    K Number
    K161012

    Validate with FDA (Live)

    Date Cleared
    2016-07-06

    (86 days)

    Product Code
    Regulation Number
    886.4300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Easy-Load Lens Delivery System is indicated for the folding and injection of Bausch + Lomb intraocular lenses identifying the Easy-Load Lens Delivery System in their approved labeling.

    Device Description

    The Easy-Load Lens Delivery System is used for folding and delivering a Bausch + Lomb three-piece silicone IOL into the eye. An IOL is placed into the loading area and the drawer is closed. This compresses the IOL. The plunger is advanced until it stops at a detent position. The distal end is filled with viscoelastic or balanced salt solution and placed through an incision into the eye. The haptic puller is used to place the leading haptic in the correct loading position. Once the tip is in the eye, the plunger is advanced until the lens is fully expressed into the capsular bag.

    AI/ML Overview

    This document describes the premarket notification for the "Easy-Load Lens Delivery System" (Model EZ-24). It's an intraocular lens guide used for folding and injecting intraocular lenses.

    Here's the information regarding the acceptance criteria and study, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance:

    ItemTESTAcceptance CriteriaRESULTS
    Function Performance(Before and after insertion) IOL surface & bulk homogeneityFree from surface and bulk defects and all edges should appear smooth when viewed at 10X magnificationPass
    (Before and after insertion) IOL lens powerThe diopter power of the lenses shall be within the allowed tolerancePass
    (Before and after insertion) IOL image qualityThe image quality of the lenses shall meet or exceed the minimum resolution group element of the Air Force target specified in the protocol.Pass
    (Before and after insertion) IOL dimensionsThe lenses must meet the dimensional requirements specified in the protocol.Pass
    IOL delivery outcomeThe IOLs do not flip over upon delivery; The IOL must exit inserter upon completion of delivery; No cosmetic defects to the IOL's haptic from folding and/or deliveryPass
    Damage to insertion deviceNo damage to the insertion device due to lens deliveryPass
    Lubricant transfer studyNo detectable transferPass
    Particle studyComparable to or lower than controlPass
    Sterilization verificationComparative Resistance Performance DeterminationFor a given exposure time the average recovery of viable test organisms from biological indicators placed within the barrel of either the EZ-24 or the EZ-28 shall differ by no more than one logarithm of the average of the values determined for both inserters. Alternatively, the average recovery of viable test organisms from biological indicators placed within the barrel of the EZ-24 may be less than the average recovery of viable test organisms from biological indicators placed within the barrel of the EZ-28 insertersPass
    EO transfer test$\le$ 0.5μg EO per IOLPass
    Endotoxin test$<$ 0.2 EU/devicePass
    Packaging verificationPlunger push force and cartridge detachment testingCartridge detachment force is greater than the force to disengage the plunger from the detents in the injector body.Pass

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    The document does not explicitly state the sample size for each specific test or the data provenance. It mentions "nonclinical tests" and "accepted ISO standards," implying laboratory testing rather than human clinical trials. The data is internal to the manufacturer, Bausch + Lomb.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    This information is not provided. The tests described are primarily objective physical and chemical measurements ("Free from surface and bulk defects," "diopter power," "resolution group element," "dimensional requirements," "IOL delivery outcome," "no damage," "no detectable transfer," "comparable to or lower than control," "recovery of viable test organisms," "$\le$ 0.5μg EO," "$<$ 0.2 EU/device," "force greater than"). Expert judgment for these specific criteria would likely be embedded in the test protocol design and execution, but the document doesn't detail the number or qualifications of individuals involved in setting up or interpreting these nonclinical tests.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    Adjudication methods like 2+1 or 3+1 are typically used in clinical studies involving interpretation of medical images or subjective evaluations by multiple human readers. For the nonclinical tests described in this document, such adjudication methods are not applicable. The results are based on direct measurements and adherence to specified physical/chemical parameters.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    No MRMC study was mentioned. This device is an intraocular lens delivery system, not an AI-powered diagnostic tool, so MRMC studies and "human readers improve with AI" are not relevant to this submission.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    This is not an algorithm or AI product, so the concept of standalone performance for an algorithm is not applicable. The tests performed are on the physical device itself.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    The "ground truth" for the nonclinical tests is based on established engineering specifications, accepted ISO standards (ISO 11979-2, ISO 11979-3, and ISO 11135 where applicable), and internal protocols. For example:

    • IOL surface & bulk homogeneity: "Free from surface and bulk defects and all edges should appear smooth when viewed at 10X magnification." (Direct observation against a visual quality standard)
    • IOL lens power: "within the allowed tolerance." (Measurement against a quantitative specification)
    • Sterilization verification: "recovery of viable test organisms... differ by no more than one logarithm..." (Microbiological standard)
    • EO transfer test: "$\le$ 0.5μg EO per IOL." (Chemical quantitative standard)

    8. The sample size for the training set:

    There is no mention of a "training set" as this is not an AI/machine learning device. The tests are on the physical product and its components.

    9. How the ground truth for the training set was established:

    Not applicable, as there is no training set for an AI model. For the nonclinical tests conducted, the acceptance criteria (which serve as the "ground truth" for pass/fail) were established based on ISO standards and internal protocols.

    Ask a Question

    Ask a specific question about this device

    K Number
    K132481

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2013-12-27

    (141 days)

    Product Code
    Regulation Number
    886.4300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Easy-Load Lens Delivery System is indicated for the folding and injection of Bausch + Lomb intraocular lenses identifying the Easy-Load Lens Delivery System in their approved labeling.

    Device Description

    The Easy-Load Lens Delivery System is used for folding and delivering a Bausch + Lomb three-piece silicone IQL into the eye. An IOL is placed into the loading area and the drawer is closed. This compresses the IOL. The plunger is advanced until it stops at a detent position. The distal end is filled with viscoelastic or balanced salt solution and placed through an incision into the eye. The haptic puller is used to place the leading haptic in the correct loading position. Once the tip is in the eye, the plunger is advanced until the lens is fully expressed into the capsular bag.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Easy-Load Lens Delivery System, based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    Test TypeAcceptance Criteria (Implied)Reported Device Performance
    BiocompatibilityNew materials found to be biocompatible (Implicit: no adverse biological reactions)New materials for the Easy-Load Lens Delivery System were found to be biocompatible.
    Bench TestingMet acceptance criteria (Implicit: no damage to IOLs, successful delivery, proper visual appearance of injector)All nonclinical tests met the acceptance criteria.
    Surface Homogeneity of Delivered IOLs(Not explicitly stated, but implied as satisfactory)(Covered under "All nonclinical tests met the acceptance criteria.")
    Bulk Homogeneity of Delivered IOLs(Not explicitly stated, but implied as satisfactory)(Covered under "All nonclinical tests met the acceptance criteria.")
    Delivery Outcome(Not explicitly stated, but implied as successful and effective)(Covered under "All nonclinical tests met the acceptance criteria.")
    Injector Visual Inspection(Not explicitly stated, but implied as satisfactory)(Covered under "All nonclinical tests met the acceptance criteria.")
    Sterility Assurance Level10^-610^-6
    Shelf Life12 months12 months

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the specific sample size used for each bench test or biocompatibility test. It generally refers to "the new Easy-Load Lens Delivery System materials" and "Easy-Load Lens Delivery Systems with the new materials."

    The data provenance is retrospective, as the testing was performed on the device prior to submission and regulatory review. The country of origin for the data is not mentioned, but given the submission to the FDA, it is likely that the studies were either performed in the US or in a manner compliant with US regulatory requirements.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    This information is not provided in the document. The document describes bench and biocompatibility testing, which are typically performed in laboratories according to established protocols, rather than relying on expert consensus for "ground truth" in the same way clinical imaging studies might.

    4. Adjudication Method for the Test Set

    This information is not applicable and therefore not provided. Adjudication methods (like 2+1, 3+1) are typically used in studies where human readers are making subjective assessments, such as interpreting medical images. Bench and biocompatibility testing rely on objective measurements and predefined acceptance criteria.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is more relevant for diagnostic devices where human interpretation interacts with AI, and this device is a surgical delivery system.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    N/A (Not Applicable). This device is a manual intraocular lens delivery system, not an algorithm or AI-powered system. Therefore, the concept of "standalone algorithm performance" does not apply. The device's performance is inherently "human-in-the-loop" as it is operated by a surgeon.

    7. The Type of Ground Truth Used

    The ground truth for the performance of the Easy-Load Lens Delivery System was established through objective laboratory testing and measurements.

    • For biocompatibility, the ground truth is the absence of biological reactivity or toxicity, verified through standardized assays and tests.
    • For bench testing (surface and bulk homogeneity of delivered IOLs, delivery outcome, and injector visual inspection), the ground truth is established by design specifications and performance requirements (e.g., IOL delivered undamaged, correct placement, injector integrity), verified by direct observation, measurement, and functional testing.

    8. The Sample Size for the Training Set

    N/A (Not Applicable). As this is a mechanical medical device, there is no "training set" in the context of an algorithm or AI system. The device design and manufacturing processes are likely informed by engineering principles, material science, and previous product iterations, but not a data-driven training set in the AI sense.

    9. How the Ground Truth for the Training Set Was Established

    N/A (Not Applicable). Since there is no training set for an algorithm, there is no ground truth established for it in this context.