Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K050424

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2005-06-23

    (125 days)

    Product Code
    Regulation Number
    868.1840
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Intended to perform spirometry and may be used in hospitals and medical offices or at other locations including homes. The accessory is a Bacterial/Viral Filter designed for single patient use. For use on both adult and pediatric patients.

    Device Description

    The Ergofilter™ SP1 filter is a bacterial and viral filter with an anatomical mouthpiece, for spirometers. The filter is a single use, non-sterile filter. A new filter must be used for each new patient and for each new spirometry examination. Operating principle of Filtrete™ 3M™ filtration medium: The filtering medium is manufactured by the 3M company in a clean room; it uses an electrostatic process which attracts small mass particles, while having a very small resistance to arr flow; the density and thickness of the filtration medium itself provides an increased filtration capacity.

    AI/ML Overview

    The provided text describes the Ergofilter™ SP1, a bacterial/viral filter accessory for spirometers, and its comparison to a predicate device. Below is an analysis of its acceptance criteria and the study used to demonstrate its performance, based only on the provided text.

    Based on the provided document, the device did not undergo a standalone, human-in-the-loop, or MRMC study. The comparison is primarily based on bench testing of the filter's physical and performance characteristics against a predicate device. No clinical study details (sample size, ground truth, expert qualifications) are provided.


    1. Acceptance Criteria and Reported Device Performance

    CharacteristicAcceptance Criteria (from predicate device)Reported Device Performance (Ergofilter™ SP1)
    Airflow resistance$\leq 0.862 cm H_2O/L/Sec$ at 14 L/Sec$0.60 cm H_2O/L/Sec$ at 12 L/Sec (Note: Different flow rate, but lower resistance suggests meeting the spirit)
    Dead volumeNot explicitly stated as an "acceptance criteria," but predicate has 60 ml65 ml
    % of bacterial removal efficiency>99.7%>99.9%

    2. Sample size used for the test set and data provenance

    The document indicates "bench" and "filter effectiveness" testing, but no specific sample sizes for these tests are provided. No details on data provenance (e.g., country of origin, retrospective/prospective) are given for these tests. The document also mentions "clinical testing data," but no specifics about this data are available within the provided text.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    No information is provided regarding the number or qualifications of experts used to establish ground truth for any tests.

    4. Adjudication method for the test set

    No adjudication method is described. The assessment appears to be based on direct measurement of physical properties and performance characteristics.

    5. If a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was done

    No, there is no indication that an MRMC comparative effectiveness study was done. The comparison is between the new device and a predicate device based on technical specifications and performance metrics, not human reader performance.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    The device is a physical filter, not an algorithm. Therefore, the concept of "standalone performance" in the context of an algorithm does not apply. The performance metrics reported (airflow resistance, bacterial removal efficiency, dead volume) represent the standalone performance of the physical filter itself.

    7. The type of ground truth used

    The ground truth for the performance metrics (airflow resistance, bacterial removal efficiency, dead volume) would be the measured physical properties and filtration capabilities under controlled laboratory conditions, likely following standardized test methodologies, though these specific methodologies are not detailed in the provided text.

    8. The sample size for the training set

    The concept of a "training set" is not applicable as this is a physical medical device (filter), not a machine learning algorithm.

    9. How the ground truth for the training set was established

    Not applicable (see point 8).

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1