Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K071080

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2007-08-31

    (136 days)

    Product Code
    Regulation Number
    878.4400
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A Single use, non sterile dispersive electrode with or without a pre-attached cord to adhere to the infant patient over the entire pad surface to complete the electrosurgical circuit between the generator, the active electrode, and the patient.

    Device Description

    The E Surgical Infant Patient Return Electrode Dual Plate is a single use, non-sterile disposable electrode with and without a pre-attached cord. The use is to complete an electrical circuit during electro surgery between the generator, the active electrode, and an infant patient.

    AI/ML Overview

    The provided text describes the 510(k) summary for the E Surgical Electrosurgical Infant Patient Return Electrode, Dual Plate with Cord. Here's an analysis of the acceptance criteria and study information:

    Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Compliance with ANSI/AAMI HF-18 standard for Electrosurgical Device for Dispersive Electrode Thermal Safety"The E Surgical Pad complies with the ANSI/AAMI HF-18 standard Electrosurgical Device for Dispersive Electrode Thermal Safety and Contact Impedance."
    Compliance with ANSI/AAMI HF-18 standard for Dispersive Electrode Contact Impedance"The E Surgical Pad complies with the ANSI/AAMI HF-18 standard Electrosurgical Device for Dispersive Electrode Thermal Safety and Contact Impedance."
    Compliance with Biocompatibility standard ISO 10993"The Pad design is proven safe by compliance to biocompatibility standard ISO 10993."
    Technological characteristics comparable to predicate device (Valleylab, Inc.'s VL E7510-25 Dual Pad)"The E Surgical Infant Patient Return Dual Electrode is comparable to the Valleylab, Inc.'s VL E7510-25 Dual Pad, REM compatible electrode, a legally marketed device." (This is a summary statement, not a specific performance metric.)
    Indications for Use met (single use, non-sterile, dispersive electrode with or without pre-attached cord, adheres to infant patient over entire pad surface to complete electrosurgical circuit)The device description and indications for use align with this, inferring performance in this regard.

    Study Information

    1. Sample size used for the test set and the data provenance:

      • The document does not specify the sample size for any specific test set used to demonstrate compliance with the ANSI/AAMI HF-18 standard or ISO 10993.
      • The document does not specify the data provenance (e.g., country of origin, retrospective or prospective) for any testing.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • This information is not applicable and not provided. The studies described are related to compliance with engineering and biocompatibility standards, not clinical diagnostic performance requiring expert interpretation or "ground truth" in the traditional sense of medical image analysis or disease detection.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

      • This information is not applicable and not provided. Adjudication methods are typically used in studies where human experts are making judgments that need to be resolved, such as in clinical trials or diagnostic accuracy studies. The compliance studies mentioned (ANSI/AAMI HF-18, ISO 10993) involve objective measurements against predefined standards.
    4. If a multi-reader, multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No, an MRMC comparative effectiveness study was not done. This type of study is relevant for AI-powered diagnostic or assistive devices involving human readers. The E Surgical device is an electrosurgical accessory, not a diagnostic tool or an AI-assisted device.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

      • This information is not applicable and not provided. The device is not an algorithm; it is a physical medical device. The "studies" mentioned are compliance tests against established standards for physical device performance and safety.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • For the performance data mentioned (compliance with ANSI/AAMI HF-18 and ISO 10993), the "ground truth" is defined by the objective criteria and methodologies outlined within those respective standards. For example, for thermal safety, the ground truth would be defined by specific temperature thresholds and measurement protocols outlined in the ANSI/AAMI HF-18 standard. For biocompatibility, it would be the outcomes of specific tests as per ISO 10993 (e.g., cytotoxicity, irritation, sensitization).
    7. The sample size for the training set:

      • This information is not applicable and not provided. The E Surgical Infant Patient Return Electrode is a physical medical device, not a machine learning or AI model that requires a training set.
    8. How the ground truth for the training set was established:

      • This information is not applicable and not provided. As stated above, the device does not involve a training set.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1