510(k) Data Aggregation

    K Number
    K140745

    Validate with FDA (Live)

    Date Cleared
    2014-04-17

    (23 days)

    Product Code
    Regulation Number
    870.1650
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Encore 26 Advantage Kit is a kit of sterile disposable devices intended of ruse as accessories for percutaneous coronary angiography (PTCA) procedures. They allow for balloon inflation and wire control.
    Encore™ 26 Inflation Device: indicated for use with balloon dilatation catheters to create and monitor pressure in the balloon, and to deflate the balloon.
    Gateway™ Plus Y-Adaptor: used for providing hemostasis around balloon dilatation catheters, guidewires and other therapeutic devices during general intravascular procedures.
    Torque Device: used for guidewire manipulation during general intravascular procedures.
    Insertion Tool: used to facilitate the introduction of a quidewire during general intravascular procedures.

    Device Description

    The Encore™ 26 Advantage Kit is a Kit of sterile disposable devices intended for use as accessories and percutaneous coronary angiography (PTCA) procedures. They allow for balloon inflation and wire control.

    AI/ML Overview

    This is a 510(k) premarket notification for a medical device (Encore™ 26 Advantage Kit), which focuses on demonstrating substantial equivalence to a predicate device rather than presenting a clinical study with acceptance criteria and performance metrics in the way one might for a novel AI algorithm or diagnostic tool.

    Therefore, many of the requested sections are not applicable to the provided document. I will fill in the information that is present or directly inferable from the document.

    1. A table of acceptance criteria and the reported device performance

    The submission states, "Design Verification and Design Validation Testing was performed to verify that the performance and usability of the modified Insertion Tool remains substantially equivalent to the predicate device via K140673. In addition Sterilization, Packaging and Biocompatibility testing verifies the overall substantial equivalence to the kit predicate."

    While specific numerical acceptance criteria and performance values for each test are not listed in this summary document, the overall acceptance criterion is substantial equivalence to the predicate device(s). The reported performance is that the device met this criterion.

    Acceptance Criteria (Implied)Reported Device Performance
    Substantial equivalence in performance to predicate device (K123214 and K922410) for the modified Insertion ToolDesign Verification and Design Validation Testing showed the modified Insertion Tool remains substantially equivalent to the predicate device.
    Substantial equivalence in sterilization, packaging, and biocompatibility to the kit predicateSterilization, Packaging and Biocompatibility testing verifies overall substantial equivalence to the kit predicate.
    No new safety or performance issues raisedNo new safety or performance issues were raised during the device testing.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not provided in the summary. The document refers to "Design Verification and Design Validation Testing" but does not detail the specific sample sizes used for these engineering/performance tests. Data provenance (country of origin, retrospective/prospective) is not applicable or provided for this type of device submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This is not applicable to a device like a balloon inflation kit. Ground truth, in the context of diagnostic algorithms, involves expert labeling of data. For a hardware medical device, "ground truth" would relate to engineering specifications and performance benchmarks, not expert consensus on medical images or diagnoses.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. Adjudication methods are typically used in clinical studies involving human interpretation or subjective assessments, not for performance testing of a mechanical device.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI-assisted diagnostic device, and no MRMC study was conducted or mentioned.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an algorithm-based device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For this type of device, the "ground truth" for performance testing would be established by:

    • Engineering specifications and standards: Performance parameters (e.g., pressure capabilities, sealing integrity, material strength, dimensional accuracy) are measured against predefined engineering requirements and industry standards.
    • Predicate device performance: The performance of the predicate device serves as the benchmark for substantial equivalence.

    8. The sample size for the training set

    Not applicable. This is not a machine learning or AI device that uses a "training set."

    9. How the ground truth for the training set was established

    Not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K123214

    Validate with FDA (Live)

    Date Cleared
    2012-11-13

    (29 days)

    Product Code
    Regulation Number
    870.1650
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Encore™ 26 Advantage Kit is a kit of sterile disposable devices intended for use as accessories for percutaneous coronary angiography (PTCA) procedures. They allow for balloon inflation and wire control.

    Individual Device Intended Use:

    • Encore™ 26 Inflation Device: Inflation device are indicated for use with balloon . dilation catheters to create and monitor pressure in the balloon, and to deflate the balloon.
    • GateWay™ Plus Y-Adapter: Used for providing hemostasis around quidewires, . balloon dilatation catheters, and other therapeutic devices
    • Torque Device: Used for guidewire manipulation during general intravascular . procedures.
    • Guidewire Insertion Tool: Used for percutaneous introduction and placement of . guidewires in vascular procedures.
    Device Description

    The Encore™ Advantage Kit is a kit of sterile disposable devices intended for use as accessories for percutaneous coronary angiography (PTCA) procedures. They allow for balloon inflation and wire control.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the Encore™ 26 Advantage Kit. This submission details the device's intended use and claims substantial equivalence to a predicate device.

    However, the document does not contain any information about acceptance criteria or a study proving that the device meets such criteria. The "Summary of Non-Clinical Testing" mentions that design verification was performed, and "No additional testing was required for kit inclusion. Sterility testing was performed to verify the overall substantial equivalence to the predicates. No additional biocompatibility testing was required." The "Summary of Clinical Testing" explicitly states, "Clinical Evaluation was not required for this device."

    Therefore, I cannot provide the requested information in a table or respond to the specific questions because the provided text does not contain any data related to acceptance criteria, device performance, sample sizes, expert involvement, adjudication methods, MRMC studies, standalone performance, or ground truth establishment for either test or training sets. This device was cleared based on substantial equivalence to an existing predicate device, not through a performance study against specific acceptance criteria.

    Ask a Question

    Ask a specific question about this device

    K Number
    K120694

    Validate with FDA (Live)

    Date Cleared
    2012-04-03

    (27 days)

    Product Code
    Regulation Number
    870.1650
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Encore™ 26 Advantage Kit is a kit of sterile disposable devices intended for use as accessories for percutaneous coronary angiography (PTCA) procedures. They allow for balloon inflation and wire control.

    • Encore™ 26 Inflation Device: Inflation Devices are indicated for use with balloon dilatation catheters to create and monitor pressure in the balloon, and to deflate the balloon.
    • GateWay™ Plus Y-Adapter: Used for providing hemostasis around guidewires, balloon dilatation catheters, and other therapeutic devices.
    • TD2® Torque Device: Used for guidewire manipulation.
    • Insertion Tool: Used for percutaneous introduction and placement of guidewires in vascular procedures.
    Device Description

    The Encore™ 26 Advantage Kit is a kit of sterile disposable devices intended for use as accessories for percutaneous coronary angiography (PTCA) procedures. They allow for balloon inflation and wire control.

    The proposed Encore™ 26 Advantage Kit will include the following currently marketed devices: Boston Scientific (BSC) Encore Inflation Device (K955869); BSC GateWay™ PLUS Y-Adaptor (K951089); Navilyst Medical TD2 Torque Device (K922706); and NeedleTech Products, Inc. Guidewire Insertion Needle (K031173).

    AI/ML Overview

    This is a 510(k) premarket notification for the Encore™ 26 Advantage Kit. The submission focuses on demonstrating substantial equivalence to a predicate device through non-clinical testing. It does not involve a study of an AI/ML powered device. Therefore, many of the requested details about acceptance criteria, device performance, sample sizes for test and training sets, expert involvement, and comparative effectiveness studies are not applicable.

    Here's an breakdown of the relevant information provided:

    1. Table of Acceptance Criteria and Reported Device Performance

    Since this is a device kit and not an AI/ML powered device with specific performance metrics like sensitivity or specificity, the "acceptance criteria" pertain to mechanical, biocompatibility, and sterility testing. The "reported device performance" indicates that these tests were performed and met the necessary standards for substantial equivalence.

    Acceptance Criteria CategorySpecific TestReported Device Performance
    Mechanical Bench TestingI.D. (Guidewire Compatibility)Performed, verifying performance and usability of the guidewire insertion tool remains substantially equivalent to the predicate. (Specific numerical results are not provided in this summary but implied to meet requirements).
    Hub Tensile StrengthPerformed, verifying performance and usability of the guidewire insertion tool remains substantially equivalent to the predicate. (Specific numerical results are not provided).
    Biocompatibility TestingCytotoxicityPerformed, verifying overall safety and efficacy. (Implied to meet ISO or FDA biocompatibility standards).
    SensitizationPerformed, verifying overall safety and efficacy.
    Irritation Or Intracutaneous ReactPerformed, verifying overall safety and efficacy.
    Systemic Toxicity (Acute)Performed, verifying overall safety and efficacy.
    HemocompatibilityPerformed, verifying overall safety and efficacy.
    LatexPerformed, verifying overall safety and efficacy.
    USP PhysicochemicalPerformed, verifying overall safety and efficacy.
    Sterility TestingPerformed, verifying overall safety and efficacy. (Implied to meet relevant sterilization standards).
    Packaging TestingPerformed, verifying overall safety and efficacy. (Implied to meet relevant packaging integrity standards).

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: Not explicitly stated as this is a non-clinical, bench-testing focused submission. For mechanical and biocompatibility testing, sample sizes would typically be determined by statistical methods for device testing, but this information is not provided in the summary.
    • Data Provenance: The data is from non-clinical bench testing performed by the manufacturer, Boston Scientific Corporation. This is not medical imaging data or patient data, so concepts like "country of origin" or "retrospective/prospective" in the typical sense do not apply.

    3. Number of Experts Used to Establish Ground Truth and Qualifications of Experts

    • Not Applicable. This submission does not involve expert review/ground truth establishment for clinical data as there was no clinical study. The "ground truth" for mechanical and biocompatibility testing would be defined by established engineering and biological standards, not human expert consensus on diagnostic interpretations.

    4. Adjudication Method for the Test Set

    • Not Applicable. Since there's no diagnostic interpretation by experts, there's no adjudication method in the context of clinical or diagnostic performance.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    • No. This was not an AI/ML powered device, and no clinical studies, including MRMC studies, were performed.
    • Effect size of human readers improve with AI vs without AI assistance: Not applicable.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

    • Not Applicable. This is a medical device kit, not an algorithm.

    7. The Type of Ground Truth Used

    • The "ground truth" is based on established engineering standards, biocompatibility standards (e.g., ISO standards), and sterility validation criteria for medical devices. The testing verifies that the device meets these pre-defined physical, chemical, and biological requirements.

    8. The Sample Size for the Training Set

    • Not Applicable. There is no "training set" in the context of AI/ML for this device. The development process would involve iterative design and testing, but not machine learning model training.

    9. How the Ground Truth for the Training Set Was Established

    • Not Applicable. As there is no training set.