510(k) Data Aggregation

    K Number
    K973123

    Validate with FDA (Live)

    Date Cleared
    1998-03-26

    (218 days)

    Product Code
    Regulation Number
    866.3235
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The . Epstein-Barr Early Antigen Diffuse component (EA-D) IgG kit is an Enzyme-Linked Immunosorbent Assay (ELISA) for the qualitative determination of IgG antibodies in human serum to EA-D antigen. The Clark anti-EA-D IgG assay may be used in conjunction with other Epstein-Barr serologies (VCA IgG, VCA IgM, EBNA-1 IgG, EBNA-1 IgM and heterophile) as an aid in the diagnosis of infectious mononucleosis. For In Vitro Diagnostic Use Only.

    Device Description

    Epstein-Barr Early Antigen Diffuse component (EA-D) IgG kit is an Enzyme-Linked The Immunosorbent Assay (ELISA) for the qualitative determination of IgG antibodies in human serum to EA-D antigen. The Clark anti-EA-D IgG assay may be used in conjunction with other Epstein-Barr serologies (VCA IgG, VCA IgM, EBNA-1 IgG, EBNA-1 IgM and heterophile) as an aid in the diagnosis of infectious mononucleosis. The EA-D IgG ELISA test is an enzyme linked immunosorbent assay to detect IgG antibodies to Epstein-Barr Early Antigen Diffuse component. Purified recombinant EA-D antigen is attached to a solid phase microtiter well. Diluted test sera is added to each well. If the antibodies are present that recognize the antigen, they will bind to the antigen in the well. After incubation the wells are washed to remove unbound antibody. An enzyme labeled anti-human IgG is added to each well. If antibody is present it will bind to the antibody attached to the antigen on the well. After incubation the wells are washed to remove unbound conjugate. A substrate solution is added to each well. If enzyme is present the substrate will undergo a color change. After an incubation period the reaction is stopped and the color intensity is measured photometrically, producing an indirect measurement of specific antibody in the patient specimen.

    AI/ML Overview

    Here's an analysis of the provided text to extract the acceptance criteria and study information:

    Acceptance Criteria and Device Performance for EA-D IgG ELISA Test Kit

    1. Table of Acceptance Criteria and Reported Device Performance

    Performance CharacteristicAcceptance Criteria (Implied/Derived)Reported Device Performance
    Sensitivity (Late Acute)High (e.g., >95%)100% (95% CI: 67.9%-100%)
    Specificity (Seronegative)High (e.g., >95%)100% (95% CI: 79.1%-100%)
    Specificity (Early Acute)High (e.g., >90%)97.0% (95% CI: 91.0%-100%)
    Sensitivity (Seropositive)Not explicitly defined as a target for "sensitivity" in this context, but rather an expected low positive rate.19.7% (95% CI: 12.8%-26.6%)
    Specificity (Seropositive)High (e.g., >70-80%)80.3% (95% CI: 73.4%-87.2%)
    Overall Relative AgreementHigh (e.g., >80%)85.6% (95% CI: 80.5%-90.8%)
    Precision (Coefficient of Variation - CV)Low (e.g., <20% for individual sera, <15% for controls)Ranges from 2.18% (HPC) to 16.34% (Serum 7), with one outlier (NC) at 154.52%. Most individual sera are <12%.
    Cross-ReactivityNo cross-reactivity with common Herpes Viruses (HSV I, HSV II, CMV, VZV)Demonstrated no cross-reactivity with tested sera.

    Notes on Acceptance Criteria: The document does not explicitly state numerical acceptance criteria for each performance characteristic. The "acceptance criteria" listed in the table above are implied or derived based on generally acceptable performance ranges for diagnostic assays and the results presented, which presumably satisfied the regulatory body (FDA). For instance, 100% sensitivity and specificity are ideal and likely met an implicit high-performance requirement. For parameters where the device demonstrated lower performance (e.g., Seropositive Sensitivity), it's important to understand the context of the assay's intended use (aid in diagnosis, not a standalone definitive test).


    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: 193 selected serum samples were used for sensitivity and specificity characterization.
    • Data Provenance: The study was conducted at a "clinical lab." The country of origin is not specified, but the submission is to the U.S. FDA, implying the data is likely from the U.S. or an equivalent clinical setting. The data is retrospective, as the serum samples were "selected" and "characterized" prior to being tested with the EA-D IgG ELISA.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • Number of Experts: Not specified.
    • Qualifications of Experts: Not specified. The characterization of serum (seronegative, early acute, late acute/transitional, seropositive) was based on "serological evidence" using a panel of other Epstein-Barr serologies (VCA IgG, VCA IgM, EBNA IgG, EBNA IgM, and heterophile). While this process implies expert interpretation of these results, the specific number and qualifications of individuals performing this characterization are not detailed.

    4. Adjudication Method for the Test Set

    • Adjudication Method: Not explicitly described. The ground truth was established by characterizing the serum samples based on a combination of other EBV serologies. It's assumed that this characterization itself might have involved some form of consensus or standard interpretation protocol, but no formal adjudication process (e.g., 2+1, 3+1) is mentioned for the final classifications of the samples used as ground truth.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No, an MRMC comparative effectiveness study was not done. This device is an in-vitro diagnostic ELISA kit, not an AI-assisted diagnostic tool requiring human interpretation. Therefore, the concept of human readers improving with or without AI assistance is not applicable.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • Yes, the performance study presents standalone performance of the EA-D IgG ELISA test kit. It measures the device's ability to detect antibodies independently, without human intervention in the result determination beyond running the assay and reading the photometric output. The description specifies "measurement of specific antibody in the patient specimen" via a color change and photometric measurement.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    • Type of Ground Truth: The ground truth for the sensitivity and specificity study was an expert-derived serological classification. Serum samples were characterized as "seronegative," "early acute," "late acute or transitional," or "seropositive" based on the presence or absence of a panel of other established Epstein-Barr serologies (VCA IgM, VCA IgG, EBNA IgG, heterophile antibody). This relies on the recognized diagnostic value of these combined markers.

    8. The Sample Size for the Training Set

    • The document describes a single performance study for the device. It does not mention a distinct "training set" for algorithm development. This is typical for traditional in-vitro diagnostic assays which are often developed and then validated, rather than 'trained' in the machine learning sense. The 193 selected serum samples formed the primary test set for characterization.

    9. How the Ground Truth for the Training Set Was Established

    • As a separate "training set" is not explicitly mentioned or implied for this type of diagnostic kit development, the question of how its ground truth was established is not directly applicable. The "ground truth" for the overall device's validation (the 193 serum samples) was established through serological characterization using established EBV markers, as described in point 7.