    K Number
    K984069

    Validate with FDA (Live)

    Date Cleared
    1999-01-25

    (70 days)

    Product Code
    Regulation Number
    874.4100
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ENTaxis™ Nasal Packings are intended for nasal epistaxis and as post operative nasal packings. ENTaxis™ Nasal Packing is intended for - Epistaxis - · Post operative nasal packing

    Device Description

    ENTaxis™ Nasal Packings are highly conformable, sterile, hemostatic packings designed for managing bleeding in the nasal cavity and as post operative packings. ENTaxis™ nasal packings are 12" (30cm, 2g) "ropes" that are 100% composed of pure calcium alginate fibers.

    AI/ML Overview

    The provided text describes the 510(k) summary for the ENTaxis™ Nasal Packing device, which includes information regarding its clinical safety and efficacy study. Based on this information, here’s a summary conforming to your request:

    Acceptance Criteria and Device Performance

    The clinical study evaluated the ENTaxis™ Nasal Packing against conventional Merocel™ packing for post-operative use in sinus surgery, focusing on specific clinical endpoints. While specific numeric acceptance criteria (e.g., "bleeding must be reduced by X%") are not explicitly stated in a table format, the study's conclusions demonstrate superior performance of ENTaxis™ for the tested endpoints.

    Table 1: Acceptance Criteria (Implied) and Reported Device Performance

    Acceptance Criteria (Implied)Reported Device Performance (ENTaxis™ vs. Merocel™)
    Reduced epistaxis (degree of bleeding)Significantly less bleeding (p=0.016) upon removal on Day 2.
    Reduced pain on removalSignificantly less pain on removal (p=0.0001) upon removal on Day 2.
    Improved quality-of-healingTrend towards more complete healing with ENTaxis (10 patients) compared to Merocel (5 patients) by day 9, though not explicitly stated as statistically significant in the summary.

    Study Details

    1. Sample size used for the test set and the data provenance:

      • Test Set Sample Size: 50 patients (100 nasal cavities).
      • Data Provenance: The study was a multi-center, randomized, and controlled clinical study. The specific country of origin is not explicitly stated, but the submitter is Laboratoires Brothier S.A. from France, suggesting European origin for the clinical trial, or at least a multi-national one. It was a prospective clinical study comparing two different types of packings.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • The document refers to the study by R. Peynegre, P. Bonfils, L. Castillo, et al. from 1998, conducted in "five ENT surgery departments." This implies multiple clinicians were involved in the assessments. However, the exact number of experts establishing ground truth or their specific qualifications (e.g., "radiologist with 10 years of experience") are not detailed in the provided text. Assessments were made by clinical personnel within the ENT surgery departments.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

      • The text states, "Every patient served as their own control by being treated with both test products." This indicates a paired-comparison or within-subject design where individual patient outcomes for ENTaxis™ were compared against their own outcomes for Merocel™. Different assessments were made by clinicians during surgery, on Day 2 (epistaxis and pain on removal), and on Day 9 (quality-of-healing). There is no explicit mention of an adjudication panel or method like 2+1/3+1 consensus for resolving discrepancies in assessments.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No, this was not an MRMC comparative effectiveness study involving AI or human readers. This study evaluated two medical devices (nasal packings) directly through patient outcomes after surgical procedures, not the diagnostic performance of human readers with or without AI assistance.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • No. This study is a clinical trial evaluating the performance of a physical medical device (nasal packing) in human patients, not an algorithm, and thus standalone performance in an AI context is not applicable.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • The ground truth was based on clinical outcomes data directly observed and assessed by medical professionals during and after sinus surgery. This includes objective measures (degree of bleeding) and subjective patient experiences (pain on removal), as well as clinical inspection of healing by ENT specialists.
    7. The sample size for the training set:

      • There is no mention of a "training set" in the context of this device study, as this is a physical medical device and not an AI or machine learning algorithm.
    8. How the ground truth for the training set was established:

      • Not applicable, as there was no training set for an algorithm.