    K Number
    K023442

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2002-11-05

    (21 days)

    Product Code
    Regulation Number
    892.1000
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Signa Ovation (formerly known as MFO/I) system is an open, whole body scanner designed to support improved higher resolution imaging and shorter scan times. The Signa Ovation system is indicated for use as a diagnostic imaging device to produce transverse, sagittal, coronal and oblique images of the internal structures and organs of the entire body, including, but not limited to, the musculocskeletal, vascular, cardiac, and neuro systems. The images produced by the Signa Ovation system reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin-lattice relaxation time (T1), spin-spin relaxation time (T2) and flow. When interpreted by a trained physician, these images provide information that can be useful in determining diagnosis.

    Due to the 'open' design of the system, the Signa Ovation may also be used for imaging during interventional procedures when performed with MR compatible devices such as, in-room display, and MR safe biopsy needles.

    Device Description

    The Extra Large Body Coil is a modification to the Signa MFO/i MR System (K002611) which includes the addition of a flexible extra large body coil.

    AI/ML Overview

    This 510(k) summary describes a modification to an existing device, the GE Signa MFO/i MR System (K002611), by adding an "Extra Large Body Coil." The document focuses on demonstrating substantial equivalence to the predicate device rather than providing a detailed study of the new coil's performance against specific acceptance criteria in a clinical setting.

    Therefore, many of the requested elements for a study proving a device meets acceptance criteria are not explicitly present in the provided text. The document states that "Testing was performed to demonstrate that the design modifications to the Extra Large Body Coil meet predetermined acceptance criteria," but it does not elaborate on what those criteria were, the specific tests conducted, or the results.

    Based on the provided text, here is what can be extracted and what cannot:

    1. Table of acceptance criteria and the reported device performance

    The document does not provide a table of acceptance criteria or specific reported device performance metrics for the Extra Large Body Coil against those criteria. It broadly states that "Testing was performed to demonstrate that the design modifications to the Extra Large Body Coil meet predetermined acceptance criteria." These acceptance criteria would likely relate to image quality (resolution, signal-to-noise ratio), safety, and compatibility with the existing MR system, but details are not given.

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    This information is not provided in the document.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    This information is not provided. The study design described is not a clinical study involving expert interpretation of images for ground truth.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    This information is not provided.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC study is mentioned. This device is an MRI coil, not an AI-powered diagnostic tool, so an MRMC study comparing human readers with and without AI assistance would not be applicable here.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable, as this is hardware (an MRI coil), not a standalone algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The concept of "ground truth" in the context of clinical diagnostic accuracy studies is not directly applicable here. The testing would have focused on engineering and performance specifications of the coil, not diagnostic accuracy based on clinical outcomes or pathology.

    8. The sample size for the training set

    Not applicable, as this is a hardware device, not a machine learning algorithm requiring a training set.

    9. How the ground truth for the training set was established

    Not applicable.

    Summary of Study from the document:

    The study described is a technical verification that the "design modifications to the Extra Large Body Coil meet predetermined acceptance criteria." The purpose of this study was to support the claim of substantial equivalence to the predicate device (GE Signa MFO/i MR system, K002611). The document concludes that "Usage of the Extra Large Body Coil does not result in any new potential hazards." This type of submission (510(k)) primarily focuses on demonstrating that a new device is as safe and effective as a legally marketed predicate device, rather than providing detailed clinical performance data against specific diagnostic acceptance criteria.

    Ask a Question

    Ask a specific question about this device

    K Number
    K991057

    Validate with FDA (Live)

    Date Cleared
    1999-05-20

    (51 days)

    Product Code
    Regulation Number
    892.1000
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Imaging of:

    • -The Whole Body (including head, abdomen, breast, pelvis, limbs and extremities, spine, neck, TMJ, heart, blood vessels). [Application terms include MRCP (MR Cholangiopancreatography), MR Urography, MR Myelography, MR Fluoroscopy, SAS (Surface Anatomy Scan), Dynamic Scan and Cine Imaging.]
    • Fluid Visualization -
    • -2D/3D Imaging
    • MR Angiography/MR Vascular Imaging -
    Device Description

    The Extra Large Body coil is an enlarged version of the current OPART™ Large Body Coil. This coil is an optional coil which was developed to allow positioning of larger sized patients in the OPART™ system. The Extra Large Body coil can be used as a standalone coil, or in conjunction with the T-spine surface coil (cleared with OPART™ system K962933). Signal is received in QD mode when used in conjunction with the T-spine surface coil.
    The Extra Large Body Coil is constructed with the same materials that are currently in use for the released coil set for OPART™.

    AI/ML Overview

    This document describes the design and regulatory approval of an "Extra Large Body Coil" for an MRI system, the OPART™. The information provided primarily focuses on establishing the device's substantial equivalence to a previously cleared device, rather than a clinical study evaluating its performance against specific acceptance criteria in a medical context.

    Based on the provided text, there is no clinical study that explicitly defines and proves the device meets specific performance acceptance criteria in the way typically expected for a diagnostic AI/medical device. Instead, the submission focuses on substantial equivalence for a physical component (the coil) of an MRI system.

    Here's an analysis of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The concept of "acceptance criteria" for algorithms or diagnostic performance metrics (e.g., sensitivity, specificity, AUC) is not applicable here because this is a hardware accessory submission based on substantial equivalence. The "performance" being evaluated is related to physical and safety parameters of the MRI coil.

    Acceptance Criteria (Benchmarking/Safety/Physical)Reported Device Performance (Extra Large Body Coil)
    Safety Parameters:
    Maximum static field strength0.35 Tesla
    Rate of change of magnetic field19 T/second
    Maximum radio frequency power deposition (SAR)<0.4 Watt/kg
    Acoustic noise levels (maximum)98.4 dB (A)
    Imaging Performance Parameters:
    Head Specification volume10 cm dsv
    Body Specification volume20 cm dsv
    Material compositionSame materials as currently in use for the released coil set for OPART™
    Intended UseImaging of Whole Body (including head, abdomen, breast, pelvis, limbs and extremities, spine, neck, TMJ, heart, blood vessels), Fluid Visualization, 2D/3D Imaging, MR Angiography/MR Vascular Imaging (with no new indications from predicate)

    Note: The "acceptance criteria" here are implicitly the existing standards and performance characteristics of the predicate device (the current Large Body Coil cleared with K962933) and general MRI safety standards. The new coil is presented as meeting these same criteria or being within acceptable limits.

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Description: The "test set" in this context refers to the data used to demonstrate the functionality of the new Extra Large Body Coil. This is not a clinical study test set in the sense of a patient cohort for diagnostic algorithm validation.
    • Data Provenance: The document mentions "Sample phantom images and clinical images are presented from the Extra Large Body coil (Appendix 5 & 6)."
      • Sample Size: Not explicitly stated. The term "sample" implies a limited number of instances rather than a statistically powered clinical trial.
      • Provenance: The location (country of origin) of these phantom and clinical images is not specified.
      • Retrospective/Prospective: Not specified. Given it's a 510(k) for a hardware accessory, it's likely these images were collected as part of the engineering and verification process, which could be considered a form of prospective data collection for this purpose, but not a large-scale clinical trial.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    • Not applicable. The "ground truth" for a hardware component like an MRI coil typically relates to its physical performance (e.g., signal-to-noise ratio, image uniformity, mechanical integrity) which is measured objectively using phantoms and engineering tests, and clinical image quality (which would be reviewed by radiologists). The submission does not detail an expert review process for image quality beyond a general statement that "clinical images are presented." There is no indication of a formal ground-truth establishment process by a panel of experts for diagnostic accuracy in this submission.

    4. Adjudication Method for the Test Set

    • Not applicable. As a hardware submission focused on substantial equivalence and safety, there is no mention of an adjudication method for a test set in the context of diagnostic performance.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No. An MRMC study is typically used for evaluating the performance of diagnostic algorithms or imaging techniques, often comparing human reader performance with and without AI assistance. This submission is for a physical MRI coil, and such a study is not relevant to its regulatory approval process described here.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    • No. This is a hardware component. The concept of an "algorithm only" or "human-in-the-loop" study does not apply.

    7. Type of Ground Truth Used

    • In this context, the "ground truth" is primarily derived from:
      • Engineering Specifications: Testing the coil against predefined physical and electromagnetic parameters (e.g., static field strength, SAR, acoustic noise levels, specification volumes).
      • Phantom Imaging: Objective measurements of image quality (e.g., signal-to-noise, uniformity) using standardized phantoms.
      • Clinical Image Appearance: Subjective assessment of image quality and diagnostic utility from "clinical images" (though the criteria and expert review process are not detailed in this summary).
    • There's no mention of pathology, long-term outcomes data, or expert consensus in the diagnostic sense being used for ground truth in this specific submission.

    8. Sample Size for the Training Set

    • Not applicable. This submission is for a hardware component, not an AI/ML algorithm. Therefore, there is no "training set."

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable. As there is no training set for an AI/ML algorithm, this question is not relevant to the provided text.