Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K212834

    Validate with FDA (Live)

    Date Cleared
    2021-12-30

    (114 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner. These gloves were tested for use with chemotherapy drugs, per ASTM D6978-05 Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs.

    Device Description

    The subject device is single use, disposable gloves intended for medical purposes to be worn on the examiner's hands to prevent contamination between patient and examiner. The qloves are powder-free, ambidextrous with beaded cuff, blue colored, nitrile, and tested for use with chemotherapy drugs. The gloves are offered in five sizes: extra-small, small, medium, large, and extra-large.

    AI/ML Overview

    The provided text describes the acceptance criteria and performance of a medical device, specifically "Disposable Nitrile Medical Examination Gloves (Tested for Use with Chemotherapy Drugs)" (K212834). However, it is not an AI/ML medical device, but rather a physical product.

    Therefore, many of the requested criteria for an AI/ML study (such as sample size for test set, data provenance, number of experts for ground truth, adjudication method, MRMC study, training set details, etc.) are not applicable to this document. The document details bench testing and biocompatibility assessments, which are standard for physical medical gloves.

    Below is a table summarizing the acceptance criteria and reported device performance based on the provided document, along with a clarification of why other requested information is not present for this type of device.

    1. Table of Acceptance Criteria and Reported Device Performance

    Test MethodPurposeAcceptance CriteriaReported Device PerformanceConclusion
    ASTM D6319Physical Dimensions Test
    Length (mm)XS/S: ≥220; M/L/XL: ≥230All sizes: >230Pass
    Width (mm)XS: 70±10; S: 80±10; M: 95±10; L: 110±10; XL: 120±10XS: 77-79; S: 82-86; M: 95-97; L: 104-106; XL: 115-117Pass
    Thickness (mm)Finger: ≥0.05; Palm: ≥0.05XS: Finger 0.10-0.12, Palm 0.06-0.10; S: Finger 0.09-0.12, Palm 0.07-0.09; M: Finger 0.08-0.12, Palm 0.06-0.09; L: Finger 0.10-0.13, Palm 0.06-0.09; XL: Finger 0.11-0.12, Palm 0.06-0.08Pass
    ASTM D5151Watertightness Test for Detection of HolesMeet the requirements of ASTM D5151 AQL 2.5XS: 0/125 leaks; S: 1/125 leaks; M: 0/125 leaks; L: 2/125 leaks; XL: 2/125 leaksPass
    ASTM D6124Powder ContentMeet the requirements of ASTM D6124 < 2.0mgXS: 0.01 mg/glove; S: 0.02 mg/glove; M: 0.03 mg/glove; L: 0.01 mg/glove; XL: 0.02 mg/glovePass
    ASTM D412Physical properties (Before Aging)
    Tensile Strength≥14MPaXS: 15.3-16.6; S: 15.6-16.3; M: 15.9-17.3; L: 15.3-17.2; XL: 15.3-17.1Pass
    Ultimate Elongation≥500%XS: 525-561; S: 519-565; M: 515-574; L: 521-562; XL: 523-568Pass
    ASTM D412Physical properties (After Aging)
    Tensile Strength≥14MPaXS: 15.6-16.9; S: 15.8-17.0; M: 15.2-17.7; L: 15.3-17.2; XL: 15.0-18.7Pass
    Ultimate Elongation≥400%XS: 520-564; S: 517-570; M: 511-550; L: 533-564; XL: 528-569Pass
    ASTM D6978Chemotherapy Drugs Tested with Minimum Breakthrough Detection Time(Reference values from predicate or standard expectations, not explicitly stated as "acceptance criteria" but reported performance)Carmustine (BCNU) 3.3 mg/ml: 15.2 Minutes; Cisplatin 1.0 mg/ml: > 240 Minutes; Cyclophosphamide (Cytoxan) 20.0 mg/ml: > 240 Minutes; Dacarbazine (DTIC) 10.0 mg/ml: > 240 Minutes; Doxorubicin HCl 2.0 mg/ml: > 240 Minutes; Etoposide 20.0 mg/ml: > 240 Minutes; Fluorouracil 50.0 mg/ml: > 240 Minutes; Methotrexate 25 mg/ml: > 240 Minutes; Mitomycin C 0.5 mg/ml: > 240 Minutes; Paclitaxel 6.0 mg/ml: > 240 Minutes; Thio Tepa 10.0 mg/ml: 45.8 Minutes; Vincristine Sulfate 1.0 mg/ml: > 240 MinutesAll reported as tested and comply, noting low permeation times for Carmustine and Thio-Tepa with specific warnings.
    ISO 10993-5CytotoxicityNon-cytotoxicUnder conditions of the study, did not show potential toxicity to L-929 cells.Pass
    ISO 10993-10IrritationNon-irritatingUnder the conditions of the study, not an irritant.Pass
    ISO 10993-10SensitizationNon-sensitizingUnder conditions of the study, not a sensitizer.Pass

    Clarification regarding AI/ML specific questions:

    The provided document is a 510(k) summary for a physical medical device (examination gloves), not an Artificial Intelligence/Machine Learning (AI/ML) based medical device. Therefore, the following requested information is not applicable to this submission:

    1. Sample sizes used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective): Not applicable for physical product testing. Performance testing is done via bench methods on manufactured samples.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for physical properties (e.g., tensile strength, hole detection) is established through standardized laboratory test methods, not expert consensus.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable. Testing involves measurable physical properties.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is a physical device, not an AI assistance tool.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is not an algorithm.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Ground truth for this device is based on standardized test methods and measurements (bench testing) per ASTM and ISO standards, not medical imagery or pathology.
    7. The sample size for the training set: Not applicable. The device does not involve machine learning and therefore has no "training set."
    8. How the ground truth for the training set was established: Not applicable.

    In summary, the document details the safety and performance evaluation of a medical glove against established national and international standards for physical properties and biocompatibility. It successfully demonstrated substantial equivalence to a predicate device through these non-clinical tests.

    Ask a Question

    Ask a specific question about this device

    K Number
    K211608

    Validate with FDA (Live)

    Date Cleared
    2021-08-05

    (72 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.

    These gloves were tested for use with chemotherapy drugs, as per ASTM D6978-05 (Reapproved 2019) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

    Device Description

    The subject device is single use, disposable gloves intended for medical purposes to be worn on the examiner's hands to prevent contamination between patient and examiner. The gloves are powder-free, ambidextrous with beaded cuff, blue colored, nitrile, and tested for use with chemotherapy drugs. The gloves are offered in five sizes: extra-small, small, medium, large, and extra-large.

    AI/ML Overview

    The provided document is a 510(k) summary for a medical device: "Disposable Nitrile Medical Examination Glove (Tested for Use with Chemotherapy Drugs)". It details the non-clinical testing conducted to demonstrate substantial equivalence to a predicate device.

    Here's an analysis of the acceptance criteria and study information, based on your requested format:

    Acceptance Criteria and Device Performance

    1. Table of Acceptance Criteria and Reported Device Performance

    Test MethodPurposeAcceptance CriteriaReported Device Performance
    ASTM D6319Physical Dimensions Test
    Length (mm)XS/S: ≥220; M/L/XL: ≥230All reported lengths were >230 mm (Pass)
    Width (mm)XS: 70±10; S: 80±10; M: 95±10; L: 110±10; XL: 120±10All reported widths for three lots were within specified ranges (Pass)
    Thickness (mm): Finger≥0.05All reported finger thicknesses for three lots were ≥0.08 mm (Pass)
    Thickness (mm): Palm≥0.05All reported palm thicknesses for three lots were ≥0.08 mm (Pass)
    ASTM D5151Watertightness Test for Detection of HolesMeet the requirements of ASTM D5151 AQL 2.5XS: 0/125 leaks; S: 1/125 leaks; M: 0/125 leaks; L: 0/125 leaks; XL: 0/125 leaks (Pass)
    ASTM D6124Powder ContentMeet the requirements of ASTM D6124 < 2.0mg0.10 mg (Pass)
    ASTM D412Physical properties (Before Aging)
    Tensile Strength≥14MPaLot 1: 15.4-17.1; Lot 2: 15.6-17.3; Lot 3: 15.7-16.6 (Pass)
    Ultimate Elongation≥500%Lot 1: 542-580; Lot 2: 542-570; Lot 3: 545-562 (Pass)
    ASTM D412Physical properties (After Aging)
    Tensile Strength≥14MPaLot 1: 15.6-17.0; Lot 2: 15.8-17.0; Lot 3: 15.4-19.2 (Pass)
    Ultimate Elongation≥400%Lot 1: 546-577; Lot 2: 547-572; Lot 3: 536-571 (Pass)
    ASTM D6978Chemotherapy Drugs Tested with Minimum Breakthrough Detection Time
    Carmustine (BCNU) 3.3 mg/mlNot explicitly stated as "acceptance criteria" but presented as a measured performance. (Note: Predicate had 25.3 min, Subject has 22.9 min for this drug as a point of comparison)22.9 Minutes (Warning: Do not use with Carmustine)
    Cisplatin 1.0 mg/mlNot explicitly stated as "acceptance criteria" but presented as a measured performance. (Predicate had ≥240 min)> 240 Minutes
    Cyclophosphamide (Cytoxan) 20.0 mg/mlNot explicitly stated as "acceptance criteria" but presented as a measured performance. (Predicate had ≥240 min)> 240 Minutes
    Dacarbazine (DTIC) 10.0 mg/mlNot explicitly stated as "acceptance criteria" but presented as a measured performance. (Predicate had ≥240 min)> 240 Minutes
    Doxorubicin HCl 2.0 mg/mlNot explicitly stated as "acceptance criteria" but presented as a measured performance. (Predicate had ≥240 min)> 240 Minutes
    Etoposide 20.0 mg/mlNot explicitly stated as "acceptance criteria" but presented as a measured performance. (Predicate had ≥240 min)> 240 Minutes
    Fluorouracil 50.0 mg/mlNot explicitly stated as "acceptance criteria" but presented as a measured performance. (Predicate had ≥240 min)> 240 Minutes
    Methotrexate 25 mg/mlNot explicitly stated as "acceptance criteria" but presented as a measured performance. (Predicate had ≥240 min)> 240 Minutes
    Mitomycin C 0.5 mg/mlNot explicitly stated as "acceptance criteria" but presented as a measured performance. (Predicate had ≥240 min)> 240 Minutes
    Paclitaxel 6.0 mg/mlNot explicitly stated as "acceptance criteria" but presented as a measured performance. (Predicate had ≥240 min)> 240 Minutes
    Thio Tepa 10.0 mg/mlNot explicitly stated as "acceptance criteria" but presented as a measured performance. (Note: Predicate had 43.7 min, Subject has 44.2 min for this drug)44.2 Minutes (Caution: Testing showed an average breakthrough time of 44.2 minutes with Thio-Tepa)
    Vincristine Sulfate 1.0 mg/mlNot explicitly stated as "acceptance criteria" but presented as a measured performance. (Predicate had ≥240 min)> 240 Minutes
    ISO 10993-5CytotoxicityNon-cytotoxicUnder conditions of the study, did not show potential toxicity to L-929 cells. (Pass)
    ISO 10993-10IrritationNon-irritatingUnder the conditions of the study, not an irritant. (Pass)
    ISO 10993-10SensitizationNon-sensitizingUnder conditions of the study, not a sensitizer. (Pass)

    2. Sample size used for the test set and the data provenance

    • ASTM D6319 (Physical Dimensions, Tensile Strength, Ultimate Elongation): Data is reported for "Lot 1", "Lot 2", and "Lot 3". Each lot appears to represent multiple measurements across different sizes (XS, S, M, L, XL). The exact sample size per lot or per measurement (e.g., how many gloves were measured for length, width, thickness, or subjected to tensile tests) is not explicitly stated, but it follows the standard ASTM D6319 requirements for quality assurance.
    • ASTM D5151 (Watertightness): The sample sizes are explicitly stated as 125 gloves per size: XS: 0/125 leaks, S: 1/125 leaks, M: 0/125 leaks, L: 0/125 leaks, XL: 0/125 leaks.
    • ASTM D6124 (Powder Content): The sample size for powder content is not explicitly stated.
    • ASTM D6978 (Chemotherapy Drug Permeation): The sample size for each drug test is not explicitly stated in the provided text.
    • ISO 10993-5 (Cytotoxicity), ISO 10993-10 (Irritation, Sensitization): The sample sizes used for these biocompatibility tests are not explicitly stated, but the tests were conducted "under conditions of the study" which implies adherence to the standard's methodology.
    • Data Provenance: All testing appears to be non-clinical (bench testing and laboratory biocompatibility tests). There is no information provided about patient data or country of origin for such data, as this is a device for which clinical testing was not deemed necessary. The testing was conducted in support of a device manufactured by Jiangsu Cureguard Glove Co., Ltd. in China.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. The "ground truth" in this context is established by standardized physical, chemical, and biological testing methods (ASTM and ISO standards), not by individual expert consensus or interpretation of medical images. The results are objective measurements.

    4. Adjudication method for the test set

    Not applicable. The tests are based on objective measurements against predefined acceptance criteria from international standards. There is no element of subjective assessment requiring adjudication in the context of these non-clinical tests.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a medical examination glove, not an AI-powered diagnostic or therapeutic device that would involve human readers or AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a physical medical glove and does not involve any algorithm or AI component.

    7. The type of ground truth used

    The ground truth used for performance validation is based on standardized test methods and their defined performance criteria. Specifically:

    • ASTM Standards: ASTM D6319 (physical properties), ASTM D5151 (freedom from holes), ASTM D6124 (powder content), and ASTM D6978 (chemotherapy drug permeation). These standards define the procedures for measurement and the expected acceptable ranges or thresholds.
    • ISO Standards (Biocompatibility): ISO 10993-5 (cytotoxicity) and ISO 10993-10 (irritation and sensitization). These standards define the methods and parameters for assessing the biological safety of the device.

    8. The sample size for the training set

    Not applicable. This is a physical medical device and does not involve AI or machine learning models that require a "training set."

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1