Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Diode Laser Hair Removal Device (Model: EVOLUTION MEDICAL, M-I-X MEDICAL, GENESIS, Lotus and Ultimate) is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin.

    Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.

    Device Description

    The Diode Laser Hair Removal Device is a surgical device intended for use in dermatologic and general surgical procedure. It utilizes a diode laser as a laser source (808 nm). The laser power is delivered to the treatment area via a laser handpiece. The emission laser is activated by a foot switch and a laser handpiece. There are five models included, EVOLUTION MEDICAL, M-I-X MEDICAL, GENESIS, Lotus, and Ultimate, the five models have same mechanism of action, principle and specification, with only one difference: the adjustable range of Energy density is different, EVOLUTION MEDICAL (1-77 J/cm^2), M-I-X MEDICAL (1-70 J/cm^2), GENESIS (1-74 J/cm^2), Lotus(1-67 J/cm^2), and Ultimate(1-65 J/ cm^2).

    Changes to the device include the addition of a remote link to the device to obtain and back up the device's usage data, a remote upgrade function, and the ability to provide a URL for users to download the device's usage data, as compared to previously listed devices.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a Diode Laser Hair Removal Device. This document focuses on establishing substantial equivalence to a predicate device rather than detailing a clinical study with acceptance criteria for a new device's performance.

    Therefore, the sections below requiring information about acceptance criteria and a study proving the device meets them cannot be fully populated as they would for a de novo or PMA submission. The document primarily discusses non-clinical tests to verify design specifications and compliance with standards.

    Here's the information that can be extracted or inferred from the provided text, with clarifications where details are not present:


    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document does not contain a table of explicit acceptance criteria with corresponding device performance metrics for efficacy (e.g., specific hair reduction percentages) as would be found in a clinical study report for a new device.

    Instead, the document focuses on demonstrating that the device meets safety and performance standards and is technologically equivalent to its predicate. The "performance" discussed primarily relates to meeting design specifications and regulatory standards.

    Acceptance Criteria (Inferred from regulatory requirements and predicate comparison)Reported Device Performance
    Safety and Essential Performance (Electrical)Complies with ANS//AAM ES60601-1:2005/(R)2012 & A1:2012 C1:2009/(R)2012 & A2:2010/(R)2012 (Cons. Text) [Incl. AMD2:2021] (Consolidated Text)
    Electromagnetic DisturbancesComplies with IEC 60601-1-2 Edition 4.1 2020-09 CONSOLIDATED VERSION
    Safety and Performance (Laser Equipment)Complies with IEC 60601-2-22 Edition 3.1 2012-10
    Laser Product Safety and ClassificationComplies with IEC 60825-1:2014
    Biocompatibility (Cytotoxicity)Complies with ISO 10993-5 Third Edition 2009-06-01
    Biocompatibility (Irritation and Skin Sensitization)Complies with ISO 10993-10 Third Edition 2010-08-01
    Wireless CoexistenceComplies with IEEE ANSI USEMCSC C63.27-2021 and AAM TIR69:2017/(R2020)
    CybersecurityCybersecurity Testing conducted per FDA Guidance "Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions" (September 27, 2023)
    Energy Output AccuracyPerformance Testing for Energy Output Accuracy was conducted. (Specific metrics not provided in this summary.)
    Software Verification and ValidationTesting conducted per FDA Guidance "Content of Premarket Submissions for Device Software Functions" (June 14, 2023). Level of concern determined to be Moderate.
    Hair Removal and Permanent Hair ReductionDevice is intended for hair removal and permanent hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin, with definition of permanent hair reduction provided in the indications for use. (No direct clinical performance metrics are reported in this summary document, as this is a 510(k) based on substantial equivalence.)

    2. Sample Size Used for the Test Set and Data Provenance

    The document primarily refers to non-clinical tests for compliance with standards and design specifications. It does not describe a clinical test set with human subjects.

    • Test Set Sample Size: Not applicable as a clinical test set is not described. The tests are for device components and system performance.
    • Data Provenance: The tests are non-clinical, related to device manufacturing and engineering compliance. No country of origin for clinical data is applicable. The tests are prospective in the sense that they are conducted on the new device to demonstrate compliance.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not applicable as the document describes non-clinical engineering and compliance testing, not a clinical study requiring expert ground truth assessment.

    4. Adjudication Method for the Test Set

    Not applicable, as this refers to a clinical study with human observers or interpreters, which is not described here.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a laser hair removal system, not an AI-powered diagnostic or interpretive tool that would involve human readers or AI assistance in a diagnostic context.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. This device is a physical laser system, not an algorithm, and the concept of "standalone performance" for an algorithm is not relevant here.

    7. The Type of Ground Truth Used

    For the non-clinical tests, the "ground truth" is defined by the requirements outlined in the cited international and national standards (e.g., IEC 60601-1, ISO 10993) and the device's design specifications. These standards provide benchmarks and methodologies against which the device's performance is measured.

    8. The Sample Size for the Training Set

    Not applicable. As this document details hardware medical devices, there is no "training set" in the context of machine learning or AI.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K221312

    Validate with FDA (Live)

    Date Cleared
    2022-06-29

    (55 days)

    Product Code
    Regulation Number
    878.4810
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Diode Laser Hair Removal Device (Model: EVOLUTION MEDICAL, M-I-X MEDICAL) is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.

    Device Description

    The Diode Laser Hair Removal Device is a surgical device intended for use in dermatologic and general surgical procedure. It utilizes a diode laser as a laser source (808 nm). The laser power is delivered to the treatment area via a laser handpiece. The emission laser is activated by a foot switch and a laser handpiece. There are two models included, EVOLUTION MEDICAL, M-I-X MEDICAL, the two models have same mechanism of action, principle and specification, with only one difference: the adjustable range of Energy density is different, EVOLUTION MEDICAL (1-77 J/cm²), M-I-X MEDICAL (1-70 J/cm²).

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a Diode Laser Hair Removal Device, which means the submission is primarily focused on demonstrating substantial equivalence to a legally marketed predicate device rather than proving novel effectiveness. Therefore, the information typically associated with complex clinical studies and detailed acceptance criteria for AI/Software as a Medical Device (SaMD) is not present in this document.

    However, I can extract the non-clinical acceptance criteria and the studies that verify them, as outlined in the document. The device in question is a physical laser device, not an AI-driven one, so many of the requested categories (like MRMC studies, ground truth for training data, etc.) are not applicable here.

    Here's a breakdown of the available information regarding acceptance criteria and the studies that prove the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance (Non-Clinical)

    Acceptance Criteria CategorySpecific Standard/RequirementReported Device Performance
    Electrical SafetyANSI/AAMI ES60601-1:2005/(R) 2012 and A1:2012, C1:2009/(R)2012 and ANSI/AAMI A2:2010/(R)2012 (Consolidated Text)"Comply with ANSI/AAMI ES60601-1, IEC 60601-2-22" (The document states compliance with the Electrical Safety standard for the proposed device, aligning with the predicate device's compliance.)
    Electromagnetic Compatibility (EMC)IEC 60601-1-2:2014"Comply with IEC 60601-1-2" (The document states compliance with the EMC standard for the proposed device, aligning with the predicate device's compliance.)
    Laser SafetyIEC 60601-2-22 Edition 3.1 2012-10, IEC 60825-1:2014"Comply with IEC 60601-2-22, IEC 60825" (The document states compliance with Laser Safety standards for the proposed device, aligning with the predicate device's compliance.)
    Biocompatibility - CytotoxicityISO 10993-5 Third Edition 2009-06-01"No Cytotoxicity" (The document states no cytotoxicity for the proposed device, aligning with the predicate device's performance.)
    Biocompatibility - SensitizationISO 10993-10 Third Edition 2010-08-01"No evidence of sensitization" (The document states no evidence of sensitization for the proposed device, aligning with the predicate device's performance.)
    Biocompatibility - IrritationISO 10993-10 Third Edition 2010-08-01"No evidence of irritation" (The document states no evidence of irritation for the proposed device, aligning with the predicate device's performance.)
    Energy Output Accuracy(Implicitly, to meet design specifications)"Performance Testing for Energy Output Accuracy." (Test was conducted, and implicitly, the device met the required accuracy for its specified fluence ranges: EVOLUTION MEDICAL: 1-77 J/cm²; M-I-X MEDICAL: 1-70 J/cm²). The comparison table highlights that the proposed device's fluence range is 'Similar' to the predicate, and analysis states these differences "will not affect the safety and effectiveness".
    Software Verification & Validation"Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices""Software Verification and Validation Testing was conducted... and the level of concern was determined to be Moderate for the proposed device." (Implies the software meets the necessary requirements for safety and functionality.)

    2. Sample size used for the test set and the data provenance

    This information is not applicable as the document describes a physical laser device, not a diagnostic or AI device with a "test set" in the typical sense of evaluating performance on a dataset of images or clinical cases. The "tests" mentioned are primarily engineering safety and performance tests on the device itself and its components.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not applicable. Ground truth establishment by clinical experts is relevant for diagnostic or AI systems; for this physical device, compliance is measured against engineering and biocompatibility standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable. Adjudication methods are used in clinical studies or AI evaluations to establish a consensus ground truth from multiple readers/experts. This is not reported for the non-clinical tests of this device.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable. This describes a physical laser hair removal device, not an AI software intended to assist human readers. No MRMC study was performed or is relevant to this type of device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not applicable. The device is a physical laser system, not a standalone algorithm. Its performance is evaluated through its physical output and safety features.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For the non-clinical tests, the "ground truth" is defined by the established international standards (e.g., IEC 60601-1, ISO 10993-5) and the design specifications for parameters like energy output accuracy. For biocompatibility, the ground truth is the chemical and biological response observed in standardized tests on the material.

    8. The sample size for the training set

    This information is not applicable. This is not an AI/machine learning device that involves a "training set."

    9. How the ground truth for the training set was established

    This information is not applicable. There is no "training set" for this physical device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K180353

    Validate with FDA (Live)

    Date Cleared
    2018-04-17

    (68 days)

    Product Code
    Regulation Number
    878.4810
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Diode Laser Hair Removal device is intended for hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.

    Device Description

    The proposed device, Diode Laser Hair Removal device, is a surgical device, which is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI);

    Function module description
    a. Control Panel
    The module uses the microcontroller as the heart, utilizes the LCD screen to display all prompt information and the system state information to complete the human-machine interaction function, and realizes the device parameters settings and accurate control of the output laser energy by the operator.
    b. Main Control Module
    The module uses the microcontroller as the heart, receives the laser energy parameters and work instructions from the control panel and detects the footswitch state; Utilizes the sensors of temperature, humidity, liquid level and flow to detect the parameters such as temperature, humidity and water flow during system working, and according to the detected values to calculate the dew point temperature; Controls and detects the work state of constant current board module as well as the temperature and humidity control system; Uploads the state data and alarm information of water circulation system, cooling system, handpiece module and constant current board module during system working.
    c. Constant current board module
    The module uses the high-power MOS as the heart, receives the laser energy parameters from the main control module, supplies the semiconductor laser with constant drive current which corresponding to the received laser energy parameters to drive the semiconductor laser to emit light. The module also has the detection function of over-current, overvoltage, over-temperature and handpiece state, and uploads the detected data to the control module.
    d. Temperature and humidity control system
    The system mainly includes the condenser, cold plate, water circulation subsystem and fans. The microcontroller of main control module according to the temperature, humidity parameter detected by the sensors to control the working state of the condenser, cold plate and cooling fan to meet the temperature and humidity requirements during the semiconductor laser working.
    e. Handpiece module
    Handpiece module is the heart of the device, which is the execution unit of the device and completes the laser emission function. The module is mainly composed of semiconductor laser, sapphire, temperature and humidity sensor, data storage chips, cooling components and water flow path. The semiconductor laser emits light to output energy, temperature and humidity sensors detects the temperature and humidity parameters during handpiece working, the cooling components and water flow path take away the heat of the semiconductor laser to prevent it from being damaged caused by over-temperature, so prolongs the service life of the semiconductor laser.

    AI/ML Overview

    The provided document is a 510(k) summary for a Diode Laser Hair Removal device. It does not contain information about the acceptance criteria or a study proving the device meets those criteria in the context of an AI-powered device. This summary is focused on demonstrating substantial equivalence to a predicate device, which is a different regulatory pathway than demonstrating the performance of a novel AI/ML-based device.

    Specifically, the document states: "No clinical study is included in this submission." This means there is no study described that would involve test sets, ground truth, expert adjudication, or MRMC studies that are typically associated with evaluating AI-powered devices.

    However, I can extract the information provided about the device's performance based on non-clinical testing for substantial equivalence:

    1. A table of acceptance criteria and the reported device performance:

    The document doesn't explicitly state "acceptance criteria" in the sense of predefined thresholds for a novel AI device's performance. Instead, it presents a comparison to a predicate device to demonstrate "substantial equivalence." The "acceptance criteria" here are implicitly that the proposed device performs comparably or within acceptable deviations from the predicate device based on non-clinical tests and safety standards.

    ItemAcceptance Criteria (Predicate Device)Reported Device Performance (Proposed Device)Remark
    Product CodeGEXGEXSE (Substantially Equivalent)
    Regulation Number21 CFR 878.481021 CFR 878.4810SE
    Intended UseHair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin. Permanent hair reduction defined as long-term, stable reduction in number of hairs regrowing when measured at 6, 9, and 12 months after a treatment regime.Hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin. Permanent hair reduction defined as long-term, stable reduction in number of hairs regrowing when measured at 6, 9, and 12 months after a treatment regime.SE
    ConfigurationMain Unit, Handpiece, Foot ControlMain Unit, Handpiece, Foot ControlSE
    Principle of OperationDiode LaserDiode LaserSE
    Laser TypeDiode LaserDiode LaserSE
    Laser ClassificationClass IVClass IVSE
    Laser Wavelength808 nm808 nmSE
    Spot Size1.44 cm21.44 cm2SE
    Fluence1-120 J/cm21-100 J/cm2Discussion (Difference noted, but considered SE based on non-clinical tests)
    Irradiance347.8 W/cm214-360 W/cm2Discussion (Difference noted, but considered SE based on non-clinical tests)
    Frequency0.5-15 Hz1-20 HzDiscussion (Difference noted, but considered SE based on non-clinical tests)
    Pulse Duration5-400 ms10-400 msDiscussion (Difference noted, but considered SE based on non-clinical tests)
    Power SupplyAC 110V/60HzAC 110V-230V/50-60Hz 2000VASE
    Dimension450mm× 550mm×380mm560mmx380mmx1180mmDiscussion (Difference noted, but considered SE based on non-clinical tests)
    Weight52 Kg60 KgDiscussion (Difference noted, but considered SE based on non-clinical tests)
    Patient Contact Materials and BiocompatibilityNo Cytotoxicity, No evidence of sensitization, No evidence of irritationNo Cytotoxicity, No evidence of sensitization, No evidence of irritationSE
    Electrical SafetyComply with IEC 60601-1, IEC 60601-2-22Comply with IEC 60601-1, IEC 60601-2-22SE
    EMCComply with IEC 60601-1-2Comply with IEC 60601-1-2SE
    Laser SafetyComply with IEC 60601-2-22, IEC 60825Comply with IEC 60601-2-22, IEC 60825SE

    2. Sample size used for the test set and the data provenance:
    Not applicable. The document states "No clinical study is included in this submission." The testing described is non-clinical performance and safety testing directly on the device components or system, not on patient data or clinical samples.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
    Not applicable. See point 2.

    4. Adjudication method for the test set:
    Not applicable. See point 2.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
    Not applicable. This device is a laser hair removal system, not an AI-assisted diagnostic or imaging device. No MRMC study was performed.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
    Not applicable. This is not an AI algorithm. Its performance is evaluated through non-clinical testing against established safety and performance standards relevant to a laser surgical instrument, and by comparison to a predicate device.

    7. The type of ground truth used:
    For non-clinical tests, the "ground truth" would be the established engineering specifications for the device and compliance with recognized international standards (e.g., IEC standards for electrical, laser, and EMC safety, ISO standards for biocompatibility). For performance aspects like spot size and energy output accuracy, the ground truth is derived from validated measurement techniques.

    8. The sample size for the training set:
    Not applicable. This is not an AI/ML device. There is no training set mentioned.

    9. How the ground truth for the training set was established:
    Not applicable. This is not an AI/ML device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1