510(k) Data Aggregation

    K Number
    K252427

    Validate with FDA (Live)

    Date Cleared
    2025-10-31

    (91 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Diode Laser Hair Removal Machine (Model: GLD01) is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin.

    Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.

    Device Description

    The Diode Laser Hair Removal Machine (Model: GLD01) consists with the power system, microprocessor controller system, operation and screen system, cooling system, laser module and security alarm system. The microprocessor control the laser power, which can provide the constant current for the laser module. The inner high energy diode, from the laser module, transform the electronic to light energy, which output the constant laser with wavelength.The laser light on the skin by leaded crystal, penetrate into the deep tissue of the skin. The light energy is absorbed and then turned into the thermal energy, which can destroy the hair follicle tissue, make it gone and never grow again.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K251076

    Validate with FDA (Live)

    Date Cleared
    2025-07-30

    (113 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The KALA Therapy Wand (Model: KALA-03) is intended for the treatment of facial wrinkles, and mild to moderate inflammatory acne. The red light is intended for the treatment of wrinkles, and the blue light is intended for the treatment of mild to moderate inflammatory acne.

    Device Description

    The KALA Therapy Wand (Model: KALA-03) is indicated for over-the-counter aesthetic use. The red light is intended for the treatment of wrinkles, and the blue light is intended for the treatment of mild to moderate inflammatory acne. The device is vibrating in red light model. The device is powered by a Lithium-Ion rechargeable battery, and it has a charging cable, USB charging stand, protective goggles, storage case and instruction manual.

    The wand can be rotated 135 degrees in either direction.

    There are two switches of the device: one function is red light to wrinkle removal and vibration for relax, the other function is blue light to treat mild to inflammatory acne.

    The device will automatically shut down after 12 minutes of operation. The recommended treatment time is 3 minutes per area. After every three minutes of treatment, the device will vibrate to indicate the time. If you need to continue treatment, simply turn on the device again.

    AI/ML Overview

    The provided FDA 510(k) clearance letter and summary for the KALA Therapy Wand (Model: KALA-03) primarily focus on demonstrating substantial equivalence to predicate devices through non-clinical testing. This type of clearance generally does not require extensive clinical performance studies with detailed acceptance criteria and human reader studies as would be seen for a new or complex AI/ML-driven diagnostic device.

    Based on the provided information, the device is a light-based therapy device for over-the-counter use, intended for aesthetic purposes (facial wrinkles and mild to moderate inflammatory acne). The "study" proving the device meets acceptance criteria is a non-clinical performance testing approach demonstrating compliance with relevant electrical safety, electromagnetic compatibility, photobiological safety, battery safety, and biocompatibility standards. Software verification and validation, and usability validation were also performed.

    Here's the breakdown of the information requested, as extractable from the provided document:


    Acceptance Criteria and Device Performance

    Since this is a non-clinical submission, the "acceptance criteria" are compliance with established safety and performance standards. There isn't a table of statistical performance metrics, but rather successful adherence to defined technical and safety requirements.

    Table of Acceptance Criteria and Reported Device Performance (Non-Clinical Compliance)

    Acceptance Criteria CategorySpecific Standard/Requirement MetReported Device Performance
    Electrical SafetyIEC 60601-1 (Edition 3.2, 2020-08)Compliant
    IEC 60601-1-11 (Edition 2.1, 2020-07)Compliant
    IEC 60601-2-57 (Edition 1.0, 2011-01)Compliant
    Electromagnetic CompatibilityIEC 60601-1-2 (Edition 4.1, 2020-09)Compliant
    Photobiological SafetyIEC 62471 (First edition, 2006-07)Compliant
    Battery SafetyIEC 62133-2 (Edition 1.0, 2017-02)Compliant
    BiocompatibilityISO 10993-5 (Cytotoxicity)Compliant (based on identical materials to cleared device)
    ISO 10993-10 (Sensitization)Compliant (based on identical materials to cleared device)
    ISO 10993-10 (Irritation)Compliant (based on identical materials to cleared device)
    SoftwareFDA Guidance: "Content of Premarket Submissions for Device Software Functions"Software V&V conducted; classified as Basic Documentation Level.
    UsabilityUser testing for Usability ValidationUsability testing conducted.

    Study Details (as applicable for a non-clinical submission):

    1. Sample sizes used for the test set and the data provenance:

      • Test Set (Device Testing): The "test set" here refers to the device prototypes/units that underwent the non-clinical tests. The document does not specify the number of units tested for electrical safety, EMC, etc., which is typical for such submissions.
      • Data Provenance: The testing was conducted by or for the manufacturer (Shenzhen Kaiyan Medical Equipment Co., Ltd) in China, as indicated by the submitter's information. It's inherently "prospective" in the sense that the tests were performed on the device designed for submission.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • This is a non-clinical submission for a device used for aesthetic purposes. The concept of "ground truth" established by medical experts (like radiologists for image analysis) is not applicable here. The "ground truth" for these tests is defined by the passing criteria of the international standards themselves (e.g., specific thresholds for electrical leakage, EMC emissions, irradiance, etc.). Compliance is assessed by qualified test engineers and labs.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

      • Not applicable. Adjudication methods are typically used in clinical studies, especially those involving subjective assessments or disagreements among human readers. Non-clinical testing against objective standards does not involve expert adjudication in this manner.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No. An MRMC study is not relevant for this type of aesthetic, light-based therapy device where the claim is not for assisting human readers in diagnosis. The clearance is based on safety and functional equivalence.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Not applicable. This device is not an algorithm for diagnosis or image analysis. It is a physical device providing light therapy. Its "standalone" performance refers to its ability to meet the specified technical parameters and safety standards.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • The "ground truth" for demonstrating substantial equivalence and safety in this context is compliance with recognized international consensus standards (e.g., IEC, ISO). The performance is measured against these technical specifications, not against clinical outcomes or expert consensus on medical conditions.
    7. The sample size for the training set:

      • Not applicable. This device does not involve a machine learning algorithm that requires a "training set" in the traditional sense of AI/ML software. The software component, as described, is for controlling device functions ("Basic Documentation Level software"), not for learning from data.
    8. How the ground truth for the training set was established:

      • Not applicable, as there is no training set for an AI/ML algorithm.
    Ask a Question

    Ask a specific question about this device

    K Number
    K242951

    Validate with FDA (Live)

    Date Cleared
    2025-04-28

    (215 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Diode Laser Hair Removal Machine is intended for hair removal permanent hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.

    Device Description

    The semiconductor laser therapeutic device is mainly composed of a host, the treatment handle, foot switch, power cord, and accessories. The host consists of a control panel, cooling system, circuit control system, semiconductor laser power supply, etc.

    AI/ML Overview

    The provided document is a 510(k) clearance letter for a Diode Laser Hair Removal Machine (QDTM-02), not a study report. It states that "No clinical study is included in this submission," and therefore, it does not contain the detailed information requested regarding acceptance criteria and a study proving device performance against those criteria.

    However, based on the information available in the 510(k) summary, I can infer some aspects related to the intent of testing and acceptance criteria for demonstrating "substantial equivalence."

    Inferred Acceptance Criteria and Device Performance (Based on Substantial Equivalence Claim)

    The submission aims to demonstrate substantial equivalence to predicate devices, meaning the acceptance criteria are implicitly tied to matching or falling within acceptable ranges of the predicate devices' specifications and performance, and complying with recognized safety standards.

    1. Table of Acceptance Criteria and Reported Device Performance (Inferred)

    Feature/ParameterAcceptance Criteria (Implied by Predicate Comparison)Reported Device Performance (as per K242951)
    Intended UsePermanent hair removal/reduction on all skin types (Fitzpatrick I-VI), including tanned skin. Permanent hair reduction defined as long-term, stable reduction in regrowing hairs at 6, 9, and 12 months post-treatment.Intended for hair removal permanent hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime. (Matches predicate indication)
    Laser TypeClass IV Diode Laser (as per predicates K192735, K180353)Class IV Diode Laser
    Wavelength808nm (as per predicates K192735, K180353)808nm
    FrequencyWithin range of predicates (e.g., 1-20 Hz for K180353)1-20Hz (Matches Predicate Device 2, K180353)
    Pulse DurationWithin range of predicates (e.g., 50-400ms for K192735, 10-400ms for K180353)10-200ms (Falls within ranges of both predicates)
    Small Spot Handpiece Size1.44 cm² (as per Predicate Device 2, K180353)1.44 cm² (Identical to Predicate Device 2)
    Large Spot Handpiece SizeWithin acceptable range, possibly similar to Predicate Device 1 (2.4 cm²) or reference device Eneka Pro (4.75 cm²)4 cm² (Differs from Predicate 1, but similar to reference device Eneka Pro; deemed safe/effective)
    BiocompatibilityCompliance with ISO 10993 series standards.Test results comply with ISO 10993 series standards (ISO 10993-5, -10, -23 cited).
    Electrical Safety & EMCCompliance with relevant IEC standards (e.g., IEC 60601-1, IEC 60601-1-2)Meets IEC 60601-1 (2005), IEC 60601-1-2 (2014), IEC 60601-2-22 (2019), IEC/TR 60601-4-2.
    Laser SafetyCompliance with IEC 60825-1Meets IEC 60825-1 (2014).

    2. Sample size used for the test set and the data provenance:

    • The document explicitly states: "No clinical study is included in this submission."
    • Therefore, there is no information on a "test set" in the context of clinical performance data for hair removal.
    • The testing mentioned (biocompatibility, electrical safety, EMC, laser safety) would involve samples of the device components or the entire device, tested in a laboratory setting. This is not performance data on patients.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable, as no clinical test set for performance evaluation is reported.
    • For the non-clinical tests (biocompatibility, safety standards), the "ground truth" is defined by the requirements of the standards themselves, not by expert consensus on clinical outcomes.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • Not applicable, as no clinical test set for performance evaluation requiring expert adjudication is reported.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This device is a laser hair removal machine, not an imaging diagnostic device or AI software that would involve human readers or AI assistance in interpretation.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This device is a physical medical device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • For the non-clinical tests conducted, the "ground truth" is compliance with established international standards (IEC for electrical and laser safety, ISO for biocompatibility). There is no "outcomes data" or "pathology" cited in the context of device performance on patients.

    8. The sample size for the training set:

    • Not applicable, as "training set" typically refers to data used to train an algorithm or AI model. This is a physical laser device.

    9. How the ground truth for the training set was established:

    • Not applicable for the same reason as above.
    Ask a Question

    Ask a specific question about this device

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Diode Laser Hair Removal Device (Model: EVOLUTION MEDICAL, M-I-X MEDICAL, GENESIS, Lotus and Ultimate) is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin.

    Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.

    Device Description

    The Diode Laser Hair Removal Device is a surgical device intended for use in dermatologic and general surgical procedure. It utilizes a diode laser as a laser source (808 nm). The laser power is delivered to the treatment area via a laser handpiece. The emission laser is activated by a foot switch and a laser handpiece. There are five models included, EVOLUTION MEDICAL, M-I-X MEDICAL, GENESIS, Lotus, and Ultimate, the five models have same mechanism of action, principle and specification, with only one difference: the adjustable range of Energy density is different, EVOLUTION MEDICAL (1-77 J/cm^2), M-I-X MEDICAL (1-70 J/cm^2), GENESIS (1-74 J/cm^2), Lotus(1-67 J/cm^2), and Ultimate(1-65 J/ cm^2).

    Changes to the device include the addition of a remote link to the device to obtain and back up the device's usage data, a remote upgrade function, and the ability to provide a URL for users to download the device's usage data, as compared to previously listed devices.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a Diode Laser Hair Removal Device. This document focuses on establishing substantial equivalence to a predicate device rather than detailing a clinical study with acceptance criteria for a new device's performance.

    Therefore, the sections below requiring information about acceptance criteria and a study proving the device meets them cannot be fully populated as they would for a de novo or PMA submission. The document primarily discusses non-clinical tests to verify design specifications and compliance with standards.

    Here's the information that can be extracted or inferred from the provided text, with clarifications where details are not present:


    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document does not contain a table of explicit acceptance criteria with corresponding device performance metrics for efficacy (e.g., specific hair reduction percentages) as would be found in a clinical study report for a new device.

    Instead, the document focuses on demonstrating that the device meets safety and performance standards and is technologically equivalent to its predicate. The "performance" discussed primarily relates to meeting design specifications and regulatory standards.

    Acceptance Criteria (Inferred from regulatory requirements and predicate comparison)Reported Device Performance
    Safety and Essential Performance (Electrical)Complies with ANS//AAM ES60601-1:2005/(R)2012 & A1:2012 C1:2009/(R)2012 & A2:2010/(R)2012 (Cons. Text) [Incl. AMD2:2021] (Consolidated Text)
    Electromagnetic DisturbancesComplies with IEC 60601-1-2 Edition 4.1 2020-09 CONSOLIDATED VERSION
    Safety and Performance (Laser Equipment)Complies with IEC 60601-2-22 Edition 3.1 2012-10
    Laser Product Safety and ClassificationComplies with IEC 60825-1:2014
    Biocompatibility (Cytotoxicity)Complies with ISO 10993-5 Third Edition 2009-06-01
    Biocompatibility (Irritation and Skin Sensitization)Complies with ISO 10993-10 Third Edition 2010-08-01
    Wireless CoexistenceComplies with IEEE ANSI USEMCSC C63.27-2021 and AAM TIR69:2017/(R2020)
    CybersecurityCybersecurity Testing conducted per FDA Guidance "Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions" (September 27, 2023)
    Energy Output AccuracyPerformance Testing for Energy Output Accuracy was conducted. (Specific metrics not provided in this summary.)
    Software Verification and ValidationTesting conducted per FDA Guidance "Content of Premarket Submissions for Device Software Functions" (June 14, 2023). Level of concern determined to be Moderate.
    Hair Removal and Permanent Hair ReductionDevice is intended for hair removal and permanent hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin, with definition of permanent hair reduction provided in the indications for use. (No direct clinical performance metrics are reported in this summary document, as this is a 510(k) based on substantial equivalence.)

    2. Sample Size Used for the Test Set and Data Provenance

    The document primarily refers to non-clinical tests for compliance with standards and design specifications. It does not describe a clinical test set with human subjects.

    • Test Set Sample Size: Not applicable as a clinical test set is not described. The tests are for device components and system performance.
    • Data Provenance: The tests are non-clinical, related to device manufacturing and engineering compliance. No country of origin for clinical data is applicable. The tests are prospective in the sense that they are conducted on the new device to demonstrate compliance.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not applicable as the document describes non-clinical engineering and compliance testing, not a clinical study requiring expert ground truth assessment.

    4. Adjudication Method for the Test Set

    Not applicable, as this refers to a clinical study with human observers or interpreters, which is not described here.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a laser hair removal system, not an AI-powered diagnostic or interpretive tool that would involve human readers or AI assistance in a diagnostic context.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. This device is a physical laser system, not an algorithm, and the concept of "standalone performance" for an algorithm is not relevant here.

    7. The Type of Ground Truth Used

    For the non-clinical tests, the "ground truth" is defined by the requirements outlined in the cited international and national standards (e.g., IEC 60601-1, ISO 10993) and the device's design specifications. These standards provide benchmarks and methodologies against which the device's performance is measured.

    8. The Sample Size for the Training Set

    Not applicable. As this document details hardware medical devices, there is no "training set" in the context of machine learning or AI.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K241860

    Validate with FDA (Live)

    Date Cleared
    2024-09-25

    (90 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Diode Laser Hair Removal System is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.

    Device Description

    808nm diode laser using selective light absorption, the Diode 808 laser is preferentially absorbed by the melanin of the hair. The light energy is taken up by the hair follicle and converted to heat energy, and minimal energy is transferred to the skin. This preferentially heats the hair and its DNA, while reducing oxygen organization around hair follicle - reducing the chance of hair regrowth. Special cooling technology is applied simultaneously during treatment to cool the skin and protect skin.

    Diode Laser Hair Removal is the safest and fastest at the present. It adopts Gold standard 808nm semi-conductor laser, based on selective photothermolysis principle, through specific wavelength, penetrate epidermis into dermis, optical energy was absorbed and translated into heat energy with restraining hair follicle tissues, and produce photothermal effect. It takes laser energy in hair follicle with rich in melanin, surrounding tissues absorb less even no energy, reach to restrained melanin of hair follicles and remove hair. Meanwhile, don't hurt around tissues and virtually pain-free remove surplus hair; eventually, reach to permanent Hair Removal.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information based on the provided FDA 510(k) summary:

    This device is a Diode Laser Hair Removal System. The provided document is a 510(k) summary for a premarket notification, which focuses on demonstrating substantial equivalence to a predicate device, rather than providing detailed clinical trial results for novel devices. Therefore, a clinical study documenting device performance against specific acceptance criteria for efficacy (e.g., hair reduction percentage) is explicitly stated as not included in this submission. The acceptance criteria primarily revolve around demonstrating safety and performance equivalence through non-clinical testing.


    Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific CriteriaReported Device Performance
    SafetyElectrical Safety: Compliance with IEC 60601-1, IEC 60601-2-22.Complies with IEC 60601-1, IEC 60601-2-22.
    Electromagnetic Compatibility (EMC): Compliance with IEC 60601-1-2.Complies with IEC 60601-1-2.
    Laser Safety: Compliance with IEC 60601-2-22, IEC 60825-1.Complies with IEC 60601-2-22, IEC 60825-1.
    Biocompatibility (Patient Contact Materials):
    - Cytotoxicity: No cytotoxicity.No Cytotoxicity (compliant with ISO 10993-5).
    - Sensitization: No evidence of sensitization.No evidence of sensitization (compliant with ISO 10993-10).
    - Irritation: No evidence of irritation.No evidence of irritation (compliant with ISO 10993-23).
    PerformanceFunctionality under varying power supply: Device to work normally under specified power supply ranges (100V-240VAC / 50Hz-60Hz) despite differences from predicate device.Electrical safety and EMC tests conducted; test results show the device works normally under its specified power supply.
    Functional Equivalence despite minor differences in Fluence and Pulse Duration: The minor differences in Fluence (Proposed: 1-40J/cm² vs. Predicate: 48J/cm²) and Pulse Duration (Proposed: 3-300ms vs. Predicate: 1-300ms) are not to affect effectiveness and safety.The manufacturer states that the minor differences in fluence and pulse duration are not expected to affect effectiveness and safety. This is supported by the device having passed the various IEC safety and performance tests. The pulse duration of the proposed device is also noted to be within the range of the predicate device (though the lower limit differs). This acceptance is based on engineering rationale and compliance with safety standards rather than comparative clinical performance.
    Indications for UseHair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin. Permanent hair reduction defined as long-term, stable reduction in hair regrowth at 6, 9, 12 months post-treatment.The device claims these indications, identical to the predicate device. This is a statement of intended use, not a performance metric demonstrated in the submission.

    Study Proving Device Meets Acceptance Criteria

    This 510(k) submission explicitly states: "No clinical study is included in this submission."

    The "study" or evidence provided to meet the acceptance criteria is primarily a non-clinical testing report demonstrating compliance with recognized safety and performance standards, and a comparison to the predicate device's specifications.

    1. Sample size used for the test set and the data provenance: Not applicable for a clinical study, as none was included. For non-clinical tests, the "sample" would refer to the device itself or its components. The provenance of the data is from in-house testing and external laboratory testing for standard compliance.

    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for non-clinical tests is established by adherence to recognized international standards (e.g., IEC, ISO) and laboratory testing protocols.

    3. Adjudication method for the test set: Not applicable. Non-clinical tests typically involve objective measurements against pass/fail criteria defined by the standards.

    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done: No, not applicable for this type of device and submission (no clinical study was conducted).

    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable, as this is a physical medical device (laser system), not an AI algorithm.

    6. The type of ground truth used:

      • For Safety Criteria: Adherence to established international consensus standards (IEC 60601-1, IEC 60601-1-2, IEC 60825-1, IEC 60601-2-22 for electrical and laser safety; ISO 10993-5, ISO 10993-10, ISO 10993-23 for biocompatibility). These standards define the test methods and acceptance limits that constitute "ground truth" for safety.
      • For Performance (Functional Equivalence): Engineering rationale comparing the proposed device specifications (Fluence, Pulse Duration, Power Supply) to the predicate device, asserting that minor differences do not impact safety or effectiveness, supported by successful completion of the aforementioned non-clinical safety standards.
    7. The sample size for the training set: Not applicable, as no clinical study or machine learning component (requiring a training set) was mentioned or conducted.

    8. How the ground truth for the training set was established: Not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K241642

    Validate with FDA (Live)

    Date Cleared
    2024-08-02

    (56 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Diode Laser Hair Removal System (Model: MBT-Diode Laser) is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin. Permanent hair reduction is defined as the long-term. stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.

    Device Description

    Diode Laser Hair Removal System (Model: MBT-Diode Laser) is a surgical device, which is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI). The Dermatological Diode Laser Systems utilize a semiconductor diode with invisible infrared radiation as a laser source to emit 808nm wavelength laser which is absorbed by melanin. The laser power is delivered to the treatment area via laser handle. The emission laser is activated by footswitch or handle button.

    The Diode Laser Hair Removal System consists of the following major components

    • The main console unit that incorporates the main CPU board, power supply modules, laser power supply, laser device, cooling system and switching module
    • Handpiece
    • Footswitch
    AI/ML Overview

    The provided text is a 510(k) Summary for a Diode Laser Hair Removal System (MBT-Diode Laser). It describes the device, its intended use, and a comparison to a predicate device to demonstrate substantial equivalence.

    Here's an analysis of the acceptance criteria and study information, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The submission does not outline specific, quantified acceptance criteria for the new device as would be seen in a clinical study. Instead, it demonstrates substantial equivalence to a predicate device by comparing technical specifications and adherence to recognized standards.

    The table below summarizes the technical comparison provided, where "acceptance criteria" can be interpreted as demonstrating equivalence or justification for differences compared to the predicate/reference devices:

    FeatureAcceptance Criteria (Comparison to Predicate/Reference)Reported Device Performance (Proposed Device)
    General
    Device NameNA (for comparison)Diode Laser Hair Removal System
    Classification RegulationSAME as predicate (21 CFR 878.4810)21 CFR 878.4810
    Classification PanelSAME as predicate (General & Plastic Surgery)General & Plastic Surgery
    ClassSAME as predicate (II)II
    Product CodeSAME as predicate (GEX)GEX
    Indications for UseSAME as predicate: Hair removal, permanent hair reduction on all skin types (Fitzpatrick I-VI), including tanned skin. Permanent hair reduction defined as long-term, stable reduction measured at 6, 9, and 12 months post-treatment.Same as criterion
    Performance (Technical)
    Laser TypeSAME as predicate (Diode Laser)Diode Laser
    Laser ClassificationSAME as predicate (Class IV)Class IV
    Laser WavelengthSAME as predicate (808nm)808nm
    Spot SizeSIMILAR to predicate (15mm x 15mm vs. 2.25cm², 1.44cm²)15mm x 15mm
    FluenceDifferent from predicate (0-120 J/cm²), but justified as lower (55.96 J/cm²) and similar to reference device (48 J/cm²) for effectiveness. Safety ensured by compliance with standards.1-55.96 J/cm²
    FrequencySAME as predicate (1-10Hz)1-10Hz
    Pulse DurationDifferent from predicate (10-400ms), but within predicate range (2-240ms) and considered a minor difference not affecting effectiveness/safety. Safety ensured by compliance with standards.2-240ms
    Power SupplyDifferent from predicate (220/110 VAC/50Hz-60Hz), but electrical safety and EMC tests conducted ensure normal operation.110 VAC/50Hz-60Hz
    DimensionDifferent from predicate (112 cm x 42 cm x 60 cm), but mechanical performance accepted by IEC 60601-1.650mm x 580mm x 1210mm
    WeightDifferent from predicate (63Kg), but mechanical performance accepted by IEC 60601-1.68Kg
    Safety
    Electrical SafetyComply with IEC 60601-1, IEC 60601-2-22 (SAME as predicate)Complies
    EMCComply with IEC 60601-1-2 (SAME as predicate)Complies
    Laser SafetyComply with IEC 60601-2-22, IEC 60825 (SAME as predicate)Complies
    CytotoxicityNo Cytotoxicity (SAME as predicate; aligns with ISO 10993-5)No Cytotoxicity
    SensitizationNo evidence of sensitization (SAME as predicate; aligns with ISO 10993-10)No evidence of sensitization
    IrritationNo evidence of irritation (SAME as predicate; aligns with ISO 10993-23)No evidence of irritation

    2. Sample Size Used for the Test Set and Data Provenance

    The document explicitly states: "No clinical study is included in this submission."
    Therefore, there is no "test set" in the context of clinical performance data. The evaluation relies on non-clinical (bench) testing and comparison to predicate devices, not human subject data.

    Since no clinical study was conducted, there is no data provenance (country of origin, retrospective/prospective) for a test set.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. As no clinical study was conducted and no clinical test set was used, there was no ground truth for a test set established by experts.

    4. Adjudication Method for the Test Set

    Not applicable. As no clinical study was conducted and no clinical test set was used, there was no adjudication method.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No. The document explicitly states: "No clinical study is included in this submission." Therefore, no MRMC study was conducted.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    The device is a hardware product (Diode Laser Hair Removal System), not an algorithm or software-only device. Therefore, the concept of "standalone algorithm performance" as typically applied to AI/software as a medical device (SaMD) is not applicable here. The performance is evaluated based on its physical/electrical characteristics and safety standards.

    7. The Type of Ground Truth Used

    For the non-clinical tests regarding electrical safety, EMC, laser safety, and biocompatibility, the "ground truth" is defined by compliance with internationally recognized standards (e.g., IEC 60601-1, IEC 60601-1-2, IEC 60825-1, IEC 60601-2-22, ISO 10993 series). The device's performance is measured against the requirements of these standards.

    8. The Sample Size for the Training Set

    Not applicable. This is a hardware device submission, not an AI/ML algorithm requiring a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. No training set was used.

    Ask a Question

    Ask a specific question about this device

    K Number
    K240520

    Validate with FDA (Live)

    Date Cleared
    2024-05-20

    (87 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Diode Laser Hair Removal System is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.

    Device Description

    808nm diode laser using selective light absorption, the Diode 808 laser is preferentially absorbed by the melanin of the hair. The light energy is taken up by the hair follicle and converted to heat energy, and minimal energy is transferred to the skin. This preferentially heats the hair and its DNA, while reducing oxygen organization around hair follicle - reducing the chance of hair regrowth. Special cooling technology is applied simultaneously during treatment to cool the skin and protect skin.

    Diode Laser Hair Removal is the safest and fastest at the present. It adopts Gold standard 808nm semi-conductor laser, based on selective photothermolysis principle, through specific wavelength, penetrate epidermis into dermis, optical energy was absorbed and translated into heat energy with restraining hair follicle tissues, and produce photothermal effect. It takes laser energy in hair follicle with rich in melanin, surrounding tissues absorb less even no energy, reach to restrained melanin of hair follicles and remove hair. Meanwhile, don't hurt around tissues and virtually pain-free remove surplus hair; eventually, reach to permanent Hair Removal.

    AI/ML Overview

    The provided document is a 510(k) summary for the Diode Laser Hair Removal System (SH-VD910). It focuses on demonstrating substantial equivalence to a predicate device rather than presenting a study to prove the device meets specific acceptance criteria for performance in terms of hair removal. The document explicitly states: "No clinical study is included in this submission." Therefore, it does not contain information about:

    • A table of acceptance criteria and reported device performance (in terms of hair removal efficacy).
    • Sample size used for a test set for clinical performance.
    • Data provenance for a clinical test set.
    • Number of experts used to establish ground truth or their qualifications.
    • Adjudication method for a clinical test set.
    • Multi-reader multi-case (MRMC) comparative effectiveness study.
    • Standalone (algorithm only) performance.
    • Type of ground truth used for performance validation.
    • Sample size for a training set.
    • How ground truth for a training set was established.

    Instead, the document details non-clinical tests to demonstrate that the new device meets safety and performance standards equivalent to a predicate device, focusing on technical specifications and compliance with recognized standards.

    Here's the relevant information that is provided:

    1. A table of acceptance criteria and the reported device performance:

    The document doesn't provide specific quantitative acceptance criteria for hair removal efficacy and corresponding reported performance metrics (e.g., percentage hair reduction). Instead, it compares technical specifications of the proposed device to a predicate device and states that the proposed device complies with various safety and electrical standards.

    ItemAcceptance Criteria (Implied by Predicate & Standards Compliance)Reported Device Performance (Proposed Device)
    General Comparison
    Device NameDiode Laser Hair Removal System (Predicate)Diode Laser Hair Removal System
    Classification Regulation21 CFR 878.4810 (Predicate)21 CFR 878.4810
    Classification PanelGeneral & Plastic Surgery (Predicate)General & Plastic Surgery
    ClassII (Predicate)II
    Product CodeGEX (Predicate)GEX
    Common NamePowered Laser Surgical Instrument (Predicate)Powered Laser Surgical Instrument
    Indication for UseHair removal, permanent hair reduction on all skin types (Fitzpatrick I-VI), including tanned skin, permanent reduction measured at 6, 9, and 12 months (Predicate)Hair removal, permanent hair reduction on all skin types (Fitzpatrick I-VI), including tanned skin, permanent reduction measured at 6, 9, and 12 months
    Prescription UsePrescription use (Predicate)Prescription use
    Performance Comparison (Technical)
    Laser TypeDiode laser (Predicate)Diode laser
    Laser ClassificationClass IV (Predicate)Class IV
    Laser Wavelength808nm (Predicate)808nm
    Spot Size1.44 cm² (Predicate)10mm x 12mm (Different, deemed not to impact safety/effectiveness due to energy output and testing)
    Fluence1-120 J/cm² (Predicate)5-120 J/cm² (Different, deemed minor and not to affect effectiveness/safety)
    Frequency0.5-15Hz (Predicate)0.5-10Hz (Within predicate range, deemed not to raise new safety issues)
    Pulse Duration5-400ms (Predicate)10-300ms (Within predicate range, deemed minor and not to affect effectiveness/safety)
    Power SupplyAC 110V/60Hz (Predicate)110-120 VAC, 15A Max., 60 Hz (Different, but electrical safety and EMC tested)
    Dimension450mm× 550mm×380mm (Predicate)78(L)*65(W)*133(H)cm (Different, mechanical performance acceptable per IEC 60601-1)
    Weight52 Kg (Predicate)54 kg (Different, mechanical performance acceptable per IEC 60601-1)
    Safety Comparison
    Electrical SafetyComply with IEC 60601-1 (Predicate)Comply with IEC 60601-1
    EMCComply with IEC 60601-1-2 (Predicate)Comply with IEC 60601-1-2
    Laser SafetyComply with IEC 60601-2-22, IEC 60825-1 (Predicate)Comply with IEC 60601-2-22, IEC 60825-1
    CytotoxicityNo Cytotoxicity (Predicate)No Cytotoxicity
    IrritationNo evidence of irritation (Predicate)No evidence of irritation
    SensitizationNo evidence of sensitization (Predicate)No evidence of sensitization

    2. Sample size used for the test set and the data provenance: Not applicable, as no clinical study was conducted.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable, as no clinical study was conducted.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable, as no clinical study was conducted.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable, as this device is a laser hair removal system, not an AI-assisted diagnostic tool. No clinical study was conducted.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable, as this device is a laser hair removal system, not an algorithm. No clinical study was conducted.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not applicable, as no clinical study was conducted. The ground truth for the non-clinical tests is based on compliance with established international standards (e.g., IEC, ISO).

    8. The sample size for the training set: Not applicable, as this is a laser device, not a machine learning algorithm.

    9. How the ground truth for the training set was established: Not applicable, as this is a laser device, not a machine learning algorithm.

    Non-Clinical Test Conclusions:
    The manufacturer performed non-clinical tests to verify that the proposed device met design specifications and was substantially equivalent to the predicate device. These tests demonstrated compliance with the following standards:

    • IEC 60601-1 Edition 3.2 2020-08 (Medical electrical equipment - General requirements for basic safety and essential performance)
    • IEC 60601-1-2 Edition 4.1 2020-09 (Medical Equipment - Collateral Standard: Electromagnetic Compatibility-Requirements And Tests)
    • IEC 60825-1:2014 (Safety of laser products - Part 1: Equipment classification, and requirements)
    • IEC 60601-2-22: 2012 (Medical electrical equipment - Particular requirements for basic safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment)
    • ISO 10993-5: 2009 (Biological Evaluation of Medical Devices -- Part 5: Tests For In Vitro Cytotoxicity)
    • ISO 10993-10 Fourth edition 2021 (Biological evaluation of medical devices - Part 10: Tests for skin sensitization)
    • ISO 10993-23 First edition 2021-01 (Biological evaluation of medical devices - Part 23: Tests for irritation)

    The conclusion is that the device is as safe, as effective, and performs as well as or better than the legally marketed predicate device (K162659) based on these non-clinical tests.

    Ask a Question

    Ask a specific question about this device

    K Number
    K232117

    Validate with FDA (Live)

    Date Cleared
    2023-11-28

    (134 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Diode Laser Hair Removal (LHR-V6S-1064, LHR-V6S/B-1064) is intended for hair removal, permanent hair reduction.

    Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime. Diode Laser Hair Removal (LHR-V6S-1064, LHR-V6S/B-1064) is intended for use on all skin types (Fitzpatrick skin types I - VI), including tanned skin.

    Device Description

    The Diode Laser Hair Removal device models LHR-V6S-1064 and LHR-V6S/B-1064 emit pulses of invisible infrared laser light of 1064nm wavelength that penetrates into the skin and is selectively absorbed by melanin in the hair follicles. This creates a localized thermal effect that disrupts hair growth from the hair follicles. The device contains a sensor that detects contact with the skin so that the device will only emit the infrared laser pulses when the sensor is in contact with the skin. The device also includes a skin cooling feature and the device is powered by an external AC/DC power adaptor.

    AI/ML Overview

    Here's an analysis of the provided text regarding acceptance criteria and studies demonstrating device performance:

    Unfortunately, the provided document does not contain information about acceptance criteria for a specific performance metric (like accuracy, sensitivity, or specificity) of a diagnostic or AI-driven device, nor does it describe a study proving the device meets such criteria.

    The document details a 510(k) premarket notification for a Diode Laser Hair Removal device (LHR-V6S-1064, LHR-V6S/B-1064). The focus of this submission is on demonstrating substantial equivalence to existing predicate devices based on technical characteristics, indications for use, and non-clinical safety testing.

    Here's a breakdown of why the requested information isn't present and what information is provided:

    Key Takeaways from the Document:

    • Device Type: It's a physical medical device (laser for hair removal), not a diagnostic or AI-powered software device.
    • Regulatory Pathway: 510(k) premarket notification, which relies on demonstrating substantial equivalence to a legally marketed predicate device, rather than proving novel clinical effectiveness through extensive performance studies against specific acceptance criteria.
    • No Clinical Study: The document explicitly states: "No clinical testing has been performed." This directly indicates that there aren't clinical performance metrics or studies to assess against acceptance criteria.
    • Focus on Substantial Equivalence: The primary goal of the submission is to show that the proposed device is "as safe, as effective, and performs as well as or better than the predicate device" based on design, specifications, and non-clinical tests.

    Response to your specific questions based on the provided document:

    1. A table of acceptance criteria and the reported device performance:

      • N/A. The document does not define specific performance acceptance criteria for a diagnostic or AI algorithm, nor does it report performance metrics against such criteria. The "performance testing" mentioned (Appearance, Function, Safety, Label, Shelf Life, Cleaning and Disinfection) are general bench tests for a physical device, not performance metrics like sensitivity or accuracy for a diagnostic.
    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

      • N/A. No clinical test set or data for performance evaluation (e.g., diagnostic accuracy) is mentioned because "No clinical testing has been performed."
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

      • N/A. No clinical test set requiring ground truth established by experts is mentioned.
    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • N/A. No clinical test set requiring adjudication is mentioned.
    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • N/A. This device is for hair removal, not a diagnostic tool with AI assistance for human readers. No MRMC study was conducted.
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • N/A. This is a physical laser device, not an algorithm.
    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • N/A. No clinical ground truth is established as no clinical studies were performed.
    8. The sample size for the training set:

      • N/A. This refers to a dataset for training an algorithm. This device is not an AI algorithm.
    9. How the ground truth for the training set was established:

      • N/A. As above, this is not an AI algorithm.

    Information that is available about device equivalence and non-clinical testing:

    • Non-Clinical Tests Submitted:
      • Biocompatibility Testing: According to ISO 10993-1:2018, including in vitro cytotoxicity (ISO 10993-5:2009), skin sensitization (ISO 10993-10:2021), and irritation (ISO 10993-23:2021). These tests were performed for parts contacting intact skin.
      • Electrical Safety and Electromagnetic Compatibility Testing: Compliance with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-2-22, and IEC 60825-1.
      • Software Verification and Validation: Performed according to FDA guidance for software in medical devices.
    • Concise summary for performance testing (Bench Tests):
      • Appearance, Function, Safety, Label.
      • Shelf Life Test (accelerated aging).
      • Cleaning and Disinfection Verification.

    In summary, the provided document is a 510(k) clearance for a laser hair removal device, focusing on demonstrating substantial equivalence through technical specifications and non-clinical safety tests, not clinical performance against specific acceptance criteria for a diagnostic or AI device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K230090

    Validate with FDA (Live)

    Date Cleared
    2023-04-19

    (97 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Diode Laser Hair Removal is an over-the-counter device intended for adjunctive use with shaving for hair removal sustained with periodic treatments. Diode Laser Har Removal is also intended for permanent reduction in hair regrowth defined as a long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.

    Device Description

    The Diode Laser Hair Removal device models LHR-V8S-810, LHR-V8-810, LHR-V4S-810, and LHR-V4-810 emit pulses of invisible infrared laser light of 810mm wavelength that penetrates into the skin and is selectively absorbed by melanin in the hair follicles. This creates a localized thermal effect that disrupts hair growth from the hair follicles. The device contains a sensor that detects contact with the skin so that the device will only emit the infrared laser pulses when the sensor is in contact with the skin. The device also includes a skin cooling feature and the device is powered by an external AC/DC power adaptor.

    AI/ML Overview

    The provided document describes a 510(k) premarket notification for a Diode Laser Hair Removal device. It does not contain information about acceptance criteria and a study proving the device meets those criteria in the context of an AI/ML powered medical device. The document primarily focuses on demonstrating substantial equivalence to a predicate device through comparison of technological characteristics and safety/performance testing of physical aspects of the device, not its AI/ML performance.

    Therefore, I cannot extract the requested information regarding:

    • A table of acceptance criteria and reported device performance (for AI/ML).
    • Sample size for the test set and data provenance.
    • Number and qualifications of experts for ground truth.
    • Adjudication method for the test set.
    • Multi-reader multi-case (MRMC) comparative effectiveness study and effect size.
    • Standalone performance (AI/ML without human-in-the-loop).
    • Type of ground truth used (for AI/ML).
    • Sample size for the training set.
    • How ground truth for the training set was established.

    The document lists "Performance Data" which includes Biocompatibility Testing and Electrical, EMC, and laser output, Safety and Performance Testing. These are standard safety and engineering tests for a physical medical device, not performance studies for an AI/ML algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K223778

    Validate with FDA (Live)

    Date Cleared
    2023-03-31

    (105 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    18 - 120
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Diode Laser Hair Removal System (RD-SLD600) is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin. Permanent hair reduction is defined as the long-term. stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.

    Device Description

    The proposed device, Diode Laser Hair Removal System (RD-SLD600) is a professional platform, it is intended for the removal of unwanted hair and to effect stable, long-term hair reduction on all skin types (Fitzpatrick skin type I-VI). The complete system consists of its console, module and a footswitch. The module is pressed against the patient's skin and a light pulse is delivered when the emitted button or the footswitch are activated. The crystal contact surface is cooled by the RD-SLD600's cooling system. Output parameters and other system features are controlled from the touch-screen control panel on the console, which provides an interface to the system's microcontroller through and LCD touch-screen.

    AI/ML Overview

    The provided text is a 510(k) Summary for a Diode Laser Hair Removal System (RD-SLD600). It asserts substantial equivalence to predicate devices based on non-clinical testing. Crucially, no clinical studies were performed for this device as part of the submission. Therefore, the document does not contain information about acceptance criteria or a study proving the device meets those criteria through clinical performance.

    The document states:

    • "Clinical testing was not performed for the subject device as part of the submission." (page 8)
    • The conclusion is based on non-clinical tests (Electrical Safety and Electromagnetic Compatibility, and Biocompatibility) which determined the device to be "safe and effective for its intended use, and performs as well or better than the predicate devices" (page 8).

    Therefore, I cannot provide the requested information regarding acceptance criteria and a study proving the device meets those criteria, as such a study was explicitly not conducted according to this document.