510(k) Data Aggregation

    K Number
    K971256

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    1997-09-24

    (173 days)

    Product Code
    Regulation Number
    880.5700
    Age Range
    12 - 120
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DUO-LITE™ Phototherapy System is designed to provide treatment in the form of Phototherapy to those infants that have been identified as having elevated bilirubin levels (hyperbilirubinemia). Phototherapy as a means for reducing bilirubin levels is a well established treatment modality. The output of the light has been optimized for maximum absorption by the skin in order to reduce the bilirubin levels as quickly as possible.

    The DUO-LITE™ Phototherapy System provides the option to select white or blue light via a selector switch. This provides the caregiver with the option to give high output phototherapy for some infants (or early in the treatment cycle) followed by more moderate phototherapy for other infants (or later in the treatment cycle). By contrast, other phototherapy lamp products give the option of either White or Blue light which needs to be set up by an engineer in advance of the treatment. The DVO-LITE™ has combined these two light sources into one housing for the convenience of the caregiver.

    The DUO-LITE™ Phototherapy System can be placed on the hood of an infant incubator, mounted on the Air-Shields Rail System, or positioned over an open warming bed or crib.

    Device Description

    The DUO-LITE" Phototherapy System provides light output for maximum absorption by the infant's skin in order to reduce bilirubin levels. The DUO-LITE™ is designed for use in conjunction with a variety of other products. Perhaps the most common application is when placed on the hood of an incubator, or fitted to a rail system and appitioned over the hood of the incubator. The DUO-LITT:" can also be mounted to a mobile stand, thus extending its applications to use with Open Warmers and Open Bassinets.

    The product is L 23 x W 10 x H 5 (inches) and weighs approximately 15 Lbs. It is available in 100, 120 and 220/240V 50/60 Hz. The DUO-LITE™ is comprised of three 75W quartz halogen bulbs. The light emitted from these bulbs is filtered through IR/UV and Dichroic filters. The output of the bulbs is dependent on the driving voltage which is regulated by the internal power supply.

    The user can select the use of White light or Blue light using the selector switch on the front of the lamp housing. The different color lights give different therapeutic outputs depending on the individual needs of the patient. Adjacent to the On/Off switch is a digital counter which can be set to display exposure time and bulb life remaining.

    The DU1)-LIT!:" uses quartz halogen technology since the irradiance remains constant over the life of the bulb. By way of comparison, fluorescent bulbs degrade rapidly over the course of time and as such need to be continually monitored to ensure the output is therapeutically acceptable.

    AI/ML Overview

    The provided text describes the Air-Shields Duo-Lite™ Phototherapy System, a medical device for treating neonatal hyperbilirubinemia. The submission is a 510(k) summary, which aims to demonstrate substantial equivalence to legally marketed predicate devices, rather than proving a device meets novel acceptance criteria through a clinical study. Therefore, some of the requested information, such as sample size for test and training sets, number of experts for ground truth, adjudication methods, and MRMC studies, is not applicable to this type of regulatory submission.

    1. Table of Acceptance Criteria and Reported Device Performance

    For a 510(k) submission, the "acceptance criteria" are typically defined by demonstrating that the device performs comparably to predicate devices in key specifications, rather than setting specific numerical thresholds for a novel device. The Duo-Lite™ compares its performance against several predicate devices. The critical performance metric highlighted is irradiance.

    Acceptance Criteria (Implied by Predicate Comparison)Reported Duo-Lite™ Performance
    Achieves therapeutic irradiance for phototherapy.Irradiance (400 - 520 Nm) @ 43 cm: ≥10 µW/cm²/nm (White light) ≥20 µW/cm²/nm (Blue light)
    Low UV and IR output."also indicate low UV and IR output."
    Meets specified bulb life.Bulb Life: 1,000 Hrs
    Line Voltage compatibility.Line Voltage: 120 Vac
    Power Consumption within acceptable limits.Power Consumption: 350W
    Leakage Current within acceptable limits.Leakage Current: <300µA

    2. Sample Size Used for the Test Set and Data Provenance

    This is a non-clinical test of the device's physical performance, not a study involving patient data. Therefore, typical "sample size" and "data provenance" as applied to clinical studies are not applicable.
    The tests involved measurements of the Duo-Lite™ Phototherapy System under various configurations (filter color, voltage, distance). The data provenance is from an "outside test house" for spectral irradiance measurements and "in-house laboratory" for mattress mapping.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. The "ground truth" here is the physical performance of the device (e.g., spectral irradiance) as measured by specialized equipment and methodologies, not human expert interpretation of clinical data.

    4. Adjudication Method for the Test Set

    Not applicable. This was a non-clinical device performance test, not a clinical study requiring adjudication of expert opinions.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done

    No, a Multi Reader Multi Case (MRMC) comparative effectiveness study was not done. This 510(k) submission focuses on demonstrating the physical and functional equivalence of the device to existing predicate devices through engineering specifications and non-clinical testing.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    Yes, in a sense. The "standalone performance" in this context refers to the intrinsic physical performance of the device itself (e.g., its irradiance output). The "spectral irradiance measurements conducted by an outside test house" and "Mattress mapping done in our in-house laboratory" represent this standalone assessment of the device's output. There is no algorithm or human-in-the-loop component relevant to these performance tests.

    7. The Type of Ground Truth Used

    The "ground truth" for this non-clinical testing is the objective, measurable physical properties and output of the device, obtained through:

    • Spectral irradiance measurements: Using specialized photometers/spectroradiometers.
    • Mattress mapping: To assess the distribution of irradiance.
    • Engineering specifications: (e.g., bulb life, voltage, power consumption) as verified by testing or design.

    8. The Sample Size for the Training Set

    Not applicable. This is not an AI/ML device that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. This is not an AI/ML device that requires a training set.