510(k) Data Aggregation
(155 days)
The EarliPoint Assessment device is indicated as a tool to aid qualified clinicians in the diagnosis and assessment of Autism Spectrum Disorder (ASD) in children ages 16 months through 95 months (7 years), who are at risk based on concerns identified by a parent, caregiver, or healthcare provider.
The EarliPoint assessment uses an eye tracker to measure the patient's looking behavior while viewing a series of videos. The system then remotely analyzes the looking behavior data using software and outputs a diagnosis of the patient's ASD status and associated developmental delay indices.
The EarliPoint System device consists of the following:
- Eye-tracking module and a separate Operator Module that can control the Eye-tracking module remotely. The patient sits on a chair and the Eye-tracking module is adjusted by the operator such that the patient's eyes are within the specification of the eye-tracking window.
- Eye-tracking module captures the patient visual response to social information provided in the form of a series of age-appropriate videos.
- Operator's module is used to initiate and monitors the session remotely
- WebPortal securely stores all patient information, analyzes the eye-tracking data, and outputs the results. Users can retrieve the results directly from the web portal.
- Software analyzes the eye-tracking data and provides a diagnosis for ASD. In addition, it also outputs 3 developmental delay indices (called EarliPoint Severity Indices) that proxy validated ASD instruments
Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided FDA 510(k) clearance letter for the EarliPoint Assessment device.
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are implicitly derived from the performance of the predicate device and the new clinical study for the expanded age range. The primary metrics are Sensitivity, Specificity, Positive Predictive Value (PPV), Negative Predictive Value (NPV), and Accuracy, compared to the expert clinical diagnosis (reference standard). The "Comparison" column in the provided table states "Substantially Equivalent" for all metrics, indicating that the performance of the subject device in the expanded age range (and the combined age range) meets the implicit acceptance criteria based on its predicate.
| Performance Measure | Acceptance Criteria (Implicit, based on Predicate K243891 16-30 months) | Reported Device Performance (EarliPoint System 16 to <96 months) |
|---|---|---|
| Sensitivity | 78.0% (70.5 - 84.3) | 80.2% (75.6 - 84.3) |
| Specificity | 85.4% (79.5 - 90.2) | 81.3% (76.7 - 85.3) |
| Positive Predictive Value | 81.3% (73.9 - 87.3) | 81.2% (76.6 - 85.2) |
| Negative Predictive Value | 82.7% (76.6 - 87.8) | 80.4% (75.7 - 84.4) |
| Accuracy | 82.1% (77.6 - 86.0) | 80.8% (77.6 - 83.7) |
Note: The predicate values for the 16-30 month age range, when reanalyzed with the borderline range, showed slightly different (and in the case of sensitivity, higher) performance. The performance of the subject device across the combined age range appears to maintain substantial equivalence. The document explicitly states, "clinical data from the expanded age range demonstrates the safety and effectiveness of the device without raising new or different questions of safety and effectiveness."
2. Sample Size Used for the Test Set and Data Provenance
- Test Set Sample Sizes:
- 16-30 months (original study): The table "EarliPoint System 16-30 months" shows a total of 335 participants for the predicate device. For the subject device (with borderline range implemented), the performance figures are based on 307 participants after excluding borderline cases from sensitivity/specificity calculations.
- 31 to <96 months (new study): The table "EarliPoint System 31 to <96 months" shows performance figures based on 369 participants (when borderline cases are excluded). Initial count for the entire population was 477 (when borderline was considered ASD-negative).
- Combined 16 to <96 months: A total of 676 participants (when borderline cases are excluded). Initial count for the entire population was 812 (when borderline was considered ASD-negative).
- Data Provenance: The studies were prospective, double-blind, multi-center, within-subject comparison, US, pivotal studies. This means the data was collected specifically for this evaluation, from multiple centers within the United States.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts
The document states:
- "All patients were evaluated for ASD by ... expert clinician diagnosis (current best practice for diagnosis of ASD)."
- "robust convergent validity in comparison with Reference Standard assessments of social disability, verbal ability, and nonverbal cognitive ability made by highly trained experts at leading expert centers."
The exact number of experts is not specified. Their qualifications are described as "expert clinician" and "highly trained experts at leading expert centers," implying specialized training and experience in ASD diagnosis, likely including child psychiatrists, developmental pediatricians, or licensed psychologists with expertise in ASD.
4. Adjudication Method for the Test Set
The document does not explicitly state the adjudication method (e.g., 2+1, 3+1). However, it highlights the challenge of "clinical uncertainty" in ASD diagnosis, noting that even clinicians are sometimes less than fully certain of their diagnoses. It states: "EarliPoint clinical trial procedures worked effectively to reduce and minimize, to the extent possible, clinician uncertainty in the reference standard diagnosis." This suggests that efforts were made to achieve a high degree of confidence in the ground truth, but the specific process (e.g., multiple independent diagnoses followed by a consensus meeting or a tie-breaker by a more senior expert) is not detailed.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was Done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
The study described is a standalone (algorithm only) performance study, comparing the device's output directly against expert clinical diagnosis (reference standard). It is not an MRMC comparative effectiveness study where human readers' performance with and without AI assistance is measured. Therefore, there is no information provided about the effect size of human readers improving with AI assistance.
6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was Done
Yes, a standalone study was done. The entire premise of the clinical study is to evaluate the "EyeTracking System diagnosis relative to the expert clinical diagnosis" and quantify its sensitivity, specificity, and other metrics. The device "outputs a diagnosis of the patient's ASD status," implying a direct algorithmic output without human intervention for the primary outcome.
7. The Type of Ground Truth Used
The primary ground truth used is expert clinician diagnosis, specifically described as "current best practice for diagnosis of ASD." This diagnosis also includes assessments using "validated ASD instruments" like ADOS-2, and Mullen/DAS-II, as referenced for correlating the EarliPoint Severity Indices.
8. The Sample Size for the Training Set
The document does not provide any information regarding the size or characteristics of the training set used to develop the EarliPoint Assessment algorithm. The clinical studies described are for validation/testing of the device's performance, not for its training.
9. How the Ground Truth for the Training Set Was Established
Since information on the training set's sample size is not provided, the method for establishing its ground truth is also not specified in this document.
Ask a specific question about this device
(98 days)
The EarliPoint System device is indicated as a tool to aid qualified clinicians in the diagnosis and assessment of Autism Spectrum Disorder (ASD) in children ages 16 months through 30 months, who are at risk based on concerns identified by a parent, caregiver, or healthcare provider.
The EarliPoint system uses an eye tracker to capture the patient's looking behavior while viewing a series of videos. The system then remotely analyzes the looking behavior data using software and outputs a diagnosis of the patient's ASD status and associated developmental delay indicies.
The EarliPoint System device consists of the following:
• Eye-tracking module and a separate Operator Module that can control the Eye-tracking module remotely. The patient sits on a chair and the Eye-tracking module is adjusted by the operator such that the patient's eyes are within the specification of the eye tracking window.
• Eye-tracking module captures the patient visual response to social information provided in the form of a series of age-appropriate videos.
• Operator's module is used to initiate and monitors the session remotely
• WebPortal securely stores all patient information, analyzes the eye tracking data, and outputs the results. Users can retrieve the results directly from the web-portal.
• Artificial intelligence software analyzes the eye-tracking data and provides a diagnosis for ASD. In addition, it also outputs 3 developmental delay indices (called EarliPoint Severity Indices) that proxy the ADOS-2 and Mullen validated ASD instruments Social Disability Index correlates and proxies ADOS-2 Verbal Ability Index correlates and proxies the age equivalent Mullen Verbal Ability score Non-verbal Ability Index correlates and proxies the age equivalent non-verbal Mullen Ability score
The EarliPoint System device's acceptance criteria and the study proving its performance are detailed below. It's important to note that the provided text is an FDA 510(k) clearance letter and summary, which primarily focuses on demonstrating substantial equivalence to a predicate device rather than presenting a full clinical study report. Therefore, some information, particularly granular details about study design, expert qualifications, or the exact training process, might not be explicitly stated to the level one would find in a peer-reviewed publication.
Acceptance Criteria and Reported Device Performance
The core acceptance criteria for the EarliPoint System, as demonstrated in the pivotal study, revolve around its ability to accurately diagnose Autism Spectrum Disorder (ASD) in comparison to expert clinical diagnosis. The key metrics are Sensitivity and Specificity.
Table of Acceptance Criteria and Reported Device Performance
| Metric | Acceptance Criteria (Implied/Expected for a Diagnostic Aid) | Reported Device Performance (Full Study Population) | Reported Device Performance (CertainDx Subpopulation) |
|---|---|---|---|
| Sensitivity | High (to correctly identify individuals with ASD) | 71% (157/221) | 78.0% (117/150) |
| Specificity | High (to correctly identify individuals without ASD) | 80.7% (205/254) | 85.4% (158/185) |
Note: The document does not explicitly state pre-defined acceptance thresholds for sensitivity and specificity. The reported performance suggests the levels that were considered acceptable for clearance.
Study Details
2. Sample Size and Data Provenance
- Test Set Sample Size: 475 evaluable patients. 25 patients had missing data for either the device or the control diagnosis, making the initial enrollment 500 patients.
- Data Provenance:
- Country of Origin: United States (six sites).
- Retrospective or Prospective: Prospective.
3. Number of Experts and Qualifications for Ground Truth
- Number of Experts: Not explicitly stated how many individual experts were used across the six sites, but the ground truth was established by "expert clinicians" as the "current best practice for diagnosis of ASD."
- Qualifications of Experts: Not explicitly detailed, but they are referred to as "expert clinicians" in "specialized developmental disabilities centers," implying specialized training and experience in diagnosing ASD.
4. Adjudication Method for the Test Set
- The document does not explicitly describe an adjudication method for the expert clinical diagnosis (ground truth). It refers to it as the "current best practice," suggesting that the consensus or standard diagnostic process by qualified clinicians was deemed sufficient as the reference standard.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- Was an MRMC study done? No, a multi-reader multi-case (MRMC) comparative effectiveness study was not explicitly mentioned or performed in the context of human readers improving with AI vs. without AI assistance. The study described is a direct comparison of the EarliPoint system's diagnosis (algorithm only) against expert clinical diagnosis, not an AI-assisted human reader study.
6. Standalone (Algorithm Only) Performance
- Was a standalone performance study done? Yes, the presented sensitivity and specificity values are for the EarliPoint System's diagnosis alone (algorithm only), without human-in-the-loop assistance for interpretation of the device's output itself for the primary outcome. The system produces a diagnosis for ASD.
7. Type of Ground Truth Used
- Type of Ground Truth: "Expert clinician diagnosis (current best practice for diagnosis of ASD)." This is a form of expert consensus or clinical standard of care, rather than pathology or long-term outcomes data.
8. Sample Size for the Training Set
- The document does not provide the sample size for the training set. The clinical study details refer to a "pivotal study" used for evaluating safety and effectiveness, which serves as the test set for the device, rather than data used for initial model training.
9. How Ground Truth for the Training Set was Established
- The document does not provide details on how the ground truth for the training set was established. Since the pivotal study's data is described as the test set (evaluable N=475), any training data and its associated ground truth establishment would have occurred prior to this specific study and are not disclosed in this regulatory submission summary.
Ask a specific question about this device
(142 days)
The EarliPoint System is indicated for use in specialized developmental disabilities centers as a tool to aid clinicians in the diagnosis and assessment of ASD patients ages 16 months through 30 months.
The device is a more compact version of the predicate device but otherwise has similar functions and features. The system uses an eye tracker to capture the patient's looking behavior while viewing a series of videos. The system then remotely analyzes the looking behavior data using software and outputs a diagnosis of the patient's ASD status and assesses the symptoms associated with ASD.
The system has two modules:
EarliPoint System consists of the following:
- Eye-tracking module and a separate Operator Module that can control the Eye-tracking module remotely. The patient sits on a chair and the Eye-tracking module is adjusted by the operator such that the patient's eyes are within the specification of the eye tracking window
- -Eye-tracking module captures the patient visual response to social information provided in the form of a series of age-appropriate videos
- Operator's module is used to initiate and monitors the session remotely -
- -WebPortal securely stores all patient information, analyzes the eye tracking data, and outputs the results. Users can retrieve the results directly from the web-portal.
- -Artificial intelligence software analyzes the eye-tracking data and provides a diagnosis for ASD. In addition, it also outputs 3 indices (called EarliPoint Severity Indices) that proxy the ADOS-2 and Mullen validated ASD instruments
- o Social Disability Index correlates and proxies ADOS-2
- O Verbal Ability Index correlates and proxies the age equivalent Mullen Verbal Ability score
- O Non-verbal Ability Index correlates and proxies the age equivalent non-verbal Mullen Ability score
The eye-tracker used in the EarliPoint device has similar capability as the eye tracker used in the predicate device.
The provided text describes a 510(k) submission for a modified EarliPoint System, which is an aid for diagnosing and assessing Autism Spectrum Disorder (ASD). The submission details performance testing and a pivotal clinical study.
Here's an analysis of the acceptance criteria and the study proving the device meets them:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria in this submission are derived from the clinical study's performance metrics against a reference standard (expert clinical diagnosis for ASD).
| Performance Metric | Acceptance Criteria (from study results) | Reported Device Performance (EarliPoint mITD) | Reported Device Performance (EarliPoint CertainDx) |
|---|---|---|---|
| Sensitivity | Not explicitly stated as a pre-defined acceptance criteria, but observed performance | 71% (95% CI: 64.6% - 76.9%) | 78.0% (95% CI: 70.5% - 84.3%) |
| Specificity | Not explicitly stated as a pre-defined acceptance criteria, but observed performance | 80.7% (95% CI: 75.3% - 85.4%) | 85.4% (95% CI: 79.5% - 90.2%) |
| Safety | No serious adverse events related to device use | No reported serious adverse events | No reported serious adverse events |
| Correlation with ADOS-2 | Positive correlation with EarliPoint Social Disability Index | Correlates and proxies ADOS-2 | Correlates and proxies ADOS-2 |
| Correlation with Mullen Verbal Ability | Positive correlation with EarliPoint Verbal Ability Index | Correlates and proxies age equivalent Mullen Verbal Ability score | Correlates and proxies age equivalent Mullen Verbal Ability score |
| Correlation with Mullen Non-verbal Ability | Positive correlation with EarliPoint Non-verbal Ability Index | Correlates and proxies age equivalent non-verbal Mullen Ability score | Correlates and proxies age equivalent non-verbal Mullen Ability score |
Note: The document does not explicitly state pre-defined numerical acceptance criteria for sensitivity and specificity. The reported performance is the outcome of the study, and its acceptance implies these values were deemed sufficient by the FDA for substantial equivalence.
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: 475 evaluable patients for primary and secondary endpoint analysis. 25 patients had missing data for either the device or control diagnosis, bringing the total enrolled to 500.
- Data Provenance:
- Country of Origin: United States
- Retrospective or Prospective: Prospective
- Multi-center: Six sites
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
- Number of Experts: Not specified. The text mentions "expert clinicians" but does not quantify the number of individual experts or how many were involved per case.
- Qualifications of Experts: Referred to as "expert clinicians" and their diagnosis is considered the "current best practice for diagnosis of ASD." No further specific qualifications like years of experience or board certifications are provided in the document.
4. Adjudication Method for the Test Set
- Adjudication Method: Not explicitly stated. The text mentions "expert clinician diagnosis" as the reference standard but does not detail how consensus was reached if multiple experts were involved or if there was a single expert per case.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size
- MRMC Study: No, a multi-reader multi-case (MRMC) comparative effectiveness study was not conducted in the traditional sense of comparing human readers with AI assistance versus human readers without AI assistance.
- Effect Size: Therefore, no effect size for human reader improvement with AI assistance is reported. The study's design was a within-subject comparison where all patients were evaluated by both the EarliPoint system and expert clinicians, comparing the device's diagnostic output against the expert diagnosis.
6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done
- Standalone Performance: Yes, the study evaluated the "EarliPoint System diagnosis" relative to the expert clinical diagnosis. The device's output (ASD diagnosis and severity indices) is generated by "Artificial intelligence software [that] analyzes the eye-tracking data and provides a diagnosis for ASD." This indicates a standalone performance evaluation of the algorithm. The "EarliPoint CertainDx" analysis further refined this by focusing on cases where clinicians were certain of the diagnosis, still representing the device's standalone output being compared to this subset of expert diagnoses.
7. The Type of Ground Truth Used
- Type of Ground Truth: "Expert clinician diagnosis (current best practice for diagnosis of ASD)." This is a form of expert consensus or clinical standard of care. It is further supported by correlation with validated ASD instruments (ADOS-2 and Mullen).
8. The Sample Size for the Training Set
- The document does not explicitly state the sample size for the training set. The clinical study described (the pivotal study) is an evaluation study (test set) for the device's performance, not the dataset used to train the AI algorithm.
9. How the Ground Truth for the Training Set Was Established
- The document does not explicitly state how the ground truth for the training set was established. While it details the ground truth for the test set (expert clinician diagnosis), it does not provide information on the training data.
Ask a specific question about this device
(177 days)
The EarliPoint System is indicated for use in specialized developmental disabilities centers as a tool to aid clinicians in the diagnosis and assessment of Autism Spectrum Disorder (ASD) for patients ages 16 months through 30 months.
The EarliPoint System is a medical device for diagnosis of Autism Spectrum Disorder (ASD) in children.
EarliPoint System consists of the following:
- EarliPoint WebPortal to enter the patient information and for access to the patient evaluation results,
- EarliPoint Device with eye-tracking capability captures the patient visual response to social information provided in the form of a series of age-appropriate videos
- Artificial intelligence software analyzes the eye-tracking data and provides a diagnosis for ASD.
- Eye-tracking data also outputs 3 indices (called EarliPoint Severity Indices) that proxy the ADOS-2 and Mullen validated ASD instruments
- Social Disability Index correlates and proxies ADOS-2
- Verbal Ability Index correlates and proxies the age equivalent Mullen Verbal Ability score
- Non-verbal Ability Index correlates and proxies the age equivalent non-verbal Mullen Ability score
The EarliPoint System, a device for diagnosing Autism Spectrum Disorder (ASD), underwent a pivotal study to demonstrate its safety and effectiveness.
Here's a breakdown of the acceptance criteria and study details:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly present a table of "acceptance criteria" but rather shows the device's performance (sensitivity and specificity) relative to a predicate device (CanvasDx) and a "gold standard" expert clinical diagnosis. The implication is that meeting or exceeding the predicate's performance, especially for certain populations, was a key aspect of acceptance.
| Population/Metric | Sensitivity Mean (n/N) [95% CI] | Specificity Mean (n/N) [95% CI] |
|---|---|---|
| CanvasDx (mITD), N=425 | 51.6% (63/122) [42.8% - 60.5%] | 18.5% (56/303) [14.3% - 23.3%] |
| EarliPoint (mITD), N=475 | 71% (157/221) [64.6% - 76.9%] | 80.7% (205/254) [75.3% - 85.4%] |
| EarliPoint CertainDx (N=335) | 78.0% (117/150) [70.5%, 84.3%] | 85.4% (158/185) [79.5% - 90.2%] |
| EarliPoint UncertainDx (N=140) | 56.3% (p=1.00) | 68.1% (p=0.69) |
Note: The document states: "When compared to the predicate. CanvasDx, the sensitivity and specificity of the EarliPoint device are higher than the CanvasDx and hence the two devices are substantially equivalent." This implies that outperforming the predicate was a key acceptance criterion.
2. Sample Size and Data Provenance
- Test Set Sample Size: 475 evaluable patients for primary and secondary endpoint analysis (from an initial enrollment of 500).
- Data Provenance: The study was "multi-center" with data collected from "six sites in the United States." The study was "prospective."
3. Number of Experts and Qualifications for Ground Truth
- Number of Experts: Not explicitly stated, but the ground truth was established by "expert clinicians" at "six sites." The document refers to "expert clinician diagnosis" and "clinicians' reference diagnoses."
- Qualifications of Experts: Described as "expert clinician diagnosis (current best practice for diagnosis of ASD)." Specific qualifications (e.g., years of experience, specialization) are not detailed.
4. Adjudication Method for the Test Set
The document does not explicitly describe an adjudication method for conflicting expert opinions. It refers to a "gold standard of the expert clinician diagnosis." It does, however, categorize the ground truth into "Certain Dx Population" (clinicians' certainty rating > 80%) and "Uncertain Dx Population" (clinicians' certainty rating ≤ 80%), indicating some level of consideration for diagnostic confidence.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- No MRMC comparative effectiveness study was explicitly mentioned in the provided text, comparing human readers with AI assistance vs. without AI assistance. The study focuses on the device's performance compared to human expert diagnosis as the ground truth. The device is intended as an "aid clinicians," suggesting a human-in-the-loop scenario, but a specific MRMC study design is not detailed.
6. Standalone (Algorithm Only) Performance
- The reported sensitivity and specificity values represent the standalone performance of the EarliPoint System's algorithm in diagnosing ASD, as it provides a "diagnosis for ASD" from analyzing eye-tracking data. The study "evaluated for ASD by both the EarliPoint system and by expert clinician diagnosis."
7. Type of Ground Truth Used
- The primary ground truth used was "expert clinician diagnosis (current best practice for diagnosis of ASD)."
- Additionally, the study correlated the three EarliPoint Severity Indices (Social Disability, Verbal Ability, Non-verbal Ability) against corresponding "expert clinical instruments of ADOS-2 and Mullen." This indicates a reliance on established clinical diagnostic tools as part of the expert ground truth.
8. Sample Size for the Training Set
- The document does not provide the sample size used for the training set. It only describes the design and results of the pivotal clinical study (test set).
9. How the Ground Truth for the Training Set was Established
- The document does not provide information on how the ground truth for the training set was established. It focuses solely on the test set's ground truth.