510(k) Data Aggregation

    K Number
    K200991

    Validate with FDA (Live)

    Date Cleared
    2020-05-13

    (28 days)

    Product Code
    Regulation Number
    888.3080
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The EVOL® Spinal Interbody System is an intervertebral body device intended for use in skeletally mature patients with Degenerative Disc Disease (DDD) of the lumbar spine with up to Grade 1 spondylolisthesis at one or two contiguous levels from L2-S1. Degenerative disc disease is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies.

    The EVOL® Spinal Interbody System is intended to be used with autograft comprised of cancellous and/ or corticocancellous bone graft to facilitate fusion. It is to be used in patients who have had six months of non-operative treatment and is to be implanted via a direct posterior, transforaminal or anterior approach. The CES-LIF is implanted in pairs in the lumbosacral spine. The EVOL® Spinal Interbody System is intended to be used with supplemental fixation.

    Device Description

    The EVOL® Spinal Interbody System is a family of PEEK-OPTIMA® HA Enhanced spacers offered in a variety of widths and lengths. The CES-LIF are generally rectangular in shape. The CES-LIF is used in pairs.

    The EVOL® Spinal Interbody System implants are available in a range of sizes to accommodate variations in patients' anatomy. In addition, tantalum beads are embedded in the spacers to help allow for radiographic visualization. The hollow implants have holes through four sides for bone graft and an inserter instrument interface on the face. The CES-LIF HA implants themselves are made of biocompatible PEEK-Optima® HA Enhanced (LT120) polymer material and provided non-sterile for single-use.

    Teeth on top and bottom of the implants improve fixation. They are symmetrical so that the implant will resist motion equally in either direction (anterior or posterior).

    The purpose of this submission is to add a new sterilization method.

    AI/ML Overview

    This document is a 510(k) premarket notification for a medical device called the "EVOL® Spinal Interbody System." This type of submission is for demonstrating substantial equivalence to a legally marketed predicate device, rather than proving safety and effectiveness through extensive clinical trials with specific acceptance criteria that a new, novel device would require.

    Therefore, the document does not contain the kind of information you are asking for regarding acceptance criteria and a study that proves the device meets those criteria, as typically seen for AI/ML-driven medical devices or devices requiring clinical performance studies to establish performance claims.

    Here's why the requested information cannot be extracted from this document:

    • Device Type: The EVOL® Spinal Interbody System is a spinal implant (a physical medical device), not an independent diagnostic or AI/ML-driven device that would output performance metrics like sensitivity, specificity, or image analysis results.
    • Regulatory Pathway (510(k)): A 510(k) simply requires demonstrating that the new device is "substantially equivalent" to an already legally marketed device (predicate device). This is primarily done through a comparison of technological characteristics, materials, intended use, and sometimes non-clinical performance (e.g., mechanical testing, biocompatibility). It generally doesn't involve clinical performance studies with pre-defined acceptance criteria, expert ground truth establishment, or MRMC studies for quantifying human-AI collaboration.
    • Purpose of Submission: The specific purpose of this K200991 submission is to add a new sterilization method (steam sterilization) for an already cleared device line (EVOL® Spinal Interbody System). It explicitly states: "No additional testing is required to determine if Steam as a sterilization method will impact the safety or efficacy of this implant."

    In summary, this document does not describe a study that proves the device meets specific acceptance criteria related to its diagnostic or performance accuracy, as it is a physical implant and the submission focuses on a change in sterilization method.

    If this were an AI/ML device, the detailed information requested would be crucial for its 510(k) clearance or de novo authorization.

    Ask a Question

    Ask a specific question about this device

    K Number
    K180891

    Validate with FDA (Live)

    Date Cleared
    2018-05-04

    (29 days)

    Product Code
    Regulation Number
    888.3080
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The EVOL® Spinal Interbody System is an intervertebral body device intended for use in skeletally mature patients with Degenerative Disc Disease (DDD) of the lumbar spine with up to Grade 1 spondylolisthesis at one or two contiguous levels from L2-S1. Degenerative disc disease is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies.

    The EVOL® Spinal Interbody System is intended to be used with autograft comprised of cancellous and/ or corticocancellous bone graft to facilitate fusion. It is to be used in patients who have had six months of non-operative treatment and is to be implanted via a direct posterior, transforaminal, retroperitoneal or anterior approach. The CES-PLIF and CES-LIF are implanted in pairs, while the CES-TLIF and CES-RLIF devices may be implanted singly or in pairs in the lumbosacral spine. The EVOL® Spinal Interbody System is intended to be used with supplemental fixation.

    The EVOS Lumbar Interbody System is an intervertebral body device intended for use in skeletally mature patients with Degenerative Disc Disease (DDD) of the lumbar spine with up to Grade 1 spondylolisthesis at one or two contiguous levels from L2-S1. Degenerative disc disease is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies.

    The EVOS Lumbar Interbody System is intended to be used with autograft comprised of cancellous and or corticocancellous bone graft to facilitate fusion. It is to be used in patients who have had six months of non-operative treatment and is to be implanted via a direct posterior or transforaminal approach. The EVOS CURVED devices are implanted singly, while the EVOS ROTATE and EVOS STRAIGHT devices may be implanted singly or in pairs in the lumbosacral spine. The EVOS Lumbar Interbody System is intended to be used with supplemental fixation.

    Device Description

    The EVOL® Spinal Interbody System is a family of PEEK-OPTIMA or PEEK-OPTIMA HA spacers offered in a variety of widths and lengths. There are five main configurations: CES-PLIF, CES-LIF, CES-TLIF, CES-CLIF and CES-RLIF. The different configurations allow for multiple surgical technique options. The CES-PLIF, CES-LIF, CES-CLIF and CES-RLIF are generally rectangular in shape while the CES-TLIF is curved. The CES-PLIF and CES-LIF are used in pairs while the other options are used singly or in pairs.

    The EVOL® Spinal Interbody System implants are available in a range of sizes, as well as parallel and lordotic angled implants, to accommodate variations in patients' anatomy. In addition, tantalum beads or pins are embedded in the spacers as an option to help allow for radiographic visualization. The hollow implants have holes through four sides for bone graft and an inserter instrument interface on the face. The implants themselves are made of biocompatible PEEK-Optima Natural (LT1) or [PEEK-Optima HA Enhanced (LT120) - CES-LIF only] polymer material (per ASTM F2026).

    Teeth on top and bottom of the implants improve fixation. They are symmetrical so that the implant will resist motion equally in either direction (anterior or posterior).

    The EVOS Lumbar Interbody System is comprised of a variety of Implants manufactured from PEEK (Polyetheretherketone) -Optima LT1 and PEEK Optima LT120HA per ASTM F-2026, with Tantalum bead markers per ASTM F-560. There are three main configurations; STRAIGHT, CURVED, and ROTATE. The different configurations allow for multiple surgical technique options. The EVOS STRAIGHT and the EVOS ROTATE are generally rectangular in shape while the EVOS CURVED is curved. The EVOS CURVED devices are implanted singly, while the EVOS ROTATE and EVOS STRAIGHT devices may be implanted singly or in pairs.

    The EVOS device(s) are available in a range of sizes, as well as flat and biconvex endplates, and with various degrees of lordosis to accommodate variations in patients' anatomy. The angle of lordosis on each lordotic EVOS device is oriented in a way that will provide a true anterior/posterior lordotic orientation once the device has been implanted in its final intervertebral location. For example, the True Oblique Lordotic (TOL) devices have a lordotic orientation to the device that will provide a correct anterior/posterior lordotic orientation once they have been inserted at 30 degrees from the anterior/posterior midline.

    Implant heights range from 6mm to 16mm in maximum height (minus the height of the teeth). Widths range from 8mm to 12mm and lengths range from 22mm to 30mm. The variety of implant shapes and sizes accommodate various surgical technique options. The hollow implants have holes through four sides for bone graft and an inserter instrument interface on the face. Teeth on top and bottom of the spacers improve fixation. The EVOS Lumbar Interbody System includes the instrumentation to facilitate the implantation of the implants.

    The approach and the discectomy are conducted using standard instruments while subsequent steps are conducted using standard and/or custom instruments. The system is comprised of instruments and perforated instrument cases that are generally comprised of aluminum, stainless steel and/or polymeric materials.

    The EVOS rotate and non-rotate Inserter(s) are instruments intended for use with the EVOS Lumbar Interbody System. The inserter(s) are designed specifically to implant the EVOS device into a prepared disc space. The EVOS non-rotate inserters are designed to work with the EVOS straight and curved devices while the EVOS rotate inserter is designed to work with the EVOS rotate devices.

    AI/ML Overview

    The provided text is a 510(k) summary for the EVOL® Spinal Interbody System and EVOS Lumbar Interbody System. It does not describe acceptance criteria, a specific study proving the device meets acceptance criteria, or any details related to AI/algorithm performance.

    The document focuses on demonstrating substantial equivalence to predicate devices based on technological characteristics, intended use, design features, and mechanical performance. It mentions non-clinical performance testing for the EVOL® Spinal Interbody System, specifically:

    • Static Compression
    • Static Compression Shear
    • Dynamic Compression
    • Expulsion testing

    These tests were performed per ASTM F2026, and the conclusion states that the "Testing performed showed substantial equivalence to the predicate." However, specific acceptance criteria values and the reported device performance values are not provided within this summary for these tests.

    For the EVOS Lumbar Interbody System, the document states, "No design changes were made to the existing devices, nor were any new components added to the system. Therefore, no additional testing was required or performed." This implies that prior testing for the original EVOS system would have satisfied the substantial equivalence requirements, but those details are not in this document.

    Therefore, I cannot fulfill the request for information regarding acceptance criteria, device performance, sample sizes, data provenance, expert qualifications, adjudication methods, MRMC studies, standalone algorithm performance, or ground truth details because these particulars are not present in the provided text. The document is a regulatory submission focused on substantial equivalence based on mechanical and material properties, not an AI or diagnostic device study.