Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K221241

    Validate with FDA (Live)

    Date Cleared
    2022-09-01

    (125 days)

    Product Code
    Regulation Number
    892.2080
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DrAid™ for Radiology v1 is a radiological computer-assisted triage & notification software to aid the clinical assessment of adult Chest X-Ray cases with features suggestive of pneumothorax in medical care environment. DrAid™ analyzes cases using an artificial intelligence algorithm to features suggestive of suspected findings. It makes case-level output available to a PACS for worklist prioritization or triage.

    As a passive notification for prioritization-only software tool with standard of care workflow, DrAid™ does not send a proactive alert directly to appropriately trained medical specialists. DrAid™ is not intended to direct attention to specific portions or anomalies of an image. Its results are not intended to be used on a stand-alone basis for clinical decisionmaking nor is it intended to rule out pneumothorax or otherwise preclude clinical assessment of X-Ray cases.

    Device Description

    DrAid™ for Radiology v1 (hereafter called DrAid™ or DrAid) is a a radiological computer-assisted triage & notification software product that automatically identifies suspected pneumothorax on frontal chest x-ray images and notifies PACS of the presence of pneumothorax in the scan. This notification enables prioritized review by the appropriately trained medical specialists who are qualified to interpret chest radiographs. The software does not alter the order or remove cases from the reading queue. The device's aim is to aid in the prioritization and triage of radiological medical images only.

    Chest radiographs are automatically received from the user's image storage system (e.g. Picture Archiving and Communication System (PACS)) or other radiological imaging equipment (e.g. Xray systems) and processed by DrAid™ for analysis. Following receipt of chest radiographs, the software device de-identifies a copy of each chest radiographs in DICOM format (.dcm) and automatically analyzes each image to identify features suggestive of pneumothorax. Based on the analysis result, the software notifies PACS/workstation for the presence of Pneumothorax as indicating either "flag" or "(blank)". This would allow the appropriately trained medical specialists to group suspicious exams together that may potentially benefit for their prioritization. Chest radiographs without an identified anomaly are placed in the worklist for routine review, which is the current standard of care.

    The DrAid™ device works in parallel to and in conjunction with the standard care of workflow. After a chest x-ray has been performed, a copy of the study is automatically retrieved and processed by the DrAid™ device, therefore, the analysis result can also be provided in the form of DICOM files containing information on the presence of suspicious Pneumothorax. In parallel, the algorithms produce an on-device notification indicating which cases were prioritized by DrAid™ in PACS. The on-device notification does not provide any diagnostic information and it is not intended to inform any clinical decision, prioritization, or action to who are qualified to interpret chest radiographs. It is meant as a tool to assist in improving workload prioritization of critical cases. The final diagnosis is provided by the radiologist after reviewing the scan itself.

    The following modules compose the DrAid™:

    Data input and validation: Following retrieval of a study, the validation feature assessed the input data (e.g. age, modality, view) to ensure compatibility for processing by the algorithm.

    AI algorithm: Once a study has been validated, the AI algorithm analyzes the frontal chest x-ray for detection of suspected pneumothorax.

    API Cognitive service: The study analysis and the results of a successful study analysis are provided through an API service, to then be sent to the PACS for triaging & notification.

    Error codes feature: In the case of a study failure during data validation or the analysis by the algorithm, an error is provided to the system.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study proving DrAid for Radiology v1 meets them, based on the provided text:

    1. Acceptance Criteria and Reported Device Performance

    The document does not explicitly state "acceptance criteria" as a separate, pre-defined set of thresholds that the device must meet. Instead, it compares the performance of the DrAid™ for Radiology v1 device to its predicate device (HealthPNX, K190362) to demonstrate substantial equivalence. The predicate device's performance metrics effectively serve as the implicit "acceptance criteria" for demonstrating comparable safety and effectiveness.

    Here's a table comparing the performance of DrAid™ for Radiology v1 (aggregate results) against its predicate:

    MetricsDrAid™ for Radiology v1 Performance (Mean)DrAid™ for Radiology v1 (95% CI)Predicate HealthPNX Performance (Mean)Predicate HealthPNX (95% CI)
    Sensitivity0.9461 (94.61%)[0.9216, 0.9676]93.15%[87.76%, 96.67%]
    Specificity0.9758 (97.58%)[0.9636, 0.9865]92.99%[90.19%, 95.19%]
    AUC0.9610 (96.10%)[0.9473, 0.9730]98.3%[97.40%, 99.02%]
    Timing of Notification3.83 minutesN/A22.1 secondsN/A

    The document concludes that the performance of DrAid™ for Radiology v1 is "substantially equivalent" to that of the predicate and satisfies the requirements for product code QFM. It notes that the timing performance, despite being longer, is also considered "substantially equivalent."

    2. Sample Sizes Used for the Test Set and Data Provenance

    The test set was composed of two separate datasets:

    • NIH Data Set:

      • Sample Size: 565 radiographs (386 negative, 179 positive pneumothorax cases).
      • Data Provenance: National Institute of Health (NIH), implicitly US. This dataset was used to demonstrate "generalizability of the device to the demographics of the US population."
      • Retrospective/Prospective: Not explicitly stated, but typically large public datasets like NIH are retrospective.
    • Vietnamese Data Set:

      • Sample Size: 285 radiographs (110 negative, 175 positive pneumothorax cases).
      • Data Provenance: Four Vietnamese hospitals (University Medical Center Hospital, Nam Dinh Lung Hospital, Hai Phong Lung Hospital, and Vinmec Hospital).
      • Retrospective/Prospective: Not explicitly stated, but likely retrospective from existing hospital archives.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    • Number of Experts: 3
    • Qualifications of Experts: US board-certified radiologists.

    4. Adjudication Method for the Test Set

    The adjudication method is not explicitly described beyond stating that the datasets were "truthed by a panel of 3 US board certified radiologists." This implies a consensus-based approach, but the specific dynamics (e.g., majority vote, discussion to achieve consensus, a designated tie-breaker) are not detailed. It is often referred to as an "expert consensus" ground truth.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No Multi-Reader Multi-Case (MRMC) comparative effectiveness study was mentioned. The study focused on the standalone performance of the AI algorithm. Therefore, there is no reported effect size of how much human readers improve with AI vs without AI assistance from this document.

    6. Standalone Performance Study

    Yes, a standalone performance study was done. The performance metrics (Sensitivity, Specificity, AUC) reported in the tables are for the algorithm only, without human-in-the-loop performance. This is further clarified by the section title "Performance Testing - Stand-Alone."

    7. Type of Ground Truth Used

    The ground truth used for the test sets (both NIH and Vietnamese) was expert consensus from a panel of 3 US board-certified radiologists.

    8. Sample Size for the Training Set

    The document mentions that the training data came from "a hospital system in Vietnam and the publicly available CheXpert data set." However, the specific sample size for the training set is not provided in the text.

    9. How the Ground Truth for the Training Set Was Established

    • Hospital system in Vietnam: The method for establishing ground truth for this dataset is not explicitly stated.
    • Publicly available CheXpert data set: The CheXpert dataset typically derives its labels from automated natural language processing (NLP) of radiology reports, which are then often reviewed and further annotated, though the precise methodology for this specific training use is not detailed here.

    The document emphasizes that there was no overlap between training and validation data sets, with data from different hospitals and no patient overlap confirmed.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1