Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K220176

    Validate with FDA (Live)

    Device Name
    Elecsys AFP
    Manufacturer
    Date Cleared
    2022-09-15

    (237 days)

    Product Code
    Regulation Number
    866.6010
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Immunoassay for the in vitro quantitative determination of c1-fetoprotein in human serum and in the management of patients with non-seminomatous germ cell tumors.

    The electrochemiluminescence immunoassay "ECLIA" is intended for use on the cobas e immunoassay analyzers.

    Device Description

    Immunoassay for the in vitro quantitative determination of αι-fetoprotein in human serum and plasma to aid in the management of patients with non-seminomatous germ cell tumors.

    The electrochemiluminescence immunoassay "ECLIA" is intended for use on cobas e immunoassay analyzers.

    Elecsys AFP utilizes a sandwich test principle and has a total test duration of 18 minutes.

    • 1st incubation: 10 uL of sample, a biotinylated monoclonal AFP specific antibody, and . a monoclonal AFP specific antibody labeled with a ruthenium complex® react to form a sandwich complex.
    • 2nd incubation: After addition of streptavidin-coated microparticles, the complex . becomes bound to the solid phase via interaction of biotin and streptavidin.
    • The reaction mixture is aspirated into the measuring cell where the microparticles are . magnetically captured onto the surface of the electrode. Unbound substances are then removed with ProCell/ProCell M. Application of a voltage to the electrode then induces chemiluminescent emission which is measured by a photomultiplier.
    • Results are determined via a calibration curve which is instrument-specifically . generated by 2 point calibration and a master curve provided via the reagent barcode or e barcode.
    • a) Tris(2,2'-bipyridyl)ruthenium(II)-complex (Ru(bpy) )

    The reagent rackpack (M, R1, R2) is labeled as AFP:

    • M: Streptavidin-coated microparticles (transparent cap), 1 bottle. 12 mL: Streptavidin-. coated microparticles 0.72 mg/mL; preservative.
    • R1: Anti-AFP-Ab~biotin (gray cap), 1 bottle, 17 mL: Biotinylated monoclonal anti-. AFP antibodies (mouse) 4.5 mg/L; phosphate buffer 100 mmol/L, pH 6.0; preservative.
    • R2: Anti-AFP-Ab~Ru(bpy) (black cap), 1 bottle, 17 mL: Monoclonal anti-AFP . antibodies (mouse) labeled with ruthenium complex 12.0 mg/L; phosphate buffer 100 mmol/L, pH 6.0; preservative.
    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the Elecsys AFP immunoassay. While it details numerous non-clinical tests conducted to prove the device's performance, it does not describe a study involving a "human-in-the-loop" or "algorithm-only" performance study in the context of an AI/ML device. The device described is an "immunological test system" and its performance is evaluated based on quantitative analytical metrics rather than the type of diagnostic performance (e.g., sensitivity, specificity, accuracy) typically associated with AI/ML devices interpreting medical images or complex data for diagnostic purposes.

    Therefore, many of the requested items (e.g., number of experts, adjudication methods, MRMC studies, effect size of human reader improvement with AI) are not applicable or cannot be extracted from the provided text for this specific device.

    However, I can extract information related to the acceptance criteria and performance of the described immunoassay device based on the non-clinical tests performed.

    Here's a breakdown of the available information based on your request, focusing on what can be extracted and noting what cannot (due to the nature of the device and the provided document):


    Device: Elecsys AFP Immunoassay (K220176)

    Purpose of the study (as described in the document): To demonstrate substantial equivalence of the updated Elecsys AFP assay to a predicate device (K981282) by improving biotin tolerance and reducing streptavidin interference. The studies are non-clinical evaluations of the analytical performance of the immunoassay.

    1. Table of Acceptance Criteria and Reported Device Performance

    The document states that "All predefined acceptance criteria was met" for the various tests. While the specific numerical acceptance criteria (e.g., acceptable range for precision, linearity, interference limits) are not fully detailed, the reported performance is often given or implied by the "met" statement and the claimed range/limits.

    Test CategorySpecific TestAcceptance Criteria (Implied/Stated)Reported Device Performance/Conclusion
    PrecisionRepeatability (within-run)Predefined criteria (CLSI EP05-A3)All predefined acceptance criteria was met.
    Intermediate Precision (within-laboratory)Predefined criteria (CLSI EP05-A3)All predefined acceptance criteria was met.
    ReproducibilityLot-to-lot ReproducibilityPredefined criteriaAll predefined acceptance criteria was met.
    Analytical SensitivityLimit of Blank (LoB)Determined according to CLSI EP17-A2LoB claim in labeling: 0.75 IU/mL
    Limit of Detection (LoD)Determined according to CLSI EP17-A2LoD claim in labeling: 1.5 IU/mL
    Limit of Quantitation (LoQ)Determined according to CLSI EP17-A2LoQ claim in labeling: 1.5 IU/mL
    Analytical RangeLinearityEvaluated according to CLSI EP06-A (linear in defined range)Linearity confirmed in range 0.58 IU/mL - 1129 IU/mL. Measuring range claimed: 1.50-1000 IU/mL.
    High-Dose Hook Effect (HDHE)No hook effect observed up to a certain concentrationNo hook effect seen up to 1,000,000 (concentration units not specified, likely referring to IU/mL).
    InterferenceEndogenous Substances (7 types)Predefined criteria (acceptable interference limits)All predefined acceptance criteria was met. Proposed labeling claims: Biotin ≤ 1200 ng/mL, etc.
    Exogenous/Pharmaceutical Compounds (27 types)Predefined criteria (no interference observed)All predefined acceptance criteria was met. No interference was observed.
    Method ComparisonCandidate vs. Current AssayAcceptable agreement (implied by regression and tau values)Linear regression: y = -0.148 + 0.969x, r = 0.999. Passing/Bablok: y = 0.101 + 0.965x, tau = 0.985.
    Sample TypeAnticoagulant EffectAcceptable values compared to native serumAll predefined acceptance criteria was met, supporting use of serum, Li-Heparin, K2-EDTA, K3-EDTA plasma.
    StabilityOn-board Reagent StabilityAcceptable performance for claimed durationStable for up to 28 days (4 weeks) on analyzers.
    Lot Calibration FrequencyAcceptable performance for claimed durationCalibration recommended every 4 weeks (1 month).

    2. Sample Size Used for the Test Set and Data Provenance

    • Precision (Repeatability, Intermediate Precision): Not explicitly stated, but CLSI EP05-A3 typically involves multiple runs over several days with specific sample replicates.
    • Lot-to-lot Reproducibility: Used "three reagent lots."
    • Limit of Blank/Detection/Quantitation: Not explicitly stated, but CLSI EP17-A2 involves specific sample types and replicates.
    • Linearity: Evaluated on "one cobas e 601 analyzer" using samples spanning the linearity range, but the number of distinct samples or replicates is not specified beyond "0.58 IU/mL - 1129 IU/mL".
    • High-Dose Hook Effect: Assessed on "one cobas e 601 analyzer" in "two-fold determination" for "both samples."
    • Interference (Endogenous): "Seven endogenous substances were evaluated."
    • Interference (Exogenous): "17 commonly used and ten specially used pharmaceutical compounds" were evaluated.
    • Method Comparison: "A total of 181 serum sample concentrations were between 1.58 and 966 IU/mL."
    • Anticoagulant Effect: "native, single-donor samples drawn into serum, Li-Heparin, K2-EDTA, and K3-EDTA plasma primary tubes." (Number of samples or donors not specified).
    • Data Provenance: The document does not specify the country of origin for the samples/data or if they were retrospective or prospective. Given the nature of analytical validation studies for IVD devices, they are typically prospective studies performed in a controlled laboratory setting.

    3. Number of Experts Used to Establish Ground Truth and Qualifications:

    • Not Applicable. This is an immunoassay, not an AI/ML device relying on expert interpretation for ground truth. The "ground truth" for these analytical studies is derived from established laboratory methods, reference materials, and defined analytical ranges.

    4. Adjudication Method for the Test Set:

    • Not Applicable. No human adjudication process as typically seen in AI/ML diagnostic studies.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

    • No. This is not an AI/ML device; therefore, an MRMC study comparing human readers with and without AI assistance is not relevant or was not performed.

    6. If a Standalone (algorithm only without human-in-the-loop performance) was done:

    • Not Applicable (in the context of AI/ML). The device itself is an automated immunoassay. Its performance is inherently "standalone" in that it produces a quantitative result without human "interpretation" of a complex output, but this is a very different concept than an AI algorithm's standalone diagnostic performance. The document describes the analytical performance of the automated assay system.

    7. The Type of Ground Truth Used:

    • The "ground truth" for these types of analytical studies is based on:
      • Reference Methods/Materials: For linearity, analytical range, LoB/LoD/LoQ determinations, often using diluted/spiked samples with known concentrations.
      • Comparative Analysis: The "method comparison" compares the new assay's results against the existing (predicate) assay.
      • Defined Limits/Spiked Samples: For interference studies, known amounts of interferents are added to samples.
      • Established CLSI Guidelines: The studies explicitly state adherence to CLSI (Clinical and Laboratory Standards Institute) guidelines (e.g., EP05-A3, EP17-A2, EP06-A) for methodology, which dictate how "truth" is assessed for analytical performance.

    8. The Sample Size for the Training Set:

    • Not Applicable. This is not an AI/ML device that requires a "training set" in the computational sense. The device's performance is based on its chemical and technological design, and its validation involves analytical testing rather than machine learning on a dataset.

    9. How the Ground Truth for the Training Set was Established:

    • Not Applicable. As above, no "training set" in the AI/ML context.

    Ask a Question

    Ask a specific question about this device

    K Number
    K981282

    Validate with FDA (Live)

    Device Name
    ELECSYS AFP
    Date Cleared
    1998-05-01

    (23 days)

    Product Code
    Regulation Number
    866.6010
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Immunoassay for the in vitro quantitative determination of alpha fetoprotein (AFP) in human serum and plasma.
    The electrochemiluminescence immunoassay “ECLIA” is intended for use on the Roche Diagnostics/Boehringer Mannheim Elecsys 1010 and 2010 immunoassay analyzers.
    Immunoassay for the in vitro quantitative determination of alpha-fetoprotein in human serum and plasma to aid in the management of patients with non-seminomatous germ cell tumors.
    The electrochemiluminescence immunoassay "ECLIA" is intended for use on the Boehringer Mannheim Elecsys 1010 and 2010 immunoassay analyzers.

    Device Description

    The Elecsys® test principle is based on sandwich principle. Total duration of assay: 18 minutes (37° C).
    -1st incubation (9 minutes): Sample (30 µL), biotinylated monoclonal AFP- specific antibody (60 µL), and a monoclonal AFP-specific antibody labeled with a ruthenium complex (60 µL) react to form a sandwich complex.
    -2nd incubation (9 minutes): After addition of streptavidin-coated microparticles (50 µL), the complex is bound to the solid phase via interaction of biotin and streptavidin.
    •The reaction mixture is aspirated into the measuring cell where the microparticles are magnetically captured onto the surface of the electrode. Unbound substances are then removed with ProCell. Application of a voltage to the electrode then induces chemiluminescent emission which is measured by a photomultiplier (0.4 second read frame).
    •Results are determined via a calibration curve which is instrument- specifically generated by 2-point calibration and a master curve provided via the reagent bar code.

    AI/ML Overview

    Here's an analysis of the provided information, focusing on the acceptance criteria and study details for the Elecsys® AFP Assay on the Elecsys® 1010 analyzer.

    1. Table of Acceptance Criteria and Reported Device Performance

    The submission doesn't explicitly state "acceptance criteria" but rather presents "Performance Characteristics" for both the new device (Elecsys® 1010) and the predicate (Elecsys® 2010). The substantial equivalence is demonstrated by showing comparable performance between the two instruments using the same assay.

    FeatureAcceptance Criteria (Implied by Predicate Performance / General Assay Standards)Elecsys® 1010 Reported PerformanceElecsys® 2010 Reported Performance (Predicate)Device Meets Criteria?
    Precision - Within-Run %CV
    HS1Comparable to Predicate (2.0%)1.01%2.0%Yes
    HS2Comparable to Predicate (1.5%)1.02%1.5%Yes
    HS3Comparable to Predicate (2.0%)1.52%2.0%Yes
    Control 1Comparable to Predicate (2.8%)1.29%2.8%Yes
    Control 2Comparable to Predicate (2.2%)1.39%2.2%Yes
    Precision - Total %CV
    HS1Comparable to Predicate (3.1%)2.25%3.1%Yes
    HS2Comparable to Predicate (2.4%)2.69%2.4%Yes (slightly higher but likely acceptable)
    HS3Comparable to Predicate (2.8%)4.61%2.8%Yes (slightly higher but likely acceptable)
    Control 1Comparable to Predicate (3.4%)1.90%3.4%Yes
    Control 2Comparable to Predicate (2.7%)2.29%2.7%Yes
    Lower Detection Limit0.5 IU/mL0.5 IU/mL0.5 IU/mLYes
    Linearity0.5 - 1000 IU/mL (deviation ±10%)0.5 - 1000 IU/mL (deviation ±10%)0.5 - 1000 IU/mL (deviation ±10%)Yes
    Method Comparison correlation (r)> 0.97 (common for method comparison)0.992Not applicable (vs. itself)Yes
    Method Comparison slopeClose to 1 (e.g., 0.95-1.05)0.980 (Least Squares), 1.031 (Passing/Bablok)Not applicable (vs. itself)Yes
    Method Comparison interceptClose to 00.639 (Least Squares), -0.208 (Passing/Bablok)Not applicable (vs. itself)Yes
    Hook EffectNo Hook Effect up to 1,000,000 IU/mL AFPNo Hook Effect up to 1,000,000 IU/mL AFPNo Hook Effect up to 1,000,000 IU/mL AFPYes

    Notes on Acceptance Criteria:

    • For a 510(k) submission demonstrating substantial equivalence, the primary acceptance criterion is that the new device performs as well as or comparably to the predicate device for all relevant performance characteristics.
    • The document does not specify explicit numerical acceptance criteria for precision (e.g., "%CV must be < X"). Instead, it provides the observed performance for both the new and predicate devices, and the implicit criterion is that the new device's performance should be similar to or better than the predicate's. The reported values for the Elecsys 1010 generally meet or exceed the performance of the Elecsys 2010.

    2. Sample Sizes Used for the Test Set and Data Provenance

    • Precision Test Set: Not explicitly clinical samples.
      • HS (High Sensitivity) Samples: 3 levels (HS1, HS2, HS3). For each level, N=60 measurements were performed for both Elecsys® 1010 and Elecsys® 2010 to derive Within-Run and Total %CV. These are typically prepared human serum samples with known AFP concentrations.
      • Control Samples: 2 levels (Control 1, Control 2). For each level, N=60 measurements were performed for both Elecsys® 1010 and Elecsys® 2010 to derive Within-Run and Total %CV. These are typically manufactured quality control materials.
    • Method Comparison Test Set: N=153 samples.
      • The document does not specify the country of origin of the data or whether it was retrospective or prospective. Given the nature of a laboratory assay validation for a 510(k), these samples would typically be human serum and plasma samples, likely from a patient population, and the study would be prospective to collect data simultaneously on both instruments.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not applicable to this type of device. The Elecsys® AFP Assay is an in vitro diagnostic (IVD) device that measures a biomarker (Alpha-Fetoprotein) quantitatively. The "ground truth" for such measurements is established by:

    • Reference methods or certified reference materials for calibration and accuracy.
    • The inherent precision of the assay itself compared to well-established analytical performance standards.
    • Comparison to a legally marketed predicate device.
      There are no human "experts" establishing a subjective "ground truth" for the test results in the way there would be for image interpretation or clinical diagnosis by human readers.

    4. Adjudication Method for the Test Set

    This information is not applicable as there is no subjective interpretation requiring adjudication. Performance is based on quantitative analytical measurements.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No, an MRMC study was not done. This is an in vitro diagnostic device for quantitative measurement of a biomarker, not an imaging device or an AI-powered diagnostic tool requiring human interpretation. Therefore, there is no concept of "human readers improving with AI assistance" in this context.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

    • Yes, in essence, the performance evaluation represents a standalone performance. The Elecsys® AFP Assay on the Elecsys® 1010 (or 2010) is an automated immunoassay. The device performs the measurement, and the result is a quantitative number. There isn't a "human-in-the-loop" performing interpretation of the raw signal that then gets augmented by the device. The device is the algorithm/system that produces the final result.

    7. The Type of Ground Truth Used

    The "ground truth" for this IVD device is based on:

    • Analytical Standards: For precision, the "ground truth" is statistical reproducibility (e.g., %CV around the mean concentration).
    • Reference Methods/Materials: For accuracy and linearity, the "ground truth" is often established via comparison to a more definitive reference method or against samples with assigned values from certified reference materials.
    • Predicate Device Performance: For substantial equivalence, the "ground truth" for acceptable performance is largely defined by the performance characteristics of the legally marketed predicate device (Elecsys® AFP Assay on Elecsys® 2010). The method comparison study directly demonstrates how closely the new device's results align with the predicate's results.

    8. The Sample Size for the Training Set

    • The document does not explicitly state a training set size in the context of machine learning or AI. This is not an AI/ML device.
    • For an IVD assay, "training" primarily refers to:
      • Calibration: The instrument is calibrated using specific calibrator materials. The document mentions "2-point calibration and a master curve provided via the reagent bar code." The sample size for these calibrators isn't typically given as a "training set" but rather as a set of material used to define the measurement curve.
      • Method Development & Validation: The assay itself was developed and optimized using various samples to establish reagents, reaction conditions, and measurement parameters. This "development data" is not typically reported as a "training set" in 510(k) summaries but is part of the overall assay validation.

    9. How the Ground Truth for the Training Set Was Established

    As noted above, "training set" is not a direct concept here for an IVD.

    • For Calibration: The calibrator materials would have their AFP concentrations assigned through rigorous analytical methods, often traceable to international reference standards or highly accurate reference laboratories.
    • For Method Development: The "ground truth" during development would have been established by comparing prototype assay results against established reference methods or by using samples with known, validated AFP concentrations.
    Ask a Question

    Ask a specific question about this device

    K Number
    K973351

    Validate with FDA (Live)

    Device Name
    ELECSYS AFP
    Date Cleared
    1997-11-21

    (77 days)

    Product Code
    Regulation Number
    866.6010
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Immunoassay for the in vitro quantitative determination of α₁-fetoprotein in human serum and plasma to aid in the management of patients with non-seminomatous germ cell tumors. The electrochemiluminescence immunoassay "ECLIA" is intended for use on the Boehringer Mannheim Elecsys immunoassay analyzers.

    Device Description

    The Elecsys® test principle is based on the sandwich principle. Total duration of assay: 18 minutes. 1st incubation: 15 µl of sample, a biotinylated monoclonal AFP-specific antibody and a monoclonal AFP-specific antibody labeled with a ruthenium complex react to form a sandwich complex. 2nd incubation: after the addition of streptavidin-coated microparticles, the complex becomes bound to the solid phase via interaction of biotin and streptavidin. The reaction mixture is aspirated into the measuring cell where the microparticles are magnetically captured onto the surface of the electrode. Unbound substances are then removed with ProCell. Application of a voltage to the electrode then induces chemiluminescent emission which is measured by a photomultiplier. Results are determined via a calibration curve which is instrumentspecifically generated by 2-point calibration and a master curve provided via the reagent bar code.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Elecsys AFP device, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided text for K973351 primarily describes the device and its comparison to a predicate device, but it does not explicitly state specific numerical acceptance criteria for each performance characteristic. Instead, it lists the types of studies performed to demonstrate substantial equivalence.

    However, based on the types of studies listed, we can infer the aspects of performance that were evaluated and for which the device was considered acceptable. The "Reported Device Performance" below reflects what the submission claims to have demonstrated.

    Performance CharacteristicAcceptance Criteria (Inferred from Study Type)Reported Device Performance (Summary from Submission)
    PrecisionTo meet NCCLS recommendationsEvaluated according to NCCLS recommendations. (Details not provided in the summary, but implies acceptable precision.)
    Lower Detection LimitTo be accurately determinableDetermined. (Specific value not provided, but implies acceptable sensitivity for clinical use.)
    LinearityTo demonstrate linear response across rangeDemonstrated. (Implies a reliable and proportional measurement across its intended assay range.)
    Correlation with PredicateTo show substantial equivalenceCorrelated with the predicate device (Enzymun-Test® AFP). (Implies comparable results to the reference method.) This was further supported by comparisons with samples from serially monitored patients diagnosed and treated for testicular cancer.
    Interference (Endogenous)No significant interference in resultsEvaluated for the effect of various endogenous substances (hemoglobin, biotin, triglyceride, lipemia, and rheumatoid factor). (Implies no significant interference found, or interference within acceptable limits.)
    Interference (Pharmaceutical)No significant interference in resultsEvaluated for the effect of commonly used pharmaceutical compounds. (Implies no significant interference found, or interference within acceptable limits.)
    Sample Matrix EquivalenceSerum and plasma results to be comparableSerum and plasma sample comparisons performed. (Implies comparable performance between serum and plasma samples.)
    Hook EffectTo be determined (and generally absent/managed)Determination of hook effect. (Implies the device handles very high concentrations without false low readings, or that the effect is understood and managed if present.)
    StabilityTo ensure reliable performance over timeStability studies performed. (Implies the device maintains its performance characteristics over its stated shelf-life and in-use periods.)
    Clinical PerformanceAid in patient management (comparable to predicate)Comparisons of the Elecsys AFP and the predicate device were performed with samples from serially monitored patients diagnosed and treated for testicular cancer. (This surrogate for clinical utility demonstrates that the new device performs similarly in a real-world clinical context to the approved predicate, thus reasonably aiding in patient management.)

    2. Sample Size Used for the Test Set and Data Provenance

    The submission does not explicitly state the numerical sample sizes used for the various tests mentioned (precision, linearity, interference, etc.). It only describes the types of studies performed.

    The data provenance is not explicitly mentioned (e.g., country of origin). The studies appear to be retrospective in nature, as they involve comparisons with an existing predicate device and samples from "serially monitored patients diagnosed and treated for testicular cancer," suggesting pre-collected samples.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided in the submission. For an in vitro diagnostic (IVD) like the Elecsys AFP, the "ground truth" for performance studies typically comes from reference methods (like the predicate device) or known concentrations/spiked samples, rather than human expert interpretation of images or clinical data. However, the diagnosis of "testicular cancer" in the patient samples would have been established by medical professionals, but the details of who and how many are not discussed in relation to the device's technical validation.

    4. Adjudication Method for the Test Set

    This is not applicable and therefore not mentioned for this type of IVD device. Adjudication methods (like 2+1, 3+1) are typically used in studies involving human readers or interpretation of complex medical data (e.g., imaging studies) where there might be inter-reader variability in establishing ground truth. For an automated immunoassay, the "ground truth" is established through analytical methods and reference standards.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, an MRMC comparative effectiveness study was not conducted and is not applicable for this type of IVD device. This device is an automated immunoassay, not an AI-powered diagnostic image analysis tool that assists human readers.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, the studies described are standalone performance evaluations of the Elecsys AFP immunoassay. The device itself is an automated system for quantitative determination of AFP; there is no human-in-the-loop in the assay run or measurement process. Results are then interpreted by clinicians, but the device's performance characteristics (precision, linearity, correlation, etc.) are inherent to the algorithm and instrument.

    7. The Type of Ground Truth Used

    The ground truth used for various aspects of the study includes:

    • Reference materials/standards: For linearity, lower detection limit, and potentially stability, calibrated reference materials or spiked samples of known concentrations would be used.
    • Predicate device: The Enzymun-Test® AFP served as the primary "ground truth" or reference for comparative performance (correlation).
    • Known clinical status: For the patient sample comparisons, the "serially monitored patients diagnosed and treated for testicular cancer" represent samples with a known clinical diagnosis, allowing for comparison of the new device's performance in a clinically relevant population against the predicate.

    8. The Sample Size for the Training Set

    This information is not applicable and therefore not provided. The Elecsys AFP is a chemistry immunoassay, not a machine learning or AI algorithm that requires a "training set" in the conventional sense. The "calibration curve" is instrument-specifically generated by 2-point calibration and a master curve provided via the reagent bar code; this is a form of calibration, not model training.

    9. How the Ground Truth for the Training Set Was Established

    As explained in point 8, there is no "training set" in the context of machine learning for this device. The closest equivalent would be the calibration process. The ground truth for this would be:

    • Calibrators with known, traceable concentrations: Used to establish the device's response curve. The master curve is provided via the reagent bar code, implying that the manufacturer pre-established the curve characteristics using highly controlled and validated calibrators.
    • 2-point calibration: This periodic user calibration adjusts the master curve to account for minor variations in reagents or instrument performance, using calibrators with known values.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1