Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K173541

    Validate with FDA (Live)

    Device Name
    EPINAUT
    Manufacturer
    Date Cleared
    2018-08-24

    (282 days)

    Product Code
    Regulation Number
    868.5120
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For administration of anesthetic agents into the epidural space to provide epidural or caudal anesthesia.

    Device Description

    EPINAUT Spring Guide Epidural Catheter is constructed of a stainless steel continuous spring. EPINUAT Spring Guide Epidural Catheter offers two types of catheter. The catheter types, L330 & L600 have an uncoated distal tip which is flexible, smooth and rounded with coils slightly spread for maximum flexibility and lateral distribution of injectant. The other catheter types, C310 & C600 have a distal end which is entirely coated by tube for maximum smoothness. There are 16 models in EPINAUT: L33011320, L33013820, L33011315, L33013815, L60011320, L60013820, L60011315, L60013815, C31011320, C31013820, C31011315, C31013815, C60011320, C60013820, C60011315 and C60013815. Each model is characterized by different length of catheter, length and outer diameter of wire, inner diameter of adapter, length and diameter of introducer needle and needle cap.

    EPINUAT Spring Guided Epidural Catheter have enhanced tensile/ break strength and restrict longitudinal catheter stretch, while the spring coils make the catheter resistant to kinking and collapsing. Wire is also provided with the device which consists of a stainless steel and a molded plastic hub.

    The introducer needle is inserted into the epidural space through the sacral hiatus under intermittent fluoroscopy. After appropriate determination of the epidurogram and target area, the catheter with the wire in it is advanced through the introducer needle to the concerned area as determined by MRI or symptomatology. When the catheter is approached to target site, the wire is removed from the catheter and then the medication can be injected by attaching the adapter to a syringe.

    AI/ML Overview

    The medical device in question, EPINAUT, is an Anesthesia Conduction Catheter. The provided context does not include information about clinical studies involving human patients or AI-based performance metrics. The information focuses on non-clinical testing to establish substantial equivalence to a predicate device.

    Here's an analysis of the provided text in relation to the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document describes various non-clinical tests performed to evaluate the performance of EPINAUT and to demonstrate substantial equivalence to the predicate device, the Racz Epidural Catheter. The acceptance criteria are implicitly met if the test results "met the pre-set criteria" or "supported that the subject device is substantially equivalent." Specific numerical acceptance criteria are not explicitly stated for all tests; rather, the successful completion of the test according to a standard often serves as the acceptance.

    No.TestAcceptance Criteria (Implicit)Reported Device Performance
    1Visual Inspection (Appearance)Device appears as intended, free from defects.Test performed, results met pre-set criteria.
    2Size MeasurementDimensions conform to specifications.Test performed, results met pre-set criteria.
    3Leakage Test (Catheter)No leakage.Test performed, results met pre-set criteria.
    4Tensile strength (Catheter)Meets specified tensile strength (e.g., ISO 10555-1 Annex B for catheter, BS 6196:1989 Appendix D for catheter-hub/Stylet/Needle Stylet).Test performed, results met pre-set criteria. (Also see side-by-side testing results below).
    5Assemble compatibility (Catheter)Components fit together correctly.Test performed, results met pre-set criteria.
    6Radiopacity testDevice is visible under fluoroscopy.Test performed, results met pre-set criteria.
    7Corrosion resistance testNo significant corrosion.Test performed, results met pre-set criteria.
    8Flexural Rigidity (Catheter introducer and needle)Meets specified flexibility requirements.Test performed, results met pre-set criteria.
    9Pulling (Drawing) Test (Catheter introducer and needle)Meets specified pulling force requirements.Test performed, results met pre-set criteria.
    10Flexural Rigidity (Wire)Meets specified flexibility requirements.Test performed, results met pre-set criteria.
    11EO Gas residueResidue levels below accepted limits.Test performed, results met pre-set criteria.
    12Sterility TestSterility Assurance Level (SAL) of 10^-6 (in accordance with ISO 11135 and ISO 11737-1, 2).Test performed, results met pre-set criteria.
    13Endotoxin TestEndotoxin levels below specified limits.Test performed, results met pre-set criteria.
    14Particulate Matter TestParticulate matter levels below specified limits.Test performed, results met pre-set criteria.
    Side-by-Side Comparative Tests with Predicate Device:
    1Tensile strength (Catheter, Catheter-Hub, Stylet, Needle Stylet)Performance comparable to predicate device, meeting relevant standards (ISO 10555-1 Annex B, BS 6196:1989 Appendix D for various components, BS 6196:1989 Appendix F for Needle Stylet)."The test results supported that the subject device is substantially equivalent to the predicate devices." Implies comparable or better performance.
    2Dullness testPerformance comparable to predicate device, meeting BS 6196:1989 Appendix G."The test results supported that the subject device is substantially equivalent to the predicate devices." Implies comparable or better performance.
    3Fluid pressure of the catheter/adapter connectionPerformance comparable to predicate device, meeting ISO 10555-1 Annex G."The test results supported that the subject device is substantially equivalent to the predicate devices." Implies comparable or better performance.
    4Flow ratePerformance comparable to predicate device, meeting ISO 10555-1 Annex E."The test results supported that the subject device is substantially equivalent to the predicate devices." Implies comparable or better performance.
    5Luminal integrityPerformance comparable to predicate device, meeting ISO 10555-1 Annex C."The test results supported that the subject device is substantially equivalent to the predicate devices." Implies comparable or better performance.
    Biocompatibility:Biocompatible as per ISO 10993-4, 5, 10, 11 and USP 39 <151> for External communicating device (tissue/bone/dentin communicating), Limited exposure (<24 hours)."Biocompatibility tests were performed... and the test results supported that the subject devices are biocompatible."
    Shelf Life:Validated 3-year shelf life using accelerated aging (ASTM F1980)."The test results validated 3 year shelf life."

    2. Sample size used for the test set and the data provenance:

    The document describes non-clinical bench testing. It does not refer to "test sets" in the context of clinical data, human subjects, or AI performance evaluation. Therefore, information on sample sizes for a clinical test set or data provenance (country of origin, retrospective/prospective) is not applicable or provided. The tests mentioned adhere to international standards (e.g., ISO, ASTM, BS) which typically specify the number of samples required for each test.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not applicable. The device is an Anesthesia Conduction Catheter, and its performance evaluation relies on engineering, material, and biocompatibility testing according to established standards, not on expert-adjudicated ground truth from clinical image interpretation or similar AI applications.

    4. Adjudication method for the test set:

    Not applicable, as there is no "test set" in the context of clinical data requiring expert adjudication.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. The EPINAUT device is an anesthesia conduction catheter, a physical medical device, not an AI software. There is no mention of AI assistance or MRMC studies.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    Not applicable. This is a physical medical device, not an algorithm.

    7. The type of ground truth used:

    For non-clinical performance tests, the "ground truth" is defined by the requirements of the relevant international standards (e.g., ISO, ASTM, BS) and the pre-set internal specifications of the manufacturer. For biocompatibility, the ground truth is established by the results of tests performed according to ISO 10993 and USP standards, demonstrating the material not to be toxic or irritating. For sterilization, the ground truth is a Sterility Assurance Level (SAL) of 10^-6 demonstrated through validated methods according to ISO 11135 and ISO 11737.

    8. The sample size for the training set:

    Not applicable. This is a physical medical device, not an AI system that requires a "training set."

    9. How the ground truth for the training set was established:

    Not applicable, as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1