510(k) Data Aggregation

    K Number
    K042966

    Validate with FDA (Live)

    Date Cleared
    2004-11-19

    (22 days)

    Product Code
    Regulation Number
    888.3030
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Duet™ and Impact™ Suture Anchors are intended for use to reattach soft tissue to bone in orthopaedic surgical procedures. The device may be used in either arthroscopic or open surgical procedures. After the suture is anchored to the bone, it may be used to reattach soft tissue, such as ligaments, tendons, or joint capsules, to the bone. The suture anchor system thereby stabilizes the damaged soft tissue, in conjunction with appropriate postoperative immobilization, throughout the healing period.

    The Duet™ and Impact™ Suture Anchors are contraindicated in 1) Surgical procedures other than those listed, 2) Conditions that may compromise Anchor fixation (osteopenic, comminuted bone, pathologic conditions in the soft tissues to be attached, etc., 3) Conditions that may retard healing (poor blood supply, past or potential infection, etc), 4) Active infection, 5) Conditions that may limit the patients ability or willingness to restrict activities or follow directions during the healing period, 6) Foreign body sensitivity to materials, 7) Patients with suspected or known allergy with implant or suture materials.

    Device Description

    Duet™ Suture Anchor and Impact™ Suture Anchors are bioabsorbale suture anchors that are preloaded on a disposable inserter device with two non-absorbable sutures. Originally they were preloaded with non-absorbable, braided, polyester #2 sutures, one of them is green and another one is white.

    Purpose of this special 510(k) premarket notification is amendment of new preloaded suture material into Duet™ and Impact™ suture anchor product lines. The new suture, Herculine is non-absorbable, braided, ultra-high molecular weight polyethylene #2 suture. In coloured version 2 filaments are replaced by blue polypropylene monofilaments Herculine suture meets USP requirements for knot tensile strength and needle attachment strength.

    AI/ML Overview

    The provided text is a 510(k) summary for a medical device (Duet™ and Impact™ Suture Anchor) and a letter from the FDA regarding its clearance. It does not contain information about acceptance criteria, device performance, study design details (sample size, data provenance, expert consensus, adjudication, MRMC, or standalone studies), or ground truth establishment for either training or test sets.

    The document primarily describes a special 510(k) premarket notification for an amendment to the existing suture anchors, specifically changing the preloaded suture material from polyester to ultra-high molecular weight polyethylene (Herculine). The core argument for substantial equivalence is that this change has no effect on the intended use, principles of operation, production, raw materials, or sterilization of the device, and that the new suture meets USP requirements for knot tensile strength and needle attachment strength.

    Therefore, I cannot fulfill the request to provide a table of acceptance criteria and reported device performance or details about the study that proves the device meets the acceptance criteria, as this information is not present in the given text.

    The text focuses on demonstrating substantial equivalence to a predicate device based on material change and compliance with general standards rather than presenting a performance study against specific acceptance criteria.

    Ask a Question

    Ask a specific question about this device

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Bionx Duet™ Suture Anchor is intended for use to reattach soft tissue to bone in orthopaedic surgical procedures. The device may be used in either arthroscopic or open surgical procedures. After the suture is anchored to the bone, it may be used to reattach soft tissue, such as ligaments, tendons, or joint capsules, to the bone. The suture anchor system thereby stabilizes the damaged soft tissue, in conjunction with appropriate postoperative immobilization, throughout the healing period.

    Device Description

    Duet™ Suture Anchor is threaded, bioabsorbable suture anchor, provided as preloaded with two non-absorbable, braided, polyester #2 sutures, one of them is green and another one is white. Diameter of the anchor is 6.0mm and length 16.5mm. It is assembled with sutures with needles into single use, sterile, disposable driver and packaged into blister with a punch. The Duet™ Suture Anchor is made of poly-L/D-lactide copolymer.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a medical device called the Duet™ Suture Anchor. This document focuses on demonstrating substantial equivalence to predicate devices rather than proving performance against specific acceptance criteria through a standalone study.

    Therefore, the requested information elements related to acceptance criteria, device performance, study design (test/training sets, ground truth, expert opinions, MRMC studies) are not explicitly described in the provided text.

    Here's a breakdown of what can be extracted and what information is missing:

    1. A table of acceptance criteria and the reported device performance

    • Not explicitly provided. The document focuses on demonstrating that the Duet™ Suture Anchor has the "same intended use and principles of operation and very similar technological characteristics" as existing predicate devices. It states that "the minor technological differences between the Bionx Implants Inc. Duet™ Suture Anchor and the predicate devices do not raise any new issues of safety or effectiveness." This implies that the device is expected to perform comparably to its predicates, but no specific performance metrics or acceptance criteria are listed for the Duet™ itself.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Not applicable/Not provided. This document is a 510(k) submission, not a clinical study report. It does not describe a test set or data provenance in the context of device performance evaluation.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable/Not provided. Ground truth establishment is not discussed in the context of device performance in this document.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable/Not provided. Adjudication is not mentioned as no performance study is described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This device is a bioabsorbable surgical anchor, not an AI-powered diagnostic tool. Therefore, an MRMC study is irrelevant to its evaluation.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    • Not applicable. This is a physical implant, not an algorithm.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    • Not applicable/Not provided. Ground truth is not discussed.

    8. The sample size for the training set

    • Not applicable/Not provided. Training sets are relevant for AI/machine learning models, which this device is not.

    9. How the ground truth for the training set was established

    • Not applicable/Not provided. Ground truth for a training set is not discussed.

    Summary of what the document does provide:

    • Intended Use: Reattachment of soft tissue to bone by arthroscopy or direct visualization.
    • Predicate Devices: A long list of existing bioabsorbable suture anchors with similar intended uses and principles of operation.
    • Material: Self-Reinforced (96/4D) PLA Copolymer, with the material provenance being previously cleared Bionx Implants Inc. devices.
    • Mechanism of action: Retains 90% strength through 20 weeks and completely resorbs over several years, allowing gradual strength loss as collagen fibers heal.
    • Conclusion: The FDA determined the device is "substantially equivalent" to legally marketed predicate devices, meaning it does not raise new issues of safety or effectiveness. This is the primary "proof" of meeting regulatory requirements for this type of submission.