Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K171307

    Validate with FDA (Live)

    Date Cleared
    2017-06-22

    (50 days)

    Product Code
    Regulation Number
    876.1500
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    EsophyX2 HD Device with SerosaFuse Fasteners and Accessories: The EndoGastric Solutions EsophyX2 HD Device with SerosaFuse Fasteners and Accessories is indicated for use in transoral tissue approximation, full thickness plication and ligation in the GI tract and is indicated for the treatment of symptomatic chronic gastroesophageal reflux disease in patients who require and respond to pharmacological therapy. It is also indicated to narrow the gastroesophageal junction and reduce hiatal hernia < 2cm in size in patients with symptomatic chronic gastroesophageal reflux disease. Patients with hiatal hernias larger than 2cm may be included, when a laparoscopic hiatal hernia repair reduces the hernia to 2cm or less.

    EsophyX Z Device with SerosaFuse Fasteners and Accessories: The EndoGastric Solutions EsophyX Z Device with SerosaFuse Fasteners and Accessories is indicated for use in transoral tissue approximation, full thickness plication and ligation in the GI tract and is indicated for the treatment of symptomatic chronic gastroesophageal reflux disease in patients who require and respond to pharmacological therapy. It is also indicated to narrow the gastroesophageal junction and reduce hiatal hernia ≤ 2cm in size in patients with symptomatic chronic gastroesophageal reflux disease. Patients with hiatal hernias larger than 2cm may be included, when a laparoscopic hiatal hernia repair reduces the hernia to 2cm or less.

    Device Description

    The subject EsophyX devices are unchanged from the currently cleared devices. The principles of operation for the devices are the same and also remain unchanged from the prior clearances. The indications for use for both models are identical.

    The EndoGastric Solutions EsophyX Device models (EsophyX2 HD and EsophyX Z, K142113 and K160960, respectively) with SerosaFuse Fasteners and Accessories are a prescription use only, disposable, single use system consisting of an all mechanical, flexible fastener delivery device with user controls outside the patient's body. The devices are used in an operating room environment, using the Transoral Incisionless Fundoplication (TIF) procedure.

    The device is classified as a surface device, mucosal membrane contacting with limited duration. The fastener classification is an implant, with permanent tissue/bone contact. Materials include acrylic/polycarbonate plastics, machined aluminum, stainless steel, nitinol wire, UV cured adhesives, pellethane and nylon.

    The implantable tissue fasteners are constructed of injection molded polypropylene.

    These transoral devices and tissue fasteners are provided sterile (EO). A separate, commercial endoscope operating independently down the center of the devices' flexible shaft lumen provides visualization of the procedure at all times, from device insertion through extraction. Two fasteners are loaded from a replaceable fastener cartridge containing ten fasteners for each channel. The loaded fasteners are pushed from the proximal device end. to the distal end down two separate channels via lumens in a flexible shaft where they are then in position ready for deployment into tissue. The operator, controlling the distal end of the devices through the mechanical controls at the proximal end of the devices, captures and positions a fold of tissue at the gastroesophageal junction. The fastener controls deploy both fasteners simultaneously in the EsophyX Z model, or individually in the EsophyXo HD model, at the captured tissue position. This creates a permanent surgical partial fundoplication. Additional fasteners are used as needed to complete the valve restoration.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the "EsophyX2 HD Device with SerosaFuse Fasteners and Accessories" and "EsophyX Z Device with SerosaFuse Fasteners and Accessories." This submission focuses on demonstrating substantial equivalence to previously cleared predicate devices, primarily due to an expanded indication for use rather than presenting novel performance data for the devices themselves.

    Therefore, the typical structure of acceptance criteria and a study proving a device meets them (especially for AI/algorithmic medical devices) is not directly applicable in the way you've outlined. This document is a regulatory submission for a mechanical device that has existing clearances. The "study" here refers to a literature review supporting an expanded indication.

    However, I can extract information related to the closest aspects of your request based on the provided text, focusing on how the expanded indication was supported:


    Understanding the Regulatory Context

    This is a 510(k) submission, meaning the manufacturer is demonstrating that their device is "substantially equivalent" to a legally marketed predicate device. This is not a de novo submission or a PMA, which would typically require more extensive clinical trials to prove efficacy from scratch.

    In this specific 510(k), the key element is an expanded indication for use. The devices themselves are stated to be "unchanged from the currently cleared devices" and "identical to the predicates devices." Therefore, the submission is not about demonstrating new performance for the device's core function but rather showing that the expanded use case is safe and effective when combined with the device's established performance.

    Response to Your Specific Questions:

    1. Table of Acceptance Criteria and Reported Device Performance

    Given that this is a mechanical device whose performance is already cleared, and the new aspect is an expanded indication supported by literature, there isn't a direct "acceptance criteria" table in the AI/algorithm sense.

    However, we can infer the "criteria" for the expanded indication were based on clinical outcomes reported in existing literature for the combined procedure (laparoscopic hiatal hernia repair then TIF with the EsophyX device).

    Acceptance Criteria Category (Inferred)Reported Device Performance / Clinical Outcomes (from Literature)
    SafetyNo long-term postoperative complications (Ihde et al.)
    Most complications were minor (dysphagia, nausea, vomiting) (Chang et al.)
    No adverse effects reported (Janu et al.)
    Effectiveness (GERD-HRQL Score)73% of patients normalized GERD-HRQL score (Ihde et al.)
    Greatest GERD-HRQL score improvement (average 20.5 points) observed in HHR + fundoplication (Chang et al.)
    HRQL scores improved 85% for heartburn and regurgitation questions (Janu et al.)
    Effectiveness (Symptom Elimination)73% of patients eliminated heartburn (Ihde et al.)
    76% of patients eliminated regurgitation (Ihde et al.)
    Effectiveness (Medication Use)76% of patients were off daily PPIs (Ihde et al.)
    Effectiveness (Other Symptoms)67% of patients reported elimination of atypical symptoms (Ihde et al.)
    RSI scores for hoarseness, throat clearing, excess mucus, coughing, chest pain improved 50-80% (Janu et al.)
    GSRS questions on heartburn & regurgitation showed 80% improvement (Janu et al.)
    DurabilityAll results durable at 6 and 12 months follow up (Janu et al.)
    Patient Satisfaction88% of patients satisfied with current health condition (Ihde et al.)

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The clinical outcomes supporting the expanded indication were drawn from a literature review summarizing outcomes from 163 patients across multiple studies.
      • Ihde et al.: 48 patients.
      • Chang et al.: 221 patients (sub-group underwent HHR + TIF, N=46).
      • Janu et al.: ~99 patients.
    • Data Provenance: The studies are published clinical literature. The specific countries of origin are not specified, but the journals (e.g., "The American Journal of Surgery," "Journal of the Society of Laparoendoscopic Surgeons") suggest North American or international scope. The studies are retrospective analyses of collected patient data.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    • This information is not provided in the text. The "ground truth" here is essentially the reported clinical outcomes in peer-reviewed literature. The authors of these papers conducted the studies and presented their findings, which are then summarized in this submission. There is no mention of external experts specifically establishing "ground truth" for a test set in the context of this 510(k) submission.

    4. Adjudication Method for the Test Set

    • Not applicable. There was no independent "test set" in the sense of a dataset requiring expert adjudication for ground truth. The submission relies on already published clinical findings.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    • No. This is not an AI device, and the submission does not describe an MRMC study. It describes a literature review supporting an expanded surgical indication for a mechanical device.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • Not applicable. This is a mechanical device, not an algorithm.

    7. The Type of Ground Truth Used

    • Clinical Outcomes Data from Published Literature: The "ground truth" for the expanded indication is established through the reported efficacy and safety data from published clinical studies on patients undergoing the combined procedure (laparoscopic hiatal hernia repair followed by TIF). This includes objective measures (e.g., GERD-HRQL scores, PPI discontinuation rates) and patient-reported outcomes.

    8. The Sample Size for the Training Set

    • Not applicable. This is not an AI/machine learning device that uses a "training set."

    9. How the Ground Truth for the Training Set was Established

    • Not applicable. As above, no training set for an algorithm is involved. The "ground truth" for the device's primary function was established during its original clearances (K142113 and K160960). For the expanded indication, it was established by the clinical data collected and reported in the cited scientific literature.
    Ask a Question

    Ask a specific question about this device

    K Number
    K142113

    Validate with FDA (Live)

    Device Name
    ESOPHYX2 HD
    Date Cleared
    2014-09-30

    (57 days)

    Product Code
    Regulation Number
    876.1500
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The EndoGastric Solutions EsophyX® 2 HD Device with SerosaFuse is indicated for use in transoral tissue approximation, full thickness plication in the GI tract and is indicated for the treatment of symptomatic chronic gastroesophageal reflux disease in patients who require and respond to pharmacological therapy. It is also indicated to narrow the gastroesophageal junction and reduce hiatal hernia < 2cm in size in patients with symptomatic chronic gastroesophageal reflux disease.

    Device Description

    The EndoGastric Solutions EsophyX® 2 HD Device with SerosaFuse Fasteners and Accessories consists of an all mechanical, flexible fastener delivery device with user controls outside the patient's body and sterile polypropylene fasteners delivered transorally through a flexible shaft into the GI tract via a common delivery mechanism comprised of three elements: a stylet, a pusher rod, and a delivery tube. All three fastener delivery elements run the length of the delivery device. The stylet runs down the inside of the lumen of the delivery tube and the pusher rod rides over the length of the stylet. The stylet is designed to pierce and hold the desired tissue plication in place and guide the fastener into position. There are two separate tubes or channels in the device, referred to as the posterior and anterior channels. A fastener is loaded mechanically onto the stylet via a replaceable fastener cartridge containing ten fasteners for each channel. To load, the fastener cartridge lever is depressed, which snaps a fastener onto the stylet. The fastener is then pushed down the stylet from the proximal handle assembly to the distal tissue port via the pusher rod where it is then ready to be deployed into the tissue. When the desired tissue approximation/ plication is achieved, the pusher rod is used to push the fastener into the tissue, quided by the stylet which precedes it into the tissue. The leading leq of the fastener slides along the stylet into the tissue. Slight additional pressure advances the pusher causing the leading leg of the fastener to disengage from the stylet and the fastener then seats across the desired tissue plication. A new fastener is loaded proximally in the device at the loading port, after retracting the pusher rod. The loading and fastening procedure can be repeated. The device is provided sterile and is a single use device.

    AI/ML Overview

    The provided document is a 510(k) summary for the EndoGastric Solutions EsophyX®2 HD Device with SerosaFuse Fasteners and Accessories. It focuses on demonstrating substantial equivalence to a predicate device, rather than providing detailed acceptance criteria and study results for algorithm performance as would be expected for an AI/ML device.

    Therefore, much of the requested information cannot be extracted from this document, as it pertains to AI/ML device evaluation, which is not the subject of this 510(k) submission. This submission is for a physical medical device.

    Here's what can be extracted and what cannot:

    1. Table of acceptance criteria and the reported device performance:

    Acceptance CriteriaReported Device Performance
    Meet design specifications and user needs.Verification and validation testing demonstrates the device meets product specifications and user requirements.
    Meet biocompatibility requirements.Biocompatibility testing provides evidence the device meets biocompatibility requirements.
    Compatible with high-definition endoscopes.Verified and validated to be compatible with high definition endoscopes.
    Maintain the same indications for use as the predicate device.Same indications for use as predicate.
    Maintain the same operating principle as the predicate device.Same operating principle as predicate.
    Maintain the basic design with a larger shaft diameter (60Fr versus 54Fr) compared to the predicate device.Incorporates the same basic design with a larger shaft diameter (60Fr versus 54Fr).

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective):

    This information is not available in the provided document. The document refers to "verification and validation testing" which included "dimensional, mechanical, and performance testing" but does not specify sample sizes for these tests, nor the nature of the data or its provenance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):

    This information is not applicable as this is not an AI/ML device requiring expert-established ground truth for a test set.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    This information is not applicable as this is not an AI/ML device requiring adjudication of a test set.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This information is not applicable as this is not an AI/ML device.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    This information is not applicable as this is not an AI/ML device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    This information is not explicitly stated in terms of "ground truth" as it would be for an AI/ML device. For a physical device, performance evaluation typically involves engineering specifications, compatibility testing, and simulated use.

    8. The sample size for the training set:

    This information is not applicable as this is not an AI/ML device.

    9. How the ground truth for the training set was established:

    This information is not applicable as this is not an AI/ML device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1