Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K132571

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2013-10-18

    (64 days)

    Product Code
    Regulation Number
    862.1215
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Immunoassay for the in vitro quantitative determination of the MB isoenzyme of creatine kinase in human serum and plasma. Measurements of the MB isoenzyme of creatinine kinase are used as an aid in the diagnosis of myocardial infarction. The electrochemiluminescence immunoassay "ECLIA" is intended for use on the indicated Elecsys and cobas e immunoassay analyzers.

    Device Description

    The CK-MB STAT Assay and the CK-MB Assay are two-step sandwich immunoassays with streptavidin microparticles and electrochemiluminescence detection. Results are determined using a calibration curve that is generated specifically on each instrument by a 2-point calibration and a master curve (5-point-calibration) provided with the reagent bar code. The CK-MB STAT application is identical to the CK-MB assay, with the only difference being the length of incubation (9 minutes vs. 18 minutes).

    AI/ML Overview

    This document describes the 510(k) Summary for the Elecsys CK-MB STAT and Elecsys CK-MB Immunoassays (4th Generation reagent). The purpose of the submission is to market a new reagent developed by Roche Diagnostics, claiming substantial equivalence to the previous 3rd Generation CK-MB STAT assay (K022654). The new reagent is intended for the in vitro quantitative determination of the MB isoenzyme of creatine kinase in human serum and plasma, used as an aid in diagnosing myocardial infarction on Elecsys and cobas e immunoassay analyzers.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are generally implied by the performance characteristics of the predicate device and the goals for the new 4th Generation assay. The study demonstrated the performance of the new device (referred to as "4th Generation CK-MB STAT Assay (modified)" or "4th Generation CK-MB 18-Minute assay (modified)") against these implied benchmarks and established its own performance characteristics.

    Here's a table summarizing the performance characteristics for both the predicate (3rd Generation Elecsys CK-MB STAT Assay) and the 4th Generation CK-MB STAT Assay (the primary focus of the comparison):

    Feature / Performance CharacteristicPredicate Device: Elecsys 3rd Generation CK-MB STAT Assay (K022654)4th Generation CK-MB STAT Assay (modified)Acceptance Criteria (implied/demonstrated)
    Measuring Range0.1-500 ng/mL1-300 ng/mLThe new device has a slightly more conservative upper limit for the measuring range, but the overall range aligns for clinical utility.
    Precision (Within-run / Repeatability) - e.g., @ 5.77 ng/mL or 5.46 ng/mLElecsys 2010: e.g., 1.5% CV @ 12.4 ng/mL, 2.1% CV @ 39.7 ng/mLcobas e 411: e.g., 1.2% CV @ 5.46 ng/mL, 1.3% CV @ 29.5 ng/mLNew device demonstrates comparable or improved precision on the cobas e 411 platform. The CVs are generally low, indicating good repeatability.
    Precision (Total / Intermediate) - e.g., @ 5.77 ng/mL or 5.46 ng/mLElecsys 2010: e.g., 2.3% CV @ 5.77 ng/mL, 2.6% CV @ 12.4 ng/mLcobas e 411: e.g., 2.5% CV @ 5.46 ng/mL, 4.2% CV @ 29.5 ng/mLNew device demonstrates comparable or improved overall precision.
    Analytical Sensitivity (Lower Detection Limit / LoD)<0.100 ng/mLLimit of Blank (LoB): = 0.1 ng/mL, Limit of Detection (LoD): = 0.3 ng/mLThe new device's LoD is slightly higher than the predicate's lower detection limit, but still within clinically acceptable low range.
    Analytical Specificity (CK-MM, CK-BB reactivity)CK-MM: None, CK-BB: 0.10%SameAchieved the same high specificity as the predicate.
    Hook EffectNo high-dose hook effect up to 5000 ng/mLSameMaintained the excellent hook effect performance of the predicate.
    Limitations (Interferences)Hemoglobin <1.5 g/dL, Bilirubin < 34 mg/dL, Intralipid < 1.500 mg/dL, Biotin < 100 ng/mL, Rheumatoid factors ≤ 1500 IU/mLHemoglobin ≤1000 mg/dL, Bilirubin ≤ 34 mg/dL, Intralipid ≤ 1.500 mg/dL, Biotin ≤ 30 ng/mL, Rheumatoid factors ≤ 1500 IU/mL, IgG < 7 g/dL, IgM < 1 g/dL, IgA ≤ 1.6 g/dL, Serum Albumin ≤ 20 g/dLThe new device maintains similar immunity to common interferents, with a more stringent biotin tolerance (≤ 30 ng/mL vs. < 100 ng/mL) and additional robust data for immunoglobulins and serum albumin.
    Method Comparison (Predicate vs 4th Gen STAT - Serum)Not applicable (new device compared to predicate)n = 165 samples, Min = 1.08 ng/mL, Max = 283 ng/mL. Passing/Bablok: Slope = 1.053, Intercept = -0.525, Tau/r = 0.982. Linear Regression: Slope = 1.067, Intercept = -0.740, Tau/r = 0.999.The new device shows good correlation with the predicate, indicating substantial equivalence in measurement across the clinical range.
    Method Comparison (4th Gen STAT vs 4th Gen 18-minute - Serum)Not applicable (comparison between two new devices)n = 115 samples, Min = 1.44 ng/mL, Max = 283 ng/mL. Passing/Bablok: Slope = 1.003, Intercept = -0.005, Tau/r = 0.985. Linear Regression: Slope = 1.006, Intercept = 0.071, Tau/r = 0.999.Excellent correlation between the two new assays (STAT and 18-minute versions), supporting consistent performance.
    Matrix Comparison (Plasma vs. Serum)Not explicitly stated but assumed compatible.All data passed criteria for Na-heparin, Li-heparin, K2-EDTA, K3-EDTA: Slope: 0.9-1.1, Pearson r > 0.95, Intercept: ≤± 0.15, Single sample pairs: 100% ± 20%.Demonstrated compatibility with various plasma anticoagulants, showing minimal interference compared to serum.
    Reagent Stability (On Analyzers)8 weeks6 weeksSlightly reduced on-analyzer stability compared to the predicate, but still robust.

    Study Proving Acceptance Criteria:

    The studies presented in the 510(k) summary are direct comparisons demonstrating that the new 4th Generation Elecsys CK-MB STAT and Elecsys CK-MB Immunoassays meet performance criteria for substantial equivalence to the predicate device (Elecsys 3rd Generation CK-MB STAT Assay, K022654) and also illustrate the performance characteristics of the new assays themselves.

    2. Sample Sizes Used for the Test Set and Data Provenance:

    • Method Comparison (4th Gen STAT vs. Predicate): n = 165 serum samples.
    • Method Comparison (4th Gen STAT vs. 4th Gen 18-Minute): n = 115 serum samples.
    • Precision Testing (4th Gen STAT & 4th Gen 18-Minute): Performed using human serum samples and PreciControl Cardiac II over 21 days, according to CLSI EP5-A2. The number of samples for precision is not explicitly stated as 'n=' for unique patients but rather implied through the CLSI methodology.
    • Analytical Sensitivity (LoB, LoD, LoQ): Determined in accordance with CLSI EP17-A requirements, involving n ≥ 60 measurements of analyte-free samples.
    • Matrix Comparison (Plasma vs. Serum): 35 tubes collected per anticoagulant (Na-heparin, Li-heparin, K2-EDTA, K3-EDTA).
    • Clinical Study/Reference Range Study: N = 760 women, 628 men for 3rd Gen (predicate). For 4th Gen, N=523 female, 568 male for all subjects, and N=120 female, 102 male for subjects with no self-reported risk factors. Subjects were "apparently heart healthy" with exclusion of known poor cardiac health.
    • Data Provenance: The document does not explicitly state the country of origin for the patient samples. The studies were conducted internally by Roche Diagnostics. The nature of the studies (method comparisons, precision, analytical sensitivity, matrix comparisons) suggests these were primarily prospective laboratory studies designed to characterize the device's performance, rather than retrospective analysis of existing clinical data. The Clinical Study/Reference Range Study would typically involve prospective collection and analysis of de-identified healthy donor samples.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications:

    This submission concerns an in vitro diagnostic (IVD) immunoassay, not an imaging device or AI algorithm that requires human experts for ground truth establishment in the same way. The "ground truth" for this device is established through:

    • Reference Methods: Comparison against a legally marketed predicate device (Elecsys 3rd Generation CK-MB STAT Assay) and comparison between the two new assays (STAT and 18-minute versions). The predicate device itself was cleared based on its own performance relative to established methods.
    • Industry Standards: Adherence to CLSI (Clinical and Laboratory Standards Institute) guidelines (EP5-A2 for precision, EP17-A for analytical sensitivity, EP6-A for linearity). These guidelines define rigorous statistical and experimental approaches for evaluating assay performance.

    Therefore, the "ground truth" is not established by a panel of human experts reviewing individual cases but by the analytical characteristics of the device when measured against well-defined reference standards and methods in laboratory settings, overseen by qualified laboratory scientists and statisticians.

    4. Adjudication Method for the Test Set:

    Not applicable in the context of an IVD immunoassay. Adjudication methods like 2+1 or 3+1 are typically used in clinical imaging studies where multiple human readers interpret results, and a consensus or tie-breaking mechanism is needed to establish ground truth for a dataset. For this immunoassay, performance is evaluated against quantitative measurements from other assays or statistical targets derived from CLSI guidelines.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    No, an MRMC comparative effectiveness study was not done. This type of study is relevant for evaluating the impact of an AI system on human reader performance, particularly in diagnostic imaging. The Elecsys CK-MB STAT and Elecsys CK-MB Immunoassays are standalone laboratory diagnostic devices that provide quantitative measurements, not AI tools designed to assist human readers.

    6. Standalone Performance Study (Algorithm Only):

    Yes, the studies presented are essentially standalone performance studies of the reagent (the "algorithm" in a broad sense for an IVD). The various performance characteristics (measuring range, precision, analytical sensitivity, analytical specificity, hook effect, interference, and method comparisons) directly represent the performance of the immunoassay system (reagent + instrument) without human-in-the-loop assistance in interpreting the final quantitative result. The device provides a numerical value (ng/mL) directly.

    7. Type of Ground Truth Used:

    The ground truth for evaluating this immunoassay is primarily based on:

    • Reference Measurement (Predicate Device): The performance of the new 4th Generation assay is compared against the established performance of the legally marketed 3rd Generation predicate device.
    • Analytical Standards: Performance metrics like precision, analytical sensitivity (LoB, LoD, LoQ), and linearity are established using statistical methods defined by CLSI guidelines (e.g., EP5-A2, EP17-A, EP6-A). This involves using control materials and samples with known or precisely characterized analyte concentrations.
    • Biological Samples: Human serum and plasma samples are used to assess performance characteristics, including method comparisons and matrix effects. The "true" value for these clinical samples is often assigned by the reference method or by a highly characterized comparative method.
    • Traceability: The CK-MB STAT assay is traceable to the Abbott IMx CK-MB assay and linearized using human recombinant CK-MB from Seradyn, indicating an effort to link its measurements to recognized standards.

    8. Sample Size for the Training Set:

    The document does not specify a distinct "training set" sample size in the context of machine learning or AI. For IVD reagents, the development process involves extensive internal testing and optimization (which can be considered analogous to "training" and "validation" phases in AI development, but with biochemical and analytical methodology), but these details are typically not provided as specific "training set sizes" in 510(k) summaries. The data presented are performance validation data.

    9. How the Ground Truth for the Training Set Was Established:

    As above, the concept of a "training set" and its "ground truth" doesn't directly apply in the machine learning sense here. For reagent development and optimization, the "ground truth" during development would be established through a combination of:

    • Known concentration standards and calibrators: Used to ensure the assay accurately measures CK-MB levels.
    • Reference methods: Comparing experimental formulations against existing, well-characterized CK-MB assays.
    • Clinical samples: Testing various formulations with a wide range of human serum and plasma samples, with consensus values likely derived from established methods or expert laboratories.
    • Biochemical principles: Ensuring the immunoassay's capture and detection mechanisms accurately target the MB isoenzyme of creatine kinase.

    The summary details the methods for determining analytical performance characteristics, indicating a rigorous scientific approach to developing and validating the device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1