Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K231633

    Validate with FDA (Live)

    Date Cleared
    2024-01-11

    (220 days)

    Product Code
    Regulation Number
    876.4400
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Disposable Hemoclip is indicated for clip placement within the gastrointestinal (GI) tract for the purpose of:

    1. Endoscopic marking
    2. Hemostasis for:
      • Mucosal/sub-mucosal defects < 3 cm
      • Bleeding ulcers
      • Arteries < 2 mm
      • Polyps < 1.5 cm in diameter
      • Diverticula in the colon
    3. Anchoring to affix jejunal feeding tubes to the wall of the small bowel;
    4. As a supplemental closure method of luminal perforations < 20 mm that can be treated conservatively.
    Device Description

    The Disposable Hemoclip is the key and common accessory in the diagnosis and treatment procedure of digestive endoscope. As a single use instrument, the disposable hemoclip does not allow to be used by twice, and cannot be separated by components except to be used as well. It is also a sterile device consisting of a pre-loaded, radiopaque, single-use, endoscopic clipping device consisting of two main components: the delivery system and the clip. The delivery system consists of a handle and tube. The delivery system is constructed using stainless steel and polyester materials. The delivery system will allow for the device to rotate at the distal end. The delivery system is offered in 1600mm, 1800mm, 2100mm and 2300mm working lengths.
    The clip is made of stainless-steel material and deployed from the delivery system during use. The clip jaws are designed that they can be opened and closed up to five times prior to deployment, aiding in repositioning of the clip at the lesion site. Re-opening, closing, and rotation capability may be limited by clinical circumstances and patient anatomy.

    AI/ML Overview

    The provided text is a 510(k) Summary for a medical device called "Disposable Hemoclip (AF series)". It describes the device, its intended use, a comparison with a predicate device, and the non-clinical data used to demonstrate substantial equivalence.

    However, the document does not contain information about:

    • A table of acceptance criteria and the reported device performance in the sense of predefined thresholds for clinical efficacy or accuracy. Instead, it lists various non-clinical tests performed.
    • Sample sizes used for a "test set" in a clinical study because no clinical studies were performed.
    • Data provenance, number of experts, adjudication methods, or MRMC studies, as these relate to clinical data which is explicitly stated as N/A.
    • Standalone algorithm performance or details about ground truth for training data, as this is a physical medical device, not an AI/software device.

    Therefore, many of the requested items cannot be provided from the given input.

    Here's a breakdown of what can be extracted based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not present "acceptance criteria" in the format of specific thresholds for clinical performance (e.g., sensitivity, specificity, accuracy) because it relies on non-clinical testing for substantial equivalence. Instead, it lists various non-clinical tests performed to demonstrate that the subject device performs similarly to the predicate device. The "reported device performance" is essentially that the device "met all design specifications" and that "test results demonstrate substantial equivalence between the subject device and the predicate device" for each test.

    Test ItemAcceptance Criteria (Implied)Reported Device Performance
    BiocompatibilityCompliance with FDA guidance "Use of International Standard ISO 10993-1, 'Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process'" September 4, 2020, and ISO 10993-1. Specific tests for cytotoxicity, sensitization, systemic toxicity (acute, subacute, subchronic), pyrogenicity, genotoxicity, and local effects after implantation.All specified tests (MTT Cytotoxicity, Skin Sensitization, Intracutaneous, Acute Systemic Toxicity, Subacute Systemic Toxicity, Subchronic Systemic Toxicity, Pyrogen, Genotoxicity, Muscle Implant) were considered and passed according to the relevant ISO 10993 standards.
    Sterilization ValidationCompliance with ISO 11135:2014+A1:2018 (EO sterilization), ISO 11737-2:2019 (sterility tests), ISO 10993-7:2008 (EO residuals), and USP <85> (bacterial endotoxins).Validated according to applicable standards.
    Shelf Life & Sterile BarrierCompliance with ASTM F1980-16 (accelerated aging), ISO11607-1:2019 (materials/systems), ISO11607-2:2019 (validation of process), ASTM F 1929-15 (dye penetration), ASTM D 3078-02(2013) (bubble emission), DIN 58593-6: 2016 (microbial barrier), ASTM F88/F88M-15 (seal strength), ASTM D4169-16 (shipping performance).Validated according to applicable standards.
    Mechanical PerformancePerformance consistent with the specific product technical specification and demonstrating substantial equivalence to the predicate device.Tests for Surface roughness/hardness, Clip Assembly Repeated Open/Close, Clip Retention Force, Clip Releasing Force, 360° Rotation, Clip Close Force, Clamp strength-durability, Clamp strength-stability, Separation Force, Clip approach, Metal Part Against Corrosion were performed. The test results "demonstrate substantial equivalence between the subject device and the predicate device."

    2. Sample size used for the test set and the data provenance

    • Test Set Sample Size: Not applicable. The submission relies on non-clinical (bench) testing, not clinical studies with patients or human data to form a "test set" in the context of clinical performance.
    • Data Provenance: The non-clinical tests were conducted internally or by contract labs following recognized standards.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not applicable. No clinical studies were conducted, and therefore no ground truth was established by medical experts for a test set.

    4. Adjudication method for the test set

    • Not applicable. No clinical studies requiring adjudication were performed.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This is a physical medical device (hemoclip), not an AI/software device, and no MRMC studies were performed.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This is a physical medical device, not an algorithm.

    7. The type of ground truth used

    • The "ground truth" for the non-clinical tests was established by adhering to recognized international and FDA standards (e.g., ISO, ASTM, USP) for biocompatibility, sterilization, shelf life, and mechanical properties. The performance of the predicate device also served as a benchmark for demonstrating substantial equivalence.

    8. The sample size for the training set

    • Not applicable. This is a physical medical device, not an AI/software device that requires a training set.

    9. How the ground truth for the training set was established

    • Not applicable, for the same reason as above.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1