Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K232529

    Validate with FDA (Live)

    Date Cleared
    2024-05-13

    (266 days)

    Product Code
    Regulation Number
    868.5740
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Disposable Double Lumen Endobronchial Tube is used to isolate the left or right lung of a patient for surgery, one lung ventilation or one lung anesthesia.

    It is indicated for patients with pathological lung conditions or other medical conditions that require endobronchial intubation, mechanical ventilation and isolation of one lung from the other, e.g. for thoracic surgery.

    The Disposable Double Lumen Endobronchial Tube is indicated for adults use only.

    Device Description

    The Disposable Double Lumen Endobronchial Tube is a sterile, single patient use PVC Double-Lumen Endobronchial Tube (also referred to as a DLT) that is inserted into the trachea via the mouth in order to selectively ventilate one lung.

    The Disposable Double Lumen Endobronchial Tube is consisted with double lumen tube with the body including 2 cuffs (Bronchial Cuff and Tracheal Cuff).

    The tracheal cuff provides sealing against tracheal wall and the corresponding pilot balloon indicates state of cuff inflation/deflation. Similarly, the bronchial cuff provides sealing against bronchial wall and the corresponding pilot balloon indicates state of cuff inflation/deflation. It is fitted with a stylet to enable shaping of the tube for navigation during intubation. When the suction is necessary, this device can be possible to be assemble with suction catheter which is 8Fr or 10Fr with PVC and this device is packaged with two suction catheters.

    The Disposable Double Lumen Endobronchial Tube comes in three models, each model include two types: Ordinary Left Type, Ordinary Right Type, Visible Left Type B, Visible Right Type B, Visible Left Type, Visible Right Type. Each type will be available in six sizes: 28 Fr, 32 Fr, 35 Fr, 37 Fr, 39 Fr and 41 Fr.

    The device is composed of biologically safe materials. It is indicated for patients with pathological lung conditions or other medical conditions that require endobronchial intubation, mechanical ventilation and isolation of one lung from the other, e.g. for thoracic surgery. The Disposable Double Lumen Endobronchial Tube is indicated for adults use only. Environments of use: Hospital-OR and ICU.

    AI/ML Overview

    The provided text is a 510(k) summary for a medical device (Disposable Double Lumen Endobronchial Tube) and focuses on non-clinical performance testing and biocompatibility. It does not describe a study involving an AI algorithm or human readers.

    Therefore, many of the requested details about acceptance criteria and studies (e.g., sample size for test set, data provenance, number of experts, adjudication method, MRMC study, standalone performance, ground truth establishment for training set, effect size of AI assistance) are not applicable to this document.

    The document primarily describes non-clinical performance testing to demonstrate functional and material equivalence to a predicate device.

    Here's an analysis based on the information provided:


    Acceptance Criteria and Device Performance (Non-Clinical)

    The submission provides a table of non-clinical performance test items, their reference standards, acceptance criteria, and reported test results.

    1. Table of Acceptance Criteria and Reported Device Performance:

    No.Performance Test ItemsReference StandardsAcceptance CriteriaTest Result
    1DimensionsISO 5361:2016 / ISO 16628:2008When tested in accordance with the test method, the L1, L2, L3, OD, ID shall meet the requirements on size and tolerances.Pass
    2Inflating tube O.D. & AngleISO 5361:2016Inflating tube O.D.≤3.0mm; The angle between the inflating tube and the Endobronchial Tube at the point of separation shall not exceed 45°.Pass
    3Cuff seal (Sealing of cuff inflating system)/When tested in accordance with the test method, no bubble shall be noted over the 10-s interval.Pass
    4Cuffed tube collapseISO 5361:2016When tested in accordance with the test method in ISO 5361 Annex C, the steel ball shall pass freely through the lumen of the tube.Pass
    5Cuff herniationISO 5361:2016When tested in accordance with the test method in ISO 5361 Annex D, no part of the inflated cuff shall reach beyond the nearest edge of the bevel.Pass
    6Cuff Resting DiameterISO 5361:2016When tested in accordance with the test method in ISO 5361 Annex B, at a pressure of 2 kPa, the cuff diameter shall meet the requirements on size and tolerances.Pass
    7Air leakage of one way valve/When tested according to the test method, the pilot balloon should be able to naturally fill up and flat deflate.Pass
    8Liquid leakage of one way valveISO 80369-7:2021When tested in accordance with the test method, there shall be no leakage sufficient to form a falling drop of water.Pass
    9Size of the Murphy EyeISO 5361:2016The area of the Murphy eye shall be not less than 80 % of the cross-sectional area derived from the minimum inside diameter for that size tube.Pass
    10Security of Construction of connection components/When tested in accordance with the test method, the force required to detach any component permanently attached to the shaft shall be not less than that specified in standard.Pass
    1115mm connectorISO5356-1:2004When tested in accordance with the test method, 15mm conical connectors shall comply with the ISO5356-1.Pass
    12Colour CodingISO16628:2008When tested in accordance with the test method, the color coding of bronchial cuff and its pilot balloon were entirely coloured blue.Pass
    13Segment differentiation of the tracheal segment and the bronchial segmentsISO16628:2008When tested in accordance with the test method, the segment differentiation of the tracheal segment and the bronchial segments were clearly distinguished.Pass
    14Test for balloon diameter to inflation pressureISO10555-4:2013When tested in accordance with the test method, can reflect the relationship between balloon diameter and inflation pressure.Pass
    15Cuff thickness/When tested in accordance with the test method, the cuff thickness shall meet the requirements on size and tolerances.Pass
    15Cuff burst pressure/When tested in accordance with the test method, the burst pressure of the cuff shall meet the RBP 30kPa.Pass

    Biocompatibility Testing:

    The biocompatibility evaluation was conducted based on the device's classification as a surface device in contact with mucosal membrane for limited exposure (<24 hours) and as a breathing gas pathway device.

    No.Test Items (General Biocompatibility)Standards
    1CytotoxicityISO 10993-5:2009
    2Intracutaneous testISO 10993-10:2010
    3Skin SensitizationISO 10993-10:2010
    No.Test Items (Biocompatibility of Breathing Gas Pathways)Standards
    1Particulate Matter EmissionsISO 18562-2:2017
    2Volatile Organic Compounds (VOCs) EmissionsISO 18562-3:2017
    3Leachable in CondensateISO 18562-4:2017
    4Biological Risk Evaluation ReportISO 18562-1:2017

    All biocompatibility evaluation acceptance criteria were met.


    Regarding the other requested information, which pertains to AI/algorithm-based medical devices, the provided document explicitly states that it is not applicable:

    2. Sample size used for the test set and the data provenance: Not applicable. This submission is for a physical medical device, not an AI/algorithm. Performance testing was non-clinical (laboratory tests) rather than on a clinical test set.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for clinical data is not a concept relevant to the testing performed for this device.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is not an AI-assisted device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc): Not applicable. "Ground truth" in the context of clinical data/AI is not relevant here. The "ground truth" for the non-clinical tests would be the established engineering/physical principles and standards (e.g., ISO standards defining dimensions, pressures, etc.).

    8. The sample size for the training set: Not applicable. This is not an AI/machine learning device that requires a training set.

    9. How the ground truth for the training set was established: Not applicable.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1