, armrest, backrest, seat belt, footrest, battery box and charger.

    The device is powered by Li-ion Battery pack (25V 10Ah) with 11.8 Km range, which can be recharged by an off-board battery charger that can be plugged into an AC socket outlet(100-240V, 50/60Hz) when the device is not in use.

    The user can activate the controller handle (joystick) to control the speed and direction of the wheelchair movement. When the user releases the joystick will automatically go back to the central position and the wheelchair will be automatically stopped soon due to automatic electromagnetic brake system. Once the joystick is activated again move to other position, the wheelchair will be re-energized.

    AI/ML Overview

    This document is a 510(k) summary for an electrically powered wheelchair (HP202). It focuses on demonstrating "substantial equivalence" to a legally marketed predicate device (K230964) rather than proving the device meets specific acceptance criteria through a clinical study of an AI/software component.

    Therefore, the request for details related to AI/software performance evaluation (like acceptance criteria for an AI model, sample sizes for test/training sets, expert consensus, MRMC studies, standalone performance, etc.) cannot be fully answered from the provided text.

    The document states: "No clinical study is included in this submission." and primarily relies on non-clinical testing against established ISO standards and direct comparison of specifications to the predicate device to argue for substantial equivalence.

    Here's an attempt to answer the questions based only on the provided text, highlighting where information is absent for AI-specific criteria:


    1. A table of acceptance criteria and the reported device performance

    The acceptance criteria are generally implied by conformity to specific ISO standards and the performance metrics of the predicate device. The document presents a comparison table (Table 1: General Comparison, and implicitly Table 3: Performance Comparison) rather than a formal acceptance criteria table with a "Pass/Fail" or "Meets" column. The "Remark" or "Results" columns in these tables serve a similar purpose by indicating "S.E." (Substantially Equivalent) or explaining minor differences that do not raise new safety or effectiveness concerns.

    Feature/TestAcceptance Criteria (Implied by Predicate/Standards)Reported Device Performance (Subject Device)Result/Remark
    BiocompatibilityCompliance with ISO 10993-5, -10, -23Compliance with ISO 10993-5, -10, -23S.E.
    EMCCompliance with ISO 7176-21Compliance with ISO 7176-21 & IEC 60601-2-1S.E.
    Braking distance≤ 1.5 m (as per predicate)≤ 1 mMinor difference, shorter distance is safer.
    Max. safe operational incline9° (as per predicate)10°Minor difference, doesn't raise new safety/effectiveness concerns as stability evaluated per ISO 7176 series.
    Max loading weight110kg (as per predicate)120kgDifference on loading weight will not cause different performance; validated with max loading weight.
    Max distance of travel10 km (as per predicate)11.8 kmCaused by wheel size; farther is better.
    Turning Radius900 mm (as per predicate)1000 mmMinor difference due to size, doesn't raise new safety/effectiveness concerns.
    Maximum obstacle climbing40 mm (as per predicate)20 mmMinor difference, doesn't raise new safety/effectiveness concerns.
    Static stabilityMeets ISO 7176-1 requirementsMeets ISO 7176-1 requirementsS.E.
    Dynamic stabilityMeets ISO 7176-2 requirementsMeets ISO 7176-2 requirementsS.E.
    Brake effectivenessMeets ISO 7176-3 requirementsMeets ISO 7176-3 requirementsS.E.
    Energy consumption/Distance rangeMeets ISO 7176-4 requirementsMeets ISO 7176-4 requirementsS.E.
    Overall dimensions, massMeets ISO 7176-5 requirementsMeets ISO 7176-5 requirementsS.E.
    Max speed, acceleration, decelerationMeets ISO 7176-6 requirementsMeets ISO 7176-6 requirementsS.E.
    Seating and wheel dimensionsMeets ISO 7176-7 requirementsMeets ISO 7176-7 requirementsS.E.
    Static, impact, fatigue strengthsMeets ISO 7176-8 requirementsMeets ISO 7176-8 requirementsS.E.
    Climatic testsMeets ISO 7176-9 requirementsMeets ISO 7176-9 requirementsS.E.
    Obstacle-climbing abilityMeets ISO 7176-10 requirementsMeets ISO 7176-10 requirementsS.E.
    Test dummies requirementsMeets ISO 7176-11 requirementsMeets ISO 7176-11 requirementsS.E.
    Coefficient of friction of test surfacesMeets ISO 7176-13 requirementsMeets ISO 7176-13 requirementsS.E.
    Power and control systemsMeets ISO 7176-14 requirementsMeets ISO 7176-14 requirementsS.E.
    Information disclosure, documentation, labelingMeets ISO 7176-15 requirementsMeets ISO 7176-15 requirementsS.E.
    Resistance to ignition of postural support devicesMeets ISO 16840-10 requirementsMeets ISO 16840-10 requirementsS.E.
    Batteries and chargersMeets ISO 7176-25 requirementsMeets ISO 7176-25 requirementsS.E.

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    This information is not provided as no clinical study was conducted. The testing described is primarily physical device testing against engineering standards. The manufacturer is based in China.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    This information is not provided as no clinical study or expert-based ground truth establishment (relevant for AI/image analysis) was performed.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    This information is not provided as no clinical study or expert-based ground truth establishment was performed.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC study was done. The device is an electrically powered wheelchair, not an AI-assisted diagnostic tool for human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    A standalone performance test for an AI algorithm was not performed because the device itself is an electrically powered wheelchair and the submission does not describe an AI algorithm component requiring such testing. The "software validation" mentioned relates to the wheelchair's control system, not an AI for interpretation or diagnosis.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for this device's performance is established by conformance to internationally recognized engineering standards (ISO 7176 series, ISO 10993 series) and direct comparison of technical specifications and safety features to a legally marketed predicate device. This is a non-clinical "ground truth."

    8. The sample size for the training set

    This information is not provided as no AI training set was described or used in this submission.

    9. How the ground truth for the training set was established

    This information is not provided as no AI training set was described or used in this submission.


    In summary: The provided FDA 510(k) summary for the "Electrically powered wheelchair (HP202)" demonstrates substantial equivalence through non-clinical testing against ISO standards and direct comparison to a predicate device. It does not involve a clinical study or the evaluation of an artificial intelligence/software component where the requested details about acceptance criteria, sample sizes, expert ground truth, or human-AI interaction would be relevant.

    Ask a Question

    Ask a specific question about this device

    K Number
    K241632

    Validate with FDA (Live)

    Date Cleared
    2024-10-16

    (132 days)

    Product Code
    Regulation Number
    890.3860
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    It is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.

    Device Description

    This product consists of frame, wheels, seat, armrest, lithium battery, motor and controller with a lightweight and compact design. The whole wheelchair can be folded by one button and it can be easily carried or rolled after folding, The seat cushion is detachable. The armrest can be flipped backward, which is convenient for the wheelchair by the wheelchair by themselves through the control device.

    The wheelchair uses lithium batteries as its power source. The drive left/right motor to realize the wheelchair forward, backward and turn functions.

    The frame of the device is carbon fiber. The front wheels suitable for rotation, acceleration, retrograde and other actions of the wheelchair. The front wheels mover will be achieved by thrust generated from the rear wheels are driving wheels to control the speed and direction. The wheels are Solid PU tires.

    When in use, the operator drives the motor of the rear wheel by operating the controller handle (joystick) to achieve the rear wheels movement.

    The DC Brushless motor and Brake system are fixed on the rear wheels. The max loading of the device is 125KG. Only for one person sit.

    AI/ML Overview

    This is an FDA 510(k) premarket notification for an Electrically Powered Wheelchair (K241632).

    For devices like this (powered wheelchairs), the acceptance criteria and study that proves the device meets them typically involve demonstrating substantial equivalence to a legally marketed predicate device, primarily through non-clinical performance testing to relevant international standards. Unlike AI/ML-based medical devices, there isn't a "test set" in the sense of a dataset for algorithm evaluation, nor are there "expert readers" establishing ground truth in the same way. The "performance" refers to the physical and functional aspects of the wheelchair.

    Here's a breakdown based on the provided document:

    Acceptance Criteria and Reported Device Performance

    The document states that the device's acceptance criteria are met by demonstrating compliance with various international ISO and IEC standards relevant to electrically powered wheelchairs. The reported "performance" is inherently tied to passing these standards.

    Table 1: Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Standard Compliance)Reported Device Performance (Compliance Status)
    Mechanical & Performance Standards:
    ISO 7176-2:2001 (Wheelchair Testing - Mechanical Properties)Complied
    ISO 7176-3:2012 (Wheelchair Testing - Electrical Requirements)Complied
    ISO 7176-4:2008 (Wheelchair Testing - Climatic Requirements)Complied
    ISO 7176-5:2008 (Wheelchair Testing - Dimensions, Mass, and Maneuvering Space)Complied
    ISO 7176-6:2018 (Wheelchair Testing - Maximum Speed, Acceleration, Deceleration)Complied
    ISO 7176-7:1998 (Wheelchair Testing - Wheelchair Measurements)Complied
    ISO 7176-9:2009 (Wheelchair Testing - Climatic Chambers for Testing)Complied
    ISO 7176-10:2008 (Wheelchair Testing - Obstacle Climbing)Complied
    ISO 7176-11:2008 (Wheelchair Testing - Rolling Resistance)Complied
    ISO 7176-13:1989 (Wheelchair Testing - Static, Impact, and Fatigue Strengths)Complied
    ISO 7176-14:2008 (Wheelchair Testing - Power & Control Systems)Complied
    ISO 7176-15:1996 (Wheelchair Testing - Dimensions, Mass, and Volume)Complied
    ISO 7176-21:2014 (Wheelchair Testing - Electromagnetic Compatibility)Complied
    ISO 7176-22:2013 (Wheelchair Testing - Setup Procedures)Complied
    Material/Biocompatibility Standard:
    ISO 10993 series (Biological evaluation of medical devices)Complied (for parts in contact with user)
    Seating/Safety Standard:
    ISO 16840-10: 2021 (Wheelchair Seating - Flammability)Complied (for seat cushion/backrest)
    Electrical Safety/EMC Standard:
    IEC 60601-1-2: 2014 (Medical Electrical Equipment - EMC)Complied
    IEC 62133-2:2017 (Secondary Cells and Batteries - Safety)Complied (for lithium battery)

    The document explicitly states: "The conclusions drawn from the nonclinical tests demonstrate that the subject device is as safe, as effective, and performs as well as the legally marketed predicate device K231508."

    Study Details

    Given the nature of the device (electrically powered wheelchair) and the information provided in the 510(k) summary, the "study" is a series of non-clinical, bench-top, and possibly simulated-use tests to demonstrate compliance with the referenced international standards.

    1. Sample size used for the test set and the data provenance:

      • Sample Size: The document does not specify a numerical sample size (e.g., number of wheelchairs tested). For non-clinical performance testing of physical devices, testing typically involves a representative sample or a single unit (depending on the test standard and design).
      • Data Provenance: The document does not explicitly state the country of origin for the testing data itself. The applicant is Kunshan Hi-Fortune Health Products Co., Ltd in China, and their consultant is in Shanghai, China. It is highly probable that the testing was conducted in laboratories in China, often by accredited testing houses. The testing would be considered prospective in the sense that it was conducted specifically to demonstrate compliance for this 510(k) submission.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • This question is not applicable in the context of this device and 510(k) submission. "Ground truth" established by experts (like radiologists for AI algorithms) is not a component of demonstrating substantial equivalence for an electrically powered wheelchair. The "ground truth" here is the pass/fail criteria defined by the requirements of the international standards themselves. Compliance is verified by engineers and technicians performing the tests according to standard protocols.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • This is not applicable. Adjudication methods like 2+1 or 3+1 are used for expert consensus on image interpretation or clinical outcomes, typically in studies involving human readers or clinical trials. For non-clinical performance testing against engineering standards, compliance is objectively measured.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • This is not applicable. MRMC studies are used to evaluate AI-assisted diagnostic devices in a clinical reading setting. This 510(k) is for a powered wheelchair, not a diagnostic device.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • This refers to AI algorithm performance. This is not applicable as this is not an AI/ML device. The "standalone performance" refers to the physical functionality of the wheelchair according to the specified standards.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • The "ground truth" for this device's performance is objective compliance with the pass/fail criteria and specifications outlined in the referenced international standards (e.g., maximum speed, turning radius, battery performance, mechanical strength, electromagnetic compatibility limits, flammability). It is a standards-based performance evaluation, not clinical outcomes or expert interpretation.
    7. The sample size for the training set:

      • This is not applicable. There is no "training set" in the context of a physical device like a powered wheelchair unless it uses an AI/ML component (which is not described or implied here). The design and engineering of the wheelchair are based on established principles, not a data training set.
    8. How the ground truth for the training set was established:

      • This is not applicable for the same reasons as above.
    Ask a Question

    Ask a specific question about this device

    K Number
    K220156

    Validate with FDA (Live)

    Date Cleared
    2022-10-07

    (261 days)

    Product Code
    Regulation Number
    890.3860
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device is a motor-driven, and indoor transportation vehicle with the intended use to provide mobility to a disabled or an elderly person limited to a seated position.

    Device Description

    The TUNG KENG Electrically Powered Wheelchair, Joy Rider, is suitable for indoor uses. It is characterized by high portability in comfortable travel, small storage and convenient delivery in daily life. As folding of Joy Rider only takes just a few seconds, so this model increases opportunities for travel and provides small storage. To use Joy Rider makes your travel in any place and at any moment! It is an individual transportation vehicle for disabled and elderly people that experience difficulties in walking. The Joy Rider comes as preassembled, designed to be lightweight, easily maneuvered, highly portable, and most importantly, safe and comfortable.

    It is driven by two brushless DC motors, and contains two foldable armrests, a seat belt, a backrest, a seat cushion and small with light texture, a foldable frame, two rear driving wheels with hub motor/electromagnetic brake assemblies, two pivoting casters, one Lithium-ion battery, an off-board battery charger, a control panel, and an electrical controller.

    The device is powered by one 24 VDC/ 10.5 Ah Lithium-ion battery with 9.3 miles cruising range that can be recharged by an off-board battery charger, rated at input: 100-240 VAC / 50-60 Hz, and output: 24 VDC/ 2 A, that can be plugged into an AC outlet, when the device is not in use. Its overall dimensions are 33.6' x 22.6' x 35.8" (855 x 580 x 910 mm). The user can activate the joystick to move in the direction of the joystick is actuated. When the user releases the joystick, the device slows to stop and the brakes are automatically re-engaged. The maximum weight capacity of Joy Rider is 264 lbs. (120 kg) and its maximum speed is 3.75 mph (6 km/h).

    AI/ML Overview

    I am sorry, but the provided text does not contain information about acceptance criteria and a study proving a medical AI device meets those criteria. The document is a 510(k) submission for an Electrically Powered Wheelchair, named Joy Rider, which is a physical medical device, not an AI-powered diagnostic or therapeutic tool.

    The document discusses:

    • A 510(k) premarket notification for a powered wheelchair.
    • Indications for use, intended user population, and device description.
    • Non-clinical performance tests conducted to comply with various ISO and ANSI/RESNA standards relevant to wheelchairs (e.g., static stability, dynamic stability, effectiveness of brakes, speed, obstacle climbing, electromagnetic compatibility).
    • Biocompatibility tests for patient-contacting parts.
    • A comparison of the subject device (Joy Rider) with a predicate device (Powered Wheelchair, DYW30A(D09)) to demonstrate substantial equivalence, discussing differences in components like motors, wheel sizes, maximum safe operational incline, dimensions, cruising range, and controllers, and asserting that these differences do not raise new safety or effectiveness concerns.

    Therefore, I cannot provide the requested information about acceptance criteria and a study for an AI device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K201855

    Validate with FDA (Live)

    Date Cleared
    2021-09-28

    (449 days)

    Product Code
    Regulation Number
    890.3860
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device is a motor-driven, and indoor transportation vehicle with the intended use to provide mobility to a disabled or an elderly person limited to a seated position.

    Device Description

    The Electrically Powered Wheelchair - Model HP358E is powered by Li-ion battery, driven by DC motor. Users control direction and adjust speed by controller. The main frame of the wheelchair is made of aluminum alloy. It has a seat base with four-wheeled and two with a back cover on the frame. Wheels are connected to the stainless steel axle receivers via stainless steel axles. On the front end of the frame are assembled two caster housings. Caster forks are mounted to these housings via steel axles. There are two rear caster wheels in back and anti-tip wheels in front to prevent the power chair from tipping. On the front end of the frame, upon the caster, a foot platform is assembled for foot holding. Two non-adjustable armrests are mounted on the frame for user's arm holding. Two grips mounted on the top end of the frame with a brake, and another brake is designed under the seat base. The backrest can be adjusted by squeezing yellow catches on the back together. A storage compartment located between the seat and foot platform. The joystick can be mounted on the left or right side using the quick release mechanism. The subject device has two 2 inch Anti-tippers, two 8 inch front wheels and two 12 inch rear wheels. The motor is DC24V 250W; the battery is 25.2 V 10.4Ah Li-ion battery; the charger is DC 24V, 2A. Max. loading can not be over than 125Kg, the maximum distance of travel on the fully charged battery is 10km and its maximum speed is 6km/h. The braking time is about 2~3s, and the braking distance is ≤1.0m.

    AI/ML Overview

    The provided text is a 510(k) Summary for an Electrically Powered Wheelchair (Model HP358E). It focuses on demonstrating substantial equivalence to a predicate device rather than presenting a study of a device's performance against specific acceptance criteria in the context of AI or advanced diagnostic algorithms. Therefore, much of the requested information regarding AI device evaluation (e.g., sample size for test set, provenance, expert adjudication, MRMC studies, standalone performance, training set details) is not applicable or present in this document.

    However, I can extract the relevant "acceptance criteria" (which in this context are performance and safety standards) and the "device performance" as reported for the proposed device, along with other applicable details.

    1. A table of acceptance criteria and the reported device performance

    The "acceptance criteria" for this wheeled device are derived from various ISO and IEC standards that it claims to comply with, and the "reported device performance" are the measured specifications of the HP358E. These are primarily found in the "Non-Clinical Test Summary" and the "Performance Comparison" table.

    Acceptance Criteria & Reported Device Performance for Electrically Powered Wheelchair - Model HP358E

    Acceptance Criteria (from Standards/Predicate Device)Reported Device Performance (Model HP358E)
    Material/Construction Standards (Implied Compliance)
    Main frame material: Aluminum alloy (per ASTM B221 for predicate)Main frame: Aluminium alloy
    Performance Standards (ISO & EN)
    Static Stability: Conforms to ISO 7176-1:2014Static stability forward: 22.4°
    Static stability rearward: 18.2°
    Static stability sideways: 23°
    Dynamic Stability: Conforms to ISO 7176-2:2017(Not explicitly stated as a single value, but compliance implies meeting the standard's criteria)
    Effectiveness of Brakes: Conforms to ISO 7176-3:2012Minimum braking distance: 1 m (Same as predicate)
    Energy Consumption & Distance Range: Conforms to ISO 7176-4:2008Cruising Range: 10 km (Predicate: 18 km - shorter cruising range noted, but implies compliance with standard)
    Dimensions, Mass & Maneuvering Space: Conforms to ISO 7176-5:2008Dimensions: 960mm635mm880mm (37.8"x25.0"x34.6")
    Weight, w/ Battery: 59.15 lbs / 27.2 kg
    Min. Turning Radius: 57.5" (1460mm) (Larger than predicate 32.5", but compliance implies meeting the standard)
    Max Speed, Acceleration & Deceleration: Conforms to ISO 7176-6:2018Max Speed: 6 km/h (Same as predicate)
    Max Speed Forwards: 3.75 mph (6 km/h) (Same as predicate)
    Max Speed Backward: 1.80 mph (2.9 km/h) (Predicate: 1.86 mph - noted as different but implies compliance with standard)
    Seating & Wheel Dimensions: Conforms to ISO 7176-7:1998Front wheel: 8" (PU solid tire)
    Rear tire: 12" (Pneumatic tire)
    Static, Impact & Fatigue Strengths: Conforms to ISO 7176-8:2014Max. loading: 275 lbs (125 kg) (Larger than predicate 264 lbs)
    Climatic Tests: Conforms to ISO 7176-9:2009(Implied compliance)
    Obstacle-climbing ability: Conforms to ISO 7176-10:2008Obstacle climbing: 2.36" (60mm) (Larger than predicate 1.36")
    Maximum obstacle climbing: 2.36" (60mm) (Larger than predicate 1.36")
    Test Dummies: Conforms to ISO 7176-11:2012(Implied compliance)
    Coefficient of Friction of Test Surfaces: Conforms to ISO 7176-13:1989(Implied compliance)
    Power & Control Systems: Conforms to ISO 7176-14:2008Controller: Shanghai Micon Mechanical&Electrical Co.,Ltd., M7084 (Evaluated per standard and software validation)
    Information Disclosure, Documentation & Labeling: Conforms to ISO 7176-15:1996(Implied compliance, and stated "Comply with FDA Regulatory Requirements" in 'Safety Comparison' table)
    Resistance to Ignition of Postural Support Device: Conforms to ISO 7176-16:2012(Implied compliance, and "flame retardant test of the seat cushion/back cushion and armrest" mentioned in summary discussion)
    Electromagnetic Compatibility: Conforms to ISO 7176-21:2009(Implied compliance)
    Setup Procedures: Conforms to ISO 7176-22:2014(Implied compliance)
    Electrically powered wheelchairs, scooters and their chargers: EN 12184:2014(Implied compliance)
    Basic Safety and Essential Performance: IEC 60601-1: 2012(Implied compliance)
    Electromagnetic Disturbances: IEC 60601-1-2 Edition 4.0 2014-02(Implied compliance)
    Safety of secondary cells and batteries: IEC 62133:2012Battery: Lithium-ion, FH2410 10.4Ah x25.2VDC x1 pc (Implied compliance)
    Biocompatibility Standards
    Biocompatibility: Comply with ISO 10993-1, FDA GuidanceCytotoxicity (ISO 10993-5:2009)
    Irritation and Skin Sensitization (ISO 10993-10:2010) (for other patient-contacting materials; Joystick materials refer to predicate K070501)
    Software Standards
    Software Validation: In accordance with FDA Nov 2005 guidanceSoftware validation conducted; Moderate level of concern designation
    General Operational Limits
    Maximum safe operational incline: (Predicate 8 degrees)Maximum safe operational incline: 10 degrees (Larger than predicate)

    2. Sample sized used for the test set and the data provenance:

    This document describes the testing of a physical medical device (an electrically powered wheelchair) against engineering and safety standards. The "test set" here refers to the actual device prototypes or samples subjected to the described non-clinical tests. The number of individual devices tested is not specified, but it's standard practice in such non-clinical testing to test a representative sample size to ensure reliability and repeatability according to the testing standards (e.g., ISO and IEC).

    • Sample Size: Not explicitly stated for each test, but implied to be a sufficient number of device units to perform the required non-clinical tests according to the cited ISO/IEC standards.
    • Data Provenance: The tests are non-clinical engineering and safety performance tests, not clinical data from patients. The document does not specify the location where these tests were performed, but the manufacturer is based in China. The data would be derived from the physical testing of the device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This is not applicable as the "ground truth" for non-clinical performance and safety of a wheelchair is established by the objective measurements against established international engineering and safety standards (ISO, IEC, EN). There are no "experts" establishing a "ground truth" in the way radiologists establish ground truth for an AI diagnostic algorithm; rather, the device's performance is measured and compared against the limits and criteria defined by the standards.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    This is not applicable for non-clinical engineering and safety tests. Adjudication methods like 2+1 or 3+1 are used for establishing ground truth in clinical imaging studies, particularly for AI device evaluations where human expert consensus is required for labeling data.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This is not applicable. The device is an electrically powered wheelchair, not an AI-powered diagnostic tool, and therefore, no MRMC study or AI assistance evaluation was performed or is relevant.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    This is not applicable. The device is a physical medical device, not an algorithm, so the concept of standalone algorithm performance is irrelevant.

    7. The type of ground truth used:

    The "ground truth" for the device's performance and safety is derived from established international engineering and safety standards (ISO, IEC, EN). The device's physical and functional attributes were measured and assessed to ensure they comply with the stipulated requirements of these standards. For biocompatibility, the ground truth is established by adherence to ISO 10993 series standards and related FDA guidance, with specific tests like cytotoxicity, irritation, and skin sensitization. For software, the ground truth for its safety and functionality is established through validation against FDA guidance for "Moderate Level of Concern" software.

    8. The sample size for the training set:

    This is not applicable. The document does not describe an AI/ML device that requires a training set.

    9. How the ground truth for the training set was established:

    This is not applicable, as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1