Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K213647

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-02-17

    (90 days)

    Product Code
    Regulation Number
    884.5300
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Durex Patronus Wide condoms are used for contraception and for prophylactic purposes (to help reduce the risk of pregnancy and the transmission of sexually transmitted infections, STIs).

    Device Description

    Durex Patronus Wide are natural rubber latex-based condoms that completely cover the penis with a closely fitted membrane. Durex Patronus Wide is made of a non-colored natural rubber latex with silicone lubricant and is a taper shaft, teat ended smooth shaped condom. Durex Patronus Wide has a nominal length of 200 mm, width of 60 mm, and thickness of 62 um. It will be packaged in individually sealed flexible laminate foils made of polyethylene terephthalate, polyethylene, and aluminum. The foils will come packaged in an outer consumer cardboard carton. The number of condoms in the carton may vary. Durex Patronus Wide condoms are intended for over-the-counter (OTC) use. These condoms conform with FDA-recognized standards ASTM D3492-16 and ISO 4074:2015.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and study information provided in the document for the "Durex Patronus Wide" condom.

    It is important to note that this document is a 510(k) summary for a condom, not a medical device driven by an algorithm or AI. Therefore, many of the typical questions regarding AI-driven medical devices (like sample sizes for test/training sets, expert ground truth, adjudication methods, MRMC studies, and standalone algorithm performance) are not applicable or will not be found in this type of submission.


    Acceptance Criteria and Study Information: Durex Patronus Wide Condom (K213647)

    1. Table of acceptance criteria and the reported device performance

    The acceptance criteria are generally based on recognized international and national standards for male natural rubber latex condoms. The reported device performance indicates that the Durex Patronus Wide met all requirements of these standards.

    Acceptance Criteria (Standard Reference)Reported Device Performance (Durex Patronus Wide)
    Biocompatibility:
    Cytotoxicity (ISO 10993-5:2009/R 2014)Met requirement: Non-cytotoxic
    Sensitization (ISO 10993-10:2010/R 2014)Met requirement: Non-sensitizing
    Vaginal Irritation (ISO 10993-10:2010/R 2014)Met requirement: Non-irritating
    Acute Systemic Toxicity (ISO 10993-11:2017)Met requirement: Not acutely, systemically toxic
    Physical Performance:
    ISO 4074:2015 - Natural rubber latex male condoms – Requirements and test methodsMet all requirements
    ASTM D3492-16 - Standard Specification for Rubber Contraceptives (Male Condoms)Met all requirements
    Shelf Life: (21 CFR 801.435)
    Predefined acceptance criteria for accelerated stabilityMet predefined acceptance criteria
    Specific Device Specifications (from Comparison Table - Section 8):
    Nominal Width$60 \pm 2$ mm
    Nominal Length$200 \pm 10$ mm
    Nominal Thickness$0.062 \pm 0.01$ mm
    Air Burst Pressure> 1.0 kPa (met)
    Air Burst Volume22.0 L (met)

    2. Sample size used for the test set and the data provenance

    • Sample Size for Test Set: The document does not specify a distinct "test set" sample size in the way an AI/algorithm study would. Instead, performance testing was conducted according to the specified standards (ISO 4074:2015 and ASTM D3492-16) and biocompatibility test methods (ISO 10993 series). These standards prescribe specific sampling plans and test methods for evaluating parameters like burst pressure, burst volume, dimensions, and biocompatibility. The exact number of units tested for each specific criterion is not provided in this summary but would be detailed in the full test reports.
    • Data Provenance: The studies were conducted by the manufacturer (RB Health (US) LLC) or contracted labs as part of the regulatory submission process. This data is generally prospective, generated specifically for the device under evaluation. The country of origin of the data is not explicitly stated but would be derived from the testing facilities used by the applicant.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This question is not applicable to a condom's performance evaluation as described. The "ground truth" for condom testing is based on objective physical measurements and biological assays against established, quantitative pass/fail criteria defined in international and national standards. It does not involve expert interpretation or consensus in the way a diagnostic imaging device might.

    4. Adjudication method for the test set

    Not applicable. Condom testing is based on objective measurements and established standards, not judgmental adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. The Durex Patronus Wide is a physical medical device (condom), not an AI-driven system, and therefore, MRMC studies are irrelevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a physical device, not an algorithm.

    7. The type of ground truth used

    The ground truth for physical and biological testing of the Durex Patronus Wide condom is based on:

    • Quantitative metrics defined in recognized standards: e.g., minimum air burst pressure, minimum air burst volume, dimensional tolerances (Nominal Width, Length, Thickness).
    • Biological assay outcomes: Demonstrating non-cytotoxicity, non-sensitization, non-irritation, and absence of acute systemic toxicity, as per ISO 10993 series standards.

    8. The sample size for the training set

    Not applicable. There is no "training set" for a physical device like a condom. Manufacturing processes are validated through quality control and process validation, not machine learning training.

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1