Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K100594

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2010-04-01

    (30 days)

    Product Code
    Regulation Number
    862.1660
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Elecsys Troponin I CalCheck 5 is an assayed control for use in calibration verification and for use in the verification of the assay range established by the Elecsys Troponin I reagent on the indicated Elecsys and cobas e immunoassay analyzers.

    Device Description

    The Elecsys Troponin I CalCheck 5 is a lyophilized product consisting of recombinant human cardiac Troponin I in human serum matrix. During manufacture, the analyte is spiked into the matrix at the desired concentration levels.

    AI/ML Overview

    This document describes the Elecsys Troponin I CalCheck 5, a lyophilized product containing recombinant human cardiac Troponin I in a human serum matrix. It is intended for use in calibration verification and to verify the assay range established by the Elecsys Troponin I reagent on Elecsys and cobas e immunoassay analyzers.

    1. Table of Acceptance Criteria and Reported Device Performance

    Performance CharacteristicAcceptance CriteriaReported Device Performance
    Intended Use"an assayed control for use in calibration verification and for use in the verification of the assay range established by the Elecsys Troponin I reagent"Substantially equivalent to predicate device with the same intended use for Troponin I.
    LevelsFiveFive
    FormatLyophilizedLyophilized
    MatrixHuman serum matrixHuman serum matrix
    Handling Instructions"Reconstitute the contents of each vial with exactly 1.0 mL distilled or deionized water; Allow the bottle to stand closed for 15 minutes. Mix gently by inversion to ensure homogeneity."Same as predicate device.
    Stability (Unopened)Store at 2-8°C until expiration dateSame as predicate device.
    Stability (Reconstituted)20 – 25°C : 4 hrsSame as predicate device.
    Value AssignmentNot explicitly stated, implied to be comparable to predicate.Evaluated. Data indicate substantial equivalence.
    Stability (general)Not explicitly stated, implied to be comparable to predicate.Evaluated. Data indicate substantial equivalence.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the sample size used for the test set in the conventional sense of a clinical study. The evaluation focused on "value assignment and stability" of the control material itself. Therefore, the "sample size" would refer to the number of vials/batches tested for these characteristics. This information is not provided in the summary.

    Data Provenance: Not specified, but generally, studies for IVD control materials are conducted internally by the manufacturer, typically in a controlled laboratory setting.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. This is a characteristic for a diagnostic device that interprets or classifies. For an assayed control material, "ground truth" relates to the assigned value and stability of the control and is established through rigorous internal testing and calibration processes, not expert consensus on individual cases.

    4. Adjudication Method for the Test Set

    Not applicable for a control material. Ground truth is established through laboratory measurement and metrological traceability, not expert adjudication of individual results.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No. An MRMC study is relevant for diagnostic devices that involve human interpretation (e.g., imaging devices) to assess reader performance with and without AI assistance. This document describes an assayed control material, which does not involve human readers interpreting results in the same way.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was Done

    This concept is not directly applicable. The device is a "control" material, not an "algorithm." Its performance is inherent in its chemical and physical properties (concentration, stability). The "study" for this device evaluated its intrinsic characteristics (value assignment and stability) against its predicate, not its performance as a standalone diagnostic algorithm.

    7. The Type of Ground Truth Used

    For a control material like Elecsys Troponin I CalCheck 5, the "ground truth" would be:

    • Assigned Values: The target concentrations of Troponin I in each of the five levels, established through a rigorous value assignment process, usually involving reference methods or highly characterized internal standards.
    • Stability Specifications: The predefined parameters for how long the control material maintains its assigned values under specified storage and handling conditions.

    The document states, "The Elecsys Troponin I CalCheck 5 was evaluated for value assignment and stability." This indicates these intrinsic properties formed the basis of the performance evaluation.

    8. The Sample Size for the Training Set

    Not applicable. This device is a control material, not a machine learning algorithm that requires a "training set."

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. As this is not a machine learning device, there is no training set or associated ground truth in that context.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1