    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For the qualitative determination of IgM antibodies in human serum to Epsien Barr Virus (recombinant) Nuclear antigen (EBNA-1) antigen. The 78 EBNA-1 IgM assay should be used in conjunction with other Epstein-Barr serologies (VCA IgM, VCA IgG, EA-D IgG, EBNA-1 IgG and heterophile) as an aid in the diagnosis of infectious mononucleosis. The test can be performed either manually or in conjunction with the MAGO PLUS TM Automated EIA Processor.

    Device Description

    The & EBNA-1 IgM ELISA Kit is an enzyme-linked immunosorbent assay (ELISA) for the detection of IgM antibodies to Epstein-Barr Nuclear antigen-1 in human serum.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    Acceptance Criteria and Device Performance

    Acceptance Criteria CategorySpecific CriteriaReported Device Performance
    Clinical PerformanceRelative Specificity (Convalescent): High accuracy in identifying individuals who have had a past EBV infection (negative for current infection markers).97.0% (95% CI: 91.6-99.4)
    Relative Sensitivity (Current Infection): High accuracy in identifying individuals with a current (recent) EBV infection (positive for current infection markers).59.5% (95% CI: 42.1-75.3)
    Relative Specificity (Seronegative): High accuracy in identifying individuals who have never been infected with EBV (negative for all EBV markers).100.0% (95% CI: 89.1-100.0)
    Overall Agreement: General concordance with the established EBV serological status across all characterized sera.89.4% (95% CI: 84.8-94.0)
    PrecisionIntra-Assay and Interassay Precision (Site #1, 2, 3): Consistent results within a single run and across multiple runs at individual sites for positive and negative sera, indicating reliability and reproducibility of the assay. (Implicitly, CV% values should be within an acceptable range, although specific numerical acceptance criteria are not explicitly stated. The provided CV% values are reported.)Site #1: Interassay CV% range for positive sera: 4.85-10.00%; negative sera: 16.74-44.75%. Site #2: Interassay CV% range for positive sera: 5.75-6.51%; negative sera: 15.67-34.34%. Site #3: Interassay CV% range for positive sera: 4.43-22.09%; negative sera: 23.91-62.45%.
    Inter-Site Precision: Consistent results across different testing sites, indicating robustness of the assay across various laboratory environments. (Implicitly, CV% values should be within an acceptable range.)Inter-site CV% range for positive sera: 6.80-14.33%; negative sera: 10.37-39.85%.
    Cross-ReactivitySpecificity with Potentially Cross-Reactive Sera: Minimal false positives when tested against samples known to be positive for other common viral infections (Varicella Zoster, Cytomegalovirus, Herpes Simplex Virus IgM). The expectation is that the EBNA-1 IgM Test Kit should not react with these other antibodies. (Implicitly, a low number of cross-reactive results is desired.)Some cross-reactivity expected: 2/5 anti-CMV IgM positive sera were non-reactive for anti-EBNA-1 IgM, 4/4 anti-VZV IgM positive sera were non-reactive, and 4/4 anti-HSV positive sera were non-reactive. This phrasing suggests some reactivity with CMV was observed, but the non-reactive counts imply a level of specificity.
    Automation EquivalenceCorrelation of Manual and MAGO Plus Results: Strong positive correlation between results obtained manually and those obtained using the MAGO Plus™ Automated EIA Processor, demonstrating that the automated system provides equivalent results to the manual method. (Implicitly, high Pearson Correlation Coefficient and close agreement.)Pearson Correlation Coefficient of 0.985 with a linear regression of Y = 0.0186 + 0.9826 X.
    MAGO Plus Precision: Consistent results (intra-assay and interassay precision) when the assay is performed on the MAGO Plus Automated EIA Processor. (Implicitly, CV% values should be within an acceptable range for automated use).Interassay CV% range for positive sera: 8.11-10.15%; negative sera: 6.88-66.58%.

    Study Details

    1. Test Set Sample Size and Data Provenance:

      • Sample Size: 176 sera.
      • Data Provenance: Frozen retrospective sera from patients.
      • Country of Origin: Not specified.
    2. Experts Used for Ground Truth and Qualifications:

      • The text does not specify the number of experts used to establish the ground truth or their qualifications.
      • The "ground truth" was established by characterizing the sera using commercially available kits for VCA IgM, VCA IgG, EBNA-1 IgG, and heterophile antibodies, and then classifying them into "convalescent," "current infection," or "seronegative" based on the results of this testing.
    3. Adjudication Method for the Test Set:

      • The text does not describe an adjudication method for the test set. The characterization of sera appears to be based on the results of multiple commercial assays rather than a human consensus process.
    4. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

      • No, an MRMC comparative effectiveness study was not done. This device is an in-vitro diagnostic test, and the study evaluated its performance against established serological markers, not against human readers with and without AI assistance.
    5. Standalone (Algorithm Only) Performance:

      • Yes, a standalone performance study was done. The entire clinical performance section evaluates the device (the EBNA-1 IgM ELISA Kit) directly against the characterized sera without human-in-the-loop assistance for the interpretation of the EBNA-1 IgM ELISA results. Human interpretation was part of the "ground truth" establishment using other assays, but the device's output (positive, negative, equivocal) was directly compared.
    6. Type of Ground Truth Used:

      • Expert Consensus (indirectly) / Reference Comparator Assay Results: The ground truth was established by classifying patient sera based on the results of multiple commercially available EBV serology kits (VCA IgM, VCA IgG, EBNA-1 IgG, and heterophile antibodies). This is a form of reference standard based on accepted laboratory diagnostics.
    7. Training Set Sample Size:

      • The document does not specify a training set sample size. This is typical for traditional ELISA kits, where optimization and assay development are performed, but a distinct "training set" for an algorithm in the modern AI sense is not applicable. The device's parameters (e.g., cut-offs) would have been established during its development phase.
    8. How Ground Truth for Training Set Was Established:

      • As there is no distinct "training set" in the AI sense for this device, the concept of establishing ground truth for it is not directly applicable. The development of the assay (e.g., determining optimal concentrations, incubation times, and cut-off values) would have involved extensive experimentation and standardization using characterized samples, similar to how the ground truth for the test set was established using comparator assays.
    Ask a Question

    Ask a specific question about this device

    K Number
    K980598

    Validate with FDA (Live)

    Date Cleared
    1998-05-01

    (73 days)

    Product Code
    Regulation Number
    866.3235
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Epstein Barr Nuclear Antigen 1 (EBNA-1) IgM kit is an Enzyme-Linked Immunosorbent Assay (ELISA) for the qualitative determination of IgM antibodies in human serum to EBNA-1 antigen. The Clark anti-EBNA-1 IgM assay may be used in conjunction with other Epstein-Barr serologies (VCA IgG, VCA IgM, EA-D IgG, EBNA-1 IgG and heterophile) as an aid in the diagnosis of infectious mononucleosis in adult populations.

    Device Description

    The Wampole Epstein-Barr Virus Nuclear Antigen 1 (EBNA-1) IgM kit is an Enzyme-Linked Immunosorbent Assays (ELISA) for the qualitative determination of IgM antibodies in human serum to EBNA-1 antigen. The Wampole anti-EBNA-1 IgM assay may be used in conjunction with other Epstein-Barr serologies (VCA IgG, VCA IgM, EA-D IgG, EBNA-1 IgG and heterophile) as an aid in the diagnosis of infectious mononucleosis.

    For In Vitro Diagnostic Use Only.

    The EBNA-1 IgM ELISA test is an enzyme linked immunosorbent assay to detect IgM antibodies to Epstein-Barr Nuclear Antigen -1. Recombinant EBNA-1 antigen is attached to a solid phase microtiter well. Diluted test sera is added to each well. If the antibodies are present that recognize the antigen, they will bind to the antigen in the well. After incubation the wells are washed to remove unbound antibody. An enzyme labeled anti-human IgM is added to each well. If antibody is present it will bind to the antibody attached to the antigen on the well. After incubation the wells are washed to remove unbound conjugate. A substrate solution is added to each well. If enzyme is present the substrate will undergo a color change. After an incubation period the reaction is stopped and the color intensity is measured photometrically, producing an indirect measurement of specific antibody in the patient specimen.

    AI/ML Overview

    Here's an analysis of the provided information regarding the EBNA-1 IgM ELISA Test Kit, presented with the requested structure:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are not explicitly stated in the provided text. However, based on the presented performance characteristics, we can infer the implied targets for sensitivity and specificity. The study demonstrates performance against these inferred criteria.

    Performance MetricImplied Acceptance Criteria (Inferred)Reported Device Performance
    SensitivityHigh for acute cases (target > ~50%)51.4% (95% CI: 34.9%-67.8%)
    Specificity (Seronegative)High (target = 100%)100% (95% CI: 89.4%-100%)
    Specificity (Seropositive)High (target > ~90%)94.5% (95% CI: 89.7%-99.3%)
    Relative AgreementHigh (target > ~80%)85.3% (95% CI: 79.6%-90.9%)
    Precision (Inter-Site CV)Low variability (e.g., < 15-20% for most samples)Variable (e.g., 8.60% to 26.36%, with HPC 4.45%, CAL 6.48%, NC 181.66%)
    Cross-ReactivityNo significant cross-reactivity with common interfering antibodies (e.g., RF, VZV)No cross-reactivity with RF or VZV. Some cross-reactivity with HSV and CMV.

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: 166 selected serum samples
    • Data Provenance: The document states, "One hundred and sixty six selected serum were tested at a clinical lab." The country of origin is not explicitly stated, but given the company's US presence and the FDA submission, it is likely US-based. The data is retrospective, as the samples were "selected" and characterized prior to testing with the Wampole EBNA-1 IgM kit.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    The ground truth was established by "characteriz[ing]" the serum samples as seronegative, acute, or seropositive based on other established EBV serologies (VCA IgM, VCA IgG, EA-D IgG, EBNA-1 IgG, and heterophile). The text does not specify the number of experts or their qualifications for this characterization. It implies standard laboratory methods were used.

    4. Adjudication Method for the Test Set

    No specific adjudication method is mentioned. The ground truth (characterization of samples) appears to be based on results from other established serological tests, implying a consensus based on these diagnostic markers rather than a panel of experts adjudicating individual cases. Equivocal results from the Wampole assay were "not included in the calculations" and were simply "reported as equivocal," indicating no adjudication for these specific outcomes.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve With AI vs Without AI Assistance

    This is an ELISA (Enzyme-Linked Immunosorbent Assay) test kit, not an AI-powered diagnostic device. Therefore, an MRMC comparative effectiveness study involving human readers and AI assistance is not applicable and was not performed.

    6. If a Standalone (i.e. algorithm only, without human-in-the-loop performance) Was Done

    Yes, the performance characteristics described (Sensitivity and Specificity, Precision, Cross-Reactivity) represent a standalone evaluation of the EBNA-1 IgM ELISA kit. It is an in vitro diagnostic device that produces a quantitative (ISR Value) or qualitative (Positive/Negative/Equivocal) result based on biochemical reactions, without a human-in-the-loop for interpretation beyond reading the instrument's output.

    7. The Type of Ground Truth Used

    The ground truth used was based on expert consensus of established serological markers. Samples were categorized as "seronegative," "acute," or "seropositive" using a panel of other EBV serologies: VCA IgM, VCA IgG, EA-D IgG, EBNA-1 IgG, and heterophile anti-body status. This combines multiple diagnostic test results to define the true disease status for comparison.

    8. The Sample Size for the Training Set

    The document does not explicitly mention a "training set" for the ELISA kit. ELISA kits are typically developed and validated using biochemical principles and reagents. While there would have been internal R&D and analytical validation experiments, the clinical "Performance Characteristics" section focuses on testing against known samples, which serves as the validation or test set.

    9. How the Ground Truth for the Training Set Was Established

    As no specific training set for a machine learning algorithm is discussed, this question is not applicable to an ELISA test kit. The "ground truth" for the overall development and validation of the assay would stem from the known characteristics of the antigens, antibodies, and patient samples used in its development and analytical testing.