Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K170110

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2018-08-10

    (575 days)

    Product Code
    Regulation Number
    880.5440
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Diana ChemoLock Transfer Set is a sterile, single-use closed system transfer device used for drug preparation to transfer drug from a drug vial to an IV bag for intravenous drug administration. It prohibits the transfer of environmental contaminants into the system and the escape of drug or vapor concentrations outside the system.

    Device Description

    Diana ChemoLock Transfer Set is intended for use with the Diana Compounding System. It is a disposable, single use ancillary device for Pharmacy Compounding Devices (PCD).

    The transfer set is comprised of multiple components bonded together to form a single device. These components include the following: stopcock, tubing, ChemoLock, syringe, and cassette handle. When placed in the Diana Compounding System, the Diana ChemoLock Transfer Set allows the transfer of fluids from one container to the other container for reconstitution of lyophilized drug or transfer of stock drug solution to prepare medications.

    The purpose of this submission is to release a new closed system transfer device (CSTD) drug transfer set to be used with Diana Pharmacy Compounding System

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the Diana ChemoLock Transfer Set, which is a medical device. This document focuses on demonstrating substantial equivalence to a predicate device, rather than presenting a study design with acceptance criteria and performance data for a standalone AI/algorithm device.

    Therefore, the requested information regarding acceptance criteria, study details, sample sizes, expert involvement, adjudication methods, MRMC studies, standalone performance, and ground truth establishment is not available in the provided text. The document details various functional and safety tests performed on the device to ensure its performance and safety, but these are not presented as "acceptance criteria" in the context of an AI/algorithmic performance study.

    Here's what can be extracted, framed as general performance data rather than specific acceptance criteria for an AI/algorithm:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document doesn't explicitly list acceptance criteria in a tabular format with corresponding reported performance for AI/algorithmic measures. Instead, it lists various performance tests conducted.

    Test CategoryDescription of Test Conducted
    Functional TestingDemonstrated functionality of ChemoLock, stopcock, tubing, syringe, and cassette handle. Included: Fluid Flow, Accuracy Testing, Positive Pressure, Chemical Compatibility, Internal Seal Integrity, Stopcock Handle Torque, Component Bond Strength Testing, Visual Inspection. (Follows FDA guidance "Intravascular Administration Sets Premarket Notification Submissions [510(k)]")
    Packaging Integrity and Shelf LifeConducted according to ASTM D4169, ASTM F1980, ASTM F2096, and ASTM F88 standards.
    BiocompatibilityEvaluated according to FDA Guidance and ISO 10993-1. Included: Hemocompatibility, Cytotoxicity, Sensitization, Intracutaneous Irritation, Acute Systemic Toxicity, Material Mediated Pyrogenicity.
    Microbial Ingress TestingDemonstrated that the needless access site maintains physical integrity and prohibits environmental contaminant transfer after access. (Consistent with FDA guidance "Intravascular Administration Sets Premarket Notification Submissions [510(k)]")
    Emission, Dry Disconnection, and CSTD Hazardous Drug Exposure EvaluationPerformed to validate no escape of drug from the device outside the closed system.
    ParticulatesTesting performed following USP <788> (method 1) to demonstrate particulate levels meet USP 788 requirements.
    Sterility TestingE-beam sterilization process validation conducted according to ISO 11137-1 and ISO 11137-2. Bacterial endotoxin testing conducted based on AAMI ST72 and USP <85>, and followed FDA Guidance for Industry - Pyrogen and Endotoxins Testing: Questions and Answers. Achieved Sterility Assurance Level (SAL) of 10-6.

    2. Sample size used for the test set and the data provenance: Not applicable. The document describes physical and chemical testing, not a test set for an AI/algorithm.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. This pertains to AI/algorithm validation, not to the functional testing of a transfer set.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable. This pertains to AI/algorithm validation.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is a drug transfer set, not an AI-assisted diagnostic or therapeutic device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc): Not applicable. The testing described relies on established engineering and medical device standards (e.g., ASTM, ISO, USP, FDA guidance), rather than ground truth in the context of an AI model.

    8. The sample size for the training set: Not applicable. This device does not involve a training set for an AI/algorithm.

    9. How the ground truth for the training set was established: Not applicable. This device does not involve a training set for an AI/algorithm.

    In summary, the provided document is a 510(k) submission for a physical medical device, not for an AI/algorithmic device. Therefore, the questions related to AI/algorithm performance and validation are not addressed within this text. The "performance data" section outlines various physical, chemical, and biological tests conducted to demonstrate the device's safety and functionality in accordance with relevant standards.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1