Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K232672

    Validate with FDA (Live)

    Date Cleared
    2024-05-29

    (271 days)

    Product Code
    Regulation Number
    866.3990
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Genetic Signatures EasyScreen™ Gastrointestinal Parasite Detection Kit is a rapid in vitro nucleic acid amplification assay for the qualitative detection of pathogenic gastrointestinal parasite nucleic acid from the stool of patients with signs and/or symptoms of gastroenteritis. The test, based on real-time PCR, detects the nucleic acid of the following organisms:

    • · Cryptosporidium spp.
    • · Giardia intestinalis
    • Dientamoeba fragilis
    • · Entamoeba histolytica
    • Blastocystis hominis
    • · Enterocytozoon bieneusi
    • · Encephalitozoon intestinalis
    • · Cyclospora cayetanensis

    The kit is compatible with stool specimens that are unpreserved or frozen or in transport media including Cary Blair or C&S media from symptomatic patients with suspected gastroenteritis. It is required that the stool is first processed using the EasyScreen™ Sample Processing Kit. Nucleic acid extraction and real-time PCR set up are performed on the automated Genetic Signatures GS1 platform.

    This device is an in vitro diagnostic (IVD) intended to be used by trained personnel in clinical, pathology or hospital laboratories as an aid in the diagnosis of gastrointestinal illness. This test is intended for use, in conjunction with clinical presentation, laboratory findings, and epidemiological information, as an aid in the differential diagnosis of infections by Dientamoeba fragilis, Blastocystis hominis, Enterocytozoon bieneusis, Entamoeba histolytica, Encephalitozoon intestinalis, Cryptosporidium spp. (including C. parvum), and Giardia intestinalis. Results of this test should not be used as the sole basis for diagnosis, treatment, or other patient management decision. Positive results do not rule out co-infection with other organisms that are not detected by this test, and may not indicate the sole or definitive cause of patient illness. Negative results in the setting of clinical illness compatible with gastroenteritis and/or colitis may be due to infection by pathogens that are not detected by this test or non-infectious causes such as ulcerative colitis, irritable bowel syndrome, or Crohn's disease.

    Device Description

    The EasyScreen™ Gastrointestinal Parasite Detection Kit (EP005) is designed to simultaneously identify 8 potential pathogens of the gastrointestinal tract, from human stool samples. The device is only compatible with nucleic acids prepared using an EasyScreen™ Sample Processing Kit (SP008B).

    A stool sample from a patient suspected of having gastroenteritis (usually liquid or soft stool) is collected and transported to the testing laboratory. A portion of the stool material is taken using a swab or pipette and processed with the EasyScreen™ Sample Processing Kit (SP008B), which lyses cells and converts the nucleic acid to a 3base™ form.

    An aliquot of purified eluate is then added to the PCR reagents supplied in the EP005 kit, which selectively amplify the genetic targets of Cryptosporidium spp., Giardia intestinalis, Entamoeba Dientamoeba fragilis, Blastocystis hominis, Enterocytozoon histolytica. bieneusi. Encephalitozoon intestinalis and Cyclospora cayetanensis. The reaction mix is manufactured to detect an Extraction Control (EC) and features an incorporated Internal Positive Control (IPC) to determine the reliability of the extracted nucleic acid and to detect the presence of any inhibitors after extraction from the primary sample.

    Amplified targets are detected with probes labeled with fluorophores as detected by the real-time PCR platform. The PCR amplification takes approximately 150 minutes, depending on the PCR platform used. A positive control is included to ascertain that the detection reagents and analyzer are functioning correctly.

    The amplified nucleic acid targets are detected by probes labeled with fluorophores, as detected by the real-time PCR platform. If no amplification occurs for a given target, then there will not be any significant increase in fluorescence. Each probe fluoresces at a given wavelength and the signals are measured and distinguished from each other by the real-time PCR platform. The realtime PCR software interprets all data collection and provides the information for automated or manual result analysis. The assay is semi-automated.

    AI/ML Overview

    The provided text describes the performance of the Genetic Signatures EasyScreen™ Gastrointestinal Parasite Detection Kit. This device is a rapid in vitro nucleic acid amplification assay for the qualitative detection of pathogenic gastrointestinal parasite nucleic acid from human stool samples.

    Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided document:


    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a consolidated table of acceptance criteria for all aspects of the study alongside the reported performance in a single, clear format. However, acceptance criteria are stated within each section of the performance studies, and the results are then presented against those criteria. Below is a reconstructed table based on the explicit statements regarding acceptance and observed performance.

    Acceptance Criteria and Reported Device Performance

    Study AspectAcceptance Criteria (Stated)Reported Device Performance
    Analytical Sensitivity (LoD)≥95% detection of the specified target AND <95% detection at 0.5X LoD (with a minimum of 20 extraction replicates).LoD values were established for all targets in unpreserved stool or Cary Blair media. "LoD studies showed comparable performance with minimal variability observed between LoD values obtained across different isolates, PCR analyzers (...) and EP005 reagent batches with all targets showing an LoD within a ±2-fold dilution across all variables." The final LoD for each organism is provided in Table 3.
    Multisite ReproducibilityQualitative reproducibility percent agreement for targets evaluated at >1 sites (excluding C. cayetanensis) was expected to be high.For targets evaluated at >1 sites, overall site-to-site qualitative reproducibility percent agreement was 100% for all targets at 2x LoD (Low Positive) and for all targets except C. parvum at 4x LoD (Medium Positive). C. parvum at 4x LoD showed 98.9% (95% CI: 93.8-99.9). Within-site reproducibility for C. cayetanensis was 97.1% for LP and 100% for MP. All True Negative samples were 100% correctly identified. Concluded as "acceptably consistent performance."
    Analytical Specificity (Cross-Reactivity)No detection for any given target for whole organism/genome wet testing. For in silico analysis, potential cross-reactivity was categorized.Wet Testing: No cross-reactivity observed with 94 organisms and 7 media, except for three congeneric protozoa (C. muris, E. cuniculi, E. hellem) that showed positive signals. In Silico Analysis: Identified several targets with high/moderate potential for cross-reactivity. Confirmatory wet testing of synthetic RNA targets from clinically relevant protozoa showed that C. meleagridis, C. tyzzeri, C. canis, C. felis, C. muris, E. cuniculi, and E. hellem cross-reacted. (Note: Chilomastix mesnili and Entamoeba dispar did not cross-react in wet testing despite in silico prediction).
    Analytical Reactivity (Inclusivity)All isolates detected at all tested concentrations (1X-3X LoD).Eighty-two isolates representing eight target parasites were tested. The kit detected all isolates at all tested concentrations.
    Interfering SubstancesNo interference if <100% target positivity (≤9/10) or >15% change in average Ct values in test (with interferent) samples relative to control (no interferent).Two substances showed potential interference: Whole Blood at >0.63% (v/v) and Mucin at >0.75 mg/mL. All other 21 substances showed no interference.
    Microbial InterferenceNo reportable interference if 100% (10/10) target positivity with test Ct changes at or below 10%.All targets showed <10% change in average Ct values and 100% (10/10) target positivity. Concluded "no microbial interference was observed."
    Competitive InhibitionNo reportable interference if 100% (10/10) target positivity with test Ct changes at or below 10%.Moderate competitive interference was observed for D. fragilis and G. intestinalis at 10^8 copies/mL (C. cayetanensis) and 10^5 org/mL (E. histolytica) respectively, with ~12% Ct change. Interference (20% positivity) also seen for D. fragilis/B. hominis with C. parvum/E. histolytica respectively. This interference was resolved by reducing competitor concentrations (e.g., to 5x10^4 org/mL or 10^4 org/mL). All other combinations showed no reportable interference.
    Cross-Contamination (Carry Over)100% detection for pooled high positive samples and 0% detection for negative samples and negative processing control.100% (240/240) detection for pooled high positive samples. 0% (0/210) detection for negative samples. 0% (0/15) detection for negative processing control. Concluded "no reportable carry over/cross contamination."
    Specimen Stability (2-8ºC)100% positivity (10/10 replicates) for at least three weeks.All targets are stable for three weeks in unpreserved negative stool. In Cary Blair matrix, all targets except C. cayetanensis and E. histolytica are stable for three weeks; these two are stable for two weeks.
    Specimen Stability (Fresh vs. Frozen)4X LoD, 100% (10/10) positive; 2X LoD, ≥95% (≥19/20) positive; Negative, 0% (0/10) positive; Average Ct values within ±10% of baseline.All eight targets showed 100% detection at 4X LoD and ≥95% detection at 2X LoD at 4 weeks (3 weeks for B. hominis). Average Ct values were within ±10% of baseline. Supports frozen stability claim of three weeks (two weeks for B. hominis).
    Reagent Stability/Shelf-Life100% positivity of all targets (5/5 replicates) at 2X LoD.SP008B kit: All targets detected at 5/5 replicates after 38-48 months. Confirmed 24 months at 15-25°C. EP005 kit: All targets detected at 5/5 replicates after 19-32 months. Confirmed 24 months frozen (-25°C to -15°C).
    In-Use (Freeze/Thaw) Stability100% analytical targets (5/5 extraction replicates) testing positive at 2X LoD over 5 freeze-thaw cycles.At the 5th freeze-thaw, all PCR replicates at 2X LoD achieved 5/5 positivity (100% detection). Confirmed stability for up to four (4) freeze-thaw cycles.
    C&S Matrix Equivalency5X LoD, 100% (5/5 replicates) positive; 1-2X LoD, ≥95% (≥24/25) positive; Negative, 0% (0/10) positive, relative to LoD in Cary-Blair.For all targets and concentrations tested, targets diluted in C&S matrix met the acceptance criteria, demonstrating equivalence.
    Clinical Performance (Positive Percent Agreement - PPA)PPA for individual target parasites ranging between 91-99% with lower limit of 95% Cl at ≥80% compared to reference method.PPA for Cryptosporidium spp. (Prospective: 100%; Retrospective: 90.7%) and E. histolytica (Prospective: N/A (0 TP); Retrospective: 96.88%) meet this criterion. D. fragilis (Prospective: 86.67%), B. hominis (Prospective: 95.92%), G. intestinalis (Prospective: 100%; Retrospective: 97.06%), E. bieneusi (Prospective: 100%; Retrospective: 100%; Contrived: 98.68%), E. intestinalis (Retrospective: 100%; Contrived: 92.41%) are within the expected range or higher. C. cayetanensis (Prospective: 100%; Retrospective: 97.73%) is also in range.
    Clinical Performance (Negative Percent Agreement - NPA)NPA ≥99% when compared to reference method.NPA for D. fragilis (Prospective: 99.65%; Retrospective: 99.56%), C. cayetanensis (Prospective: 99.59%; Retrospective: 97.74%), Cryptosporidium spp. (Prospective: 99.66%; Retrospective: 97.3%), B. hominis (Prospective: 99.22%; Retrospective: 98.94%), E. histolytica (Prospective: 99.73%; Retrospective: 99.14%), G. intestinalis (Prospective: 99.52%; Retrospective: 94.81%), E. bieneusi (Prospective: 99.52%; Retrospective: 99.24%; Contrived: 100%), and E. intestinalis (Prospective: 99.52%; Retrospective: 100%; Contrived: 100%) are generally ≥99% or close to it, meeting or closely approaching criteria.

    2. Sample size used for the test set and the data provenance

    • Clinical Study Test Set (Analyzable Samples): A total of 1,926 analyzable specimens (out of 2,806 collected).
      • Prospective Samples (Category I & II): 1,461 samples.
      • Retrospective Clinical Samples (Category III): 265 samples.
      • Contrived Samples (Category IV): 165 samples (for E. bieneusi and E. intestinalis, which had low prevalence or availability).
    • Data Provenance:
      • Clinical Samples: Collected from four geographically diverse US clinical sites (US1, US2, US3, US4). Some retrospective procurement was also done at one US site (US4).
      • Contrived Samples: Prepared/enrolled at an Outside the US (OUS) internal site for testing at the US sites.
      • Retrospective or Prospective: Both retrospective and prospective samples were used. Prospective samples (Category I: fresh, and Category II: frozen) were collected in an "All-Comers" mode. Retrospective, positive-identified (Category III) samples were also included, along with randomly distributed negative samples, tested in a masked/blinded manner.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    The document does not specify the number or qualifications of experts used to establish the ground truth for the test set. It mentions that the reference method for the clinical studies was "two (2) well-characterized and validated Nucleic Acid Amplification Tests (NAAT) followed by bi-directional sequencing (referred to as "alternative NAAT")." The alternative NAATs were performed at an OUS internal site. The results from the alternative NAAT were compared against sequence data that met "acceptability criteria listed in Section VII(D)(1) of the Class II Special Controls guidelines," but it does not detail human expert adjudication for ground truth. The adjudication of sequence data to determine ground truth status appears to be based on pre-defined criteria rather than a panel of human experts.


    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    The document states that the true analyte status ("positive" or "negative") of each clinical sample was established for each target parasite by comparison with the reference method of alternative NAATs. These NAATs consisted of "two separate single-plex, PCR amplification tests." Amplicons from PCR-positive reactions were then subjected to "bi-directional Sanger sequencing." Samples producing sequence data that met "acceptability criteria" were reported as positive by alternative NAAT.

    There is no mention of a human expert adjudication method (like 2+1 or 3+1 consensus) for the resolution of discrepancies between multiple readers or ground truth methods. The ground truth appears to be established algorithmically or through a predefined analytical process based on the output of the reference NAATs and sequencing results.


    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not an AI/human-in-the-loop study. The device described is an in vitro diagnostic (IVD) molecular test (real-time PCR kit) for detecting pathogenic gastrointestinal parasite nucleic acid. Therefore, an MRMC comparative effectiveness study involving human readers assisted by AI is not applicable here. The device output is qualitative (detection or non-detection of specific nucleic acids), interpreted directly by the instrument software or through a validated macro-enabled Excel sheet, not by human image readers.


    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, the primary performance evaluation of the EasyScreen™ Gastrointestinal Parasite Detection Kit is a standalone (algorithm only) performance assessment. The device is a diagnostic kit that performs nucleic acid amplification and detection. The interpretation of results is automated, with the real-time PCR software interpreting data and providing information for automated or manual result analysis. The device itself (the kit and associated platform) is the "algorithm" in this context, directly generating a result based on the processed sample without human interpretation of raw signals like an image.


    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The ground truth for the clinical study was established using a molecular reference method: "two (2) well-characterized and validated Nucleic Acid Amplification Tests (NAAT) followed by bi-directional sequencing (referred to as "alternative NAAT")". Sequence data meeting predefined acceptability criteria were considered positive. This is a form of molecular ground truth, not expert consensus, pathology, or outcomes data.


    8. The sample size for the training set

    The document does not specify a separate training set size or mention a training phase for the device described. As an in vitro diagnostic (IVD) PCR kit, its "training" or "development" would involve optimizing primers, probes, and reaction conditions rather than machine learning algorithm training on a dataset in the typical sense. The studies described are validation studies (analytical and clinical performance) on test sets.


    9. How the ground truth for the training set was established

    Since a specific training set or "training phase" in the context of machine learning is not described for this IVD device, the method for establishing ground truth for a training set is not applicable or described in the provided document. The document focuses on the validation of the finalized device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1