Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K080636

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2009-02-13

    (344 days)

    Product Code
    Regulation Number
    862.1345
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Duo-Max Blood pressure and Glucose Monitor, Model HMF-100 consists of a meter with wrist cuff and test strips. The system is intended for use in the quantitative measurement of glucose in whole blood taken from the fingertip. Testing is done outside the body (In Vitro diagnostic use). It is indicated for use at home (over the counter [OTC]) by persons with diabetes mellitus, or in clinical settings by healthcare professionals, as an aid to monitor the effectiveness of diabetes control. It is not intended for use on neonates. Also the system measures systolic and diastolic blood pressure and pulse rate from adult's wrist in the home care environment. The device employs a wrist cuff and the oscillometric metric method of measurement.

    The Blood pressure and glucose monitor. Duo-Max™ consists of a meter with wrist cuff and glucose test strips. The glucose test system is intended for use in the quantitative measurement of glucose in capillary whole blood taken from the fingertip, dorsal hand, ventral palm, upper arm, forearm, calf and thigh. Glucose testing is done outside the body (In Vitro diabetes mellitus, or in clinical settings by healthcare professionals. Glucose testing system is not to be used for the diagnosis or screening of diabetes. It is not intended for use on neonates. Also the blood pressure test system measures systolic and diastolic blood pressure and pulse rate from adult's wrist in the home care environment. The device employs a wrist cuff and the oscillomentric metric method of measurement.

    Device Description

    HuBDIC's Duo- Max Blood Pressure and Glucose Monitor System, Model HMF-100 combines the function of a blood pressure meter and a blood glucose monitoring system in one unit. Supplies with the meter are test strips, lancets, Lancing device, storage case, batteries and log book.

    HMF-100 adopts the wrist type cuff for blood pressure meter part. The cuff and control unit are combined into a single wrist-mounted assembly. The user interface pane has power switch, timer switch, memory switch for blood pressure meter part and LCD for displaying the systolic blood pressure, diastolic blood pressure, pulse rate, date and time. This device has the memory function that permits memory and display of the 180 most recent measurement results.

    AI/ML Overview

    The provided text describes a medical device called the "Duo-Max Blood Pressure and Glucose Monitor, Model HMF-100" and its 510(k) submission. However, the document does not contain specific acceptance criteria, detailed study designs, or reported device performance metrics in the format requested.

    The document primarily focuses on:

    • Device Description: What the device is and its components.
    • Intended Use: For whom and how the device is to be used (blood pressure and glucose monitoring).
    • Comparison to Predicate Device: Highlighting similarities and minor physical differences (dimensions, weight, memory capacity) with an existing device (Sein BGP-100).
    • Standards Conformance: Stating that the device "conforms to applicable standards including ANSI/AAMI SP-10, IEC 60601-1 and IEC 60601-1-2 requirements" and "met all relevant requirements of the aforementioned tests." This implies that the device achieved performance in line with these standards, but the specific numerical acceptance criteria or test results are not provided in this summary.
    • Conclusion of Substantial Equivalence: The manufacturer's assertion that the device is as safe and effective as predicate devices.
    • FDA Clearance Letter: Confirming the FDA's decision that the device is substantially equivalent.

    Therefore, many of the requested details cannot be extracted from the provided text.

    Here's a breakdown of what can and cannot be answered based on the provided text:


    1. A table of acceptance criteria and the reported device performance

    Metric / ParameterAcceptance Criteria (as per standards)Reported Device Performance
    Blood PressureConforms to ANSI/AAMI SP-10 standards"met all relevant requirements"
    Blood GlucoseConforms to unspecified standards related to glucose monitoring"met all relevant requirements"
    Safety & EMCConforms to IEC 60601-1 and IEC 60601-1-2"met all relevant requirements"
    • Explanation: The document states the device "conforms to applicable standards including ANSI/AAMI SP-10, IEC 60601-1 and IEC 60601-1-2 requirements" and "met all relevant requirements of the aforementioned tests." However, it does not explicitly list the numerical acceptance criteria from these standards or provide the specific numerical results of the device's performance against those criteria. It only makes a general statement of compliance.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Not provided. The summary mentions "performance testing results" but does not detail the sample size (number of participants or measurements), the nature of the study (retrospective or prospective), or the data provenance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable / Not provided. For a blood pressure and glucose monitor, the ground truth is typically established by comparing the device readings to a reference method (e.g., a clinically validated sphygmomanometer for BP, or a laboratory analyzer for glucose). This does not involve human expert consensus in the same way as image interpretation. The document does not describe how ground truth was established, beyond implying comparison to reference methods as part of standardized testing.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable / Not provided. Adjudication methods like 2+1 or 3+1 are typically used in studies involving subjective assessment (e.g., medical image interpretation by multiple readers). For an objective measurement device like this, ground truth is established by a reference method, not by expert adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This device is a direct measurement tool (blood pressure and glucose monitor), not an AI-powered diagnostic imaging tool that assists human readers. Therefore, an MRMC study or AI assistance is not relevant to this product.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable for "algorithm only" in the typical sense. The device itself is a standalone system for measurement. The performance described "met all relevant requirements of the aforementioned tests," implying that the device was tested as a complete system (hardware and internal algorithms) without human intervention in the measurement process itself. However, it's not an "algorithm only" study in the context of AI models.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Implicitly Reference Methods. For blood pressure, the ground truth would typically be established using a carefully calibrated and validated reference sphygmomanometer (e.g., mercury or an oscillometric device that has passed a higher level of validation). For blood glucose, the ground truth would be established by laboratory-grade glucose analyzers. The document does not explicitly state this, but it is standard practice for validating such devices.

    8. The sample size for the training set

    • Not applicable / Not provided. As this is a traditional medical device for objective measurement (not an AI/machine learning model in the usual sense that requires a "training set"), the concept of a training set as understood in AI development does not apply here. The device's internal algorithms for measurement are likely developed through engineering and calibration, not statistical training on a dataset.

    9. How the ground truth for the training set was established

    • Not applicable / Not provided. See explanation for question 8.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1