Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K983776

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    1999-04-26

    (182 days)

    Product Code
    Regulation Number
    884.5160
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Evenflo Comfort Control Breast Pump and Comfort Control Dual Breast Pump are personal use battery/electric powered suction devices used to express milk from the breast of lactating women. These devices are not intended for hospital use.

    Device Description

    The Evenflo Comfort Control Adjustable Speed Breast Pump with Automatic Cycling, and the Evenflo Comfort Control Dual Adjustable Speed Breast Pump with Automatic Cycling are diaphragm type pumps, driven by small DC motors, that create cycling negative pressure. The negative pressure generated by the pump is sufficient to overcome the resistance to the outflow of milk that is created by the muscles, yet gentle enough to prevent tissue damage to the user. The cycling speed and the strength of the negative pressure are controlled by individual rotary controls located on the pump unit. The combination of cycling speed and negative pressure control allow for a range of pressures to be selected. When used as a single collection pump, the maximum negative pressure ranges from 173 mm Hg to 56 mm Hg, and when used in the dual mode, it ranges from 173 mm Hg to 43 mm Hg.

    AI/ML Overview

    The provided text describes a 510(k) submission for the Evenflo Comfort Control Breast Pump and Evenflo Comfort Control Dual Breast Pump. As such, the "acceptance criteria" and "device performance" are framed in terms of substantial equivalence to predicate devices, rather than strict quantitative performance metrics against a defined standard typical of a new device's clinical trial.

    Here's the information extracted and organized as requested:

    Acceptance Criteria and Device Performance for Evenflo Comfort Control Breast Pump

    1. Table of Acceptance Criteria and Reported Device Performance

    The "acceptance criteria" here are implicitly defined by the performance characteristics of the predicate devices. The Evenflo Comfort Control Breast Pump aims to demonstrate substantial equivalence by matching or exceeding the performance of these predicates in key areas.

    Performance CharacteristicAcceptance Criteria (Predicate Range/Value)Evenflo Comfort Control Performance (Reported Value)
    Intended UseTo Express MilkTo Express Milk
    Power SourceAC Adapter / BatteriesAC Adapter or 2 C Batteries
    Single or Double PumpingYes / Single OnlyYes (for Dual model)
    Adjustable Suction LevelsYesYes
    Adjustable Cycle SpeedYes / NoYes
    Overflow ProtectionYesYes, valve in horn
    Highest Vacuum Setting (mm Hg)165 - 185 mm Hg173 mm Hg
    Lowest Vacuum Setting (mm Hg)30 - 109 mm Hg43 mm Hg
    Range of Cycle Speeds (Cycles/Minute)29 - 60 Cycles/Minute36 - 54 Cycles/Minute
    Nipple AdapterYesYes
    Time to 75% Initial Vacuum on Batteries (hours)1 - 20 hours7 hours
    Time to 0% Initial Vacuum on Batteries (hours)5 - 37 hours10 hours

    Note: The acceptance criteria are derived from the range of performance observed across the predicate devices listed. The Evenflo device's reported performance falls within, or is comparable to, these ranges, supporting its claim of substantial equivalence.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: Not explicitly stated for specific performance tests. The testing was conducted on "its Comfort Control Pumps and each of the predicate devices" implying at least one unit of each model.
    • Data Provenance: Not specified, but likely internal testing conducted by Evenflo Company, Inc. It is not mentioned whether it was retrospective or prospective, or the country of origin of the data.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • Number of Experts: Not applicable. The "ground truth" for this device's performance is objective physical measurements (vacuum pressure, cycle speed, battery life), not expert consensus on medical outcomes.
    • Qualifications of Experts: Not applicable.

    4. Adjudication Method for the Test Set

    • Adjudication Method: Not applicable. Performance was measured objectively, not interpreted by adjudicators.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    • MRMC Study: No, an MRMC comparative effectiveness study was not done. This type of study is more common for diagnostic or imaging devices where human interpretation is a key factor.
    • Effect Size: Not applicable.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    • Standalone Performance: Yes, the performance testing described as "Non-Clinical Testing" on the Evenflo Comfort Control Pumps and predicate devices would be considered standalone performance, as it involves objective measurement of device parameters (negative pressure, cycle speed, battery life) without a human operator's influence on the outcome beyond operating the device for the test.

    7. The Type of Ground Truth Used

    • Type of Ground Truth: Objective physical measurements and specifications. The "ground truth" for this device's performance evaluation is based on direct measurements of mechanical parameters such as vacuum pressure (mm Hg), cycle speed (cycles/minute), and battery life (hours).

    8. The Sample Size for the Training Set

    • Sample Size: Not applicable. This device is a mechanical breast pump, not an AI/ML algorithm that requires a "training set" in the computational sense. The design and engineering would be based on general principles and regulatory requirements, not a data-driven training set.

    9. How the Ground Truth for the Training Set Was Established

    • Ground Truth Establishment: Not applicable, as there is no "training set" for a mechanical device like this. The device design and safety would be established through engineering principles, material science, and regulatory standards for medical devices.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1