510(k) Data Aggregation

    K Number
    DEN150057

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2016-12-19

    (376 days)

    Product Code
    Regulation Number
    862.1092
    Type
    Direct
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Elecsys AMH system, consisting of the Elecsys AMH CalSet, PreciControl AMH, and AMH CalCheck 5, is intended for use in the in vitro quantitative determination of anti-Müllerian hormone (AMH) in human serum and lithium heparin plasma. The determination of AMH is used for the assessment of ovarian reserve in women presenting to fertility clinics. This system is intended to distinguish between women presenting with AFC (antral follicle count) values >15 (high ovarian reserve) and women with AFC values ≤15 (normal or diminished ovarian reserve). This system is intended to be used for assessing the ovarian reserve in conjunction with other clinical and laboratory findings before starting any fertility therapy. The Elecsys AMH system is not intended to be used for monitoring of women undergoing controlled ovarian stimulation in an Assisted Reproduction Technology program.

    The Elecsys AMH system is intended for use on cobas e 411 analyzer.

    AMH CalSet is used for calibrating the quantitative Elecsys AMH assay.

    PreciControl AMH is used for quality control of the Elecsys AMH assay.

    AMH CalCheck 5 is an assayed control for use in calibration verification and for use in the verification of the assay range established for the Elecsys AMH assay.

    Device Description

    The Elecsys AMH reagent working solutions are packed in Rack Pack (kit placed on analyzer), which include:

    • Streptavidin-coated microparticles (transparent cap). 1 bottle. 6.5 mL: Streptavidin-. coated microparticles 0.72 mg/mL; preservative.
    • . Reagent 1: Anti-AMH-Ab~biotin (gray cap), 1 bottle, 8 mL: Biotinylated monoclonal anti-AMH antibody (mouse) 1.0 mg/L, phosphate buffer 50 mmol/L, pH 7.5: preservative.
    • Reagent 2: Anti-AMH-Ab~Ru(bpy)32+ (black cap), 1 bottle, 8 mL: Monoclonal . anti-AMH antibody (mouse) labeled with ruthenium complex 1.0 mg/L, phosphate buffer 50 mmol/L, pH 7.5; preservative.

    The AMH CalSet is a lyophilized equine serum matrix with endogenous AMH (Cal 1) and a lyophilized equine serum matrix with added bovine AMH (Cal 2). The CalSet includes:

    • AMH Cal 1: approximately 0.04 ng/mL endogenous AMH in an equine serum matrix, . preservative.
    • . AMH Cal 2: approximately 8 ng/mL bovine AMH in an equine serum matrix, preservative.

    PreciControl AMH is a lyophilized equine serum matrix with added bovine AMH (male fetal bovine serum) in two concentration ranges. The controls are used for monitoring the accuracy and precision of the Elecsys AMH assay. PreciControl AMH includes:

    • . PC AMH 1: approximately 1 ng/mL bovine AMH in an equine serum matrix, preservative.
    • PC AMH 2: approximately 5 ng/mL bovine AMH in an equine serum matrix, ● preservative.

    AMH CalCheck 5 is a (b) (4) (b) (4) (b) (4)

    AI/ML Overview

    Here's an analysis of the acceptance criteria and study details for the Elecsys AMH system, based on the provided text:

    Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific CriteriaReported Device Performance
    PrecisionDefined by Repeatability (Within Run) and Intermediate Precision (%CV) for various AMH levels.Internal Precision Study:- Control 1 (1.14 ng/mL): Repeatability 1.4% CV, Intermediate Precision 1.6% CV- Control 2 (5.61 ng/mL): Repeatability 1.6% CV, Intermediate Precision 1.8% CV- Human Serum 1 (0.055 ng/mL): Repeatability 1.6% CV, Intermediate Precision 2.3% CV- Human Serum 2 (1.05 ng/mL): Repeatability 1.3% CV, Intermediate Precision 1.8% CV- Human Serum 3 (3.56 ng/mL): Repeatability 1.5% CV, Intermediate Precision 1.6% CV- Human Serum 4 (11.7 ng/mL): Repeatability 1.0% CV, Intermediate Precision 1.2% CV- Human Serum 5 (19.0 ng/mL): Repeatability 1.4% CV, Intermediate Precision 1.5% CVReproducibility (External Sites):- Total %CV ranged from 3.45% (Serum 3, 3.55 ng/mL) to 5.24% (Control 1, 0.93 ng/mL)- Repeatability %CV ranged from 1.41% (Serum 3) to 1.89% (Control 2)
    Linearity/Reportable RangeEvaluation according to CLSI EP6-A; linear regression with high correlation (r) to support claimed measuring range.Three sample sets showed slopes close to 1 (1.0091, 1.0054, 1.0063) and correlation coefficients (r) of 0.999. Supports claimed measuring range of 0.03 ng/mL - 23 ng/mL.
    DilutionDemonstrated ability to dilute samples with AMH concentrations above the measuring range, both manually and automatically, with Diluent Universal 2 (1:2 ratio).Data supports dilution with Diluent Universal 2 (1:2 ratio) for samples > 10 ng/mL, with automatic adjustment by cobas e software or manual multiplication.
    TraceabilityAdequate traceability plan to minimize the risk of drift; traceability to manufacturer's internal reference standards. Target values for Calibrators.AMH CalSet traceable to manufacturer's internal reference standards (Master Calibrators). Target values for Calibrator 1 (<0.1 ng/mL) and Calibrator 2 (8.0 ng/mL, range 7.2-8.8 ng/mL) provided. Assessment found acceptable.
    Stability (Shelf Life)Assay reagents, calibrators, and controls stable for at least 15 months unopened at 2-8°C. Open Stability: 12 weeks for reagents (2-8°C), 8 weeks on board for reagents, single use for CalSet/PreciControl AMH, 3 hours on board for CalCheck 5.Reagents, calibrators, and controls stable for 15 months unopened at 2-8°C. Opened AMH reagents stable for 12 weeks at 2-8°C, 8 weeks on board. Lyophilized AMH CalSet and PreciControl AMH are for single use only on board. Reconstituted AMH CalCheck 5 stable for 3 hours on board. Protocols and acceptance criteria reviewed and found adequate.
    Stability (Sample)Serum and Li-Heparin plasma stable for specified conditions (temperature, freeze/thaw).Serum and Li-Heparin plasma stable for 3 days at 15-25°C, 5 days at 2-8°C, 6 months at -25°C to -15°C, freeze only once.
    Detection LimitsDetermined in accordance with CLSI EP17-A2 (LoB, LoD, LoQ).LoB: 0.007 ng/mL, LoD: 0.01 ng/mL, LoQ: 0.03 ng/mL (with within-laboratory precision ≤ 20%).
    Analytical Specificity (Endogenous Interference)Interference considered significant if analyte recovery is outside of ± 10%. Highest concentrations of various endogenous substances tested with no significant interference.No significant interference (recovery within ± 10%) observed for Hemoglobin (1000 mg/dL), Biotin* (30 ng/mL), Intralipid (1000 mg/dL), Bilirubin (66 mg/dL), Rheumatoid Factor (1000 IU/mL), HAMA (805 ug/L), Human IgG (2.5 g/dL), Human IgM (0.5 g/dL), Human IgA (1.8 g/dL). *Labeling includes a limitation for high biotin doses.
    Analytical Specificity (Exogenous Interference)Interference considered significant if analyte recovery is outside of ± 10%. Highest concentrations of various exogenous substances tested with no significant interference.No significant interference observed for a range of pharmaceutical compounds at various concentrations (e.g., Acetylcysteine 150 mg/L, Ampicillin-Na 1000 mg/L, Ascorbic acid 300 mg/L, etc.). Labeling includes specific limitations for Cetrotide, Ovitrelle, Endometrin, and Follistatin.
    Cross-reactivityCross-reactivity observed to be <0.1% at 0 ng/mL AMH for specified cross-reactants.<0.1% cross-reactivity observed for Inhibin A (100 ng/mL), Activin A (100 ng/mL), LH (500 mIU/mL), FSH (500 mIU/mL).
    Hook EffectNo hook effect observed up to 14000 ng/mL AMH.No hook effect observed up to 14000 ng/mL AMH.
    Clinical Performance (Prediction of AFC > 15)Prospective clinical study demonstrating clinical performance; adequate clinical performance relative to a well-accepted comparator (AFC). AFC cutoff (15) correlated with AMH cutoff (1.77 ng/mL). PPV, NPV, Sensitivity, Specificity reported with 95% CI.A multicenter, prospective, non-interventional study (N=856 women) correlated AMH values with AFC. Used an AMH cutoff of 1.77 ng/mL for predicting AFC > 15. Results: PPV 75.2% (71.3-78.8%), NPV 84.3% (80.0-88.1%), Sensitivity 88.3% (85.0-91.2%), Specificity 68.3% (63.6-72.8%).
    Matrix EquivalenceDemonstration that Li-Heparin plasma samples perform equivalently to serum samples.75 matched serum/Li-Heparin plasma samples tested. Passing/Bablok regression analysis yielded a slope of 1.017, intercept of -0.00186, and correlation coefficient (r) of 0.999, demonstrating equivalence.
    Reference IntervalsReference intervals generated by testing an adequate number of samples from apparently healthy normal individuals in the intended use population.Age-dependent reference ranges established in 718 healthy females (20-44 years). Detailed percentile values (2.5%, 5%, Median, 95%, 97.5%) with 95% CIs provided for 5-year age groups.
    LabelingSufficient and satisfies 21 CFR Parts 801 and 809 and special controls, including warning statement about use in conjunction with other clinical and laboratory findings and AFC.Labeling deemed sufficient. Includes requirements for use in conjunction with other clinical/laboratory findings and AFC before fertility therapy.

    Study Details for Clinical Performance (Prediction of AFC > 15)

    1. Test Set Sample Size: N = 856 women
    2. Data Provenance:
      • Country of Origin: United States (samples collected at 13 geographically diverse sites).
      • Retrospective or Prospective: Prospective, non-interventional study.
    3. Number of Experts and Qualifications for Ground Truth: The document does not specify the number of experts or their qualifications for determining the Antral Follicle Count (AFC) which served as the ground truth. It states that "AFC was determined by transvaginal ultrasonography (TVUS)". It's implied that trained medical professionals (e.g., sonographers, gynecologists, reproductive endocrinologists) performed the TVUS, but specific numbers and experience levels are not provided.
    4. Adjudication Method: Not specified. It only states that AFC was determined by TVUS.
    5. Multi Reader Multi Case (MRMC) Comparative Effectiveness Study (AI vs. without AI assistance): Not applicable. This device is an immunoassay system, not an AI-assisted diagnostic imaging or interpretation tool for human readers. It provides a quantitative AMH value.
    6. Standalone Performance (Algorithm only without human-in-the-loop): Yes, the study evaluates the standalone performance of the Elecsys AMH system. The AMH values determined by the system are correlated directly with the AFC values, operating without a human interpretation loop of the AMH value itself (though the AFC ground truth is human-derived). The reported Sensitivity, Specificity, PPV, and NPV are for the device's ability to predict AFC > 15 at a given cutoff.
    7. Type of Ground Truth: "Antral Follicle Count (AFCS)" determined by transvaginal ultrasonography (TVUS).
    8. Training Set Sample Size: The document does not explicitly mention a separate "training set" for the AMH assay itself in the context of predicting AFC. The N=856 women study is referred to as the "Clinical Studies" and the results from this study are presented as the device's clinical performance. Immunoassay development typically involves extensive analytical validation (precision, linearity, specificity, etc.) and calibration, which would use dedicated sample sets, but these are not explicitly labeled as "training sets" in the AI sense. The cut-off of 1.77 ng/mL for predicting AFC > 15 would have been established during the development and optimization phase of the assay, likely using a dataset from which this clinical study was drawn or informed.
    9. How Ground Truth for Training Set was Established: Not explicitly detailed in the provided text for a specific "training set." For the purpose of establishing the AMH cutoff (1.77 ng/mL) which aligns with AFC > 15, it would have involved analyzing AMH levels and corresponding AFC values from a cohort of patients. The AFC ground truth would have been established via transvaginal ultrasonography, similar to the method used in the main clinical study, guided by a clinical definition of antral follicles (2-10 mm).