Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K221977

    Validate with FDA (Live)

    Date Cleared
    2022-10-20

    (107 days)

    Product Code
    Regulation Number
    878.4040
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Disposable Surgical Gowns are intended to be worn by operating room personnel during surgical procedures to protect the surgical patient and operating room personnel from the transfer of microorganisms, body fluids and particulate matter.

    In addition, this surgical gown meets the requirements of AAMI Level 3 barrier protection for a surgical gown per ANSI/AAMI PB70:2012 Liquid barrier performance and classification of protective apparel and drapes intended for use in health care facilities (AAMI PB70). This is a single use, disposable device, provided sterile. The surgical gown has six sizes - S, M, L, XL, XXL and XXXL.

    Device Description

    The Disposable surgical gown is composed of collar, body, sleeve and tie. The back is full opening, the neck and waist are laced, the sleeve are made of cotton closure by sewing, and the rest are made of heat sealing. It has been tested according to AAMI PB70:2012 and meet AAMI Level 3 barrier level protection for a surgical gown.

    AI/ML Overview

    The document details the non-clinical testing performed for the Disposable Surgical Gown (G4003) to support its substantial equivalence claim.

    1. Table of acceptance criteria and the reported device performance

    Test MethodologyPurposeAcceptance CriteriaReported Device Performance
    Tearing StrengthEvaluate the tearing strength of the test sample.≥ 10NPass: MD: average 52.03N, CD: average 29.15N
    Tensile StrengthEvaluate the breaking strength of the test sample.≥ 30NPass: MD: average 135.83N, CD: average 98.25N
    Seam StrengthEvaluate the seam strength of the test sample.≥ 30NPass: Average 45.18N
    Impact PenetrationEvaluate the water impact of the test sample.≤1.0 gPass: 0.16g max
    Hydrostatic PressureDetermine the hydrostatic pressure of the test sample.>50 cmPass: 52~98cm
    Thermal and Evaporative ResistanceEvaluate the evaporative resistance of the test sample.< 3 Pa.m²/W.Pass: 0.00542 Pa.m²/W
    LintingEvaluate the linting of the test sample.Log10 <4Pass: Average 2.2
    FlammabilityEvaluate the flammability of the test sample.Class 1Pass: Class 1
    EO/ECH residueEvaluate the level of sterilant residues.EO: < 10 ug/cm²ECH: < 10 mg/deviceEO: not detectedECH: Average 6.8 ug/device
    In Vitro Cytotoxicity (ISO 10993-5)Demonstrate the safety of the subject device.Non-cytotoxicUnder the conditions of the study, the device is non-cytotoxic.
    Irritation and Skin Sensitization (ISO 10993-10)Demonstrate the safety of the subject device (purpose not explicitly stated for irritation/sensitization but implied for biocompatibility).Non-irritatingNon-sensitizingUnder the conditions of the study, the device is non-irritating and non-sensitizing.

    2. Sample size used for the test set and the data provenance

    The document does not explicitly state the sample sizes for each specific non-clinical test. However, the tests are performed according to recognized international and national standards (e.g., ASTM, AATCC, ISO, CPSC). Data provenance is from Xiantao Topmed Nonwoven Protective Products Co., Ltd. located in Hubei Province, China, likely from prospective testing specifically conducted for this submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This section is not applicable as the document describes non-clinical performance and biocompatibility testing of a physical product (surgical gown), not a diagnostic or AI device that requires expert ground truth establishment. The "ground truth" for these tests is defined by the objective measurement criteria within the specified standards.

    4. Adjudication method for the test set

    This section is not applicable. Adjudication methods like 2+1 or 3+1 are typically used for establishing ground truth in clinical studies involving multiple readers, not for objective non-clinical performance testing.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This section is not applicable. The device is a physical surgical gown, not an AI or diagnostic tool, so no MRMC study was conducted or is relevant.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    This section is not applicable. The device is a physical surgical gown, not an algorithm or AI system.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For these non-clinical tests, the "ground truth" is established by objective measurements and results obtained from standardized testing methodologies, as defined by the various ASTM, AATCC, ISO, and CPSC standards cited (e.g., tearing strength in Newtons, impact penetration in grams, hydrostatic pressure in cm).

    8. The sample size for the training set

    This section is not applicable. The device is a physical surgical gown. There is no concept of a "training set" as there would be for an AI or machine learning model.

    9. How the ground truth for the training set was established

    This section is not applicable for the same reason as point 8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1