510(k) Data Aggregation

    K Number
    K142426

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2015-03-27

    (210 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    EBI Dental Implant System is intended for immediate, delayed, or conventional placement in the maxillary and/or mandibular arches to support crowns, bridges, or overdentures in edentulous or partially edentulous patients.

    Device Description

    An endosseous dental implant is a device made of a material such as Pure Titanium Grade 4 and Titanium alloy (Ti-6A1-4V ELI). EBI Implant system has been designed to accommodate the following dental implant restoration protocols; Immediate or Early loading, immediate placement or one or two stage placement. EBI Implant systems help patients who have partial or whole teeth loss mastication to chew as dental implant. The EBI External Implant System has an external connection.

    The surface of the system has been treated with RBM (Resorbable Blasted media).

    The Fixture diameters are 3.25, 3.3, 3.75, 4, 4.1, 4.8, 5, 5.5, 6, 6.5, 7mm and lengths are 7, 8, 8.1, 8.5, 9, 9.6, 10, 11, 11.5, 12, 12.6, 13, 14, 14.6, 15mm in this system.

    The contained various abutments and accessories in the system are cover screw, healing abutment, cemented abutment, temporary abutment, healing cap, impression coping, analog, fixture mount, Ti-screw, angled abutment, impression guide pin, gold UCLA abutment and locator.

    Fixtures and abutments are packaged separately. The Fixtures are supplied sterile and the abutments and accessories are provided non-sterile. The abutments and accessories should be sterilized before use.

    AI/ML Overview

    The EBI External Implant System is a dental implant system. The provided document is a 510(k) premarket notification, which demonstrates substantial equivalence to predicate devices rather than proving specific acceptance criteria with detailed studies in the manner typically seen for novel AI/software devices. Therefore, the information typically requested for AI/software-based devices regarding acceptance criteria, ground truth, sample sizes, and expert adjudication for performance studies is not directly applicable or available in this document.

    However, based on the provided text, we can glean information about the performance testing conducted and how it demonstrates "safety and effectiveness" in the context of a dental implant.

    Here's an analysis based on the available information:

    1. A table of acceptance criteria and the reported device performance

    The document does not specify quantitative acceptance criteria in terms of performance metrics like sensitivity, specificity, accuracy, or F1-score, as would be expected for a diagnostic or AI-driven decision support system. Instead, for a dental implant, performance is assessed through bench testing demonstrating structural integrity and compliance with recognized standards.

    Acceptance Criteria (Implied)Reported Device Performance
    Fatigue Resistance (to ensure durability under physiological loads)Fatigue Testing was performed in accordance with ISO 14801:2007
    Biocompatibility (materials used are safe for implantation)Fabricated from CP Titanium (Grade 4) conforming to ASTM F67 and Titanium alloy (Ti-6Al-4V ELI) conforming to ASTM F136
    Sterility (implants are sterile)Fixtures are supplied sterile; abutments and accessories are provided non-sterile but should be sterilized before use; Gamma Sterilization: Yes
    Shelf Life (maintaining integrity over time)5 years

    2. Sample size used for the test set and the data provenance

    The document doesn't provide specific sample sizes for the fatigue testing in terms of the number of tested devices. It only states that "Fatigue Testing was performed in accordance with ISO 14801:2007." ISO 14801 specifies test methods for dental implants, which would inherently include a defined number of samples to ensure statistical validity, but this detail is not explicitly stated in the 510(k) summary.

    Data Provenance: The testing was conducted by EBI Inc., a South Korean company.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This question is not applicable in the context of this device and testing. The "ground truth" for a dental implant's mechanical performance is defined by adherence to engineering standards and material specifications, not expert consensus on an image or diagnosis.

    4. Adjudication method for the test set

    Not applicable. Adjudication methods (like 2+1, 3+1) are used for resolving disagreements among human experts in clinical performance studies, typically for diagnostic devices. Mechanical bench testing of an implant does not involve such human-based adjudication.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a medical device (dental implant), not an AI or software-assisted diagnostic system. Therefore, MRMC studies examining human reader performance with or without AI are not relevant.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. There is no algorithm. The device is a physical dental implant.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for the performance of the EBI External Implant System is based on established engineering and material standards, specifically:

    • ISO 14801:2007 for fatigue testing of dental implants.
    • ASTM F67 for unalloyed titanium (CP Titanium Grade 4) used in dental implants.
    • ASTM F136 for wrought titanium-6aluminum-4vanadium alloy (Ti-6Al-4V ELI) for surgical implant applications.
    • General device safety and effectiveness principles outlined in FDA's Class II special controls guidance document for root-form endosseous dental implants and abutments.

    8. The sample size for the training set

    Not applicable. There is no "training set" as this is not a learning algorithm or AI/ML device.

    9. How the ground truth for the training set was established

    Not applicable. As there is no training set for an AI/ML algorithm, the concept of establishing ground truth for it does not apply.