Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K151487

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2016-01-14

    (226 days)

    Product Code
    Regulation Number
    862.1345
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Diatron Glucose Hexokinase Method is for the in vitro quantitative determination of Glucose in serum for use with Diatron Pictus 700 Chemistry Analyzer. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes.

    Diatron ISE is for the in vitro quantitative determination of Sodium (Na+), Potassium (K+), and Chloride (Cl-) concentrations in serum on the Diatron Pictus 700 Chemistry Analyzer. Sodium measurements are used in the diagnosis and treatment of diseases involving electrolyte imbalance. Potassium measurements are used in monitoring electrolyte balance and in the diagnosis and treatment of diseases/conditions characterized by low or high blood potassium levels. Chloride measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders.

    The Diatron Pictus 700 Clinical Chemistry Analyzer is a wet-chemistry analyzer for the direct determination of sodium, potassium, chloride and glucose concentrations in serum, and to measure a variety of analytes that may be adaptable to the analyzer depending on the reagent used. It is for in vitro diagnostic use.

    Device Description

    The Pictus 700 Clinical Chemistry Analyzer is an automatic, floor-standing, wet chemistry system, designed for the qualitative and quantitative analyses of various diagnostic test systems. This premarket notification is for the direct quantitative measurements of Na (sodium), K (potassium), CI (chloride) and glucose in serum samples. Additionally, other types of chemistry assays may be performed on the analyzer, provided that suitable colorgenerating reactions or reactions with variation of color are used. The system is intended for use in clinical laboratories.

    The instrument consists of an analyzer unit and an operations computer with a screen that allows the user to input commands for system operation and data display. The analyzer unit includes two temperature-controlled incubation rotors and a multi-wavelength photometer, a cooled carousel for loading barcoded sample tubes or micro cups and reagent cartridges, and two probes that deliver reagents and samples to the incubation rotors and the ISE measurement flow cell. The analyzer unit also houses containers for wash solution and waste. An ISE module, installed on the analyzer unit, is used to measure sodium, potassium and chloride ionic activity in serum.

    AI/ML Overview

    The provided text describes the acceptance criteria and the study performance for the Diatron ISE, Diatron Glucose Hexokinase Method, and Diatron Pictus 700 Clinical Chemistry Analyzer. This is a submission for a 510(k) premarket notification for an in vitro diagnostic device, specifically a clinical chemistry analyzer and associated assay methods. The studies presented are nonclinical performance studies to demonstrate substantial equivalence to a predicate device, not clinical effectiveness studies involving human subjects in an MRMC setting.

    Here's a breakdown of the requested information based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are generally implied by the claim of substantial equivalence to the predicate device and the successful demonstration of various analytical performance characteristics. The document explicitly states acceptance criteria for linearity.

    Study/ParameterAnalyte(s)Acceptance CriteriaReported Device Performance
    Functional Sensitivity/LoD/LoQSodiumHighest dilution concentration (lowest concentration) with < 10%CV was lower than, or equal to, the low-end linearity claim.1:9 dilution resulted in 27.6 mmol/L with 3.02%CV. Low-end linearity claim: 110 mmol/L (verified).
    PotassiumHighest dilution concentration (lowest concentration) with < 10%CV was lower than, or equal to, the low-end linearity claim.1:10 dilution resulted in 1.05 mmol/L with 0.00%CV. Low-end linearity claim: 1.0 mmol/L (verified).
    ChlorideHighest dilution concentration (lowest concentration) with < 10%CV was lower than, or equal to, the low-end linearity claim.1:9 dilution resulted in 14.8 mmol/L with 3.02%CV. Low-end linearity claim: 50 mmol/L (verified).
    GlucoseLoQ measured as 7 mg/dL.LoQ CV was 1.3%. Verified low end LOD and LOQ claim of 7 mg/dL.
    LinearitySodium+/- 10% and or +/- 5 mmol/LLinear from 13 mmol/L to 196 mmol/L. Claimed range: 115 to 196 mmol/L.
    Potassium+/- 10% and or +/- 1.0 mmol/LLinear from 0.9 to 12.8 mmol/L. Claimed range: 1.1 to 8.8 mmol/L.
    Chloride+/- 10% and or +/- 5 mmol/LLinear from 41 mmol/L to 200 mmol/L. Claimed range: 49 mmol/L to 152 mmol/L.
    Glucose+/- 10% and or +/- 5 mg/dLLinear from 13 mg/dL to 632 mg/dL. Claimed range: 13 to 500 mg/dL.
    PrecisionSodium, Potassium, Chloride, Glucose%CVs within acceptable ranges (comparison to predicate's range and general analytical expectations). Specific criteria not explicitly stated but implied by successful testing according to CLSI EP5-A2 and presentation of low %CVs.Sodium: Within-run %CV: 0.5-0.7%, Total %CV: 0.9-1.4% Potassium: Within-run %CV: 0.00-0.78%, Total %CV: 1.05-1.65% Chloride: Within-run %CV: 0.5-0.6%, Total %CV: 0.8-1.2% Glucose: Within-run %CV: 0.9-1.6%, Total %CV: 3.0-3.2%
    InterferencesSodium, Potassium, Chloride, Glucose≤ 10 % Change or ≤ 3 mmol/L (or mg/dL for glucose) up to specified interferent concentrations.Sodium: Hemolysis (600 mg/dL), Lipemia (2500 mg/dL), Bilirubin (24 mg/dL) Potassium: Not for hemolyzed serum, Lipemia (2500 mg/dL), Bilirubin (24 mg/dL) Chloride: Hemolysis (600 mg/dL), Lipemia (2500 mg/dL), Bilirubin (24 mg/dL) Glucose: Hemolysis (600 mg/dL), Lipemia (700 mg/dL), Bilirubin (12 mg/dL) all passed inference evaluation.
    Method ComparisonSodium, Potassium, Chloride, GlucoseCorrelation coefficients and 95% CIs for slope and intercept should demonstrate agreement with the predicate device. Specific numerical acceptance criteria for correlation, slope, and intercept are not explicitly stated as hard cutoffs but implied by demonstrating substantial equivalence to the predicate.Sodium: N=69, r=0.995, Slope=1.002 (0.978-1.027), Intercept=-1.02 (-4.7 to 2.2) Potassium: N=69, r=0.998, Slope=0.999 (0.984-1.014), Intercept=0.02 (-0.05 to 0.09) Chloride: N=69, r=0.997, Slope=1.015 (0.996-1.035), Intercept=-1.90 (-3.9 to 0.1) Glucose: N=136, r=0.999, Slope=0.993 (0.987 to 1.000), Intercept=1.2 (-0.1 to 2.4)

    2. Sample Sizes and Data Provenance for Test Sets

    • Glucose Functional Sensitivity/LoQ: Glucose depleted samples, tested in 5 replicates per dilution for functional sensitivity. LoQ testing was performed according to CLSI EP17-A2. The document mentions "Testing the neat sample, depleted of Glucose y standing > 24 hrs exposed to red cells in the sample draw."
    • Linearity: Eleven levels of each analyte were tested, in duplicate.
    • Precision: Control samples (low, middle, and high levels) were tested in duplicate, twice a day, for 20 days, for a total of 80 results per level.
    • Interferences: A two-level control set (low and high analyte levels) was spiked to seven or eight levels of interferent. Spiked samples plus neat samples were tested in duplicates.
    • Method Comparison:
      • Sodium: N=69 clinical specimens
      • Potassium: N=69 clinical specimens (listed under Sodium, but context implies N=69 for K as well given the table structure)
      • Chloride: N=69 clinical specimens
      • Glucose: N=136 clinical specimens
      • Data Provenance: Clinical specimens were "previously-collected serum samples that had been stored frozen and then thawed prior to analysis." The country of origin is not specified but typically for a US FDA submission, these would be collected in the US or under similar regulatory standards. The data is retrospective in that it uses "previously-collected" and "stored frozen" samples.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    Not applicable. This is an analytical performance study for an in vitro diagnostic device, not an AI/imaging device requiring expert interpretation for ground truth. The "ground truth" for the test samples in this context relates to their known concentrations as determined by a reference method or the predicate device.

    4. Adjudication Method for the Test Set

    Not applicable. This is an analytical performance study for an in vitro diagnostic device, where quantitative results are compared to reference methods/predicate, not subjective interpretations requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size on Human Readers

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not performed. This type of study is typically done for AI-driven diagnostic imaging devices where human readers' performance (with or without AI assistance) is evaluated. This submission is for an automated clinical chemistry analyzer.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, the studies presented are essentially "standalone" performance evaluations of the Diatron Pictus 700 Clinical Chemistry Analyzer and its associated methods (Diatron Glucose Hexokinase Method and Diatron ISE). The performance measured is that of the automated instrument itself, without direct human intervention in the analytical process being tested for. The output is a quantitative measurement of analytes.

    7. The Type of Ground Truth Used

    The ground truth for the analytical studies are based on:

    • Reference methods/assigned values: For linearity, samples were assigned their reference values arithmetically from labeled values. For method comparison, the predicate system served as the comparative method against which the new device's performance was evaluated.
    • Known concentrations: For precision, control samples with known low, medium, and high concentrations were used.
    • Spiked samples: For interference studies, samples were spiked with known concentrations of interferents to assess their impact.

    8. The Sample Size for the Training Set

    Not applicable. This is not a study involving machine learning with distinct training and test sets in the typical sense of AI algorithm development. This is a traditional analytical validation for an in vitro diagnostic device.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable for the reason stated in point 8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1