Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K121132

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2012-05-10

    (27 days)

    Product Code
    Regulation Number
    892.1680
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    EzSensor P, Intra-oral Imaging System, is intended to collect dental x-ray photons and convert them into electronic impulses that may be stored, viewed, and manipulated for diagnostic use by dentists.

    Device Description

    The EzSensor P digital intra oral sensor is a solid state x-ray imager designed for dental radiographic applications. The EzSensor P digital intra oral sensor provides digital diagnostic image capture to replace radiographic film/screen systems in general dental diagnostic procedures. The captured digital image is transferred to personal computer via USB interface port.

    AI/ML Overview

    The acceptance criteria and the study proving the EzSensor P device meets them are described below:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided text does not contain specific acceptance criteria with numerical targets. Instead, it claims substantial equivalence to a predicate device, EzSensor T, by comparing technical characteristics. Performance is implicitly evaluated by demonstrating these characteristics are either the same or superior.

    CharacteristicAcceptance Criteria (Implied)Reported EzSensor P Performance
    Indication for UseSame as predicate device (EzSensor T)EzSensor P, Intra-oral Imaging System, is intended to collect dental x-ray photons and convert them into electronic impulses that may be stored, viewed, and manipulated for diagnostic use by dentists. (Identical to predicate)
    Device DescriptionSame as predicate device (EzSensor T)EzSensor P is a solid state x-ray imager designed for dental radiographic applications. The EzSensor P digital intra oral sensor provides digital image capture to replace radiographic film/screen systems in general dental diagnostic procedures. The captured digital image is transferred to Personal Computer via USB interface port. (Similar to predicate)
    Sensor Dimension (mm) Size "1.5"Within acceptable range, compared to predicate (EzSensor T)38.7 x 29.2 (Identical to predicate)
    Sensor Dimension (mm) Size "2.0"Within acceptable range, compared to predicate (EzSensor T)42.8 x 31.5 (Identical to predicate)
    Sensor ThicknessSame as predicate device (EzSensor T)4.95 (Identical to predicate)
    Active Area (mm) Size "1.5"Within acceptable range, compared to predicate (EzSensor T)24.01 x 33.04 (Identical to predicate)
    Active Area (mm) Size "2.0"Within acceptable range, compared to predicate (EzSensor T)26.04 x 36.05 (Identical to predicate)
    USB ModuleSame as predicate device (EzSensor T)Integrated USB 2.0 module (Identical to predicate)
    Pixel Size (µm)Equal to or improved over predicate device20x20 (Improved compared to 35x35 of predicate)
    Pixel Matrix Size "1.5"Equal to or improved over predicate device1200x1650 pixel (Improved compared to 686x944 of predicate)
    Pixel Matrix Size "2.0"Equal to or improved over predicate device1300x1800 pixel (Improved compared to 744x1030 of predicate)
    Pixel Pitch (Spacing)Equal to or improved over predicate device0.02 mm x 0.02 mm (Improved compared to 0.035 mm x 0.035 mm of predicate)
    Theoretical ResolutionEqual to or improved over predicate device25 lp/mm (Improved compared to 14.1 lp/mm of predicate)
    Scintillator materialsAcceptable as a design difference that does not alter fundamental scientific technology or safety/effectiveness (Implied)Cesium iodide (CsI) (Different from GOS(Gd2O2S:Tb) of predicate, but deemed substantially equivalent)
    Safety and EMCCompliance with relevant standards (IEC 60601-1, IEC 60601-1-1, IEC 60601-1-2)Electrical, mechanical, environmental safety and performance testing according to standard IEC 60601-1 (A1+A2, 1995), IEC 60601-1-1(2nd edition, 2000) for use with IEC60601-1(Al+A2, 1995) was performed, and EMC testing were conducted in accordance with standard IEC 60601-1-2:2007. All test results were satisfactory.
    Non-clinical & Clinical ConsiderationsCompliance with FDA Guidance for Solid State X-ray Imaging Devices submissionsNon-clinical & Clinical considerations according to FDA Guidance for the Submissions of 510(k)'s for Solid State X-ray Imaging Devices" was performed. All test results were satisfactory.

    2. Sample Size Used for the Test Set and Data Provenance

    The provided text does not specify a sample size for a test set for evaluating the imaging performance of the EzSensor P directly through clinical or non-clinical image analysis against a ground truth.

    The "Non-clinical & Clinical considerations according to FDA Guidance for the Submissions of 510(k)'s for Solid State X-ray Imaging Devices" were performed, and "All test results were satisfactory." However, details on the specific nature of these "clinical considerations," the data provenance (e.g., country of origin, retrospective or prospective), or the number of images/cases used are not provided. The study primarily relies on a comparison of technical specifications to a predicate device.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    This information is not provided in the document. As detailed above, a specific test set requiring expert ground truth establishment for a diagnostic performance study is not explicitly described.

    4. Adjudication Method for the Test Set

    This information is not provided in the document.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of Human Readers Improvement with AI vs. Without AI Assistance

    A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not conducted (or at least, not reported in this summary). The device is an imaging sensor, not an AI-assisted diagnostic tool. Therefore, a study on human reader improvement with AI assistance is not applicable to this submission.

    6. If a Standalone (i.e., Algorithm Only Without Human-in-the-Loop Performance) Was Done

    This device is an imaging sensor, not an algorithm, so a "standalone" algorithmic performance study as typically understood for AI/ML devices is not applicable. The performance is inherent to the sensor's ability to capture images based on its technical specifications.

    7. The Type of Ground Truth Used

    The document does not explicitly state the type of ground truth used for evaluating the imaging performance beyond physical and electrical measurements. The "Non-clinical & Clinical considerations" mention that "all test results were satisfactory," implying some form of objective assessment was conducted, but the specific nature of the ground truth (e.g., expert consensus, pathology, outcomes data) for image quality or diagnostic accuracy is not detailed. The primary ground for acceptance appears to be substantial equivalence to a predicate device based on technical specifications and conformance to safety and EMC standards.

    8. The Sample Size for the Training Set

    This information is not applicable as the EzSensor P is a digital X-ray sensor, not an AI/ML device that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable as the EzSensor P is a digital X-ray sensor, not an AI/ML device that requires a training set.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1