510(k) Data Aggregation

    K Number
    K171339

    Validate with FDA (Live)

    Date Cleared
    2017-07-07

    (60 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A powder-free patient examination glove is a disposable device intended for use medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and use for use with chemotherapy drugs.

    In addition these gloves were tested for use with chemotherapy drugs in accordance ASTM D6978-05 standards- Practice for assessment of Medical Glove to Permeation by chemotherapy drugs

    Device Description

    The subject device in this 510(k) Notification is EMG Blue Nitrile Examination Gloves Powder Free with tested for use with Chemotherapy drugs. The subject device is a patient examination glove made from nitrile latex compound, Blue color, powder free and non-sterile (Per 21 CFR 880.6250, class I). The device meets all the specifications in ASTM D6319-10, Standard specification for Nitrile Examination Gloves. Additionally, the gloves have been tested for biocompatibility and permeability to chemotherapy drugs.

    The Blue Nitrile Medical Examination Gloves ,Powder Free, is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner (product Code LZA) and is used with chemotherapy drugs (Product code LZC).

    AI/ML Overview

    The document describes the acceptance criteria and performance of the "EMG Blue Nitrile Examination Gloves Powder Free with tested for use with Chemotherapy Drugs." This is a medical device, not an AI/ML device, so many of the requested fields are not applicable.

    Here's the information extracted and formatted as requested, with "N/A" for fields that are not relevant to this type of device:

    Acceptance Criteria and Device Performance for EMG Blue Nitrile Examination Gloves

    1. Table of Acceptance Criteria and Reported Device Performance:

    CharacteristicsAcceptance CriteriaReported Device Performance (K171339)Comparison to Predicate (K161187)
    Product CodeLZA and LZCLZA and LZCSame
    Intended UseA powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. The device is for over-the-counter use. Intended use with chemotherapy drugs in accordance with ASTM D6978-05.A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. The device is for over-the-counter use. Intended use with chemotherapy drugs in accordance with ASTM D6978-05.Same
    Material UseNitrile latex compoundNitrile latex compoundSame
    ColorBlueBlueSame
    SterilityNon sterileNon sterileSame
    Single UseSingle usedSingle usedSame
    DimensionsOverall Length (mm): Min 230 mm Width (± 5mm): Size S = 85mm, Size M = 95mm, Size L = 105mm, Size XL = 115mm Thickness at Palm (mm): Min 0.05 mm Thickness at Finger Tip (mm): Min 0.05 mmSize: Medium Palm Width: 98 mm Length: 247 mm Finger Thickness: 0.08mm Palm Thickness: 0.05mm (Meets ASTM D6319-10)Size: Medium Palm Width: 99 mm Length: 247 mm Finger Thickness: 0.11mm Palm Thickness: 0.07mm (Meets ASTM D6319-10) Overall dimensions are considered "Same" for substantial equivalence based on meeting the ASTM standard.
    Physical PropertiesBefore Ageing: Tensile Strength (MPa) = Min 14 Mpa, Ultimate Elongation (%) = Min 500 % After Aging at 70°C for 168 hrs @ 100°C for 22 hrs: Tensile Strength (MPa) = Min 14 Mpa, Ultimate Elongation (%) = Min 400 %Tensile strength (Mpa): Before aging: 25.45Mpa, After Accelerated aging: 26.33Mpa Ultimate Elongation(%): Before aging: 600%, After aging: 550% (Meets ASTM D6319-10)Tensile strength (Mpa): Before aging: 25.45Mpa, After Accelerated aging: 26.33Mpa Ultimate Elongation(%): Before aging: 600%, After aging: 550% (Meets ASTM D6319-10) This indicates the predicate also met these criteria, confirming "Same".
    Freedom from PinholesAQL 2.5 (Inspection Level G-1)AQL 2.5 (Inspection Level G-1)AQL 2.5 (Meets ASTM D5151-06) The device meets the same AQL and inspection level.
    Residual Powder< 2.0 mg/gloveAverage 0.17 mg/glove (Meets ASTM D6124-06)Average 0.19 mg/glove (Meets ASTM D6124-06) Both devices meet the acceptance criteria and demonstrate similar performance.
    Biological Evaluation (Primary Skin Irritation)ISO 10993-10:2010 compliance for non-irritantUnder the conditions of this study, the test article was a non-irritant.Under the conditions of this study, the test article was a non-irritant. Both devices show non-irritant properties.
    Biological Evaluation (Dermal Sensitization Assay)ISO 10993-10:2010 compliance for non-sensitizerUnder the conditions of this study, the test article was a non-sensitizer.Under the conditions of this study, the test article was a non-sensitizer. Both devices show non-sensitizer properties.
    Chemotherapy Drug Permeation (ASTM D6978-05) - Selected ExamplesCarmustine (BCNU)(3.3 mg/ml): Minimum Breakthrough detection time: > 24.0 MinutesCyclophosphamide (20 mg/ml): Minimum Breakthrough detection time: > 240 MinutesThiotepa (10mg/ml): Minimum Breakthrough detection time: > 56.9 Minutes(Note: Max testing time is 240 minutes)Carmustine (BCNU)(3.3 mg/ml): 24 minutesCyclophosphamide (20 mg/ml): > 240 minutesThiotepa (10mg/ml): 56.9 minutesList includes 15 drugs, with most showing >240 minutes breakthrough time. Specific values for Carmustine and Thiotepa are noted as having low permeation time.The Indications for Use section (pages 2-3 and 10-11) lists the same 15 chemotherapy drugs and concentrations with identical "Minimum Breakthrough Detection Time" values for both the subject device (K171339) and the predicate (K161187) (as detailed on page 2 and stated in the comparison for substantial equivalence).

    2. Sample size used for the test set and the data provenance:

    • Sample Size for Physical/Chemical Tests: Not explicitly stated as a single number. The testing is based on ASTM standards (D6319-10, D5151-06, D6124-06) which prescribe sampling plans for quality control (e.g., AQL for pinholes).
    • Sample Size for Chemotherapy Permeation: Not explicitly stated. ASTM D6978-05, the standard referenced, outlines the methodology for testing permeation, which would involve multiple samples.
    • Data Provenance: The testing was conducted to meet established ASTM and ISO standards. The manufacturer is ECO MEDI GLOVE SDN. BHD. based in Malaysia. The tests appear to be prospective for the purpose of demonstrating substantial equivalence.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • N/A: This is a physical device, and its performance is measured against objective standards (ASTM, ISO) and laboratory tests, not subjective expert judgment for ground truth establishment.

    4. Adjudication method for the test set:

    • N/A: Not applicable for objective physical and chemical testing.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • N/A: This is not an AI/ML device, and no MRMC study was performed.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • N/A: This is not an AI/ML device. The "standalone" performance refers to the glove's physical and chemical properties.

    7. The type of ground truth used:

    • Objective Standards and Laboratory Measurements: The "ground truth" for this device is based on its ability to meet the specifications outlined in recognized international standards (ASTM D6319-10 for nitrile gloves, ASTM D6978-05 for chemotherapy drug permeation, ASTM D5151-06 for pinholes, ASTM D6124-06 for residual powder, and ISO 10993-10 for biocompatibility). These are quantifiable, objective measurements.

    8. The sample size for the training set:

    • N/A: There is no "training set" as this is not an AI/ML device.

    9. How the ground truth for the training set was established:

    • N/A: There is no "training set" for this type of device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K161187

    Validate with FDA (Live)

    Date Cleared
    2016-09-09

    (135 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A powder-free patient eaxmination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and for use with chemotherapy drugs.In addition these gloves were tested for use with Chemotherapy drugs in accordance with ASTM D6978-05 standards Practice for assessment of Medical Glove to Permeation by chemotherapy drugs.

    Device Description

    The subject device in this 510(k) Notification is EMG Blue Nitrile Examination Gloves Powder Free with tested for use with Chemotherapy drugs. The subject device is a patient examination glove made from nitrile latex compound, Blue colour, powder free and non-sterile (Per 21 CFR 880.6250, class I). The device meets all the specifications in ASTM D6319-10, Standard specification for Nitrile Examination Gloves. Additionally, the gloves have been tested for biocompatibility and permeability to chemotherapy drugs.

    The Blue Nitrile Medical Examination Gloves ,Powder Free, is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner (product Code LZA) and is used with chemotherapy drugs (Product code LZC). The subject device is substantially equivalent to the legally marketed Nitrile Medical Examination Gloves (product Code LZA and LZC).

    AI/ML Overview

    The provided text describes the acceptance criteria and performance of EMG Blue Nitrile Examination Gloves Powder Free with Tested for use with Chemotherapy Drugs (K161187). This device is a patient examination glove and is assessed against the ASTM D6319-10 standard for Nitrile Examination Gloves and ASTM D6978-05 for permeation by chemotherapy drugs.

    Here's the breakdown of the information requested:


    1. Table of Acceptance Criteria and Reported Device Performance

    The core test for this device, beyond general glove properties, is its resistance to chemotherapy drugs. The acceptance criteria are implicit in the performance shown for the predicate device (K141982) and the subject device's indication to be "tested for use with Chemotherapy Drugs in accordance with ASTM D6978-05 standards." The performance is reported as the "Minimum Breakthrough Detection Time (Min) in µg/cm²/minute." A longer breakthrough time indicates better resistance.

    Chemotherapy Drugs and ConcentrationAcceptance Criteria (Implicit from ASTM D6978-05 compliance for gloves intended for chemotherapy drug use)Reported Device Performance (EMG Blue Nitrile Examination Gloves K161187)
    1) Carmustine (BCNU) (3.3mg/ml)Generally, ideally >240 minutes for many drugs, but lower for aggressive drugs is noted. For predicate, 15 µg/cm²/minute was acceptable.1.3 minutes
    2) Cyclophosphamide (20mg/ml)> 240 minutes> 240 minutes
    3) Cytarabine (100mg/ml)> 240 minutes> 240 minutes
    4) Doxorubicin Hydrochloride (2mg/ml)> 240 minutes> 240 minutes
    5) Etopside (20mg/ml)> 240 minutes> 240 minutes
    6) Fluorouracil (50mg/ml)> 240 minutes> 240 minutes
    7) Methotrexate (25mg/ml)> 240 minutes> 240 minutes
    8) Paclitaxel (6mg/ml)> 240 minutes> 240 minutes
    9) Thiotepa (10mg/ml)Generally, ideally >240 minutes, but lower for aggressive drugs is noted. For predicate, 2 µg/cm²/minute was acceptable.67.8 minutes

    Note: The document explicitly states for the subject device: "The maximum testing time is 240 minutes. Please note that the following drugs have extremely low permeation time: Carmustine (BCNU) (3.3mg/ml) - Minimum Breakthrough detection time 1.3 μg/cm²/minute. Thiotepa (10mg/ml) - Minimum Breakthrough detection time 67.8 μg/cm²/minute." This highlights that even with lower breakthrough times for these specific drugs, if tested according to the standard, the device is considered to meet the "tested for use with chemotherapy drugs" indication. The predicate device also showed similar, and in one case even lower (Thiotepa, 2 minutes), breakthrough times for these aggressive drugs.

    Beyond chemotherapy drug resistance, the document also details other specifications that the gloves meet, primarily referencing ASTM D6319-10 (Standard specification for Nitrile Examination Gloves):

    CharacteristicAcceptance Criteria (from ASTM D6319-10)Reported Device Performance (EMG Blue Nitrile Examination Gloves K161187)
    Overall LengthMin 230 mmMeets ASTM D6319-10 (Specific values for different sizes are S, M, L, XL all 230mm minimum)
    WidthSize S = 85mm, M = 95mm, L = 105mm, XL = 115mm (± 5mm)Meets ASTM D6319-10 (Specific values for different sizes are S=85, M=95, L=105, XL=115mm)
    Thickness at PalmMin 0.05 mmMeets ASTM D6319-10 (Specific value is 0.05mm minimum)
    Thickness at Finger TipMin 0.05 mmMeets ASTM D6319-10 (Specific value is 0.05mm minimum)
    Tensile Strength (Before Ageing)14 MPa minMeets ASTM D6319-10 (Reported as "14min")
    Ultimate Elongation (Before Ageing)500% minMeets ASTM D6319-10 (Reported as "500min")
    Tensile Strength (After Ageing)14 MPa minMeets ASTM D6319-10 (Reported as "14min" after 70°C for 168 hrs @ 100°C for 22 hrs)
    Ultimate Elongation (After Ageing)400% minMeets ASTM D6319-10 (Reported as "400min" after 70°C for 168 hrs @ 100°C for 22 hrs)
    Freedom from pinholesAQL 2.5 (Inspection Level G-1)Meets ASTM D5151-06 (This ASTM standard is referenced, not D6319-10 for pinholes, but given the "Meets" statement, it implies compliance with the stated AQL and inspection level, which are standard for examination gloves. The document states "AQL 2.5, Inspection Level G-1")
    Residual Powder≤ 2.0 mg/pcMeets ASTM D6124-06 (This ASTM standard is referenced, not D6319-10 for residual powder. Given the "Meets" statement, it implies compliance with the stated limit.)
    Primary Skin Irritation TestNon-irritant statusUnder the conditions of this study, the test article was a non-irritant.
    Dermal Sensitization AssayNon-sensitizer statusUnder the conditions of this study, the test article was a non-sensitizer.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the sample sizes for the various tests (e.g., how many gloves were tested for chemotherapy permeation, physical properties, or biocompatibility). However, it consistently refers to compliance with established ASTM standards (D6319-10, D6978-05, D5151-06, D6124-06). These standards inherently define the sampling plans and testing methodologies.

    • Data Provenance: The studies were performed by the manufacturer, Eco Medi Glove Sdn. Bhd., located in Malaysia. The type of study would be considered prospective testing in a laboratory setting to evaluate the device's performance against defined standards.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    This is a performance test for a physical medical device (gloves) against established engineering and material science standards (ASTM). It does not involve "ground truth" derived from expert medical opinion in the way an AI diagnostic algorithm would.

    Instead, the "ground truth" (or more accurately, the pass/fail criteria) is established by the ASTM standards themselves, which are developed by experts in materials science, engineering, and medical device manufacturing. The testing would be performed by qualified laboratory technicians and engineers who are proficient in executing these standardized tests. The report does not specify the number or qualifications of these individuals directly, as their adherence to the standard methodologies is sufficient.


    4. Adjudication Method for the Test Set

    Not applicable. As described above, this is a direct performance test against quantitative standards, not an assessment requiring expert adjudication of a qualitative outcome. The outcomes are objective measurements (e.g., breakthrough time, tensile strength, elongation, presence of pinholes) or standardized biological evaluations (irritation, sensitization).


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No, an MRMC study was not done. This type of study is relevant for diagnostic devices that involve human interpretation (e.g., radiologists reading images) and the impact of AI assistance on their performance. This product is a physical medical device (examination gloves).


    6. If a Standalone (i.e. algorithm only, without human-in-the-loop performance) was done

    Yes, in a sense, a standalone performance assessment was done. The glove's performance (e.g., resistance to chemotherapy drugs, physical properties, biocompatibility) was assessed independently of human interaction during its use. The "algorithm" here is the physical glove itself and its material properties, and its performance was measured directly.


    7. The Type of Ground Truth Used

    The "ground truth" for the performance of these gloves is based on established standard specifications and test methods.

    • For physical properties and general requirements for nitrile examination gloves, the ground truth is defined by ASTM D6319-10.
    • For resistance to chemotherapy drug permeation, the ground truth and test methodology are defined by ASTM D6978-05.
    • For pinholes, ASTM D5151-06.
    • For residual powder, ASTM D6124-06.
    • For biocompatibility, the tests (Primary Skin Irritation, Dermal Sensitization) follow recognized biological evaluation standards (often ISO 10993 series, though not explicitly stated for these specific tests, but generally implied for such evaluations).

    These are all objective, quantitatively measurable criteria as defined by consensus standards, not subjective expert consensus, pathology, or outcomes data in the medical sense.


    8. The Sample Size for the Training Set

    Not applicable. This device is not an AI algorithm or a diagnostic tool that requires a "training set" in the context of machine learning. The "training" in manufacturing comes from process control, material selection, and quality systems, not data-driven model training.


    9. How the Ground Truth for the Training Set Was Established

    Not applicable, for the same reasons as in point 8.