510(k) Data Aggregation

    K Number
    K212325

    Validate with FDA (Live)

    Device Name
    EEG-acp
    Manufacturer
    Date Cleared
    2022-06-15

    (324 days)

    Product Code
    Regulation Number
    882.1275
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    EEG adhesive conductive paste (hereafter EEG-acp) is intended for use in clinical and research EEG recordings from humans under professional guidance. EEG-acp is an electroconductive paste which should be used in combination with EEG cup electrodes on the scalp.

    Device Description

    EEG adhesive conductive paste (hereafter EEG-acp) is intended for use in clinical and research EEG recordings from humans (from the age of 2) under professional guidance. EEG-acp is an electroconductive paste which should be used in combination with EEG cup electrodes on the scalp. It acts as a conductor between the scalp and the cup electrodes to reduce impedance between the electrode surface and the skin. The reduction in impedance facilitates better recording of the signal. It is intended to be used for up to 10 hours after application to healthy, intact skin. After drying, it is self-adhesive and can be washed off with water. The cream is provided non-sterile and in a 100g aluminum tube. It is white and has no adverse smell. Its texture is smooth and creamy directly after withdrawal from the tube, and solid after drying. EEG-acp is intended to be used as an electroconductive medium between the electrode and the patient's skin, i.e., it amplifies the electrical signal and thereby improves the electrophysiological recording. The high salt content of the cream is responsible for its conductive properties. Its impedance is -1.9 kOhm and its conductivity is 20 mS/cm.

    AI/ML Overview

    The provided text describes the 510(k) summary for the EEG-acp device. This document is a premarket notification to the FDA to demonstrate that the device is at least as safe and effective as a legally marketed predicate device. The information details the device, its intended use, and comparative data against a predicate device.

    Here's an analysis of the acceptance criteria and study proving device performance based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are not explicitly stated as typical performance metrics (e.g., sensitivity, specificity, AUC) for an AI/ML diagnostic device, because EEG-acp is an "Electroconductive Media" (a physical cream), not a diagnostic algorithm. Instead, the "acceptance criteria" for this device are established by demonstrating substantial equivalence to a legally marketed predicate device (SAC2 Electrode Cream, K192606) in terms of its physical and performance characteristics.

    The table below summarizes the key comparisons made for "acceptance" of EEG-acp:

    CharacteristicAcceptance Criteria (Predicate: SAC2 Electrode Cream, K192606)Reported Device Performance (EEG-acp, K212325)Comparison/Outcome
    ClassificationClass II, Product Code GYB, 882.1275 Electroconductive MediaClass II, Product Code GYB, 882.1275 Electroconductive MediaSame
    Indications for UseClinical and research EEG/EP recordings from humans; used with external electrodes as conductor to reduce impedance.Clinical and research EEG recordings from humans (age >2); used with EEG cup electrodes on scalp as electroconductive paste to reduce impedance.Substantially Equivalent differences noted: EEG-acp specifies "cup electrodes" (more specific than "recessed electrodes") and removes EP recordings. It also specifies use for children older than 2 years. These differences are deemed not to raise new safety/effectiveness concerns.
    Intended Use Time>24h, < 30 days< 10 hoursDifference: Shorter intended use time. Deemed not to negatively impact safety or effectiveness.
    Body ContactTopically on intact skinTopically on intact skinSame
    Thickening AgentCarboxymethylcellulose, GlycerolHydroxyethylcellulose, GlycerinDifference: Different agents with similar properties; texture, density, and viscosity found to be the same.
    PreservativePhenoxyethanol, EthylhexylglycerinPhenoxyethanolDifference: Predicate has an additional preservative. Deemed not to raise new questions; the predicate's additional preservative allows for a longer shelf life (3 years vs 2 years for EEG-acp).
    Conductive MaterialSalts (NaCl and KCl)Salt (NaCl)Difference: EEG-acp only uses NaCl. Deemed not to affect effectiveness as conductivity is the same.
    BiocompatibilityEvaluation in accordance with ISO 10993-1Evaluation in accordance with ISO 10993-1Same; Cytotoxicity, Irritation, and Sensitization tests were performed and passed.
    SterilizationProvided non-sterileProvided non-sterileSame
    Shelf Life3 years2 yearsDifference: Shorter shelf life for EEG-acp. Clearly communicated on labeling; deemed not to raise new questions. Validated by accelerated aging testing (ASTM F1980-16), meeting pre-defined pass/fail criteria (color, consistency, impedance).
    Impedance$50 \pm 10$ Ohm$0.3 \pm 0.1$ kOhm (or 300 $\pm$ 100 Ohm)Difference: EEG-acp's direct reported impedance value ($0.3 \pm 0.1$ kOhm) is higher than the predicate's ($50 \pm 10$ Ohm). However, the document states, "Differences are due to different test setup. Impedance of the subject device when placed on skin is equivalent to the predicate devices. No new questions regarding safety or effectiveness are raised." This implies that under comparable real-world conditions (on skin), the impedance is considered equivalent.
    Conductivity20 mS/cm20 mS/cmSame
    Time to DryNot Publicly Available> 15 minNo new questions raised.
    pH Range8-10 (alkaline)6-7.5 (acidic to neutral)Difference: EEG-acp has a more skin-neutral pH. Deemed not to impact safety/effectiveness and potentially preferable for patients with skin diseases.
    Weight100g per tube100g per tubeSame
    PackagingAluminum tubeAluminum tubeSame
    CharacteristicsSalt base, Non-irritating, Non toxic, Light-Grey color adhesiveSalt base, Non-irritating, Non toxic, White color adhesiveDifference: Color different (white vs light-grey). Deemed not to affect safety/effectiveness.
    Additional IngredientsNone mentioned (for SAC2)Allantoin, Bentonite, Glycerin, Lactate, Panthenol, PEG-40, Phenoxyethanol, Polyquaternium-10, Sodium Chloride, Sodium Hydroxide, Skin protection solution, Talcum, Water, Zinc oxideDifference: EEG-acp includes components like Allantoin, Polyquaternium-10, and Panthenol for skin care/protection, aiming to reduce adverse effects. This is presented as an improvement for safety.

    2. Sample Size Used for the Test Set and the Data Provenance

    The document does not describe a clinical study in humans with a "test set" in the traditional sense of an AI/ML algorithm evaluation (i.e., a dataset of patient data used to evaluate algorithm performance). Instead, the studies performed are bench testing and biocompatibility testing of the physical cream itself.

    • Bench Testing:
      • Impedance and Conductivity: Measured to confirm equivalence with the predicate device. The sample size for these measurements is not specified (e.g., number of cream samples, number of measurements).
      • Qualitative characteristics (Color, odor, adhesiveness, time to dry): Verified to be as specified and equivalent to the predicate device. Sample size not specified.
      • Shelf-life: Validated by accelerated aging testing in accordance with ASTM F1980-16. Sample size not specified (e.g., number of batches, number of samples tested over time).
    • Biocompatibility Testing:
      • The tests performed were:
        • ISO 10993-5: Cytotoxicity in vitro Skin Irritation (Human Skin Model Test)
        • ISO 10993-10: Acute Dermal Irritation/Corrosion, Irritation/Skin Sensitization, Guinea Pig Maximization Test.
      • These are laboratory tests on cells/animal models, not human patient data. The sample sizes for these specific biological assays are not detailed in the summary.
    • Data Provenance: The document does not explicitly state the country of origin for the data or whether the tests were retrospective or prospective, though it implicitly describes prospective laboratory/bench testing.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not applicable to the evaluation of this device. As EEG-acp is an electroconductive paste, its performance is assessed through physical and chemical property measurements and biocompatibility testing, not through expert interpretation of diagnostic images or clinical outcomes. Therefore, there is no "ground truth" expert adjudication process in the context of diagnostic accuracy.

    4. Adjudication Method for the Test Set

    Not applicable, as there is no expert adjudication of a diagnostic performance test set.

    5. If a Multi-Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is not an AI/ML algorithm designed to assist human readers (e.g., radiologists) in diagnostic interpretation. It is a physical medical device (conductive paste).

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an algorithm. Bench tests (e.g., impedance, conductivity, shelf-life) could be considered "standalone" in the sense that they evaluate the properties of the cream itself, independent of human application for diagnosis, but this term (standalone) is typically used for AI/ML algorithms.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for this device's acceptance is based on:

    • Physical and Chemical Properties: Measured values (e.g., impedance, conductivity, pH, time to dry, color, odor, adhesiveness, consistency) are compared against scientifically established standards or performance of the predicate device.
    • Biocompatibility Standards: Results from tests defined by ISO 10993-1, -5, and -10 are used as "ground truth" for safety regarding biological compatibility (cytotoxicity, irritation, sensitization).
    • Shelf-life Standards: Accelerated aging tests following ASTM F1980-16 provide the "ground truth" for shelf-life performance.

    8. The Sample Size for the Training Set

    Not applicable. This is not an AI/ML algorithm that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for this type of device.