    Intended Use

    The EZE SIT Valvulotome is intended to render venous valves incompetent during in situ bypass procedures. This includes distal infrainguinal bypass when a non-anatomic position is required (e.g., profunda to anterior tibial artery), composite vein infrainguinal bypass, or aorto-renal bypass procedures.

    Device Description

    The EZE SIT Valvulotome is a medical device utilized for disruption of venous valves, and is used in patients that suffer from peripheral vascular disease, for which an attending physician determines the relative benefit of in situ bypass to restore blood flow to the extremities. The EZE-SIT Valvulotome is comprised of a catheter with a luer-lock hub at the proximal end and a threaded connector at the distal end. The catheter is designed with a lumen that extends throughout its entire length. The threaded connector allows attachment of 2 mm, 3 mm and 4 mm diameter cutting heads. These cutting heads are also configured with a central lumen. This design allows the physician to select the cutter head diameter that is best suited for the particular vessel being prepared. The central lumen allows irrigation during the procedure. The cutting head design minimizes vessel wall contact while effectively disrupting valve leaflet tissue. The EZE SIT Valvulotome is surgically invasive and intended for less than 24-hour use. This medical device has no software components. This device is not an in vitro diagnostic device.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for a medical device called the EZE SIT Valvulotome. It asserts substantial equivalence to a predicate device (Tru-Incise Valvulotome K930011).

    Crucially, this document states that no performance data, including clinical studies or data demonstrating adherence to acceptance criteria for an AI/algorithm-driven device, were required or provided. The submission is intended "to simply indicate a new manufacturer for a currently cleared device design that has not changed since its clearance (K930011)."

    Therefore, the acceptance criteria and study proving the device meets those criteria, as requested in the prompt, are not present in this document. This device does not have software components and is not an in vitro diagnostic device, and thus the type of AI/algorithm performance study envisioned by the prompt is not applicable to this submission.

    The document only includes:

    • Nonclinical data: Sterilization, Shelf Life/Aging, and Biocompatibility, all of which confirmed substantial equivalence to the predicate device.
    • Sterilization: Validated with half-cycle using ISO 11135-1: 2014, achieving a Sterility Assurance Level (SAL) of 10-6. Residuals measured: 0.6 mg EtO, <0.1 mg Ethylene Chlorohydrin (ECH). Endotoxin measured by LAL method.
    • Shelf Life/Aging: Extended shelf life to 6 years (from 2.5 years for the predicate) based on accelerated aging.
    • Biocompatibility: Passed all tests in the biocompatibility battery, recapitulating tests performed for the predicate device.

    In summary, none of the requested information regarding AI/algorithm performance (acceptance criteria table, sample sizes, expert ground truth, MRMC study, standalone performance, training data, etc.) can be extracted from this document, as it pertains to a non-AI/non-software medical device that underwent a 510(k) submission based on substantial equivalence without requiring new performance data for design changes.