Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K182013

    Validate with FDA (Live)

    Device Name
    EasyInstruments
    Date Cleared
    2019-03-11

    (227 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Easylnstruments are indicated for cutting, grasping, dissecting and coagulation of tissue in laparoscopic surgical procedures.

    Device Description

    The A.M.I. EasyInstruments device is a modular instrument system for use in laparoscopic highfrequency (HF) surgery and is designed for monopolar application. The Easylnstruments system comprises three-part detachable instruments, which include 1) different functional handles (EasyHandle), 2) instrument shafts of various diameters and lengths (EasyShaft), and 3) a wide range of different, application-specific attachments (EasyTips). It is intended during laparoscopic procedures to mobilize, manipulate, grasp, hold, fix, retract, dissect, cut, and coagulate tissue.

    AI/ML Overview

    This FDA 510(k) summary describes a medical device, the EasyInstruments, which are electrosurgical cutting and coagulation devices and accessories. The document focuses on demonstrating substantial equivalence to predicate devices, rather than establishing acceptance criteria and performance against those criteria in a clinical study for a new AI/software component. Therefore, much of the requested information (like number of experts, adjudication methods, MRMC studies, ground truth for test/training sets, effect size of AI improvement) is not applicable or cannot be extracted from this document because it pertains to the evaluation of AI/software in a clinical context, which is not the subject of this 510(k) for electrosurgical instruments.

    However, I can extract information related to the performance testing that was conducted to demonstrate safety and efficacy compared to predicate devices for these hardware instruments.

    Here's an attempt to answer the questions based on the provided text, noting where the requested information is not present or applicable:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state acceptance criteria in precise numerical terms or detailed performance metrics in a table. It lists numerous tests performed to demonstrate safety and performance relative to predicate devices. The "reported device performance" is summarized as "The results of these tests indicate that the EasyInstruments are substantially equivalent to the predicate devices" and "it can be concluded that the subject device does not raise new issues of safety or efficacy compared to the predicate devices."

    Below is a summary of the types of tests performed and the general outcome:

    CategoryTest PerformedReported Performance (General)
    Cleaning & SterilizationCleaning validation per AAMI TIR12 and AAMI TIR30. Sterilization validation per ISO 17665-1, ISO 11135, and ISO 11137.Demonstrated effective cleaning and sterilization.
    Longevity & IntegrityShelf life and packaging validation per ISO 11607.Device maintains integrity and sterility over its shelf life.
    BiocompatibilityBiocompatibility Risk Assessment per ISO 10993-1. Cytotoxicity testing per ISO 10993-5. Irritation and sensitization testing per ISO 10993-10.Biocompatible; no undue patient risk from materials.
    Electrical SafetyElectrical safety per IEC 60601-1 and IEC 60601-2-2. EMC per IEC 60601-1-2.Meets electrical safety and electromagnetic compatibility standards.
    Bench PerformancePerformance of disposable tips, reusable tips after reprocessing, EasyHandle reprocessing, and EasyShaft after reprocessing. Performance tests for tips included visual inspection for damage, opening angle, ease of movement in opening/closing, cutting, dissecting, grasping strength, continuity check. Performance tests for EasyShaft and EasyHandle included visual inspection (e.g., cracks, damage, warpage), continuity check, dielectric strength, and compatibility check.Maintains mechanical and electrical integrity and function.
    Thermal EffectsSide-by-side testing on ex vivo animal tissues (liver, kidney, muscle). Thermal damage assessed by measuring depth of damage at minimum, default, and maximum power settings using histology.Thermal effects are comparable and safe.
    UsabilityUsability testing per IEC 62366-1, including device assembly, disassembly, reprocessing, and application in a simulated environment.Usable and safe in simulated use scenarios.
    Overall Conclusion"The results of these tests indicate that the EasyInstruments are substantially equivalent to the predicate devices." "The equivalent indications for use, technological characteristics, and performance characteristics for the proposed EasyInstruments are assessed to be substantially equivalent to the predicate devices." "The subject device does not raise new issues of safety or efficacy compared to the predicate devices."Substantially equivalent to predicate devices in safety and efficacy.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample size for test set: Not explicitly specified for each test. For "Thermal effects on tissue," it mentions "ex vivo animal tissues (liver, kidney, and muscle)," implying a sample of tissue, but the quantity is not detailed. For bench performance, it refers to "disposable tips," "reusable tips," "EasyHandle," and "EasyShaft," suggesting multiple units were tested, but precise numbers are not provided.
    • Data provenance: Not specified. The manufacturer is based in Austria, but the location where the tests were specifically conducted or the origin of the animal tissues is not stated. The tests are non-clinical (bench, ex vivo).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. This document pertains to the 510(k) clearance of electrosurgical instruments (hardware), not an AI/software device that requires expert-established ground truth for image interpretation or diagnosis. The "ground truth" here related to performance testing (e.g., measurements of thermal damage, visual inspection criteria, electrical readings), not clinical diagnoses.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable, for the same reasons as #3. Adjudication methods are typically used in clinical studies involving interpretation of medical data, where there might be disagreement among experts.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI/software device, so no MRMC study or AI assistance evaluation was performed.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. This is not an AI/software device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the non-clinical performance and thermal effects testing, the "ground truth" would be established by:

    • Engineering specifications and standards (e.g., for electrical safety, dimensions, opening angles, grasping strength, continuity checks).
    • Histological assessment for thermal damage (directly observing tissue changes).
    • Standardized methods for cleaning/sterilization efficacy.
    • Biocompatibility test results against established safety limits.

    8. The sample size for the training set

    Not applicable. This is not an AI/software device that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable, for the same reasons as #8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1