Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K140404

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2014-05-08

    (79 days)

    Product Code
    Regulation Number
    862.1215
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Immunoassay for the in vitro quantitative determination of the MB isoenzyme of creatine kinase in human serum and plasma. Measurements of the MB isoenzyme of creatinine kinase are used as an aid in the diagnosis of myocardial infarction.
    The electrochemiluminescence immunoassay "ECLIA" is intended for use on the indicated Elecsys and cobas e immunoassay analyzers.

    Device Description

    The CK-MB STAT (one-step incubation) Assay is a sandwich immunoassay with streptavidin microparticles and electrochemiluminescence detection.
    Results are determined using a calibration curve that is generated specifically on each instrument by a 2 point calibration and a master curve (5-point-calibration) provided with the reagent bar code.
    The CK-MB STAT (one-step incubation) application is identical to the CK-MB STAT (two-step incubation) assay, the only difference being for the CK-MB STAT (one-step incubation) application, the sample, reagent 1, reagent 2 and microparticles are added at one time.

    AI/ML Overview

    This document describes the acceptance criteria and study results for the Elecsys CK-MB STAT Immunoassay, a device used for the quantitative determination of the MB isoenzyme of creatine kinase in human serum and plasma, as an aid in diagnosing myocardial infarction.

    This submission is a Special 510(k), indicating a modification to an existing device, the Elecsys CK-MB STAT (two-step incubation) Assay (K132571). The key difference in the modified device is the change to a one-step incubation protocol and its evaluation on the cobas e 601 Immunoassay Analyzer, compared to the predicate's two-step protocol on the cobas e 411.

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document primarily focuses on demonstrating substantial equivalence to the predicate device by comparing various performance characteristics. The "Acceptance Criteria" for the modified device are generally considered to be equivalent to or better than the performance of the predicate device, or within acceptable clinical laboratory standards.

    FeaturePredicate Device (Elecsys CK-MB STAT, two-step incubation, K132571 on cobas e 411) Acceptance Criteria / PerformanceModified Device (Elecsys CK-MB STAT, one-step incubation on cobas e 601) Reported Performance
    General Assay Features
    Intended Use/IndicationsQuantitative determination of MB isoenzyme of creatine kinase in human serum and plasma for myocardial infarction diagnosis on Elecsys and cobas e immunoassay analyzers.Same
    Assay ProtocolTwo-step Sandwich assayOne-step Sandwich assay
    Detection ProtocolElectrochemiluminescent ImmunoassaySame
    ApplicationsSTAT (9 minute) applicationSame
    Instrument Platformcobas e 411Roche cobas e 601
    Sample Volume15 µLSame
    Sample TypeHuman serum and plasma (K2-EDTA, K3-EDTA, lithium heparin, sodium heparin)Same
    Plasma Type Comparison(vs. Serum, normally filled tubes)
       Na Heparin PlasmaSlope: 0.9 - 1.1; Intercept: ≤ +/- 0.15; r: > 0.95; Relative dev: +/- 20%Slope: 0.994; Intercept: 0.0010; r: 0.9997; Max deviation: 14.6%
       Li Heparin PlasmaSlope: 0.9 - 1.1; Intercept: ≤ +/- 0.15; r: > 0.95; Relative dev: +/- 20%Slope: 0.996; Intercept: 0.0045; r: 0.9996; Max deviation: 12.0%
       K2-EDTA PlasmaSlope: 0.9 - 1.1; Intercept: ≤ +/- 0.15; r: > 0.95; Relative dev: +/- 20%Slope: 1.020; Intercept: 0.0082; r: 0.9999; Max deviation: 10.5%
       K3-EDTA PlasmaSlope: 0.9 - 1.1; Intercept: ≤ +/- 0.15; r: > 0.95; Relative dev: +/- 20%Slope: 0.988; Intercept: 0.0115; r: 0.9998; Max deviation: 8.7%
    ReagentsSandwich principle, 9 minutes; 2-step incubationSandwich principle, 9 minutes; 1-step incubation
    CalibratorCK-MB STAT CalSetSame
    Calibration IntervalSpecific intervals and conditions (e.g., once per lot, after 12 weeks, as required)Same
    ControlsElecsys PreciControl Cardiac IISame
    Traceability/StandardizationTraceable to Abbott IMx CK-MB assay, linearized using human recombinant CK-MB from SeradynSame
    Reagent StabilityUnopened: 2-8°C - up to expiration date; Opened: 2-8°C - 12 weeks; On Analyzers - 8 weeksSame
    LinearitySeries 1: y=0.9421 -0.0579; Series 2: y=0.9348-0.116; Series 3: y=0.942-0.0964Series 1: y=0.9571 -0.1267 (range: 0.27 to 563 ng/mL); Series 2: y=0.9499-0.0957 (range: 0.13 to 548 ng/mL); Series 3: y=0.9576-0.1188 (range: 0.16 to 328 ng/mL). Linearity was established at ± 11.3% within the stated measuring range.
    Labeled Performance Characteristics
    Measuring Range1-300 ng/mLSame
    Precision (CV%)
       Within-run (Repeatability)@ 5.46 ng/mL: 1.2%@ 29.5 ng/mL: 1.3%@ 93.5 ng/mL: 1.3%@ 301 ng/mL: 1.5%PC1 @ 4.44 ng/mL: 1.3%PC2 @ 57.9 ng/mL: 1.4%@ 5.34 ng/mL: 1.1%@ 27.3 ng/mL: 1.1%@ 89.2 ng/mL: 1.1%@ 283 ng/mL: 0.8%PC1 @ 4.27 ng/mL: 1.2%PC2 @ 54.3 ng/mL: 0.9%
       Total (Intermediate)@ 5.46 ng/mL: 2.5%@ 29.5 ng/mL: 4.2%@ 93.5 ng/mL: 4.1%@ 301 ng/mL: 3.3%PC1 @ 4.44 ng/mL: 2.6%PC2 @ 57.9 ng/mL: 3.0%@ 5.34 ng/mL: 1.4%@ 27.3 ng/mL: 3.2%@ 89.2 ng/mL: 2.5%@ 283 ng/mL: 2.2%PC1 @ 4.27 ng/mL: 1.4%PC2 @ 54.3 ng/mL: 1.3%
    Analytical SensitivityLoB: 0.1 ng/mL; LoD: 0.3 ng/mL; LoQ: 1 ng/mL (established per CLSI EP17-A)Same
    Analytical Specificity (Reactivity)CK-MM: None; CK-BB: 0.10%Same
    Hook EffectNo high-dose hook effect up to 5000 ng/mLSame
    InterferencesNo interference (recovery within ± 10% compared to unspiked reference) with various substances including: lipids (2000 mg/dL), biotin (50 ng/mL), bilirubin (40 mg/dL), hemoglobin (1000 mg/dL), RF (1700 IU/mL), albumin (14 g/dL), immunoglobulins (IgG: 7 g/dL, IgM: 1 g/dL, IgA: 1.6 g/dL), and 18 common pharmaceuticals.Same
    Special Drugs InterferencesNo interference (recovery within ± 10% compared to unspiked reference) with 33 special drugs.Same
    LimitationsRare interference from extremely high titers of antibodies; results to be assessed with patient's medical history, clinical exam, and other findings.Same
    Clinical Study/Reference Range
       99th percentile (All Subjects)Female (N=523): 5.34 ng/mL; Male (N=568): 10.36 ng/mLSame (as established by predicate)
       99th percentile (No self-reported risk factors)Female (N=120): 4.30 ng/mL; Male (N=102): 7.70 ng/mLSame (as established by predicate)
    Method Comparison (Predicate on e 411 vs. Modified on e 601)n = 115; Min = 1.47 ng/mL; Max = 269 ng/mL; Slope = 0.976; Intercept = 0.053; Tau = 0.991

    Study Proving Device Meets Acceptance Criteria:

    The study conducted is a method comparison study to demonstrate that the modified Elecsys CK-MB STAT (one-step incubation) Assay on the cobas e 601 Immunoassay Analyzer performs substantially equivalently to the legally marketed predicate device, the Elecsys CK-MB STAT (two-step incubation) Assay on the cobas e 411.

    Key Information about the Studies:

    1. Sample Size used for the Test Set and Data Provenance:

      • Plasma Type Comparison: For each plasma type (Na Heparin, Li Heparin, K2-EDTA, K3-EDTA), "at least 30" or specific numbers rounded to 34 or 35 samples were used (e.g., 30 for Na Heparin, 35 for Li Heparin, 34 for K2-EDTA, 35 for K3-EDTA).
      • Linearity: 3 series of serum samples were used, with ranges from 0.13 to 563 ng/mL.
      • Precision: Not explicitly stated, but CLSI EP5-A2 guidelines typically involve multiple replicates (e.g., 2 runs per day for 20 days) for multiple samples (e.g., controls and patient samples at different concentrations). The performance values provided (e.g., %CV) are derived from such a study.
      • Interference (Common Pharmaceuticals and Special Drugs): Not explicitly stated, but "11 numerical values" were tested for general interferents, and "18 commonly used pharmaceuticals" and "33 special drugs" were tested.
      • Clinical/Reference Range: 523 female subjects and 568 male subjects ("apparently heart healthy") were used to establish the 99th percentile for the predicate device.
      • Method Comparison: A sample size of n = 115 was used, with CK-MB concentrations ranging from 1.47 ng/mL to 269 ng/mL.
      • Data Provenance: The document does not explicitly state the country of origin for the data or whether it was retrospective or prospective. Given that Rochester Diagnostics GmbH (Germany) is the submitter, it is likely that parts of the data originated from Europe, but this is not confirmed. The clinical study for reference ranges specifies "Subjects represented 'apparently heart healthy' population," which implies prospective selection, but the details are scarce.
    2. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

      • For the analytical performance studies (precision, linearity, interference, analytical sensitivity, specificity), the "ground truth" is typically established by the reference methods/standards used for spiking, dilution, or comparison, rather than expert consensus on individual cases. For example, linearity samples are generated by diluting a high-concentration sample with a low-concentration matrix, and the expected values are the "ground truth."
      • For the clinical reference range, the "apparently heart healthy" population was selected based on "exclusion of subjects with known poor cardiac health or known peripheral vascular disease." This implies clinical assessment, but the number and qualifications of experts involved in this selection are not specified.
      • For the method comparison, the "ground truth" implicitly relies on the predicate device's established performance. No external experts are mentioned for establishing ground truth for the test set.
    3. Adjudication Method for the Test Set:

      • Not applicable as this is an in-vitro diagnostic device for quantitative measurement, and the studies described are analytical performance evaluations and method comparisons against a predicate device. There is no mention of subjective interpretation of results requiring adjudication.
    4. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

      • No, an MRMC comparative effectiveness study was not done. This type of study typically applies to imaging or diagnostic devices where human reader interpretation is a key component, often comparing AI-assisted reading to unassisted reading. The Elecsys CK-MB STAT Immunoassay is an in-vitro diagnostic test that provides a quantitative numerical result, not an image or complex data requiring human interpretation in the same way.
    5. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study:

      • Yes, this is effectively a standalone performance study of the modified assay. The device itself (the immunoassay and the analyzer) is an automated system. The studies evaluate the analytical performance of this automated system directly (e.g., precision, linearity, analytical sensitivity, specificity, interference) and compare its output to the predicate device. There is no human-in-the-loop component in the direct measurement and reporting of CK-MB values by the device itself that would require a separate "human-in-the-loop" study.
    6. Type of Ground Truth Used:

      • The ground truth for the performance studies is rooted in established laboratory standards, reference materials, and comparative methods.
        • Linearity: Gravimetric dilutions of known concentration samples.
        • Analytical Sensitivity: Statistical treatment of blank and low-concentration samples (CLSI EP17-A standard).
        • Analytical Specificity: Spiking samples with known concentrations of interfering substances (e.g., CK-MM, CK-BB, bilirubin, drugs) and measuring recovery.
        • Method Comparison: The established values measured by the predicate device (Elecsys CK-MB STAT two-step on cobas e 411) serve as the comparative standard.
        • Clinical Reference Range: Derived from measurements in a population of "apparently heart healthy" individuals, based on clinical exclusion criteria.
    7. Sample Size for the Training Set:

      • This document describes a premarket notification (510(k)) for a modified immunoassay device, not a machine learning or AI algorithm in the traditional sense that requires a "training set." The development of such an assay involves extensive R&D, reagent formulation, and optimization which can be considered "training," but it doesn't involve a distinct "training set" of patient data in the same way an AI algorithm for image recognition would have. The performance studies detailed are for validation/testing of the developed assay.
    8. How the Ground Truth for the Training Set was Established:

      • As noted above, the concept of a "training set" and associated "ground truth establishment" is not directly applicable in the context of this traditional immunoassay device's submission. The "training" for such a device is the iterative process of optimizing reagent concentrations, incubation times, antibody pairs, and detection methods during its analytical development, often guided by known chemical principles and calibrated against reference standards. The studies presented here are to validate the final optimized product.
    Ask a Question

    Ask a specific question about this device

    K Number
    K132571

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2013-10-18

    (64 days)

    Product Code
    Regulation Number
    862.1215
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Immunoassay for the in vitro quantitative determination of the MB isoenzyme of creatine kinase in human serum and plasma. Measurements of the MB isoenzyme of creatinine kinase are used as an aid in the diagnosis of myocardial infarction. The electrochemiluminescence immunoassay "ECLIA" is intended for use on the indicated Elecsys and cobas e immunoassay analyzers.

    Device Description

    The CK-MB STAT Assay and the CK-MB Assay are two-step sandwich immunoassays with streptavidin microparticles and electrochemiluminescence detection. Results are determined using a calibration curve that is generated specifically on each instrument by a 2-point calibration and a master curve (5-point-calibration) provided with the reagent bar code. The CK-MB STAT application is identical to the CK-MB assay, with the only difference being the length of incubation (9 minutes vs. 18 minutes).

    AI/ML Overview

    This document describes the 510(k) Summary for the Elecsys CK-MB STAT and Elecsys CK-MB Immunoassays (4th Generation reagent). The purpose of the submission is to market a new reagent developed by Roche Diagnostics, claiming substantial equivalence to the previous 3rd Generation CK-MB STAT assay (K022654). The new reagent is intended for the in vitro quantitative determination of the MB isoenzyme of creatine kinase in human serum and plasma, used as an aid in diagnosing myocardial infarction on Elecsys and cobas e immunoassay analyzers.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are generally implied by the performance characteristics of the predicate device and the goals for the new 4th Generation assay. The study demonstrated the performance of the new device (referred to as "4th Generation CK-MB STAT Assay (modified)" or "4th Generation CK-MB 18-Minute assay (modified)") against these implied benchmarks and established its own performance characteristics.

    Here's a table summarizing the performance characteristics for both the predicate (3rd Generation Elecsys CK-MB STAT Assay) and the 4th Generation CK-MB STAT Assay (the primary focus of the comparison):

    Feature / Performance CharacteristicPredicate Device: Elecsys 3rd Generation CK-MB STAT Assay (K022654)4th Generation CK-MB STAT Assay (modified)Acceptance Criteria (implied/demonstrated)
    Measuring Range0.1-500 ng/mL1-300 ng/mLThe new device has a slightly more conservative upper limit for the measuring range, but the overall range aligns for clinical utility.
    Precision (Within-run / Repeatability) - e.g., @ 5.77 ng/mL or 5.46 ng/mLElecsys 2010: e.g., 1.5% CV @ 12.4 ng/mL, 2.1% CV @ 39.7 ng/mLcobas e 411: e.g., 1.2% CV @ 5.46 ng/mL, 1.3% CV @ 29.5 ng/mLNew device demonstrates comparable or improved precision on the cobas e 411 platform. The CVs are generally low, indicating good repeatability.
    Precision (Total / Intermediate) - e.g., @ 5.77 ng/mL or 5.46 ng/mLElecsys 2010: e.g., 2.3% CV @ 5.77 ng/mL, 2.6% CV @ 12.4 ng/mLcobas e 411: e.g., 2.5% CV @ 5.46 ng/mL, 4.2% CV @ 29.5 ng/mLNew device demonstrates comparable or improved overall precision.
    Analytical Sensitivity (Lower Detection Limit / LoD)<0.100 ng/mLLimit of Blank (LoB): = 0.1 ng/mL, Limit of Detection (LoD): = 0.3 ng/mLThe new device's LoD is slightly higher than the predicate's lower detection limit, but still within clinically acceptable low range.
    Analytical Specificity (CK-MM, CK-BB reactivity)CK-MM: None, CK-BB: 0.10%SameAchieved the same high specificity as the predicate.
    Hook EffectNo high-dose hook effect up to 5000 ng/mLSameMaintained the excellent hook effect performance of the predicate.
    Limitations (Interferences)Hemoglobin <1.5 g/dL, Bilirubin < 34 mg/dL, Intralipid < 1.500 mg/dL, Biotin < 100 ng/mL, Rheumatoid factors ≤ 1500 IU/mLHemoglobin ≤1000 mg/dL, Bilirubin ≤ 34 mg/dL, Intralipid ≤ 1.500 mg/dL, Biotin ≤ 30 ng/mL, Rheumatoid factors ≤ 1500 IU/mL, IgG < 7 g/dL, IgM < 1 g/dL, IgA ≤ 1.6 g/dL, Serum Albumin ≤ 20 g/dLThe new device maintains similar immunity to common interferents, with a more stringent biotin tolerance (≤ 30 ng/mL vs. < 100 ng/mL) and additional robust data for immunoglobulins and serum albumin.
    Method Comparison (Predicate vs 4th Gen STAT - Serum)Not applicable (new device compared to predicate)n = 165 samples, Min = 1.08 ng/mL, Max = 283 ng/mL. Passing/Bablok: Slope = 1.053, Intercept = -0.525, Tau/r = 0.982. Linear Regression: Slope = 1.067, Intercept = -0.740, Tau/r = 0.999.The new device shows good correlation with the predicate, indicating substantial equivalence in measurement across the clinical range.
    Method Comparison (4th Gen STAT vs 4th Gen 18-minute - Serum)Not applicable (comparison between two new devices)n = 115 samples, Min = 1.44 ng/mL, Max = 283 ng/mL. Passing/Bablok: Slope = 1.003, Intercept = -0.005, Tau/r = 0.985. Linear Regression: Slope = 1.006, Intercept = 0.071, Tau/r = 0.999.Excellent correlation between the two new assays (STAT and 18-minute versions), supporting consistent performance.
    Matrix Comparison (Plasma vs. Serum)Not explicitly stated but assumed compatible.All data passed criteria for Na-heparin, Li-heparin, K2-EDTA, K3-EDTA: Slope: 0.9-1.1, Pearson r > 0.95, Intercept: ≤± 0.15, Single sample pairs: 100% ± 20%.Demonstrated compatibility with various plasma anticoagulants, showing minimal interference compared to serum.
    Reagent Stability (On Analyzers)8 weeks6 weeksSlightly reduced on-analyzer stability compared to the predicate, but still robust.

    Study Proving Acceptance Criteria:

    The studies presented in the 510(k) summary are direct comparisons demonstrating that the new 4th Generation Elecsys CK-MB STAT and Elecsys CK-MB Immunoassays meet performance criteria for substantial equivalence to the predicate device (Elecsys 3rd Generation CK-MB STAT Assay, K022654) and also illustrate the performance characteristics of the new assays themselves.

    2. Sample Sizes Used for the Test Set and Data Provenance:

    • Method Comparison (4th Gen STAT vs. Predicate): n = 165 serum samples.
    • Method Comparison (4th Gen STAT vs. 4th Gen 18-Minute): n = 115 serum samples.
    • Precision Testing (4th Gen STAT & 4th Gen 18-Minute): Performed using human serum samples and PreciControl Cardiac II over 21 days, according to CLSI EP5-A2. The number of samples for precision is not explicitly stated as 'n=' for unique patients but rather implied through the CLSI methodology.
    • Analytical Sensitivity (LoB, LoD, LoQ): Determined in accordance with CLSI EP17-A requirements, involving n ≥ 60 measurements of analyte-free samples.
    • Matrix Comparison (Plasma vs. Serum): 35 tubes collected per anticoagulant (Na-heparin, Li-heparin, K2-EDTA, K3-EDTA).
    • Clinical Study/Reference Range Study: N = 760 women, 628 men for 3rd Gen (predicate). For 4th Gen, N=523 female, 568 male for all subjects, and N=120 female, 102 male for subjects with no self-reported risk factors. Subjects were "apparently heart healthy" with exclusion of known poor cardiac health.
    • Data Provenance: The document does not explicitly state the country of origin for the patient samples. The studies were conducted internally by Roche Diagnostics. The nature of the studies (method comparisons, precision, analytical sensitivity, matrix comparisons) suggests these were primarily prospective laboratory studies designed to characterize the device's performance, rather than retrospective analysis of existing clinical data. The Clinical Study/Reference Range Study would typically involve prospective collection and analysis of de-identified healthy donor samples.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications:

    This submission concerns an in vitro diagnostic (IVD) immunoassay, not an imaging device or AI algorithm that requires human experts for ground truth establishment in the same way. The "ground truth" for this device is established through:

    • Reference Methods: Comparison against a legally marketed predicate device (Elecsys 3rd Generation CK-MB STAT Assay) and comparison between the two new assays (STAT and 18-minute versions). The predicate device itself was cleared based on its own performance relative to established methods.
    • Industry Standards: Adherence to CLSI (Clinical and Laboratory Standards Institute) guidelines (EP5-A2 for precision, EP17-A for analytical sensitivity, EP6-A for linearity). These guidelines define rigorous statistical and experimental approaches for evaluating assay performance.

    Therefore, the "ground truth" is not established by a panel of human experts reviewing individual cases but by the analytical characteristics of the device when measured against well-defined reference standards and methods in laboratory settings, overseen by qualified laboratory scientists and statisticians.

    4. Adjudication Method for the Test Set:

    Not applicable in the context of an IVD immunoassay. Adjudication methods like 2+1 or 3+1 are typically used in clinical imaging studies where multiple human readers interpret results, and a consensus or tie-breaking mechanism is needed to establish ground truth for a dataset. For this immunoassay, performance is evaluated against quantitative measurements from other assays or statistical targets derived from CLSI guidelines.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    No, an MRMC comparative effectiveness study was not done. This type of study is relevant for evaluating the impact of an AI system on human reader performance, particularly in diagnostic imaging. The Elecsys CK-MB STAT and Elecsys CK-MB Immunoassays are standalone laboratory diagnostic devices that provide quantitative measurements, not AI tools designed to assist human readers.

    6. Standalone Performance Study (Algorithm Only):

    Yes, the studies presented are essentially standalone performance studies of the reagent (the "algorithm" in a broad sense for an IVD). The various performance characteristics (measuring range, precision, analytical sensitivity, analytical specificity, hook effect, interference, and method comparisons) directly represent the performance of the immunoassay system (reagent + instrument) without human-in-the-loop assistance in interpreting the final quantitative result. The device provides a numerical value (ng/mL) directly.

    7. Type of Ground Truth Used:

    The ground truth for evaluating this immunoassay is primarily based on:

    • Reference Measurement (Predicate Device): The performance of the new 4th Generation assay is compared against the established performance of the legally marketed 3rd Generation predicate device.
    • Analytical Standards: Performance metrics like precision, analytical sensitivity (LoB, LoD, LoQ), and linearity are established using statistical methods defined by CLSI guidelines (e.g., EP5-A2, EP17-A, EP6-A). This involves using control materials and samples with known or precisely characterized analyte concentrations.
    • Biological Samples: Human serum and plasma samples are used to assess performance characteristics, including method comparisons and matrix effects. The "true" value for these clinical samples is often assigned by the reference method or by a highly characterized comparative method.
    • Traceability: The CK-MB STAT assay is traceable to the Abbott IMx CK-MB assay and linearized using human recombinant CK-MB from Seradyn, indicating an effort to link its measurements to recognized standards.

    8. Sample Size for the Training Set:

    The document does not specify a distinct "training set" sample size in the context of machine learning or AI. For IVD reagents, the development process involves extensive internal testing and optimization (which can be considered analogous to "training" and "validation" phases in AI development, but with biochemical and analytical methodology), but these details are typically not provided as specific "training set sizes" in 510(k) summaries. The data presented are performance validation data.

    9. How the Ground Truth for the Training Set Was Established:

    As above, the concept of a "training set" and its "ground truth" doesn't directly apply in the machine learning sense here. For reagent development and optimization, the "ground truth" during development would be established through a combination of:

    • Known concentration standards and calibrators: Used to ensure the assay accurately measures CK-MB levels.
    • Reference methods: Comparing experimental formulations against existing, well-characterized CK-MB assays.
    • Clinical samples: Testing various formulations with a wide range of human serum and plasma samples, with consensus values likely derived from established methods or expert laboratories.
    • Biochemical principles: Ensuring the immunoassay's capture and detection mechanisms accurately target the MB isoenzyme of creatine kinase.

    The summary details the methods for determining analytical performance characteristics, indicating a rigorous scientific approach to developing and validating the device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K974421

    Validate with FDA (Live)

    Date Cleared
    1997-12-17

    (23 days)

    Product Code
    Regulation Number
    862.1215
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Immunoassay for the in vitro quantitative determination of the MB isoenzyme of creatine kinase (CK-MB) in human serum and plasma.

    A creatine phosphokinase / creatine kinase or isoenzymes test system is a device intended to measure the activity of the enzyme creatine phosphokinase or its isoenzymes (a group of enzymes with similar biological activity) in plasma and serum. Measurements of creatine kinase and its isoenzymes are used in the diagnosis and treatment of myocardial infarction and muscle diseases such as progressive, Duchenne-type muscular dystrophy.

    Device Description

    The Elecsys® CK-MB test principle is based on a two-step sandwich with Streptavidin microparticles and electrochemiluminescence detection. Total duration of the assay: 9 minutes 1ª incubation: 15 ul of sample, a biotinylated monoclonal CK-MB-specific antibody and a monoclonal CK-MB-specific antibody labeled with a ruthenium complex 2nd incubation: after the addition of streptavidin-coated microparticles, the complex becomes bound to the solid phase via interaction of biotin and streptavidin. The reaction mixture is aspirated into the measuring cell where the microparticles are magnetically captured onto the surface of the electrode. Unbound substances are then removed with ProCell. Application of a voltage to the electrode then induces chemiluminescent emission which is measured by a photomultiplier.

    • Results are determined via a calibration curve which is instrument-specifically generated by 2-point calibration and a master curve provided via the reagent bar code.
    AI/ML Overview

    This document is a 510(k) summary for a medical device and does not contain studies proving the device meets acceptance criteria. A 510(k) submission primarily demonstrates substantial equivalence to a predicate device, rather than providing detailed studies against pre-defined acceptance criteria in the way a pharmaceutical trial or a software validation study might.

    Therefore, most of the requested information regarding study details, sample sizes, ground truth establishment, expert qualifications, and MRMC studies cannot be extracted from the provided text.

    However, I can provide the limited information that is present regarding the device and its intended use.

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state acceptance criteria or report performance against such criteria. It states that the device is "substantially equivalent" to a predicate device, meaning its performance is considered comparable, not necessarily meeting specific numerical thresholds outlined in this submission.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    Not provided in the document.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable, as this is a laboratory immunoassay device, not an imaging or diagnostic device relying on expert interpretation for ground truth in the traditional sense. Ground truth for an immunoassay typically refers to accurate measurements by a reference method or known concentrations of analytes. However, this information is not provided.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable/not provided.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is an automated immunoassay device, not an AI-powered diagnostic tool requiring human interpretation comparison.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    The device itself is a standalone immunoassay system. The document describes its mechanism as an automated process: "Results are determined via a calibration curve which is instrument-specifically generated by 2-point calibration and a master curve provided via the reagent bar code." This implies a standalone algorithm's performance in generating quantitative results. However, there's no mention of a separate "standalone study" with specific performance metrics (e.g., sensitivity, specificity, accuracy) detailed in this summary.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    Not explicitly stated. For an immunoassay, ground truth would typically be established by highly accurate reference methods or known concentrations of the analyte (CK-MB).

    8. The sample size for the training set

    Not provided. The document describes the assay principle and its components but does not detail internal validation or development datasets.

    9. How the ground truth for the training set was established

    Not provided.

    Summary of available information:

    • Device Name: Elecsys® CK-MB STAT Assay
    • Intended Use: Immunoassay for the in vitro quantitative determination of the MB isoenzyme of creatine kinase (CK-MB) in human serum and plasma. Used in the diagnosis and treatment of myocardial infarction and muscle diseases.
    • Predicate Device: Boehringer Mannheim Elecsys CK-MB STAT (K961501) (first generation).
    • Key Modification: Addition of an anti-BB antibody to Reagent 1.
    • Claimed Equivalence: Substantially equivalent to the predicate device.
    • Assay Principle: Two-step sandwich with Streptavidin microparticles and electrochemiluminescence detection.
    • Assay Duration: 9 minutes (rapid STAT assay).
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1