Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K170293

    Validate with FDA (Live)

    Date Cleared
    2017-10-25

    (267 days)

    Product Code
    Regulation Number
    862.3250
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Emit® II Plus Cocaine Metabolite Assay is a homogeneous enzyme immunoassay with a 150 ng/mL or 300 ng/mL cutoff. The assay is intended for use in the qualitative and semiquantitative analyses of benzoylecgonine (cocaine metabolite) in human urine. Emit® II Plus assays are designed for use with a number of chemistry analyzers.

    The Emit® II Plus Cocaine Metabolite Assay provides only a preliminary analytical test result. A more specific alternative chemical method must be used to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. Other chemical confirmation methods are available. Clinical consideration and professional judgment should be applied to any drug-of-abuse test result, particularly when preliminary positive results are used.

    Device Description

    The Emit II Plus Cocaine Metabolite assay is a homogeneous enzyme immunoassay that qualitatively and semiquantitatively measures benzoylecgonine. The assay has cutoffs of 150 ng/mL and 300 ng/mL benzoylecgonine.

    The assay is based on competition between drug in the specimen and drug labeled with the enzyme glucose-6-phosphate dehydrogenase (G6PDH) for antibody binding sites. Enzyme activity decreases upon binding to the antibody, so the drug concentration in the specimen can be measured in terms of enzyme activity. Active enzyme converts nicotinamide adenine dinucleotide (NAD) to NADH, resulting in an absorbance change that is measured spectrophotometrically at 340 nm.

    The Emit II Plus Cocaine Metabolite Assay consists of two ready-to-use reagents:

    Antibody/Substrate Reagent 1
    Sheep polyclonal antibodies to benzoylecgonine (2.2 µg/mL), bovine serum albumin, G6P (15 mM), NAD (12 mM), preservatives, and stabilizers

    Enzyme Reagent 2
    Benzoylecgonine labeled with bacterial G6PDH (0.46 U/mL), HEPES buffer, bovine serum albumin, preservatives, and stabilizers

    AI/ML Overview

    Here's a breakdown of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state "acceptance criteria" for the method comparison studies. However, the reported "Result" for agreement percentages can be interpreted as the performance achieved. Similarly, for precision studies, "Results" indicate the number of negative/positive determinations. For linearity, it's % recovery.

    Table 1: Acceptance Criteria and Reported Device Performance

    Performance MetricAcceptance Criteria (Implied)Reported Device PerformanceComments
    Method Comparison (Qualitative & Semi-Quantitative)
    150 ng/mL CutoffHigh agreement with GC/MS91% agreementThe document does not specify a numerical threshold for "high agreement."
    300 ng/mL CutoffHigh agreement with GC/MS88% agreementThe document does not specify a numerical threshold for "high agreement."
    Precision (Qualitative & Semi-Quantitative)
    150 ng/mL Cutoff
    0 ng/mL100% Negative80 NegativeAchieved
    38 ng/mL (-75%)100% Negative80 NegativeAchieved
    75 ng/mL (-50%)100% Negative80 NegativeAchieved
    113 ng/mL (-25%)100% Negative80 NegativeAchieved
    150 ng/mL (Cutoff)Consistent classification (~50% Positive/Negative)9 Negative / 71 PositiveAt the cutoff, some variability in classification is expected. The exact split for "acceptance" isn't defined but this is a common observation.
    188 ng/mL (+25%)100% Positive80 PositiveAchieved
    225 ng/mL (+50%)100% Positive80 PositiveAchieved
    263 ng/mL (+75%)100% Positive80 PositiveAchieved
    300 ng/mL (+100%)100% Positive80 PositiveAchieved
    300 ng/mL Cutoff
    0 ng/mL100% Negative80 NegativeAchieved
    75 ng/mL (-75%)100% Negative80 NegativeAchieved
    150 ng/mL (-50%)100% Negative80 NegativeAchieved
    225 ng/mL (-25%)100% Negative80 NegativeAchieved
    300 ng/mL (Cutoff)Consistent classification (~50% Positive/Negative)54 Negative / 26 PositiveAt the cutoff, some variability in classification is expected.
    375 ng/mL (+25%)100% Positive80 PositiveAchieved
    450 ng/mL (+50%)100% Positive80 Negative (This appears to be an error in the document, should likely be positive)Note: The document states 80 Negative for +50% of 300 ng/mL cutoff (450 ng/mL) which is contradictory to expectation for a positive result above cutoff. This might be a typo in the provided text. Based on other results, it should be 80 Positive.
    525 ng/mL (+75%)100% Positive80 PositiveAchieved
    600 ng/mL (+100%)100% Positive80 PositiveAchieved
    Recovery/Linearity (Semiquantitative)
    % RecoveryNot explicitly stated (e.g., ±X%)Ranges from -0.7% to 8.6%The document does not specify an acceptance range for % Recovery (e.g., 90-110%).
    Specificity - Structurally Related Compounds
    Cross-ReactivityLow cross-reactivity for non-benzoylecgonine compoundsEcgonine: 2-3%, Cocaine: 0.5%, others <0.01%The document lists values but does not define an explicit acceptance criteria (e.g., <5% for non-target compounds).

    2. Sample Size Used for the Test Set and Data Provenance

    • Method Comparison:
      • 150 ng/mL Cutoff: n = 102 native patient urine samples.
      • 300 ng/mL Cutoff: n = 95 native patient urine samples.
      • Data Provenance: Retrospective, as indicated by "native patient urine samples." The country of origin is not specified.
    • Precision (Repeatability and Within-Lab Precision):
      • Sample Size: 9 sample levels for each cutoff (150 ng/mL and 300 ng/mL). Each level was determined 80 times (20 days, 2 runs/day, 2 replicates/run). This implies 9 * 80 = 720 individual determinations per cutoff, across the various concentrations.
      • Data Provenance: The samples were "prepared by spiking," implying laboratory-controlled samples, not native patient samples. Data provenance (country of origin) is not specified.
    • Recovery/Linearity:
      • Sample Size: Two aliquots of negative urine pool spiked to create samples. Five replicates were tested at each of the 10 concentration levels. This implies 5 * 10 = 50 individual determinations.
      • Data Provenance: Laboratory-prepared samples, not native patient samples. Data provenance (country of origin) is not specified.
    • Specificity - Structurally Related Compounds:
      • Sample Size: Not explicitly stated as a "sample size" for a study, but rather concentrations at which various compounds were tested for cross-reactivity.
      • Data Provenance: Laboratory-controlled testing, not native patient samples. Data provenance (country of origin) is not specified.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    The ground truth for the test set (Method Comparison) was established by Gas Chromatography/Mass Spectrometry (GC/MS). GC/MS is an analytical chemistry technique, not a human expert. Therefore, the concept of "number of experts" and their "qualifications" is not applicable in this context.

    4. Adjudication Method for the Test Set

    Not applicable. The ground truth was established by an objective analytical method (GC/MS), not through human interpretation or adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, an MRMC comparative effectiveness study was not done. This device is an automated in vitro diagnostic assay (an enzyme immunoassay) designed to detect a cocaine metabolite in urine. It does not involve human readers for interpretation, nor does it incorporate AI in a way that would assist human interpretation.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Yes, the performance studies described (Method Comparison, Precision, Recovery/Linearity, Specificity) represent standalone performance of the device (assay on the analyzer). The results are generated directly by the Emit® II Plus Cocaine Metabolite Assay on the Beckman Coulter DxC 700 AU Clinical Chemistry Analyzer, without human interpretive input. The output is a qualitative (positive/negative) or semi-quantitative result.

    7. The Type of Ground Truth Used

    The primary ground truth used for the method comparison study was Gas Chromatography/Mass Spectrometry (GC/MS). For precision, linearity, and specificity studies, the ground truth was based on known concentrations of the analyte in prepared samples.

    8. The Sample Size for the Training Set

    The document does not describe a "training set" in the context of machine learning or AI. This is a traditional in vitro diagnostic device, not an AI/ML-based device.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no "training set" described for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K031512

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2003-07-23

    (70 days)

    Product Code
    Regulation Number
    862.3250
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Emit® II Plus Cocaine Metabolite Assay is a homogeneous enzyme immunoassay with a 150 ng/mL or 300 ng/mL cutoff (SAMSHA initial test cutoff level). The assay is intended for use in the qualitative and semi-quantitative analyses of benzoylecgonine (cocaine metabolite) in human urine. Emit® II Plus assays are designed for use with a number of chemistry analyzers.

    The Emit® II Plus Monoclonal Cocaine Metabolite Assay provides only a preliminary analytical test result. A more specific alternative chemical method must be used to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. Other chemical confirmation methods are available. Clinical consideration and professional judgment should be applied to any drug-ofabuse test result, particularly when preliminary positive results are used.

    Device Description

    Emit® II Plus Cocaine Metabolite Assay is a homogeneous enzyme immunoassay for qualitative and semi-quantitative analysis of cocaine metabolite (benzoylecgonine) in human urine.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Syva Company - Dade Behring Inc. Emit® II Plus Cocaine Metabolite Assay, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Implied)Reported Device Performance (150 ng/mL Cutoff)Reported Device Performance (300 ng/mL Cutoff)
    High agreement with GC/MS (reference method) for qualitative results93% agreement with GC/MS94% agreement with GC/MS

    Note: The document does not explicitly state numerical acceptance criteria, but demonstrates a high percentage agreement with the Gas Chromatography/Mass Spectrometry (GC/MS) method, implying that high concordance is the de facto acceptance criterion.

    2. Sample Size Used for the Test Set and Data Provenance

    • 150 ng/mL Cutoff: The number of samples is not explicitly stated, but the provided table shows:
      • 2 samples testing positive by GC/MS (123 and 140 ng/mL) but negative by the device.
      • 7 samples testing positive by GC/MS (150, 151, 157, 162, 164, 166, and 190 ng/mL) and positive by the device.
      • The total number of samples where an agreement or disagreement is reported (excluding the header rows) is 9. However, the full sample size for the agreement percentage calculation is not provided.
    • 300 ng/mL Cutoff: The number of samples is not explicitly stated, but the provided table shows:
      • 6 samples testing positive by GC/MS (218, 241, 261, 294, 295, and 298 ng/mL) but negative by the device.
      • 1 sample testing positive by GC/MS (354 ng/mL) and positive by the device.
      • Similar to the 150 ng/mL cutoff, the full sample size for the agreement percentage calculation is not provided, only the cases of disagreement and agreement mentioned.
    • Data Provenance: Not specified (e.g., country of origin, retrospective or prospective).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

    • Not applicable for this type of device. The ground truth is established by a reference chemical/analytical method (GC/MS), not by human experts.

    4. Adjudication Method for the Test Set

    • Not applicable as the ground truth is established by a single, definitive chemical method (GC/MS), not by human interpretation or consensus that would require adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This is an in-vitro diagnostic (IVD) device for chemical analysis, not an AI-powered diagnostic imaging or interpretation device that would involve human readers.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • Yes, this is a standalone device. The device's performance is measured directly against the GC/MS reference method. It performs the analysis of benzoylecgonine in human urine without human interpretation affecting the direct assay output. The text states: "The Emit® II Plus Monoclonal Cocaine Metabolite Assay provides only a preliminary analytical test result. A more specific alternative chemical method must be used to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method." This indicates the Emit® II Plus provides its own result which is then compared, and if positive, needs GC/MS for confirmation.

    7. The Type of Ground Truth Used

    • Reference standard/Gold Standard: Gas Chromatography/Mass Spectrometry (GC/MS) is specified as the "preferred confirmatory method" and the method against which the Emit® II Plus assay results are compared.

    8. The Sample Size for the Training Set

    • The document does not provide any information about a training set size. This assay is a homogeneous enzyme immunoassay, not a machine learning model that typically requires a distinct training set. The "method comparison" describes validation, not necessarily model training.

    9. How the Ground Truth for the Training Set was Established

    • Not applicable, as no training set for a machine learning model is mentioned or implied.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1