Found 1 results

510(k) Data Aggregation

    K Number
    K113220

    Validate with FDA (Live)

    Date Cleared
    2012-03-22

    (142 days)

    Product Code
    Regulation Number
    870.5100
    Age Range
    All
    Reference & Predicate Devices
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Emerge™ Over-The-Wire and Emerge Monorail PTCA Dilatation Catheters are indicated for the balloon catheter dilatation of the stenotic portion of a coronary artery or bypass graft stenosis for the purpose of improving myocardial perfusion. Emerge Over-The-Wire and Emerge Monorail PTCA Dilation Catheters are also indicated for the post-delivery expansion of balloon expandable stents (bare metal and drug-eluting).

    Device Description

    The Boston Scientific Emerge™ PTCA Dilatation Catheter is a sterile, single-use, intravascular medical device. The catheter consists of a shaft with a balloon near the distal tip. The balloon is designed to provide an inflatable segment of known diameter and length at recommended pressures. The Emerge™ PTCA Dilatation Catheter is offered in both Monorail (MR) and Over-the-Wire (OTW) platforms. There are radiopaque marker bands located under the balloon to aid in positioning the system during the procedure. Coatings are applied to the balloon and catheter to enhance insertion and withdrawal performance. The Emerge™ PTCA Dilatation Catheter will be available with balloon diameters 2.00 mm to 4.00 mm and balloon lengths 8 mm to 30 mm.

    AI/ML Overview

    The provided document is a 510(k) summary for the Boston Scientific Emerge™ Monorail (MR) and Over-The-Wire (OTW) PTCA Dilatation Catheter. It outlines the device description, intended use, comparison to predicate devices, and performance data to demonstrate substantial equivalence.

    This document does not contain information typically found in a study proving a device meets acceptance criteria related to AI/ML software performance. Instead, it focuses on the physical and biological performance of a medical device (a PTCA dilatation catheter), demonstrating its safety and effectiveness through bench testing and biocompatibility. Therefore, most of the questions you've asked, which are relevant to AI/ML software evaluation, cannot be answered from this document.

    However, I can extract the information that is present and indicate where requested information is not applicable or not provided.

    Device Acceptance Criteria and Performance (Physical Device)

    The acceptance criteria for this medical device (a PTCA Dilatation Catheter) are derived from the "Guidance for Industry and FDA Staff - Class II Special Controls for Certain Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheters, September 8, 2010." The reported device performance is based on successful completion of the listed tests, demonstrating conformance to these requirements.

    Acceptance Criteria (Implied by Test)Reported Device Performance
    Biocompatibility:
    CytotoxicityCompleted successfully
    SensitizationCompleted successfully
    Intracutaneous ReactivityCompleted successfully
    Acute Systemic ToxicityCompleted successfully
    Materials Mediated PyrogenicityCompleted successfully
    USP PhysicochemicalCompleted successfully
    Hemolysis (Direct Contact)Completed successfully
    Hemolysis (Extract Method)Completed successfully
    Complement ActivationCompleted successfully
    CoagulationCompleted successfully
    In Vitro HemocompatibilityCompleted successfully
    FTIR AnalysisCompleted successfully
    Residual NPGDA analysisCompleted successfully
    In-vitro Performance:
    Effective LengthCompleted successfully
    Shaft Inner and Outer DiameterCompleted successfully
    Balloon Crossing ProfileCompleted successfully
    Balloon Preparation, Deployment, and RetractionCompleted successfully
    Withdrawal into a Guide CatheterCompleted successfully
    Shaft and Bond Burst PressureCompleted successfully
    Balloon Rated Burst PressureCompleted successfully
    Balloon Fatigue (Repeat Inflations)Completed successfully
    Balloon ComplianceCompleted successfully
    Balloon Inflation/Deflation TimeCompleted successfully
    Catheter Bond Strength TensileCompleted successfully
    Tip Pull TestCompleted successfully
    Flexibility and KinkCompleted successfully
    Torque StrengthCompleted successfully
    RadiopacityCompleted successfully
    Coating IntegrityCompleted successfully
    Particulate EvaluationCompleted successfully
    Balloon Rated Burst Pressure in StentCompleted successfully
    Balloon Fatigue (Repeat Inflations) in StentCompleted successfully

    The document states: "The results of these tests provide reasonable assurance that the proposed device has been designed and tested to assure conformance to the requirements for its intended use. No new safety or performance issues were raised during the testing and, therefore, these devices may be considered substantially equivalent to the predicate devices."


    Inapplicable/Not Provided Information for AI/ML Software Evaluation

    Given that this is a submission for a physical medical device (a balloon catheter) and not an AI/ML software, the following requested information is not applicable or not provided in the document:

    1. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective): Not applicable, as testing was primarily bench testing and biocompatibility, not data-driven AI/ML performance evaluation.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for a physical device refers to its physical and biological properties.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device does not involve human readers interpreting data enhanced by AI.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. There is no algorithm.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc): For a physical device, "ground truth" would refer to the validated physical and chemical properties and biological responses, measured through standardized tests.
    7. The sample size for the training set: Not applicable. There is no training set for an AI/ML model for this device.
    8. How the ground truth for the training set was established: Not applicable.