Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K930683

    Validate with FDA (Live)

    Device Name
    EZTEST-GAS
    Date Cleared
    1997-03-04

    (1483 days)

    Product Code
    Regulation Number
    880.2800
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    EZTest® - Gas self contained biological indicators are for monitoring the efficacy of ethylene oxide gas sterilization processors. Performance characteristics are established in accordance with USP XXIII for the 600mg/l process.

    Device Description

    The biological indicator consists of a self-contained unit which includes: bacterial spores inoculated onto a paper carrier, a small glass ampule containing sterile culture medium and color indicator, and a plastic vial that serves as the culture tube. This unit complies to the performance characteristics described in the USP XXIII.

    AI/ML Overview

    The acceptance criteria and the study proving the device meets them are described below:

    1. Table of Acceptance Criteria and Reported Device Performance

    The EZTest® - Gas biological indicator is designed to meet the performance characteristics described in the USP XXIII. The document outlines several characteristics that were evaluated.

    Acceptance Criteria (Based on USP XXIII guidelines)Reported Device Performance
    Bacterial Spore Species: Bacillus subtilis, ATCC 9372 or equivalentBacillus subtilis, ATCC 9372 or equivalent (no specific performance metric, but states that this is the organism used and USP XXIII requires its use for EO sterilization monitoring).
    Bacterial Spore Concentration: Between 0.5 x 10⁶ and 5 x 10⁶ CFU for a 10⁶ labeled concentration.Population of spores on paper carrier will be between 0.5 x 10⁶ and 5 x 10⁶, labeled as a 10⁶ concentration. SGM Biotech, Inc. specification matches this guideline. (Performance is stated as meeting the spec, not a measured value from this specific study, but inherent to product design).
    Sample Purity: (Implied clean culture of target spore species)No specific performance metric given, but "Sample purity" was an evaluated characteristic.
    Resistance Characteristics:
    - D-values at 600 mg/l EO, 54°C, 60% RH."D values at 600 mg/l EO, 54°C, 60% RH" were evaluated. No specific numerical D-value is reported in this summary, but the device is concluded to "meet the USP XXIII requirements" for this parameter.
    - Survival/Kill times at 600 mg/l EO, 54°C, 60% RH."Survival/Kill times at 600 mg/l EO, 54°C, 60% RH" were evaluated. No specific numerical times are reported in this summary, but the device is concluded to "meet the USP XXIII requirements" for this parameter.
    Incubation Readout Time Validation: 48 hours incubation based on CDRH guideline."Validation of 48 hours incubation read out time following the CDRH guideline" was performed.
    Effect of Holding Time:"Resistance characteristics were evaluated to determine the effect of holding time after the exposure to the sterilant until incubation, upon the recovery of injured spores." (No specific result provided, but the evaluation was performed).
    Effect of pH Indicator:"Evaluation of the effect of the pH indicator upon the recovery of injured spores" was performed. (No specific result provided, but the evaluation was performed).
    Effect of Sterilization Process on Media Growth Support:"Validation of the effect of the sterilization process on the ability of the media to support the growth of injured spores" was performed. (No specific result provided, but the evaluation was performed).
    Product Stability: Stability over 18-month shelf life claim."Stability of the product over the eighteen (18) month shelf life claim" was evaluated. (No specific result provided, but the evaluation was performed).

    The ultimate reported performance is that the EZTest® - Gas "meets the USP XXIII requirements" for the tested parameters.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The document states that claims were evaluated using "at least three (3) spore lots using different spore crops and different media lots." This indicates that multiple units from these lots would have been tested for each characteristic. However, the exact number of individual biological indicators tested is not specified.
    • Data Provenance: The document does not explicitly state the country of origin for the data or whether the study was retrospective or prospective. Given that it's a 510(k) submission to the FDA, the studies would likely have been conducted prospectively under controlled laboratory conditions, possibly in the USA, to generate the required performance data.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This type of device (biological indicator) does not typically involve human experts establishing a "ground truth" in the way an imaging AI algorithm would. The "ground truth" for a biological indicator is determined by the growth or non-growth of bacterial spores after exposure to a sterilization process and subsequent incubation. This is an objective biological outcome, not a subjective interpretation by human experts.

    4. Adjudication Method for the Test Set

    Not applicable. As explained above, the "ground truth" is an objective biological outcome (spore growth/no growth), not a subjective assessment requiring adjudication by multiple experts.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is relevant for diagnostic devices that involve human interpretation (e.g., radiology AI), not for biological indicators that assess the efficacy of sterilization directly through objective biological endpoints.

    6. Standalone Performance Study

    Yes, a standalone study (algorithm only without human-in-the-loop performance) was done. For a biological indicator, "standalone performance" refers to the device's ability to accurately reflect the presence or absence of viable spores after a sterilization cycle. The listed evaluations (D-values, survival/kill times, incubation readout, etc.) directly assess the device's inherent performance characteristics in a controlled laboratory setting, independent of human interpretation to determine the primary outcome. The reading of the color change is a direct function of the device's biological and chemical components, not a human reader's diagnostic skill.

    7. Type of Ground Truth Used

    The ground truth used is based on biological outcomes (growth or non-growth of Bacillus subtilis spores) and established industry standards as defined by the USP XXIII (United States Pharmacopoeia). This involves direct observation of microbial growth (manifested by a color change in the medium) after defined exposure to ethylene oxide and subsequent incubation.

    8. Sample Size for the Training Set

    The document does not explicitly mention a "training set" in the context of machine learning or AI. For a biological indicator, the development and manufacturing process doesn't typically involve machine learning training data. Instead, the "training" (or rather, development and validation) of such a device relies on extensive laboratory testing and adherence to established microbiological and sterilization science principles.

    The description implies that the manufacturing process involves "a manufacturing in-process quality control system to specifications defining exacting parameters for production to assure a consistent product and good manufacturing procedures." This ongoing quality control and specification adherence serve a similar purpose to ensuring a consistent "learned" behavior in an AI model, but through traditional engineering and scientific validation.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable in the AI/ML sense. For this device, the "ground truth" for establishing its design and specifications (analogous to a training set for AI) would have been established through:

    • Established microbiological principles: Understanding the resistance of Bacillus subtilis spores to ethylene oxide.
    • USP XXIII guidelines: Adherence to defined performance characteristics for biological indicators.
    • Historical data and scientific literature: Leveraging decades of knowledge related to sterilization processes and biological indicator design.
    • Internal R&D and testing: SGM Biotech's own experimentation to develop their specific media formulation, spore carrier, and device configuration to meet the USP XXIII standards.

    The "ground truth" for manufacturing quality assurance is established by testing each lot against the USP XXIII specifications for spore count, resistance, and purity.

    Ask a Question

    Ask a specific question about this device

    K Number
    K963841

    Validate with FDA (Live)

    Device Name
    EZTEST
    Manufacturer
    Date Cleared
    1997-02-25

    (153 days)

    Product Code
    Regulation Number
    880.2800
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    EZTest® - Steam self contained biological indicators are for monitoring the efficacy of saturated steam sterilization processors. Performance characteristics are established in accordance with USP XXIII for the 121 C steam process. Additional saturated steam sterilization conditions are also included in the Certificate of Performance. EZTest® - Steam Bls are also appropriate for use in the high temperature saturated steam processes of 132 C , 134 C and 135 C.

    Device Description

    The biological indicator consists of a self-contained unit which includes: bacterial spores inoculated onto a paper carrier, a small glass ampule containing sterile culture medium and color indicator, and a plastic vial that serves as the culture tube. This unit complies to the performance characteristics described in the USP XXIII.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the EZTest® - Steam biological indicator, based on the provided 510(k) summary:

    1. Table of Acceptance Criteria and Reported Device Performance

    Performance CharacteristicAcceptance Criteria (USP XXIII)Reported Device Performance (as listed in 510(k) summary)
    Spore Population0.5 x 10^5 to 5.0 x 10^5 spores for a 10^5 BI"between 0.5 x 10^5 and 5 x 10^5" (Matches USP guideline)
    D-value at 121°CNot explicitly stated in the summary, but implied by USP XXIII."D values at 121°C...", evaluated for at least three (3) spore lots using different spore crops and different media lots. (Specific values not provided)
    Survival/Kill TimeNot explicitly stated in the summary, but implied by USP XXIII."Survival/Kill times at 121°C...", evaluated for at least three (3) spore lots using different spore crops and different media lots. (Specific values not provided)
    D-value at 132°CNot explicitly stated in the summary, but implied by USP XXIII."D values at 132°C...", evaluated for at least three (3) spore lots using different spore crops and different media lots. (Specific values not provided)
    Survival/Kill TimeNot explicitly stated in the summary, but implied by USP XXIII."Survival/Kill times at 132°C...", evaluated for at least three (3) spore lots using different spore crops and different media lots. (Specific values not provided)
    D-value at 134°CNot explicitly stated in the summary, but implied by USP XXIII."D values at 134°C...", evaluated for at least three (3) spore lots using different spore crops and different media lots. (Specific values not provided)
    Survival/Kill TimeNot explicitly stated in the summary, but implied by USP XXIII."Survival/Kill times at 134°C...", evaluated for at least three (3) spore lots using different spore crops and different media lots. (Specific values not provided)
    D-value at 135°CNot explicitly stated in the summary, but implied by USP XXIII."D values at 135°C...", evaluated for at least three (3) spore lots using different spore crops and different media lots. (Specific values not provided)
    Survival/Kill TimeNot explicitly stated in the summary, but implied by USP XXIII."Survival/Kill times at 135°C...", evaluated for at least three (3) spore lots using different spore crops and different media lots. (Specific values not provided)
    Z-valueNot explicitly stated in the summary, but implied by USP XXIII."Z values", evaluated for at least three (3) spore lots using different spore crops and different media lots. (Specific values not provided)
    Spore SpeciesBacillus stearothermophilus, ATCC 7953 or equivalent.Bacillus stearothermophilus, ATCC 7953 or equivalent. (Meets criteria)
    Sample PurityNot explicitly stated as an acceptance criterion.Evaluated as part of performance. (Details not provided)
    Incubation Readout TimeImplied by CDRH guideline (48 hours).Validation of 48 hours incubation read out time following the CDRH guideline. (Meets guideline)
    Shelf Life StabilityNot explicitly stated as an acceptance criterion.Stability of the product over the eighteen (18) month shelf life claim. (Reported as evaluated)
    Post-Exposure RecoveryNot explicitly stated as an acceptance criterion.Evaluation of the effect of holding time after exposure to sterilant until incubation, upon recovery of injured spores. (Reported as evaluated)
    pH Indicator EffectNot explicitly stated as an acceptance criterion.Evaluation of the effect of the pH indicator upon the recovery of injured spores. (Reported as evaluated)
    Media Support GrowthNot explicitly stated as an acceptance criterion.Validation of the effect of the sterilization process on the ability of the media to support the growth of injured spores. (Reported as evaluated)

    Note: The summary states that "A significant portion of the quality assurance specification is testing to the performance characteristics of biological indicators as outlined in USP XXIII." and that "Such compliance is required on each lot prior to release." This implies that the specific numeric acceptance criteria for D-values, survival/kill times, and Z-values, although not explicitly listed in the 510(k) summary, are derived from USP XXIII and are met by the device. The summary provides a list of evaluated characteristics rather than specific quantitative performance values.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The summary states that "Claims were evaluated using at least three (3) spore lots using different spore crops and different media lots." This indicates a minimum of three distinct lots were used for testing resistance characteristics. The exact number of individual biological indicators tested within each lot is not specified.
    • Data Provenance: The data is generated internally by SGM Biotech, Inc., as part of their "in-process quality control system" and "quality assurance specification." The summary does not provide information on the country of origin of the data or whether it was retrospective or prospective, but it implies a prospective testing approach as part of a release process for each lot.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

    This type of information is not applicable to this device. The ground truth for a biological indicator is inherent to its design and function: the presence or absence of viable spores after a sterilization cycle. The "ground truth" is established by direct microbiological methods (observing growth/color change), not by expert interpretation.

    4. Adjudication Method for the Test Set

    This is not applicable as the output of a biological indicator (color change, turbidity) is a direct, objective observation of microbial growth or lack thereof, rather than something requiring expert adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    This is not applicable. This device is a biological indicator, not an AI-based diagnostic tool. It does not involve human readers interpreting data or AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Study Was Done

    This is not applicable. This device is a biological indicator for monitoring sterilization, not an algorithm. Its "standalone" performance is its ability to accurately reflect sterilization efficacy based on spore survival.

    7. The Type of Ground Truth Used

    The ground truth used is based on microbiological growth/viability, specifically the survival and subsequent growth of Bacillus stearothermophilus spores. The change in the pH indicator (bromcresol purple changing from purple to yellow) is a direct, observable proxy for this microbiological activity. This is akin to a pathology/microbiology outcome.

    8. The Sample Size for the Training Set

    This is not applicable. As a biological indicator, the device does not employ a training set in the typical machine learning sense. Its performance is based on the inherent biological resistance of the spores and the chemical properties of the culture medium.

    9. How the Ground Truth for the Training Set Was Established

    This is not applicable for the same reasons as point 8. The "ground truth" for its operation is the established science of microbial heat resistance and culture medium efficacy, which is a foundational scientific principle rather than a dataset requiring ground truth establishment.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1