Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K243710

    Validate with FDA (Live)

    Date Cleared
    2025-04-11

    (130 days)

    Product Code
    Regulation Number
    876.1500
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Disposable Ureteral Access Sheath is used to establish a continuous conduit during urological endoscopic procedures facilitating the in and out passage of endoscopes and other instruments into the urinary tract.

    Device Description

    The proposed device, Disposable Ureteral Access Sheath is used to establish a continuous conduit during urological endoscopic procedures facilitating the in and out passage of endoscopes and other instruments into the urinary tract.

    This device is divided into five types:

    • Z type Straight joint access sheath
    • Y type negative suction Y joint access sheath
    • W type negative suction W joint access sheath
    • T type Straight joint access sheath
    • C type negative suction W joint access sheath

    This product consists of access sheath and dilator tube. The access sheath tube consists of access sheath tube body and access sheath joint. The surface of access sheath tube is coated with polyvidone (PVP). The dilator tube consists of dilator tube body and dilator tube connector. The access sheath tube body is made from nylon (PA), stainless steel and polytetrafluoroethylene (PTFE). The access sheath joint is made from polycarbonate (PC), acrylonitrile-butadiene-styrene copolymer (ABS) and silicone. The dilator tube body and dilator tube connector are made from PE.

    The proposed devices are sterilized by Ethylene Oxide Gas to achieve a SAL of 10-6 and supplied sterility maintenance package which could maintain the sterility of the device during the shelf life of 3 years.

    AI/ML Overview

    Based on the provided FDA 510(k) clearance letter for the Disposable Ureteral Access Sheath (K243710), the device is cleared based on non-clinical performance and biocompatibility testing, demonstrating substantial equivalence to a predicate device. There is no information about a study involving human readers or AI assistance.

    Here's a breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a table of "acceptance criteria" alongside specific numerical "reported device performance" values for each criterion. Instead, it lists the types of performance tests conducted and states that the results "met the standard requirements" or that certain characteristics "shall not exceed" or "shall not be less than" a specified value, implying these are the acceptance criteria.

    Implied Acceptance Criteria and Performance (Based on "Comparison of Technological Characteristics" and "Non-Clinical Testing" sections):

    Acceptance Criteria (Implied)Reported Device Performance (Summary)
    Whole Device
    AppearanceMet requirements (implied by "Performance testing")
    DimensionMet requirements (implied by "Performance testing")
    Access Sheath Tube
    Liquid leakageMet requirements (implied by "Performance testing")
    Peak tensile force≥ 15 N (at tube body, and at union of tube body and joint)
    Friction coefficient≤ 0.03
    ToughnessMet requirements (implied by "Performance testing")
    Resistance To FlatteningMet requirements (implied by "Performance testing")
    Dilator Tube
    Strength of union of the dilator tube connector and dilator tubeMet requirements (implied by "Performance testing")
    Resistance to flexingMet requirements (implied by "Performance testing")
    CompatibilityMet requirements (implied by "Performance testing")
    TrafficabilityMet requirements (implied by "Performance testing")
    Dilator Tube ConnectorMet requirements (implied by "Performance testing")
    Biocompatibility
    CytotoxicityMet requirements (in accordance with ISO 10993-1)
    SensitizationMet requirements (in accordance with ISO 10993-1)
    IrritationMet requirements (in accordance with ISO 10993-1)
    Acute Systemic ToxicityMet requirements (in accordance with ISO 10993-1)
    PyrogenicityMet requirements (in accordance with ISO 10993-1)
    Sterility & Shelf-life
    Sterility Assurance Level (SAL)10^-6 (achieved via EO sterilization)
    Shelf Life3 years (maintained sterility)
    Bending ResistanceNo deformations, cracks, folding, or other defective phenomena after 5cm radius bend for 1min

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The document does not specify the exact sample sizes for each non-clinical performance test. It only lists the types of tests performed.
    • Data Provenance: The testing was conducted as part of the regulatory submission by Shenzhen Trious Medical Technology Co., Ltd. The document refers to "Performance testing" and "Biocompatibility testing," which are typically done in a controlled laboratory setting. It does not mention retrospective or prospective data in the context of human studies for this device. The manufacturer is based in Shenzhen, China.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This request is not applicable to the provided document. The "ground truth" for this device's testing is based on engineering specifications, material science, and established biological safety standards (e.g., ISO and ASTM standards), not on expert clinical interpretation of medical images or data from human patients. Therefore, no human experts (e.g., radiologists) were involved in establishing "ground truth" in the way understood in diagnostic AI/image analysis studies.

    4. Adjudication Method for the Test Set

    Not applicable. Adjudication methods are typically used in clinical studies or expert review processes where there might be disagreements in interpretation (e.g., blinded reads, consensus reads). For non-clinical device performance testing, results are typically objective measurements against predefined specifications.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No. The document explicitly states: "No clinical study is included in this submission." Therefore, no MRMC study or AI assistance evaluation was performed or submitted.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    No. This device is a physical medical instrument (ureteral access sheath), not an AI algorithm or software. Therefore, the concept of "standalone algorithm performance" is not applicable.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for this device's performance is established by:

    • Engineering Specifications: Defined physical dimensions, material properties, and mechanical strengths.
    • Standardized Test Methods: Adherence to recognized international and national standards (e.g., ISO 11135, ISO 11737-1, ISO 10993-1 for biocompatibility, various ASTM standards for material properties). These standards define how performance is measured and what constitutes acceptable limits.
    • Biocompatibility Definitions: Established biological responses to materials as defined by ISO 10993-1.

    8. The Sample Size for the Training Set

    Not applicable. This is a physical device, not an AI model or software that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for a physical device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K230748

    Validate with FDA (Live)

    Date Cleared
    2023-08-02

    (138 days)

    Product Code
    Regulation Number
    876.1500
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Disposable Ureteral Access Sheath is used to establish a conduit during endoscopic urological procedures facilitating the passage of endoscopes and other instruments into the urinary tract.

    Device Description

    The product is composed of sheath tube, sheath tube base, dilator and center base of dilator. The sheath tube base contains a backwater interface. The center base of dilator consists of upper cover of dilator base, lower cover of dilator base , fiber image channel interface F, irrigation channel interface I, equipment channel interface E, obturator, two-way water valve, needle free joint. There is no guide wire. The product is sterilized by EO, it should be sterile and only for single use.

    AI/ML Overview

    This document describes the premarket notification (510(k)) for the Disposable Ureteral Access Sheath by YouCare Technology Co., Ltd. (Wuhan). The submission aims to demonstrate substantial equivalence to a predicate device, K151084 Well Lead Ureteral Access Sheath.

    No human-in-the-loop or standalone AI performance studies were conducted or are mentioned in this submission, as this device is a physical medical instrument, not an AI/software device. The "acceptance criteria" discussed here refer to the performance standards and material safety criteria for the physical device, rather than the performance of an AI algorithm.

    1. A table of acceptance criteria and the reported device performance:

    Acceptance CriteriaReported Device Performance (YouCare Disposable Ureteral Access Sheath)Predicate Device (Well Lead Ureteral Access Sheath)Conclusion on Equivalence
    Intended UseTo establish a conduit during endoscopic urological procedures facilitating the passage of endoscopes and other instruments into the urinary tract.To establish a conduit during endoscopic urological procedures facilitating the passage of endoscopes and other instruments into the urinary tract.Same
    Regulation #21 CFR § 876.150021 CFR § 876.1500Same
    Product CodeFEDFEDSame
    Classification NameEndoscope and accessoriesEndoscope and accessoriesSame
    Regulatory ClassClass IIClass IISame
    SterilityYesYesSame
    Sterilization MethodEOEOSame
    Single UseYesYesSame
    BiocompatibleYes (Passed Cytotoxicity, Sensitization, Intracutaneous reactivity, Material mediated pyrogenicity, Acute systemic toxicity tests per ISO standards).YesSame
    Shelf Life3 years3 yearsSame
    Bending ResistanceBending any section of the dilator, sheath tube or the assembly of sheath tube and dilator into a ring with a radius of 5cm for 1min, there should be no crease, crack or other undesirable phenomena.Bending any section of the dilator, sheath tube or the assembly of sheath tube and dilator into a ring with a radius of 5cm for 1min, there should be no crease, crack or other undesirable phenomena.Same
    Coefficients of FrictionWhen the sheath tube is tested for friction, the friction coefficient shall not exceed 0.03.When the sheath tube is tested for friction, the friction coefficient shall not exceed 0.03.Same
    Peak Tensile ForceThe peak tensile force of the sheath tube and dilator should not be less than 15N. And the peak tensile force of the junction between sheath tube and sheath tube base, the dilator and center base of dilator also shall meet the requirement.The Dilator to Hub Tensile Strength > 15N.Same
    Additional Performance TestsDetermining the Dimensions, Flow test, Patency, Connection, Luer taper (Details of specific thresholds/results not provided beyond "passed").(Implied to meet similar functional performance based on equivalence claim, specific thresholds not detailed in this excerpt for predicate)(Implied)
    Sheath ID12Fr10Fr, 12Fr, 14FrDifferent 1 (Considered not to raise different questions of safety/effectiveness as physicians select based on patient condition).
    Sheath length45cm, 35cm13cm, 20cm, 28cm, 35cm, 45cm, 55cmDifferent 2 (Considered not to raise different questions of safety/effectiveness as physicians select based on patient condition).
    Primary StructureComposed of sheath tube, sheath tube base, dilator and center base of dilator. Sheath tube base contains a backwater interface. Center base of dilator consists of upper/lower cover, fiber image channel interface F, irrigation channel interface I, equipment channel interface E, obturator, two-way water valve, needle free joint.Comprised of three components: sheath, dilator and connector. Outer surface of the sheath has a hydrophilic coating.Different 3 (Considered not to raise different questions of safety/effectiveness, only increases operation requirements for doctors).
    MaterialsSheath tube: Pebax6333 SA01 MED, SUS 304, PTFE, PAM. Sheath tube base: ABS, PC. Dilator: LDPE. Center base of dilator components: ABS, PC, silica gel, POM.Sheath Tube: PTFE, Nylon (PA), stainless steel. Outer surface: Nylon (PA). Inner surface: PTFE. Dilator: Polyethylene (PE).Different 4 (Pebax6333 has properties of nylon; both passed biological tests, hence not raising new questions of safety/effectiveness).

    2. Sample sizes used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective):

    The document does not specify exact sample sizes for each non-clinical test conducted (e.g., number of devices tested for bending resistance, friction, etc.). The non-clinical tests were conducted to assess the intrinsic properties of the device, likely a representative sample of manufactured units.

    The data provenance is from YouCare Technology Co., Ltd. (Wuhan), China. The tests are non-clinical (laboratory/bench testing) and thus not applicable to retrospective or prospective human patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not applicable as the device is a physical medical instrument, and the testing involves non-clinical performance and material safety assessments, not an AI or diagnostic algorithm requiring expert "ground truth" establishment for image interpretation or similar.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    This information is not applicable for the same reasons as point 3. Testing involves objective measurements against established technical standards (e.g., tensile strength in Newtons, friction coefficients, visual inspection for defects after bending).

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This information is not applicable as the device is a physical medical instrument, not an AI/software device. No MRMC study was conducted or mentioned.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    This information is not applicable as the device is a physical medical instrument, not an AI/software device. No standalone algorithm performance was assessed.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    The "ground truth" for this device's performance is based on established engineering standards, material specifications, and regulatory guidelines for mechanical properties, material biocompatibility, sterility, and physical dimensions. For example, testing for "Peak tensile force" has a "ground truth" defined as "not less than 15N." Biocompatibility is assessed against ISO 10993 standards.

    8. The sample size for the training set:

    This information is not applicable as the device is a physical medical instrument, not an AI/software device. There is no concept of a "training set" in the context of this 510(k) submission.

    9. How the ground truth for the training set was established:
    This information is not applicable for the same reasons as point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K203165

    Validate with FDA (Live)

    Date Cleared
    2021-04-30

    (189 days)

    Product Code
    Regulation Number
    876.1500
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Ureteral Access Sheath is intended to use in urologic endoscopic procedures to facilitate the passage of endoscopes.

    Device Description

    The Disposable Ureteral Access Sheath is one single use device, which is provided after EO sterilization. The shelf life of Disposable Ureteral Access Sheath is 3(three) years. The Disposable Ureteral Access Sheath consists of two major components: sheath assembly and dilator assembly. Sheath assembly consists of sheath hub, the dilator assembly consists of dilator, dilator hub and dilator clip. And there is a hydrophilic coating on the surface of the sheath and dilator. The Disposable Ureteral Access Sheath is designed to provide the physician with reliable access to the urinary tract, the ability to inject fluids, and act as a conduit for device changes. Like all ureteral access sheath sets, Beyo's Disposable Ureteral Access Sheath also protects the ureter during device exchanges, thus helping reduce tissue trauma. Both the dilator and sheath are radiopaque and have a lubricous hydrophilic coating. To guide the access sheath into the body orifice, the dilator is advanced over up to a 0.038" guidewire. The device can be visualized under X-ray(fluoroscopy) during placement to confirm location. The device can accept other urological instruments with OD's compatible with the sheath's OD of 10,12 and 14 Fr. The device has two assemblies, sheath assembly and dilator assembly, the two assemblies are combined through dilator clip, the dilator can insert and withdrawn easily from sheath, The dilator assembly can be fixed on the sheath assembly.

    AI/ML Overview

    The provided text describes a 510(k) summary for a Disposable Ureteral Access Sheath, not a medical device that uses AI or machine learning. Therefore, the document does not contain information regarding acceptance criteria and study data for an AI/ML powered medical device.

    The information I can extract pertains to the acceptance criteria for the physical ureteral access sheath device and how its performance was tested to establish substantial equivalence to a predicate device.

    Here's a breakdown of what is available in the document, which primarily focuses on non-clinical performance testing for a physical device:


    Acceptance Criteria and Device Performance (for a physical ureteral access sheath)

    1. Table of Acceptance Criteria and Reported Device Performance:

    No.CharacteristicsAcceptance CriteriaTest MethodReported Device Performance (Implied "SE" or "Complies")
    1DimensionsDimension of ureteral access sheath shall comply with the requirements of Table 3 Dimensions.Using a universal gage.SE (Substantially Equivalent)
    2Dilator & Sheath CompatibilityThe dilator should be able to enter and withdraw smoothly and unobstructed within the sheath.Pull the dilator out of the sheath, then insert it into the sheath cavity, repeat three times, the results should meet the requirements.SE
    3Hub securityDilator hub can lock onto the sheath hub, dilator can be fixed on the sheath hub, no relative slip.The dilator hub is screwed off from the sheath hub and then inserted into the sheath in the same direction. The results should meet the requirements.SE
    4Sheath Inner Lumen PassabilityThe dilator should be able to pass through the sheath with a minimum bending radius of 55 mm.The sheath is bent under the condition that the marking length is 55 mm, the dilator pass through the sheath inner cavity, the results should meet the requirements.SE
    5Dynamic Friction ForceThe surface dynamic friction of the sheath after water contact should be <0.5N.Soaking the samples with the distilled water, use the Automatic friction tester, set the grip force 200g force, speed 200mm/min, distance 15mm. Back and forth 10 times, the results should meet the requirements.SE
    6Dilator to Hub Tensile StrengthThe Dilator to Hub Strength ≥15N.Cut out a 5cm sample from the dilator (with hub part); place the sample on the clamp of the tension machine and clamp it. Set the standard distance as 20mm and the tensile speed as 400mm/min. Run the tension machine to test and record the maximum force value when the product breaks.SE
    7Sheath to Hub Tensile StrengthThe Sheath to Hub Strength ≥15N.Cut out a 5cm sample from the sheath (with hub part); place the sample on the clamp of the tension machine and clamp it. Set the standard distance as 20mm and the tensile speed as 400mm/min. Run the tension machine to test and record the maximum force value when the product breaks.SE
    8Guidewire compatibility0.038" (0.97mm) guidewire can smoothly enter and exit the inner cavity of the dilator.Inject liquid and simulate the medical guidewire lumen to verify that it can pass smoothly.SE
    9Sheath FlexibilitySheath & Dilator should be flexible; when the tip bending 90°, the sheath and dilator can recover; when bending 10 times with a minimum bending radius of 55 mm, the sheath and dilator cannot kink or break.Fix the sheath and dilator tip, rotate the hub 90°, the sheath and dilator should not permanent deformation, after the sheath and dilator recover, bending 10 times with a minimum bending radius of 55 mm, the results should meet the requirements.SE
    10Sheath Bending ResistanceThe bending force should be more than 2.1N, less than 7.6N.Cutting 25cm samples from the distal end of sheath, set the sample into the tensile machine fixture with vertical direction, set the speed 10mm/min, fix the samples, compressed down 3cm. The results should meet the requirements.SE

    2. Sample Size and Data Provenance for Test Set:
    The document does not specify the sample size used for each non-clinical performance test (e.g., how many sheaths were tested for tensile strength). The data provenance is implied to be from the manufacturer's internal testing (Suzhou Beyo Medical Technology Co., Ltd. in China), and it is non-clinical/bench testing, not patient data (retrospective or prospective).

    3. Number of Experts and Qualifications for Ground Truth:
    Not applicable. This document pertains to physical device performance testing (bench testing), not clinical image analysis or AI model evaluation that would require expert human review for ground truth establishment.

    4. Adjudication Method for Test Set:
    Not applicable, for the same reason as above.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
    Not applicable. This is not an AI-powered device, and no human reader study was conducted or is relevant for this type of submission.

    6. Standalone (Algorithm Only) Performance:
    Not applicable. This is a physical medical device, not an algorithm.

    7. Type of Ground Truth Used:
    The "ground truth" for the physical device performance tests was established through defined engineering standards, test methods, and quantitative measurements (e.g., force, dimensions, friction). It's essentially "physical measurement ground truth" based on established engineering principles and standards (e.g., EN 1618:1997, internal requirements).

    8. Sample Size for Training Set:
    Not applicable. This is a physical device, not an AI/ML model that requires a training set.

    9. How Ground Truth for Training Set Was Established:
    Not applicable.


    Summary of Non-Applicable Sections for this document:
    It is crucial to understand that the provided text is a 510(k) summary for a physical medical device (Disposable Ureteral Access Sheath), not a software or AI/ML-powered device. Therefore, many of the requested elements pertaining to AI/ML model validation (e.g., training/test sets, human experts for ground truth, MRMC studies) are not present in this type of regulatory submission and are not relevant to demonstrating the substantial equivalence of a physical instrument. The "performance data" section in this document refers to physical, chemical, and biological (biocompatibility) testing of the device itself.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1