Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K123805

    Validate with FDA (Live)

    Date Cleared
    2013-03-20

    (99 days)

    Product Code
    Regulation Number
    876.5820
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Reprocessing Products Corporation (RPC) E-Z Chek® Blood Leak Test Strips are designed to indicate the presence or absence of blood in dialysate during the hemodialysis procedure. The test strips will measure equal to 0 mg HGB/dL and 0.25 mg HGB/dL in dialysate.

    The Reprocessing Product Corporation (RPC) E-Z Chek® Blood Leak Test Strips are indicated for detecting the presence or absence of blood in dialysate during hemodialysis treatments.

    Device Description

    Device is semi-quantitative, reagent test strip comprised of a pad impregnated with chemicals which change color upon contact with blood in dialysate. The pad is attached to a plastic strip for handling.

    AI/ML Overview

    The provided document describes the Reprocessing Products Corporation (RPC) E-Z Chek® Blood Leak Test Strips. Here's an analysis of the acceptance criteria and the study outlined:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document doesn't explicitly state "acceptance criteria" in a quantitative pass/fail format typical for medical device verification. Instead, it describes the intended performance and the study's findings regarding that performance.

    Acceptance Criteria (Implied from Intended Use)Reported Device Performance
    Indicate the presence or absence of blood in dialysate.The product consistently generates color change which meets the color block for the reference solution concentration.
    Measure equal to 0 mg HGB/dL (no blood).Reference solutions were used to evaluate performance between 0 mg HGB/dL and 0.25 mg HGB/dL Bicarbonate/Acetate dialysate. The data confirm the product meets the color block for this concentration.
    Measure 0.25 mg HGB/dL (presence of blood at a specific detectable threshold).Reference solutions were used to evaluate performance between 0 mg HGB/dL and 0.25 mg HGB/dL Bicarbonate/Acetate dialysate. The data confirm the product meets the color block for this concentration.
    Appropriate performance for use in hemodialysis dialysate during treatment.These data demonstrate appropriate performance for use in hemodialysis dialysate used in treatment.
    Technological characteristics: Detection based on pseudoperoxidative activity of hemoglobin and myoglobin, catalyzing oxidation of an indicator by an organic hydroperoxide, producing a color change consistent with color blocks.Confirmed through the performance study where the device generates color change that meets the color block for reference solutions, implying the underlying technological principle is functional and produces the expected visual outcome.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: Not explicitly stated. The document mentions "Reference solutions" were used, and "data confirm the product consistently generates color change," but it doesn't quantify the number of tests performed or the sample size of distinct reference solutions/concentrations tested.
    • Data Provenance: Not explicitly stated. The study appears to be a laboratory-based performance evaluation using reference solutions, rather than human clinical data from a specific country. It is almost certainly retrospective in the sense that the test strips were manufactured and then tested.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. The ground truth for this device is established by the known concentrations of hemoglobin (HGB) in the prepared reference solutions, not by expert human interpretation. The "color block" comparison itself would be a visual assessment, but the ground truth is the chemical concentration.

    4. Adjudication Method for the Test Set

    Not applicable. The performance is evaluated by observing a color change and comparing it to a visual "color block" corresponding to known HGB concentrations. There is no mention of human adjudication for the results beyond this direct comparison.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. This device is a diagnostic test strip where performance is assessed by comparison to reference color blocks, not by interpretation by multiple human readers. Therefore, an effect size of human readers improving with AI vs. without AI assistance is not applicable.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    Yes, this appears to be a standalone performance evaluation. The device's performance is determined by its inherent color-changing reaction to blood in dialysate, and its comparison to established color blocks. There is no "human-in-the-loop" aspect to its fundamental operation or the evaluation method described beyond visually comparing the strip's color to a reference.

    7. The Type of Ground Truth Used

    The ground truth used is known concentrations of hemoglobin (HGB) in prepared reference solutions. These solutions serve as the objective standard against which the test strip's color-changing reaction is evaluated.

    8. The Sample Size for the Training Set

    Not applicable. This device is a chemical reagent test strip, not an AI or machine learning algorithm that requires a "training set." Its chemical formulation dictates its performance, which is then verified.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1