510(k) Data Aggregation

    K Number
    K232442

    Validate with FDA (Live)

    Date Cleared
    2023-11-09

    (87 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Disposable Endoscope Injection Needles are used with endoscope to perform endoscopic vascular or submucosal injection in the GI tract.

    Device Description

    The Disposable Endoscope Injection Needles consist of a Sliding handle, Handle, Cap, Sheath, Inner Tube, Guide Tube, Connecting tube, and Tip (only Type B). It is available in a variety of configurations with varying needle lengths, gauges and working lengths. Type A and Type B are included in this submission.

    AI/ML Overview

    The provided text is a 510(k) summary for the Disposable Endoscope Injection Needles. It does not contain information about the acceptance criteria or a study proving that the device meets those criteria in the context of an AI/ML medical device.

    The document primarily focuses on establishing substantial equivalence to a predicate device based on:

    • Device Description: What the device is made of and its configurations.
    • Intended Use: How the device is meant to be used.
    • Comparison of Technological Characteristics: A table comparing features of the proposed device to a predicate device.
    • Non-clinical Performance Data: Mentions conformity to ISO standards for biocompatibility and sterilization, and some bench tests (appearance, physical properties) that "passed."
    • Clinical Test Data: Explicitly states, "No Clinical Study is included in this submission."

    Therefore, I cannot provide the requested information regarding acceptance criteria, reported performance, sample sizes, expert qualifications, adjudication methods, MRMC studies, standalone performance, or ground truth details because these aspects are relevant to AI/ML device evaluations and are not present in this 510(k) for a conventional medical device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K220292

    Validate with FDA (Live)

    Date Cleared
    2022-08-19

    (199 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Disposable Endoscope Injection Needle is used for endoscopic injection into gastrointestinal mucosa and submucosa to:
    • introduce a sclerosing agent, vasoconstrictor, or other solutions into selected sites to control actual or potential bleeding lesions in the digestive system;
    • aid in Endoscopic Mucosal Resection (EMR), Endoscopic Submucosal Dissection (ESD), or polypectomy procedures;
    • control non-variceal hemorrhage.

    Device Description

    The Disposable Endoscope Injection Needle consists of Sheath, Handle buckle, Handle, Booster tube, Handle end cap, Outer tube, Inner tube, Metal end cap and injection needle tip. It is available in a variety of configurations with varying needle lengths, gauges and working lengths.

    AI/ML Overview

    The provided text is a 510(k) summary for a medical device called the "Disposable Endoscope Injection Needle." It describes the device, its intended use, comparison to a predicate device, and non-clinical performance data. However, it explicitly states that "No Clinical Study is included in this submission."

    Therefore, I cannot provide information about acceptance criteria and a study that proves the device meets clinical performance acceptance criteria, as no such clinical study was conducted or presented in this document. The submission relies solely on non-clinical (bench) testing and substantial equivalence to a predicate device.

    Here's what I can extract regarding acceptance criteria and performance based only on the non-clinical data provided:

    1. A table of acceptance criteria and the reported device performance:

    TestAcceptance Criteria (Implied by "passing")Reported Device Performance
    Appearance(Implied: Meets visual/design spec)Passing
    Dimension(Implied: Meets dimensional spec)Passing
    Performance(Implied: Meets functional spec)Passing
    Connection strength(Implied: Meets strength spec)Passing
    Leakage(Implied: No leakage)Passing
    Cleanliness(Implied: Meets cleanliness spec)Passing
    Compatibility(Implied: Compatible with endoscopy)Passing
    Puncture force(Implied: Meets force spec)Passing
    Corrosion resistance(Implied: Resists corrosion)Passing
    Patency of lumen(Implied: Lumen is clear/unobstructed)Passing
    Durability(Implied: Durable for intended use)Passing
    Stiffness test of needle(Implied: Meets stiffness spec)Passing
    Resistance of needle to breakage(Implied: Resists breakage)Passing
    Positive pressure liquid leakage(Implied: No leakage under pressure)Passing
    Sub-atmospheric pressure air Leakage(Implied: No air leakage under vacuum)Passing
    Stress cracking(Implied: Resists stress cracking)Passing
    Resistance to separation from axial load(Implied: Resists axial separation)Passing
    Resistance to overriding(Implied: Resists overriding)Passing
    Resistance to separation from unscrewing(Implied: Resists unscrewing separation)Passing
    Flow rate(Implied: Meets flow rate spec)Passing
    BiocompatibilityComply with ISO 10993-1Comply with ISO 10993-1
    Sterilization (Ethylene Oxide)Comply with ISO 11135-1, ISO 10993-7Comply with ISO 11135-1, ISO 10993-7

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):
    The document does not specify the sample sizes used for the non-clinical (bench) tests. The data provenance is implied to be from Yangzhou Fartley Medical Instrument Technology Co., Ltd. in China, where the device is manufactured. These are all prospective bench tests.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
    Not applicable. These were non-clinical, objective bench tests, not clinical evaluations requiring expert interpretation of results or establishment of ground truth in a medical context.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
    Not applicable, as these were objective bench tests, not subjective interpretations requiring adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
    Not applicable. This device is a physical medical instrument (an injection needle), not an AI-powered diagnostic or assistive technology. No MRMC study was conducted.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
    Not applicable. This is a physical medical device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):
    For the non-clinical tests, the "ground truth" was defined by engineering specifications, industrial standards (like ISO 10993, ISO 11135), and the physical properties required for safe and effective function of such a device.

    8. The sample size for the training set:
    Not applicable. This is a physical medical device, not a machine learning model requiring a training set.

    9. How the ground truth for the training set was established:
    Not applicable.