Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K092322

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2010-03-12

    (220 days)

    Product Code
    Regulation Number
    866.3510
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Elecsys Rubella IgM immunoassay is for the in vitro qualitative determination of IgM antibodies to rubella virus in human serum and Li-heparin, K3-EDTA, and sodium citrate plasma. This assay may be used as an aid in the presumptive diagnosis of an acute or recent rubella infection in individuals, including women of childbearing age. The electrochemiluminescence immunoassay "ECLIA" is intended for use on the Elecsys and cobas e immunoassay analyzers. NOTE: This assay has not been cleared/approved by the FDA for blood/plasma donor screening.

    Elecsys PreciControl Rubella IgM is used for quality control of the Elecsys Rubella IgM immunoassay on the Elecsys and cobas e immunoassay analyzers.

    Device Description

    (1) The Elecsys Rubella IgM Immunoassay is a two-step sandwich immunoassay with streptavidin microparticles and electrochemiluminescence detection. The Rubella IgM is composed of a biotin-labeled monoclonal antibody-against-human IgM, a Rubella-like particle and a ruthenium-labeled anti-Rubella antibody. A relationship exists between the concentration of the IgM antibody targets present in a patient sample and the level of signal count detected by the system. The IgM assay is a qualitative test based on a cut-off formula dependent on the negative and positive calibrators. Cut-off index (COI) is based on the ratio of assay signal to cut-off signal (also abbreviated s/co). COI values equal to or greater than 1.0 are considered positive for the presence of anti-Rubella IgM antibody. Results are determined using a 2 point calibration. The test system contains the human serum-based calibrators intended for use with the system.
    (2) The Elecsys PreciControl Rubella igM contains two levels of human serum. The positive control contains native, inactivated Rubella IgM antibodies.

    AI/ML Overview

    Here's a breakdown of the requested information based on the provided text for the Elecsys Rubella IgM Test system.

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state formal "acceptance criteria" with numerical thresholds for performance metrics. Instead, it presents performance data (e.g., agreement percentages) as part of the method comparison with predicate devices.

    Metric (Type of Agreement)Acceptance Criteria (Not explicitly stated as criteria in document)Reported Device Performance (Elecsys vs. Predicate)Notes
    Pregnant SubjectsAgainst Zeus Scientific/Abbott AxSym and DPC Immulite
    Negative AgreementN/A99.20% (131/132) [95.90%-99.98% CI]Very high agreement for negative cases
    Positive AgreementN/A0.00% (0/0) [0%-100% CI]No positive rubella IgM cases found in pregnant subjects within this specific comparison
    Non-Pregnant SubjectsAgainst Zeus Scientific/Abbott AxSym and DPC Immulite
    Negative AgreementN/A98.90% (364/368) [97.20-99.70% CI]Very high agreement for negative cases
    Positive AgreementN/A0.00% (0/1) [0.00-97.5% CI]Only one positive rubella IgM case found, and the Elecsys device did not agree. The wide CI reflects the small sample size.
    Analytical Specificity (against predicate)N/A77.6% agreement with predicate for 60 specimens representing a variety of disease statesThis indicates how well the assay differentiates Rubella IgM from other conditions.
    PrecisionVarious CVs (Coefficient of Variation) are reported for intra-assay and inter-assay precision for different control levels and plasma samples. The predicate also reports CVs. These are performance characteristics rather than explicit acceptance criteria.

    2. Sample Sizes Used for the Test Set and Data Provenance

    • Test Set Sample Sizes:
      • Method Comparison (Elecsys vs. Zeus Scientific/Abbott AxSym and DPC Immulite):
        • Pregnant Subjects: 132 (131 negative, 0 positive)
        • Non-Pregnant Subjects: 369 (368 negative, 1 positive)
        • This totals 501 samples for the method comparison.
      • Analytical Specificity: 60 specimens.
    • Data Provenance: The document does not explicitly state the country of origin of the data. It also does not explicitly state if the data was retrospective or prospective.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    The document does not provide information regarding the number of experts used to establish ground truth or their qualifications. The "ground truth" for the method comparison appears to be established by the results of the predicate devices (Zeus Scientific Rubella IgM ELISA Test System, Abbott AxSym, and DPC Immulite).

    4. Adjudication Method for the Test Set

    The document does not specify any adjudication method (e.g., 2+1, 3+1). The comparison is directly made against the results of the predicate devices.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

    No, an MRMC comparative effectiveness study involving human readers and AI assistance was not mentioned or conducted. This device is an in-vitro diagnostic (IVD) immunoassay, not an AI-assisted diagnostic tool for human readers.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    Yes, the performance data presented (e.g., method comparison, analytical specificity, precision) represents the standalone performance of the Elecsys Rubella IgM Immunoassay system. Since it's an automated immunoassay, its performance is inherently "standalone" without direct human-in-the-loop interaction in the diagnostic step (though human operators initiate and interpret the results).

    7. The Type of Ground Truth Used

    The primary "ground truth" used for evaluating the Elecsys Rubella IgM Immunoassay appears to be:

    • Predicate Device Results: For the method comparison, the results from the legally marketed predicate devices (Zeus Scientific Rubella IgM ELISA Test System, Abbott AxSym, and DPC Immulite) served as the reference for determining agreement.
    • The document also implies other laboratory methods were used for the "analytical specificity" though more detail is not provided.

    8. The Sample Size for the Training Set

    The document does not provide information regarding a "training set" size. As an immunoassay, the device is developed and validated through analytical studies and clinical comparisons, rather than a machine learning approach that typically involves distinct training and test sets.

    9. How the Ground Truth for the Training Set Was Established

    Since a "training set" is not mentioned in the context of this immunoassay's development, the document does not describe how ground truth for a training set was established. The development of such assays typically involves optimizing reagents and protocols to achieve desired analytical performance and then validating that performance against established methods and clinical samples.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1