Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K030401

    Validate with FDA (Live)

    Date Cleared
    2003-06-19

    (133 days)

    Product Code
    Regulation Number
    886.4370
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The OASIS Disposable M2-PE (Precision Edge) Microkeratome Blade (Catalog #0415) is designed as a replacement blade for the Moria M2 Microkeratome for lamellar resection of the cornea.

    Device Description

    The OASIS Disposable M2-PE Microkeratome Blades are replacement stainless steel blades for the Moria M2 Microkeratome Blade. The Disposable M2-PE Microkeratome Blades are made of 400 Series Stainless Steel, packaged and sterilized using the same methods. The OASIS Disposable M2-PE Microkeratome Blades are single-use, disposable blades.

    AI/ML Overview

    The provided document is a 510(k) Premarket Notification Summary for the OASIS Medical, Inc. Disposable M2-PE Microkeratome Blades. It focuses on demonstrating substantial equivalence to a predicate device rather than comprehensive de novo clinical studies with acceptance criteria in the typical sense of AI/ML device evaluations. Therefore, the information requested might not be fully present or will be interpreted based on the available data.

    Here's an analysis based on the document:

    1. Table of Acceptance Criteria and Reported Device Performance

    Strict "acceptance criteria" as would be defined for an AI/ML device comparing performance metrics (e.g., sensitivity, specificity thresholds) are not explicitly stated. Instead, the demonstration of substantial equivalence is based on the device having "the same" or "substantially equivalent" characteristics and performance compared to the predicate device.

    Characteristic / TestAcceptance Criteria (Implied from Substantial Equivalence Claim)Reported Device Performance (OASIS M2-PE Microkeratome Blades)
    Intended UseSame as predicate (Moria M2 Microkeratome Blade)Same
    Target PopulationSame as predicateSame
    Performance (Compatibility)Compatible with Moria M2 Microkeratome (as predicate)Same (Tested and shown to be acceptable)
    Blade MaterialSame as predicate (low carbon stainless steel)Same
    BiocompatibilitySame as predicate (for Stainless Steel Blades)Same
    Mechanical SafetyAssured, same as predicateAssured, same as predicate
    Dimensional EquivalencySubstantially equivalent to predicate in physical measurementsSubstantially equivalent to the measurements of blades manufactured by OASIS Medical, Inc.
    SharpnessPerform as well as the predicate devicePerform as well as the predicate device
    Fit into Moria M2 MicrokeratomeAcceptable fitTested and shown to be acceptable
    Non-clinical Testing (Corneal Lamellar Sections)Equivalent to the predicate devicesResulted in corneal lamellar sections equivalent to the predicate devices (on porcine eyes)

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set:
      • For Dimensional Equivalency, Sharpness, and Fit tests: The document does not specify the exact number of blades tested. It only states "physical measurements of the predicate device are substantially equivalent to the measurements of blades manufactured by OASIS Medical, Inc." and "sharpness tests show that the OASIS M2-PE blades (0415) perform as well as the predicate device."
      • For Non-clinical testing on porcine eyes: The document does not specify the number of porcine eyes used.
    • Data Provenance: The document does not explicitly state the country of origin or whether the data was retrospective or prospective. Given the nature of a 510(k) submission for a physical device like a blade, the testing would typically be prospective laboratory and possibly animal testing conducted by the manufacturer, or a contract lab on their behalf.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    There is no mention of "experts" in the context of establishing ground truth for a test set, as this is not a study evaluating human-interpretable output where expert consensus would be required (e.g., image reading). The "ground truth" for these performance tests would be objective measurements (dimensions, sharpness metrics) or outcomes from the animal model (quality of lamellar sections).

    4. Adjudication Method for the Test Set

    Not applicable. Adjudication methods (like 2+1, 3+1) are used for resolving disagreements in expert readings, which is not relevant for the types of tests (dimensional, sharpness, fit, animal model) conducted for this mechanical device.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic devices that involve human interpretation of results, often with AI assistance, to measure the impact of AI on human reader performance. This document pertains to a surgical blade, not a diagnostic imaging or AI-assisted diagnostic device.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    Yes, in a sense, the performance tests described (Dimensional Equivalency, Sharpness, Fit, Non-clinical testing on porcine eyes) can be considered "standalone" in that they assess the intrinsic characteristics and performance of the blade itself, without direct human-in-the-loop interpretation being the primary focus of the performance evaluation. However, this is not an "algorithm-only" study as described for AI/ML devices.

    7. The Type of Ground Truth Used

    The "ground truth" for the tests performed was based on:

    • Objective Measurements: For dimensional equivalency, sharpness, and fit.
    • Observed Outcomes: For non-clinical testing on porcine eyes, the "equivalence" of corneal lamellar sections to those produced by the predicate device would be the observed outcome. While not explicitly stated, this observation would likely be qualitative or semi-quantitative, assessed by experienced personnel.

    8. The Sample Size for the Training Set

    Not applicable. The concept of a "training set" is relevant for AI/ML models that learn from data. This document describes the evaluation of a physical medical device (a microkeratome blade), not an AI algorithm.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for a physical device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1