Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K192702

    Validate with FDA (Live)

    Date Cleared
    2020-09-18

    (357 days)

    Product Code
    Regulation Number
    882.5330
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Neuro:

    • Neurosurgery
    • Nasoethmoidal fractures
    • Infraorbital area fractures
    • Frontal sinus wall Fractures
    • Infant craniofacial surgery
      Dental:
    • Maxillofacial surgery

    The DePuy Synthes Mandibular Modular Fixation System is a plate and screw system, manufactured from commercially pure titanium, and is intended for use in:

    • Oral, maxillofacial surgery: trauma; surgical correction of dentofacial deformities; reconstructive surgery; and maxillofacial surgery
    • Neurosurgery: osteosynthesis of the cranial bones.

    Dental: The DePuy Synthes Single Vector Distractor with Detachable Feet is intended for use as a bone stabilizer and lengthener for conditions such as mandibular hypoplasia or posttraumatic defects of the mandible, where gradual bone distraction is required. The device is ideal for treating fonns of clefts of the lip and palate, and congenital mandbular hypoplasia, such as Hemifacial Microsomia, Treacher Collins Syndrome, Pierre Robin Syndrome, Goldenhar Syndrome, Apert Syndrome, and Crouzon Syndrome.
    The DePuy Synthes Single Vector Distractor with Detachable Feet is also ideal for treating hypoplasias of an acquired origin such as from post-traumatic growth disorders associated with injury to the temporomandibular joint, temporomandibular ankylosis, and segmental loss of bone.
    Neuro: The DePuy Synthes Single Vector Distractor with Detachable Feet can be used for stabilization and advancement of the mid-face, in which a deficiency of mid-facial bone requires gradual bone distraction. Such deficiencies include, but are not limited to Plagiocephaly, Trigonocephaly, Scaphocephaly, and Brachycephaly.

    The DePuy Synthes Cranial Tube Clamp is intended to reattach a cranial bone flap to the surrounding cranium after a craniotomy procedure.

    Dental: The DePuy Synthes Low Profile Neuro System is intended for use in selective trauma of the midface and maxillofacial skeleton; maxillofacial surgery; reconstructive procedures; and selective orthognathic surgery of the maxilla and chin.
    Neuro: The DePuy Synthes Low Profile Neuro System is intended for cranial closure and/or bone fixation, craniotomies, cranial trauma repair and reconstruction.

    Dental: The DePuy Synthes Low Profile Neuro System is intended for use in selective trauma of the midface and maxillofacial skeleton; maxillofacial surgery; reconstructive procedures; and selective orthognathic surgery of the maxilla and chin.
    Neuro: The DePuy Synthes Low Profile Neuro System is intended for cranial closure and/or bone fixation, craniotomies, cranial trauma repair and reconstruction.

    Dental: The DePuy Synthes 1.3 & 1.5 mm Contourable Titanium (Ti.) Mesh Plates are intended for use in selective trauma of the midface and maxillofacial skeleton; maxillofacial surgery; reconstructive procedures; and selective orthognathic surgery of the maxilla and chin.
    Neuro: The DePuy Synthes 1.3 & 1.5 mm Contourable Titanium (Ti.) Mesh Plates are intended for use in selective trauma of the craniofacial skeleton and craniofacial surgery.

    Dental: The DePuy Synthes External Midface Distractor is intended for use in maxillofacial surgery, reconstructive procedures, and selective orthognathic surgery of the maxilla. Specifically, it is intended for distraction of the maxilla utilizing a LeFort I osteotomy, the midface utilizing a LeFort II or III osteotomy in adult and pediations where gradual bone distraction is required.
    Neuro: The DePuy Synthes External Midface Distractor is intended for use in craniofacial surgery and reconstructive procedures. Specifically, it is intended for distraction of the cranium utilizing a monobloc osteotomy in adult and pediativ populations where gradual bone distration is required.

    The DePuy Synthes 1.0/1.2 mm Craniofacial Screws are intended for use as follows: Neuro:
    • Nasoethmoidal fractures

    • Infraorbital area fractures
    • Frontal sinus wall fractures ●
    • Infant craniofacial surgery
      Dental:
    • Maxillofacial surgery

    Dental: DePuy Synthes Neuro Plate and Screw System is intended for use in selective trauma of the midface and maxillofacial skeleton; maxillofacial surgery; reconstructive procedures; and selective orthognathic surgery of the maxilla and chin.
    Neuro: DePuy Synthes Neuro Plate and Screw System is intended for use in fixation of the cranial bones in procedures such as reconstruction, fracture repair, craniotomies, and osteotomies.

    Dental: The DePuy Synthes Low Profile Neuro System is intended for use in selective trauma of the midface and maxillofacial skeleton; maxillofacial surgery; reconstructive procedures; and selective orthognathic surgery of the maxilla and chin.
    Neuro: The DePuy Synthes Low Profile Neuro System is intended for cranial closure and/or bone fixation, craniotomies, cranial trauma repair and reconstruction.

    Dental: The DePuy Synthes Craniofacial Plate and Screw System is intended for use in selective trauma of the midface and maxillofacial skeleton; maxillofacial surgery; reconstructive procedures; and selective orthognathic surgery of the maxilla and chin.
    Neuro: The DePuy Synthes Craniofacial Plate and Screw System is intended for use in selective trauma of the craniofacial skeleton and craniofacial surgery.

    The DePuy Synthes CMF Distraction System is intended for use as a bone stabilizer and lengthening (and/or transport) device.
    The DePuy Synthes CMF Distraction System is indicated for correction of congenital deficiencies or post-traumatic defects of the cranium, where gradual bone distraction is required in adults and pediatric patients.

    Dental: The DePuy Synthes Craniofacial Plate and Screw System is intended for use in selective trauma of the midface and maxillofacial skeleton; maxillofacial surgery; reconstructive procedures; and selective orthognathic surgery of the maxilla and chin.
    Neuro: The DePuy Synthes Craniofacial Plate and Screw System is intended for use in selective trauma of the craniofacial skeleton and craniofacial surgery.

    The DePuy Synthes MatrixNEURO Cranial Plating System is intended for use in fixation of the cranial bones in procedures such as reconstruction, fracture repair, craniotomies, and osteotomies.

    The DePuy Synthes MatrixNEURO Preformed Mesh is intended for use in fixation of the cranial bones in procedures such as reconstruction, fracture repair, craniotomies, and osteotomies.

    Device Description

    The DePuy Synthes Maxillofacial Titanium Micro set consists of titanium bone plates (shapes include L,Y,H,T, Double-Y, Mesh and Straight) and self-tapping screws.

    The DePuy Synthes Mandibular Modular Fixation System is a plate and screw system manufactured from titanium. The plates are available in a variety of shapes and sizes, and attach to bone via 2.0 mm, 2.4 mm or 2.7 mm screws.

    The DePuy Synthes Single Vector Distractor with Detachable Feet is a subcutaneous bone distractor activated by a drive component. It features two telescoping components activated by a jack screw, fixed to the bone with bone screws. Bone lengthening and distraction are achieved by gradually activating the device. Upon removal, the telescoping components and jack screw are disengaged and removed, leaving the subcutaneous foot plates in the patient.

    The DePuy Synthes Cranial Flap Tube Clamp consists of two clamp discs with a clamp shaft in between the discs. The clamp shaft consists of a 1.6 mm tube which is secured to the bottom disc. The top disc has a thickness of 0.4 mm and the plate diameters range from 8.0 to 20. mm.

    The DePuy Synthes Low Profile Neuro System consists of titaniurn plates, and screws in a variety of shapes and sizes designed for various cranio-facial procedures.

    Synthes ø1.6 mm Low Profile Neuro System self-drilling, self·tapping, and ø1.9 mm emergency self-tapping screws in 3 mm lengths are to be added to the system.

    The DePuy Synthes 1.3 & 1.5 mm Contourable Ti. Mesh Plates come in a variety of shapes and sizes to meet the anatomical need of the patient. The plates are sterile and for single use only.

    The DePuy Synthes External Midface Distractor is an external distraction osteogenesis device that attaches to the cranium and midface and is used to gradually lengthen the midface at the LeFort I, II, and III levels (including monobloc). The device consists of an external headframe, a central adjustment mechanism, a veridical central rod, horizontal crosspieces containing distraction screws, and separate footplates assemblies that attach to the zygoma and maxilla.

    The DePuy Synthes 1.0/1.2 mm Craniofacial Screws are either self-drilling or self tapping, have a flat head profile with rounded edges with a cruciform recess, and are available in various lengths.

    DePuy Synthes Neuro Plate and Screw System consists of plates, burr hole covers, and meshes that come in a variety of shapes and sizes to meet the anatomical needs of the patient. This system is designed for use with 1.8 mm screws and 2.1 mm emergency screws. The screws will be used with Synthes 1.8 mm hexagonal screwdriver blades. System components are manufactured in either titanium alloy and are intended for single use only.

    The DePuy Synthes 1.6 mm burr hole covers come in various sizes to accommodate various fracture and osteotomy sites, have a low plate/screw head profile and use existing (previously cleared) 1.6 mm or 1.9 mm emergency self-tapping and self-drilling screws. The DePuy Synthes 1.6 mm Burr Hole Covers have 5 lobes to accommodate a shunt/drain and are available in 12, 15, 17, and 24 mm diameters with a plate thickness of 0.5 mm.

    The DePuy Synthes Craniofacial Plate and Screw System consist of plates and meshes that come in a variety of shapes and sizes to meet the anatomical needs of the patient. This system is designed for use with 1.8 mm screws and 2.1 mm emergency screws. The screws will be used with Synthes 1.8 mm hexagonal screwdriver blades. System components are manufactured in either titanium or titanium alloy and are intended for single use only.

    DePuy Synthes Craniomaxillofacial (CMF) Distraction System is a modular distractor system intended for correction and reconstruction of the cranium in adults, adolescents, children, infants, and neonates. The distractor construct consists of the distractor body, footplates, extension arms, and bone screws, all of which are available in multiple configurations to meet patient and surgeon needs. The distractor body is first assembled with the desired footplates, and then the footplates are secured to bone using the bone screws. After implantation the distractor is activated through the rotation of an advancement/lead screw with an activation instrument percutaneously.

    The DePuy Synthes Orbital Plates, components of the Synthes Craniofacial Plate and Screw System, consist of anatomically shaped orbital plates that come in various sizes and configurations to fit the patient anatomy. These devices are designed for use with DePuy Synthes craniofacial bone screws commercially available in the U.S. System components are manufactured in titanium and are intended for single use only.

    The DePuy Synthes MatrixNEURO Cranial Plating System consists of bone fixation implants offered in a variety of shapes and sizes to meet the anatomical needs of the patient.
    The reconstruction meshes are manufactured from titanium, are designed for use with DePuy Synthes MatrixNEURO Cranial Plating System screws, are offered sterile, and are intended for single use only.
    The screws are manufactured from titanium alloy, are designed for use with DePuy Synthes MatrixNEURO plates, burr hole covers, and meshes, maybe be offered sterile or non-sterile, and are intended for single use only.

    The DePuy Synthes MatrixNEURO Cranial Plating System consists of bone fixation implants offered in a variety of shapes and sizes to meet the anatomical needs of the patient.
    The DePuy Synthes MatrixNEURO Preformed Meshes are precontoured to cover common cranial defects manufactured from grade 2 titanium that are designed for use with MatrixNEURO Cranial Plating System screws. The Preformed Meshes are offered sterile packed and are intended for single use only.

    AI/ML Overview

    The provided text is a 510(k) Premarket Notification from the U.S. Food & Drug Administration (FDA) regarding several DePuy Synthes Craniomaxillofacial Neuro Devices. It describes changes to the labeling of these devices to indicate "MR Conditional" use.

    The document consistently states that the purpose of the submission is to add MR Conditional information to the device labeling. It explicitly mentions that the intended use and technological characteristics of the devices remain unchanged. This implies that the devices are already cleared for their stated indications and the current submission is concerned with demonstrating their safety in an MR environment.

    Therefore, the "acceptance criteria" and "study that proves the device meets the acceptance criteria" in this context refer to the non-clinical performance testing conducted to support the MR Conditional safety of the devices. This is not a study proving clinical effectiveness or diagnostic accuracy of an AI/algorithm-based device, but rather a set of engineering tests to confirm a physical device's compatibility with an MR environment.

    Based on the provided text, here's how to address your request:

    1. A table of acceptance criteria and the reported device performance:

    The document repeatedly cites the ASTM standards as the basis for performance testing, which inherently contain acceptance criteria within them. The document does not provide specific quantitative performance results for each device against these criteria, but rather states that "Non-clinical testing is provided to support the conditional safety." This implies that the devices met the acceptance criteria defined by these ASTM standards.

    Acceptance Criteria (ASTM Standards)Reported Device Performance
    Magnetically Induced Displacement Force: ASTM F2052-14Non-clinical testing provided to support conditional safety. (Implied: Met standard)
    Magnetically Induced Torque: ASTM F2213-06Non-clinical testing provided to support conditional safety. (Implied: Met standard)
    Radio Frequency (RF) Heating: ASTM F2182-11aNon-clinical testing provided to support conditional safety. The DUKE model places devices in the clinically relevant anatomic position. The DUKE results will be used for labeling of RF heating. (Implied: Met standard)
    Image Artifacts: ASTM F2119-07Non-clinical testing provided to support conditional safety. (Implied: Met standard)

    2. Sample sized used for the test set and the data provenance:

    • Sample Size: The document does not specify the number of individual devices or prototypes tested for each type of assessment (displacement, torque, heating, artifacts). These are engineering tests typically performed on representative samples.
    • Data Provenance: The data provenance is "non-clinical testing" conducted by the manufacturer, DePuy Synthes. The "DUKE model" for RF heating is mentioned, which is a widely accepted phantom model for electromagnetic simulations in medical device testing. There's no information about the country of origin of the data beyond it being part of a US FDA submission. It is a prospective test in the sense that the testing was conducted to support the MR Conditional labeling.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This question is not applicable in the context of this submission. The "ground truth" for MR Conditional safety is established by engineering tests against recognized consensus standards (ASTM), not by human expert review of clinical cases. The tests directly measure physical properties like temperature increase or magnetic force, and these measurements are compared to the limits specified in the standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    Not applicable. This is not a human reader study requiring adjudication. The assessment is based on objective physical measurements and adherence to ASTM standards.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This submission is for physical medical devices (implants), not an AI/algorithm-based diagnostic or assistive software. No MRMC study was performed because the device is not an AI reading tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    Not applicable. This is not an algorithm, so standalone performance is irrelevant. The tests are for the physical properties of the implants.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc):

    The "ground truth" for these tests is defined by the physical laws and empirically derived thresholds outlined in the ASTM standards for MR compatibility. For example, a certain temperature rise in a standard phantom model is deemed acceptable if it remains below a specified limit.

    8. The sample size for the training set:

    Not applicable. There is no "training set" as this is not a machine learning or AI model.

    9. How the ground truth for the training set was established:

    Not applicable, as there is no training set.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1