Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K100100

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2010-11-10

    (302 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DIO STEADY External Implant System is an endosseous dental implant is indicated for surgical placement in the upper and lower jaw arches, to provide a root form means for single or multiple units' prosthetic appliance attachment to restore a patient's chewing function. Implants can be placed with a conventional two stage surgical process with an option for transmucosal healing or they can be placed in a single stage surgical process for immediate loading. Immediate loading is restricted to the anterior mandible, based on four interforaminal placed implants, and not indicated for single unsplinted implants. Patients must be subject for dental treatment with endosseous implants.

    Device Description

    The DIO STEADY External Implant System is comprised of dental implants, and superstructures. The DIO STEADY External Implant System is specially designed for use in dental implant surgery. A successfully osseointegrated implant will achieve a firm implant when surgically implanted under controlled conditions, per well known clinical studies. There are intended for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations.

    AI/ML Overview

    The provided 510(k) submission for the DIO STEADY External Implant System is for a dental implant, which is a physical device, not an AI/ML-driven medical device. Therefore, a significant portion of the requested information, such as acceptance criteria for AI performance, sample sizes for test/training sets, ground truth establishment, expert qualifications, and MRMC studies, is not applicable in this context.

    However, I can extract and present the relevant information regarding the performance and testing conducted for this traditional medical device.

    Device Name: DIO STEADY External Implant System
    Predicate Device: SM Internal/External Implant System (510(k) No: K070569)
    Regulatory Class: Class II


    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Test/RequirementReported Device Performance
    Material BiocompatibilityCompliance with applicable International and US regulationsBiocompatibility performed, confirming compliance.
    Functional PerformanceFatigue testing according to "Guidance for industry and FDA staff Class II Special Controls Guidance Document Root-form Endosseous Dental Implants and Endosseous Dental Abutment"Fatigue test results similar to previously cleared predicate devices, using the worst-case scenario (DIO STEADY External Fixture and an angled abutment).
    Design ConformanceConformance to design input requirementsDevice functionality and conformance to design input requirements were determined through laboratory testing.
    Substantial EquivalenceDemonstrates equivalence to predicate device (SM Internal/External Implant System K070569) in material, intended use, and similar design/technological characteristics.Evaluation concluded that the DIO STEADY External Implant System is substantially equivalent to its predicate device.

    2. Sample size used for the test set and the data provenance

    • Test Set Sample Size: Not applicable. This is a physical device, and the performance evaluation involved laboratory testing for mechanical and material properties, rather than testing on a dataset.
    • Data Provenance: Not applicable. The testing was conducted in a laboratory setting by the manufacturer (DIO Corporation).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. Ground truth, in the context of AI/ML, refers to annotated data. For a physical dental implant, performance is validated through engineering tests and adherence to standards, not through expert consensus on a test set.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    Not applicable. Adjudication methods are relevant for resolving discrepancies in expert annotations for datasets, which is not pertinent to the mechanical and material testing of a dental implant.

    5. If a multi-reader, multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a physical dental implant, not an AI-assisted device.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a physical dental implant, not an algorithm.

    7. The type of ground truth used

    The "ground truth" for this device's performance would be established by:

    • Engineering Standards: Adherence to recognized international and US standards for medical device materials (e.g., ASTM F67 for CP Ti Gr4) and performance (e.g., "Guidance for industry and FDA staff Class II Special Controls Guidance Document Root-form Endosseous Dental Implants and Endosseous Dental Abutment").
    • Laboratory Measurements: Direct measurements of material properties and mechanical performance (e.g., fatigue strength) under controlled laboratory conditions, compared against established benchmarks or predicate device performance.

    8. The sample size for the training set

    Not applicable. This is a physical device, and no training set (in the AI/ML sense) was used. The device design and manufacturing processes are developed through traditional engineering methods and quality controls.

    9. How the ground truth for the training set was established

    Not applicable. There is no training set for this type of device. The "ground truth" for its design and manufacturing is based on established engineering principles, material science, and regulatory requirements for medical devices.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1